(77 days)
The Derma Sciences Medihoney Dressings with Active Manuka Honey provide a moist environment conducive to wound healing and is indicated for the management of light to moderately exuding wounds such as:
- . diabetic foot ulcers
- leg ulcers (venous stasis ulcers, arterial ulcers and leg ulcers of mixed . etiology)
- . pressure ulcers / sores (partial and full thickness)
- 1st and 2nd degree partial thickness burns ◆
- . donor sites, and traumatic and surgical wounds.
Derma Sciences Medihoney Dressing with Active Manuka Honey are sterile, wound care dressings for use in moist wound management. The Derma Sciences Medihoney Dressing with Active Manuka Honey are offered in several sizes including .5, 1, and 1.5 oz. sizes.
The proposed Medihoney Dressing with Active Manuka Honey contains Active Manuka Honey. As wound exudate is absorbed, the alginate forms a gel, which assists in maintaining a moist environment that aids supports the autolytic debridement for optimal wound healing.
The provided text is a 510(k) Summary for Derma Sciences Medihoney Dressings with Active Manuka Honey. It describes the device, its intended use, and claims substantial equivalence to predicate devices. However, the document does not contain a study that proves the device meets specific acceptance criteria in the way a clinical trial or performance study typically would. Instead, it refers to biocompatibility testing.
Here's an attempt to answer your questions based only on the provided text, acknowledging where information is missing:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state numerical acceptance criteria for performance in the context of efficacy for wound healing. It focuses on biocompatibility.
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Biocompatibility for Intended Use: | The biocompatibility testing performed (sensitization, irritation, cytotoxicity, implantation) supports the device's safe use. |
| Sensitization (no significant reaction) | Successfully performed. |
| Irritation (non-irritant or minimal irritant) | Successfully performed. Macroscopic reaction was not significant compared to control; slight compared to negative control. Microscopically, classified as non-irritant compared to control, slight irritant compared to negative control. |
| Cytotoxicity (expected reaction due to osmotic effect) | A moderate cytotoxic reaction was observed using 100% honey dressings. This is expected due to the osmotic effect of honey on cells in culture (maximum tolerance ~2% honey concentration for cells in culture). The document explicitly states: "However, neither sugar nor honey used on open wounds causes any adverse effects." Implantation testing was initiated to further support safe use given this in-vitro finding. |
| Substantial Equivalence (to predicate devices K072956 & K053095): | The device is considered "essentially identical" to the predicate devices, with differences limited to a "slight formulation change" (100% Manuka Honey vs. 95% Manuka Honey and 5% calcium alginate). This change "does not represent a significant change in technological characteristics" and does not "affect the safety and effectiveness of the device." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "biocompatibility testing." While it refers to "implantation testing," specifics such as the number of subjects (animal or human), data provenance (country, retrospective/prospective), or specific sample sizes for sensitization, irritation, or cytotoxicity tests are not provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided. The biocompatibility tests are laboratory-based and likely follow standardized protocols for assessment, rather than relying on expert consensus for "ground truth" in the diagnostic sense.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessments, which are not detailed here for the biocompatibility tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was mentioned. This 510(k) summary is for a physical wound dressing, not an AI-powered diagnostic device, so an MRMC study is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a wound dressing, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the biocompatibility testing:
- Sensitization and Irritation: Likely based on standardized grading scales of physiological reactions.
- Cytotoxicity: Measured in vitro using cell culture reactions.
- Implantation testing: Macroscopic and microscopic assessment of tissue reaction, likely against established criteria for irritancy. This is not "ground truth" in the sense of clinical outcomes, but rather a safety assessment.
There is no mention of clinical outcomes data or pathology results specifically for proving the efficacy of this new formulation. The device relies on substantial equivalence primarily due to minor formulation changes and the established safety of manuka honey as a wound dressing.
8. The sample size for the training set
Not applicable. There is no "training set" as this is not an AI/machine learning device.
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
page 1 of 3
510(k) Summary for Derma Sciences Medihoney Dressings APR 2 3 2008 with Active Manuka Honey
1. SPONSOR
Derma Sciences 214 Carnegie Center Suite 100 Princeton, NJ 08540
Contact Person: Barry Wolfenson Derma Sciences 214 Carnegie Center, Suite 100 Princeton, NJ 08540 Telephone: 609.514.4744 x1103
April 2, 2008 Date Prepared:
2. DEVICE NAME
| Proprietary Name: | Derma Sciences Medihoney Dressing with Active ManukaHoney |
|---|---|
| Common/Usual Name: | Wound Dressing |
| Classification Name: | Dressing |
3. PREDICATE DEVICES
- Medihoney Primary Wound Dressings with Active Manuka Honey (K072956) �
- . API-MED Active Manuka Honey Wound Dressings (K053095)
4. DEVICE DESCRIPTION
Derma Sciences Medihoney Dressing with Active Manuka Honey are sterile, wound care dressings for use in moist wound management. The Derma Sciences Medihoney Dressing with Active Manuka Honey are offered in several sizes including .5, 1, and 1.5 oz. sizes.
{1}------------------------------------------------
$$
\begin{array}{c}
\rightsquigarrow\rightsquigarrow\rightsquigarrow\rightsquigarrow\rightsquigarrow\rightsquigarrow\rightsquigarrow\rightsquigarrow\rightsquigarrow\r\pdash\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag\rash\pag$$
The proposed Medihoney Dressing with Active Manuka Honey contains Active Manuka Honey. As wound exudate is absorbed, the alginate forms a gel, which assists in maintaining a moist environment that aids supports the autolytic debridement for optimal wound healing.
5. INTENDED USE
The Derma Sciences Honey Wound Dressings provide a moist environment conducive to wound healing and is indicated for the management of light to moderately wounds such as:
- diabetic foot ulcers ●
- leg ulcers (venous stasis ulcers, arterial ulcers and leg ulcers of mixed etiology) .
- pressure ulcers / sores (partial and full thickness) �
- 1st and 2nd degree partial thickness burns .
- donor sites, and traumatic and surgical wounds. .
6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The Derma Sciences Honey Wound Dressings are essentially identical to the parent Medihoney Primary Wound Dressings with Active Manuka Honey subject of K072956 and API-MED Active Manuka Honey Wound Dressings subject of K053095. The differences between the Derma Sciences Medihoney Dressings with Active Manuka Honey and the predicate devices are limited to a slight formulation change.
The technological characteristics of the Derma Sciences Medihoney Dressings with Active Manuka Honey, the parent Medihoney Primary Dressings and the API-MED Active Manuka Honey Wound Dressings are substantially equivalent in that they are all dressings suitable for use on pressure sores, leg ulcers, post-operative wounds, superficial wounds and abrasions.
The modifications made to the Medihoney Primary Dressings with Active Manuka Honey to produce the Derma Sciences Medihoney Dressings with Active Manuka Honey are limited to a slight change in formulation. The proposed Medihoney Honey Wound Dressings are comprised of 100% w/w of Active Manuka Honey and the predicate API-MED Active Manuka Honey Dressings are comprised of 95% Active Manuka Honey and 5% calcium alginate. This slight change in formulation simply provides the user with a wider variety of honey dressings and does not represent a
{2}------------------------------------------------
K080315
page 3 of 3
significant change in technological characteristics of the Medihoney Primary Dressings with Active Manuka Honey. The only differences between the Derma Sciences Medihoney Dressing with Active Manuka Honey and the predicate devices include slightly different dressing composition which are minor and do not affect the safety and effectiveness of the device.
7. PERFORMANCE TESTING
Biocompatibility testing performed to support the formulation change for the modified dressings demonstrates that the Medihoney Dressing with Active Manuka Honey are safe for their intended use. Sensitization, and irritation testing was performed successfully using the Derma Sciences Wound Dressings. Cytotoxicity Testing using the 100% honey Medihoney Dressings demonstrated a moderate cytotoxic reaction. This reaction is expected since in research work performed, it has been found that the maximum concentration of honey that can be tolerated using cells in culture is about 2%. This is due to the osmotic effect which is the same effect as would be seen in a solution of sugars as in honey. However, neither sugar nor honey used on open wounds causes any adverse effects. In anticipation of this reaction using the Medihoney Wound dressings, Derma Sciences has initiated implantation testing to support the safe use of these dressings. This testing showed that the macroscopic reaction was not significant as compared to the control material and slight compared to the negative control implant material. Microscopically, the Medihoney Dressing was classified as a non irritant as compared to the control article and a slight irritant compared to the negative control article.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The eagle is depicted in a simple, black and white design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 3 2008
Derma Sciences, Inc. % Medical Device Consultants, Inc. Mary McNamara-Cullinane, RAC Senior Regulatory Consultant 49 Plain Street North Attleboro, Massachusetts 02760
Re: K080315
Trade/Device Name: Derma Sciences Medihoney Dressings with Active Manuka Honey Regulatory Class: Unclassified Product Code: FRO Dated: April 4, 2008 Received: April 7, 2008
Dear Mr. McNamara-Cullinane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Mary McNamara-Cullinane, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
K080315 510(k) Number (if known):
Device Name: Derma Sciences Medihoney Dressings with Active Manuka Honey
Indications for Use:
The Derma Sciences Medihoney Dressings with Active Manuka Honey provide a moist environment conducive to wound healing and is indicated for the management of light to moderately exuding wounds such as:
- . diabetic foot ulcers
- leg ulcers (venous stasis ulcers, arterial ulcers and leg ulcers of mixed . etiology)
- . pressure ulcers / sores (partial and full thickness)
- 1st and 2nd degree partial thickness burns ◆
- . donor sites, and traumatic and surgical wounds.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Ogle for mxm
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_KO US ڪا
N/A