K Number
K080380

Validate with FDA (Live)

Date Cleared
2009-07-23

(526 days)

Product Code
Regulation Number
866.3328
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SAS™ FluAlert A & B Test is a visual and rapid assay for the presumptive in-vitro qualitative detection of influenza A and influenza B viral nucleoprotein antigens from nasal washes and nasal aspirates of symptomatic patients. Negative results should be confirmed via culture. This test is not intended for the detection of Influenza Type C viral antigen. This test is for professional use only. Negative results do not preclude infection with influenza A or B and should not be used as the sole basis for treatment or other patient management decisions. It is recommended that negative results be confirmed by cell culture.

Device Description

The SAS™ FluAlert A & B Test utilizes antibodies against influenza type A and influenza type B viral nucleoproteins. After the extraction has been completed, the sample is placed into two separate sample wells. The specimen is absorbed and migrates via capillary action through membranes that contain dried gold conjugated antibody, which is specific for either influenza A or influenza B viral nucleoproteins. If these nucleoproteins are present, a "half-sandwich" immunocomplex is formed. The membrane contains immobilized antibody to influenza A or influenza B nucleoproteins, respectively, which bind the "half sandwich" complex. Thus, in the presence of influenza nucleoproteins, a "whole sandwich" immunocomplex is formed and a visible, pink-colored line develops in the specimen zone of the test device. In the absence of an influenza antigen, a "sandwich" immunocomplex is not formed and a negative result is indicated. To serve as a procedural control, a pink-colored control line will always appear in the control zone of each strip regardless of the presence or absence of influenza A or influenza B nucleoproteins.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the SAS™ FluAlert A & B Test, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets in the document. Instead, the study aims to demonstrate "substantial equivalence" to predicate devices (SAS™ Influenza A Test and SAS™ Influenza B Test). The performance benchmarks are effectively the agreement rates with these predicate devices for positive and negative influenza cases.

CharacteristicAcceptance Criteria (Implied by Substantial Equivalence to Predicate Device)Reported Device Performance (SAS™ FluAlert A & B Test)
Influenza A
Positive % AgreementHigh agreement with individual SAS™ Influenza A Test97.2% (95% CI 89-100%) (Combined Fresh NW & NA)
Negative % AgreementHigh agreement with individual SAS™ Influenza A Test99.7% (95% CI 98-100%) (Combined Fresh NW & NA)
Influenza B
Positive % AgreementHigh agreement with individual SAS™ Influenza B Test98.7% (95% CI 92-100%) (Combined Fresh NW & NA)
Negative % AgreementHigh agreement with individual SAS™ Influenza B Test99.7% (95% CI 98-100%) (Combined Fresh NW & NA)
ReproducibilityHigh agreement with expected results across sites and users88.8% - 100% agreement depending on panel member
Cross-ReactivityNo cross-reactivity with tested viral/bacterial strainsNo cross-reactivity observed
InterferenceNo interference from tested viral/bacterial strains with influenza detectionNo interference observed (Influenza still detected as "Pos")

2. Sample Size Used for the Test Set and Data Provenance

  • Prospective Clinical Study:
    • Sample Size: 461 nasal clinical specimens (nasal washes and nasal aspirates combined).
    • Data Provenance: Prospectively collected from four clinical trial sites in Texas and South Dakota, USA. An approximately equal mix of adult (>21 years) and pediatric patients (0-21 years) for nasal washes, and predominantly pediatric patients for nasal aspirates.
  • Retrospective Study:
    • Sample Size: 191 frozen, archived nasal wash samples.
    • Data Provenance: Retrospective, from a children's hospital in Texas, USA.

In the prospective study, 13 samples yielded invalid results and were excluded from the totals.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications for establishing the ground truth for the clinical specimens. The comparison is made against the predicate devices (SAS™ Influenza A Test and SAS™ Influenza B Test) and also mentions "cell culture or DFA" for those predicate devices in the "Results Summary" chart reference, but not explicitly for the new device's test set.

4. Adjudication Method for the Test Set

The document states that the clinical specimens were tested "blindly and prospectively comparing the SAS™ Influenza A Test and The SAS™ FluAlert A&B (combined) Test performance" and similarly for Influenza B. This implies that the results of the new device were compared against the established results of the predicate devices. There is no explicit mention of an adjudication method like 2+1 or 3+1 by human readers; the comparison is device-to-device.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study compares the new device's performance directly to the predicate devices, not human readers with and without AI assistance. The "reproducibility" section involved "non-professional users," but this was to assess consistent device results, not to evaluate human reader improvement with AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, this is essentially a standalone performance study. The SAS™ FluAlert A & B Test is a rapid immunoassay (visual and rapid assay) that produces a visible, pink-colored line. Its performance is evaluated directly against predicate devices based on its ability to detect influenza viral nucleoprotein antigens. There is no "human-in-the-loop" component mentioned for interpreting the device's fundamental output beyond simply reading the presence or absence of a line.

7. The Type of Ground Truth Used

The primary ground truth for evaluating the SAS™ FluAlert A & B Test's performance in the clinical studies is the results from the substantially equivalent predicate devices (SAS™ Influenza A Test and SAS™ Influenza B Test). For these predicate devices, the document refers to a comparison to "cell culture or DFA" in an older summary, implying that those methods serve as a "gold standard" for the predicate devices, and by extension, for this new device's evaluation against the predicates.

8. The Sample Size for the Training Set

The document describes performance studies for a diagnostic device, not a machine learning algorithm that typically requires a separate training set. Therefore, there is no explicit training set or sample size for training mentioned as it's not applicable in the context of this type of diagnostic device submission.

9. How the Ground Truth for the Training Set was Established

As there is no explicit training set for a machine learning algorithm, there is no description of how ground truth for a training set was established. The studies conducted are for performance evaluation against established methods and devices.

{0}------------------------------------------------

K080380

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS SAS™ FluAlert A & B Test

.

This 510(k) summary of safety and effectiveness submission is in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitted by:SA Scientific, Ltd.4919 Golden QuailSan Antonio, TX 78240210-699-8800JUL 23 2009
Establishment Reg. No:1645225
Contact Person:Robbi Perry
Date Prepared:July 20, 2009
Proprietary Name:SASTM FluAlert A & B Test
Classification Name:Antigens, CF (including CF control), Influenza virus A, B, C
Device Classification:21 CFR Part 866.3330
Regulatory Class:Class II
Classification AdvisoryCommittee:Microbiology
Product Code:GNX
Substantial Equivalence:Substantially equivalent to the SASTM individual devices-SASTM Influenza ATest (K044141) and SAS™ Influenza B Test (K041439), manufactured by SAScientific, Ltd., San Antonio, TX.
Device Description:The SAS™ FluAlert A & B Test utilizes antibodies against influenza type A andinfluenza type B viral nucleoproteins. After the extraction has been completed,the sample is placed into two separate sample wells. The specimen is absorbedand migrates via capillary action through membranes that contain dried goldconjugated antibody, which is specific for either influenza A or influenza B viralnucleoproteins. If these nucleoproteins are present, a "half-sandwich"immunocomplex is formed. The membrane contains immobilized antibody toinfluenza A or influenza B nucleoproteins, respectively, which bind the "halfsandwich" complex. Thus, in the presence of influenza nucleoproteins, a"whole sandwich" immunocomplex is formed and a visible, pink-colored linedevelops in the specimen zone of the test device. In the absence of an influenzaantigen, a "sandwich" immunocomplex is not formed and a negative result isindicated. To serve as a procedural control, a pink-colored control line willalways appear in the control zone of each strip regardless of the presence orabsence of influenza A or influenza B nucleoproteins.
Intended Use:SASTM FluAlertA&B Test is a visual and rapid assay for the presumptive in-vitro qualitative detection of influenza A and influenza B viral nucleoproteinantigens from nasal washes and nasal aspirates of symptomatic patients.Negative results should be confirmed via culture. This test is not intended for
the detection of Influenza Type C viral antigen. This test is for professional use only.
Negative results do not preclude infection with influenza A or B and should not be used as the sole basis for treatment or other patient management decisions. It is recommended that negative results be confirmed by cell culture.
Conditions for Use:For prescription use only.
Quality Controls:The SAS™ Influenza A & Influenza B Test provides two (2) internal procedural quality controls. It is recommended that external quality controls be assayed following the user's laboratory's standard quality control procedures and in conformance with local, state and federal regulations or accreditation organizations as applicable
Device comparison:The SAS™ Influenza A & Influenza B tests are rapid immunoassay tests utilizing immunochromatographic technology for the visualization of Influenza A & Influenza B viral nucleoprotein antigens. Each utilizes an antibody conjugated to colored particles and an antibody printed onto a membrane. The chemistry of the predicate devices and the proposed device is identical; the only differnce is the plastic cassette.
Performance Summary:The SASTM FluAlert A&B test performed substantially equivalent to the predicate devices, SAS™ Influenza A and SAS Influenza Flu B Tests mentioned above. This was verified by comparison to freshly collected nasal wash and nasal aspirate specimens.
Cross-reactivity and interference studies were performed on viral and bacterial strains commonly found in the human respiratory tract. None of the organisms interfered or cross-reacted with the performance of the SASTM FluAlert A& B test.
Clinical Summary:
Prospective Clinical Study:The SASTM FluAlert A&B Test combines the immunoassay test strips from the individual SASTM Influenza A and SASTM Influenza B Tests into one side-by-side plastic cassette. There are no other changes made to this test. In these studies, the users compared the combined test to the individual tests for evaluation of user interpretations. Please see: "Results Summary: SASTM Influenza A and SAS™ Influenza B Individual Devices compared to cell culture or DFA" chart for comparison of the individual tests to cell culture/DFA.
Four clinical trial sites, in Texas and South Dakota, tested four hundred sixty one (461) nasal clinical specimens blindly and prospectively comparing the SAS™ Influenza A Test and The SAS™ FluAlert A&B (combined) Test performance. The SAS™ Influenza A Test and SAS™ FluAlert A&B Test had a positive percent agreement of 97.2% and a negative percent agreement of 99.7%. Thirteen samples yielded invalid results. These data were not included in the totals.
Four clinical trial sites, in Texas and South Dakota, tested four hundred sixty one (461) nasal clinical specimens blindly and prospectively comparing the SASTM Influenza B Test and the SAS™ FluAlert A&B (combined) Test performance. The SASTM Influenza B test and SASTM FluAlert A&B Test had a positive percent agreement of 98.7% and a negative percent agreement of 99.7%.

{1}------------------------------------------------

:

{2}------------------------------------------------

Clinical sites collected the nasal wash samples from an approximately equal mix of adult (>21 years) and pediatric patients (0-21 years). The nasal aspirate samples were collected predominately from pediatric patients.

Fresh Prospective Nasal Aspirates:

SAS Influenza ATotal
SASA/BCombo+-
+44044
-1137138
Total45137182
Positive % Agreement: 97.8% (95% CI 87-100%)
Negative % Agreement: 100% (97% CI 95-100%)
SAS Influenza BTotal
SASA/BCombo+
+37037
0। 45145
Total37। 45182
Positive % Agreement: 100% (95%
CI 97-100%)
Negative % Agreement: 100% (95 % CI88-100%)

Fresh Prospective Nasal Washes:

SAS Influenza ATotal
+
SASA/B+ટર્ભl27
Combo1251252
Total27252279
Positive % Agreement: 96.3% (95% CI 79-100%)
Negative % Agreement: 99.6% (95%CI 97-100%)

{3}------------------------------------------------

SAS Influenza BTotal
SASA/BCombo+
+40141
1237238
Total41238279
Positive % Agreement: 97.6% (95% CI 86-100%)
Negative % Agreement: 99.6% (95% CI 97-100%)

Results Summary: Fresh Nasal Washes and Aspirates:

SAS Influenza ATotal
SASA/BCombo+70171
-2388390
Total72389461
Positive % Agreement: 97.2% (95% CI 89-100%)
Negative % Agreement: 99.7% (95% CI 98-100%)
SAS Influenza BTotal
SASA/BCombo+
+77178
1382383
Total78383461
Positive % Agreement: 98.7% (95% CI 92-100%)
Negative % Agreement: 99.7% (95% CI 98-100%)

Note: Performance characteristics for detecting the 2009 H1N1 influenza virus from human specimens have not been established

{4}------------------------------------------------

Demographics of Fresh Samples:

Age (years)Number of NasalWash specimens% of TotalNWSpecimensNumber of NasalAspirateSpecimens% of TotalNA Specimens
0-5165.7%7340.1%
6-21217.5%10859.3%
22 - 65259.0%10.6%
>6593.2%0
Not Determined20874.5%0
Total279182

Retrospective Study:

To supplement the prospective study, 191 frozen, archived, nasal wash samples from a children's hospital in Texas were blindly assayed comparing the individual SAS™ Influenza A Test and the individual SASTM Influenza B Test to the SAS™ FluAlert A&B (combined) Test. For these samples, the positive percent agreement with the Influenza A Test is 100% and negative percent agreement is 99.3%, while positive percent agreement with the Influenza B Test was 94.7% and negative percent agreement is 98.0%

SAS™ InfluenzaA Test
+-
SAS™ Flu Alert A&BCombined Test27128
-0163163
Total27164191
Positive % Agreement: 100% (95% CI 84-100%)

Negative % Agreement: 99.3% (95% CI 96-100%)

SASTM InfluenzaB Test
+-
SASTM FluAlert A & BCombined Test+36339
-2150152
Total38153191

Positive % Agreement: 94.7% (95% CI 81-99%) Negative % Agreement: 98.0% (95% CI 94-99%)

Reproducibility:

The reproducibility of the SASTM FluAlert A&B Test was evaluated at three clinical sites. Three or four non-professional users per site tested the SASTM FluAlert A&B Test against a panel of approximately 30 aliquots each of six (6) panel members over a two-week period. Specimens were comprised of pooled nasal aspirates and included two (2) levels of positives for influenza A and two (2) for influenza B and two (2) negatives. The low and medium positives for influenza A contained H3N2 A/Hong Kong/8/68 and the low and medium positives for influenza B contained B/Allen/45. Negative specimens for influenza A contained H3N2 A/Hong Kong/8/68 and negative specimens for influenza B contained B/Allen/45, but both were in concentrations below the

{5}------------------------------------------------

limit of detection. Although 30 aliquots of each were prepared, in some cases the total number was fewer than 30 because of spillage or pipetting errors.

Panel MemberInfluenzaAHighNegativeInfluenza ALow PositiveInfluenza AModeratePositiveInfluenzaBHighNegativeInfluenza BLow PositiveInfluenza BModeratePositive
Viral TiterFinalConcentrationTCID50/0.2 ml.$1.8 x 10^3$$7 x 10^3$$1.4 x 10^4$$2.8 x 10^2$$1.1 x 10^3$$2.2 x 10^3$
AgreementSite 129 Neg/3096.7%26 Pos/3086.6%29 Pos/3096.7%28 Neg/2996.6%29 Pos/3096.7%29 Pos/29100%
withExpectedResultSite 229 Neg/3096.7%26 Pos/2989.7%29 Pos/29100%30 Neg/30100%29 Pos/3096.7%30 Pos/30100%
Site 328 Neg/3093.3%27 Pos/3090.0%30 Pos/30100%26 Neg/2989.7%29 Pos/3096.7%30 Pos/30100%
TotalAgreement95.6%88.8%98.9%95.4%96.7%100%

Analytical Sensitivity (Limit of Detection):

The limit of detection (LOD) for the SAS™ FluAlert A&B Test was determined for five (5) each influenza A and Influenza B viral strains. Each strain was received from ATCC with a known TCID50 concentration. Each strain was serially diluted in SASTM FluAlert extraction buffer. Strains were assayed in triplicate using the SAS™ FluAlert A&B Test until no positive signal could be seen. Results are summarized below.

InfluenzaViral StrainATCCLoDTCID50/0.2 ml
H1N1 A/PR/3/34VR-95$1.2 x 10^5$
H3N2 A/Aichi/2/68VR-547$5.6 x 10^2$
H3N2 A/HongKong/8/6/8VR-544$3.5 x 10^3$
H1N1 A/FM/147VR-97$7.9 x 10^3$
H3N2A/Victoria/3/75VR-822$4.5 x 10^5$
Influenza BB/Lee/40VR-101$9.9 x 10^4$
Influenza BB/Allen/45VR-102$5.6 x 10^2$
Influenza BB/Mass/3/66VR-523$4.5 x 10^2$
Influenza BB/Taiwan/2/62VR-295$3.5 x 10^1$
Influenza BB/Maryland/1/59VR-296$1.6 x 10^2$

{6}------------------------------------------------

Cross Reactivity Study:

Twenty-two virus strains were obtained from ATCC or other commercial sources. Each cultured viral strain was tested on the SAS™ FluAlert A&B Test in these concentrations. The results are summarized below.

VirusATTC/Lot #Concentration"A" portion of theSASTM FluAlertA&B"B" portion of theSASTM FluAlertA&B
Adenovirus 510-198-000$1.2 \times 10^{10}$NegNeg
Adenovirus 7VR7$3.2 \times 10^{3}$ TCID50 /0.2 mlNegNeg
Adenovirus 10VR1087$3.2 \times 10^{3}$ TCID50 /0.2 mlNegNeg
CoxsackieA9VR186$3.2 \times 10^{2}$ TCID50 /0.2 mlNegNeg
CoxsackieB5VR185$3.2 \times 10^{6}$ TCID50 /0.2 mlNegNeg
Cytomegalovirus02130120 µg/mlNegNeg
Echovirus11VR1052NANegNeg
Echovirus3VR1040$1 \times 10^{4}$ TCID50 /0.2 mlNegNeg
Echovirus 6VR1044$3.2 \times 10^{6}$ TCID50 /0.2 mlNegNeg
HSV-12J3000015 µg/mlNegNeg
HSV-28J2950215 µg/mlNegNeg
Varicella zoster110209712 µg/mlNegNeg
Parainfluenza 1VR907$5.6 \times 10^{6}$ TCID50 /0.2 mlNegNeg
Parainfluenza 2VR92$1.8 \times 10^{5}$ TCID50 /0.2 mlNegNeg
Parainfluenza 3VR93$3.2 \times 10^{6}$ TCID50 /0.2 mlNegNeg
RSV LongVR26$0.1 \times 10^{5.5}$ TCID50 /0.2 mlNegNeg
RSV BVR1400$0.1 \times 10^{5.25}$ TCID50 /0.2 mlNegNeg
Influenza B AllenVR102$3.2 \times 10^{3}$ TCID50 /0.2 mlNegNeg
Influenza B LeeVR101$3.2 \times 10^{6}$ TCID50 /0.2 mlNegNeg
Influenza B MassVR523$1.8 \times 10^{3}$ TCID50 /0.2 mlNegNeg
Influenza B MarylandVR296$1 \times 10^{4}$ TCID50 /0.2 mlNegNeg
Influenza B TaiwanVR295$5.6 \times 10^{2}$ TCID50 /0.2 mlNegNeg
Influenza A (H1N1) PRVR95$1.8 \times 10^{4}$ TCID50 /0.2 mlNegNeg
Influenza A (H3N2) AichciVR547$1.8 \times 10^{6}$ TCID50 /0.2 mlNegNeg
Influenza A (H3N2) Hong KongVR544$5.6 \times 10^{4}$ TCID50 /0.2 mlNegNeg
Influenza A FMVR97$3.2 \times 10^{4}$ TCID50 /0.2 mlNegNeg
Influenza A (H3N2) VictoriaVR822$1.8 \times 10^{6}$ TCID50 /0.2 mlNegNeg

One yeast and fourteen bacterial strains were obtained from ATCC or other commercial sources. Each cultured bacterial or yeast strain was diluted to a concentration of 1 x 100 cfu/ml and tested on the SAS™ FluAlert A&B Test. The results are summarized below.

Bacteria or Yeast"A" portion of theSASTM FluAlertA&B"B" portion of theSASTM FluAlertA&B
Candida albicansNegNeg
Chlamydia trachomatisNegNeg
Corynebacterium diphtheriaeNegNeg
Haemophilus influenzaNegNeg

{7}------------------------------------------------

Klebsiella pneumoniaeNegNeg
Serratia marcescensNegNeg
Staphylococcus epidermidisNegNeg
Staphylococcus aureusNegNeg
Streptococcus sp gr ANegNeg
Streptococcus sp gr FNegNcg
Streptococcus sp gr GNegNeg
Streptococcus pneumoniaeNegNeg
Mycoplasma pneumoniaeNegNeg
Neisseria meningitidisNegNcg
Pseudomonas aeruginosaNegNeg

Interference Study:

The analytical specificity of the A portion of the SAS™ FluAlert A&B was evaluated by testing a panel of 22 viruses, 14 bacteria, and one yeast species which may be found in the respiratory tract. For the A portion of the test, Influenza Whole Virus Strain A/FM/147 (ATCC VR97) at a titer of 7.9 x 103 TCID30/0.2 ml. was added to viral cultures and viral antigens at the concentrations in the table below and bacterial and yeast cultures at concentrations of 1 x 108 cfu/ml. For the B portion of the test, Influenza Whole Virus Strain B/Mass/3/66 (ATCC VR523) at a titer of 3.2 x 10' TCIDsy/0.2 ml. was added to viral cultures in the concentrations in the table below and bacterial and yeast cultures at 1 x 10° cfu/ml.

Results are summarized below:

VirusATTC/Lot #Concentration"A" Portion ofthe SASTMFluAlertA&B"B" Portion ofthe SASTMFluAlertA&B
Adenovirus 510-198-000$1.2 x 10^{10}$ TCID50/0.2 mlPosPos
Adenovirus 7VR7$3.2 x 10^3$ TCID50/0.2 mlPosPos
Adenovirus 10VR1087$3.2 x 10^3$ TCID50/0.2 mlPosPos
CoxsackieA9VR186$3.2 x 10^2$ TCID50/0.2 mlPosPos
CoxsackieB5VR185$3.2 x 10^6$ TCID50/0.2 mlPosPos
Cytomegalovirus02130120 µg/mlPosPos
Echovirus11VR1052NAPosPos
Echovirus3VR1040$1 x 10^4$ TCID50/0.2 mlPosPos
Echovirus6VR1044$3.2 x 10^6$ TCID50/0.2 mlPosPos
HSV-12J3000015 µg/mlPosPos
HSV-28J2950215 µg/mlPosPos
Varicella zoster110209712 µg/mlPosPos
Parainfluenza 1VR907$5.6 x 10^6$ TCID50/0.2 mlPosPos
Parainfluenza 2VR92$1.8 x 10^5$ TCID50/0.2 mlPosPos
Parainfluenza 3VR93$3.2 x 10^6$ TCID50/0.2 mlPosPos
RSV LongVR26$0.1 x 10^{5.5}$ TCID50/0.2 mlPosPos
RSV BVR1400$0.1 x 10^{5.25}$ TCID50/0.2 mlPosPos
Influenza B AllenVR102$3.2 x 10^3$ TCID50/0.2 mlPos
Influenza B LeeVR101$3.2 x 10^6$ TCID50/0.2 mlPos
Influenza B MassVR523$1.8 x 10^3$ TCID50/0.2 mlPos
Influenza B MarylandVR296$1 x 10^4$ TCID50/0.2 mlPos
Influenza B TaiwanVR295$5.6 x 10^2$ TCID50/0.2 mlPos
Influenza A (H1N1) PRVR95$1.8 x 10^4$ TCID50/0.2 mlPos

{8}------------------------------------------------

Influenza A (H3N2) AichciVR547$1.8 x 10^6$ TCID50/0.2 mlPos
Influenza A (H3N2) HongKongVR544$5.6 x 10^4$ TCID50/0.2 mlPos
Influenza A FMVR97$3.2 x 10^4$ TCID50/0.2 mlPos
Influenza A (H3N2)VictoriaVR822$1.8 x 10^6$ TCID50/0.2 mlPos

One yeast and fourteen bacterial strains were obtained from ATCC or other commercial sources. Each cultured bacterial or yeast strain was diluted to a concentration of 1 x 10° cfu/ml and tested on the SAS™ FluAlert A&B Test. The results are summarized below.

Bacteria or Yeast"A" portion of theSAS™ FluAlertA&B"B" portion of theSAS™ FluAlertA&B
Candida albicansPosPos
Chlamydia trachomatisPosPos
Corynebacterium diphtheriaePosPos
Haemophilus influenzaPosPos
Klebsiella pneumoniaePosPos
Serratia marcescensPosPos
Staphylococcus epidermidisPosPos
Staphylococcus aureusPosPos
Streptococcus sp gr APosPos
Streptococcus sp gr FPosPos
Streptococcus sp gr GPosPos
Streptococcus pneumoniaePosPos
Mycoplasma pneumoniaePosPos
Neisseria meningitidisPosPos
Pseudomonas aeruginosaPosPos

Expected Values:

Influenza prevalence varies year to year, with the highest number of cases in the fall and winter months in the US. During the period of September 30, 2007 to April 5, 2008, prevalence in the US for both influenza A and influenza B was 18.5%, with 74% of those cases attributed to influenza A and 26% attributed to influenza B. For studies conducted on the SAS™ FluAlert A&B Test, during the 2007-2008 and 2008-2009 seasons, prevalence for fresh, prospective nasal washes and nasal aspirates was 15.1% for influenza A and 16.5% for influenza B.

Conclusion of Performance Data:

The performance agreement of all clinical samples for the predicate devices and the proposed new device, FluAlert A&B, is 98% for the positive samples for influenza A, and >99% for negative samples, and is > 97% for the positive samples for influenza B is and >99% for negative samples, indicating that the new device is substantially equivalent to the predicate devices

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three curved lines above a wavy base.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

JUL 2 3 2009

Robbi Perry Regulatory Affairs SA Scientific, Ltd. 4919 Golden Quail San Antonio, TX 78240

Re: K080380

Trade/Device Name: SASTM FluAlert A&B Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza virus serological reagents Regulatory Class: Class II Product Code: GNX Dated: June 2, 2009 Received: June 4, 2009

Dear Ms. Perry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{10}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Sally A. Ulricht, M.S., Ph.D.

A. Hoivat. M Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indication for Use

510(k) Number (if known): K080380

Device Name: SAS™ FluAlert A & B Test

Indication For Use:

SAS™ FluAlert A & B Test is a visual and rapid assay for the presumptive in-vitro qualitative detection of Influenza A and Influenza B viral nucleoprotein antigens from nasal washes and nasal aspirates of symptomatic patients. The test is not intended for the detection of Influenza Type C viral antigen. This test is for professional use only.

Negative results do not preclude infection with influenza A or influenza B and should not be used as the sole basis for treatment or other patient management decisions. It is recommended that negative results be confirmed by cell culture.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Uve Schuf
Division Sign Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

6080380510(k)_TACT Status Catalogues of Canada Career of Canada Career of Canada Compares of Canadian Compares of Canadian Compares of Canadian Compares of Canadian Compares of Canadian CoCHANDER CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CAnd States of Construction
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

§ 866.3328 Influenza virus antigen detection test system.

(a)
Identification. An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.