(72 days)
The Smith & Nephew PLLA/HA Screw is indicated for fixation of bone-tendonbone or soft tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.
The Smith & Nephew PLLA/HA Screw is an tapcred interference screw for fixation of bone-tendon-bone or soft tissue grafts during anterior or posterior cruciate ligament reconstruction surgery. The PLLA/HA Screw is a cannulated, sterile, single use bone screw, made of polylactic acid, a resorbable polymer, and hydroxylapatite, an inorganic ceramic.
The provided document is a 510(k) premarket notification for a medical device called the "PILLA/HA Screw". This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific acceptance criteria in the way a new drug or high-risk device might.
Therefore, the requested information regarding acceptance criteria, study details, and data provenance related to AI/software performance metrics (like sensitivity, specificity, MRMC studies, standalone performance) is not applicable to this document. The device is a bone fixation screw, not an AI or imaging diagnostic tool.
Instead, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" for this submission are focused on demonstrating substantial equivalence to predicate devices through product validation testing as described in Section H of the 510(k) summary.
Here's the breakdown of the information that is available in the provided text, reinterpreted in the context of a 510(k) submission for a physical medical device:
1. Table of Acceptance Criteria and Reported Device Performance
For this device, "acceptance criteria" are not reported as specific performance metrics like sensitivity or specificity. Instead, the acceptance criterion for a 510(k) submission is typically demonstrating substantial equivalence to a predicate device. The performance aspect focuses on physical and functional characteristics.
| Acceptance Criterion (Implicit for 510(k): Substantial Equivalence) | Reported Device Performance (Summary of Validation Testing) |
|---|---|
| Intended Use Equivalence: The device serves the same clinical purpose as predicate devices. | The PLLA/HA Screw is indicated for fixation of bone-tendon-bone or soft tissue grafts during ACL/PCL reconstruction, which is consistent with predicate devices. |
| Technological Characteristics Equivalence: The device has similar design and materials to predicate devices. | The PLLA/HA Screw is described as essentially identical to the HAPLA Interference Screw (K002274) and PLC Screw (K051310). It is a cannulated, sterile, single-use bone screw made of polylactic acid (resorbable polymer) and hydroxylapatite (inorganic ceramic), similar to its predicates. |
| Performance Data Demonstrates Equivalence: The device performs comparably to predicate devices in relevant tests. | "Results of product validation testing demonstrates substantial equivalence to the predicate devices; HAPLA Interference Screw (K002274) and the PLC Screw (K051310)." (K080358, page 1, Section H) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified in the provided text. For mechanical devices, testing typically involves a sufficient number of devices to demonstrate statistical significance for properties like strength, degradation, etc. This is not a clinical study involving human patients for "test sets" in the AI sense.
- Data Provenance: Not specified, but generally, validation testing for such devices would be conducted in a laboratory setting by the manufacturer, rather than being clinical data from a specific country. This is primarily in-vitro or ex-vivo mechanical testing.
- Retrospective/Prospective: Not applicable in the context of device validation testing summarized here.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable. For a bone screw, "ground truth" is established through engineering and material science standards (e.g., tensile strength, screw-out torque, degradation rates), not expert consensus on image interpretation or clinical outcomes in the same way an AI diagnostic tool would.
- Qualifications of Experts: N/A.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. This concept is relevant for clinical studies or expert reviews of data, not for direct device performance testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No, an MRMC study was not done. This type of study is specific to diagnostic imaging devices where multiple readers evaluate cases. The PILLA/HA Screw is a surgical implant.
- Effect Size of AI Improvement: Not applicable.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: Not applicable. This concept applies to AI algorithms. The PILLA/HA screw is a physical orthopedic implant.
7. The Type of Ground Truth Used
- Type of Ground Truth: Not explicitly stated as "ground truth" in the AI sense. However, for device validation, the "ground truth" would be established by engineering specifications, material standards, and biomechanical testing protocols (e.g., ISO standards for implants, ASTM standards for materials) against which the device's performance (e.g., mechanical strength, degradation properties, fixation stability) is compared to demonstrate equivalence to predicate devices. This is typically done through in vitro and ex vivo testing.
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. The PILLA/HA Screw is a physical medical device, not a software algorithm that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable.
{0}------------------------------------------------
page 1/2
SECTION IV 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
APR 2 3 2008
As required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon
which the substantial equivalence is based.
PILLA/HA Screw
Date Prepared: February 8, 2008
A. Submitter's Name:
Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road Andover, MA 01810
B. Company Contact
Deana Boushell Principal Regulatory Specialist T 508 337 4036 F 508 261 3620
C. Device Name
| Trade Name: | PLLA/HA Screw |
|---|---|
| Common Name: | Screw, Fixation, Bone |
| Classification Name: | Smooth or threaded metallic bone fixation fastener |
D. Predicate Devices
The Smith & Nephew PLLA/HA Screw is substantially equivalent in Intended Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution:
E. Description of Device
The Smith & Nephew PLLA/HA Screw is an tapcred interference screw for fixation of bone-tendon-bone or soft tissue grafts during anterior or posterior cruciate ligament reconstruction surgery.
{1}------------------------------------------------
page of
F. Intended Üse
The Smith & Nephew PLLA/HA Screw is indicated for fixation of bone-tendon-bone or soft tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.
G. Comparison of Technological Characteristics
The Smith & Nephew PLLA/HA Screw is essentially identical to the HAPLA Interference Screw and the PLC Screw described in K002274 and K051310 respectively. The PLLA/HA Screw is a cannulated, sterile, single use bone screw, made of polylactic acid, a resorbable polymer, and hydroxylapatite, an inorganic ceramic.
H. Summary Performance Data
Results of product validation testing demonstrates substantial equivalence to the predicate devices; HAPLA Interference Screw (K002274) and the PLC Screw (K051310).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing service to the nation. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 3 2008
Smith & Nephew, Inc. Endoscopy Division c/o Ms. Deana Boushell Principal Regulatory Specialist 150 Minuteman Road Andover, MA 01810
K080358 Re:
Trade/Device Name: PLLA/HA Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: February 8, 2008 Received: February 11, 2008
Dear Ms. Boushell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, areresters, while Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r loase be action and a determination that your device complies with other requirements of the Act that I Dri has intace a and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI IC I art 007), accums (21 CFR Part 820); and if applicable, the electronic forul in also quany 25thons (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -- Ms. Deana Boushell
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to (2 roy 27 et notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use 510(k) Number (if known): ¥080] 58
Device Name: PLLA/HA Screw
Indications for Use:
The Smith & Nephew PLLA/HA Screw is indicated for fixation of bone-tendonbone or soft tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.
Over-The-Counter Use Prescription Use _ X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogle Axman
(Division Sign-C Division of General, Restorative, and Neurological Devices
510(k) Number K080358
N/A