K Number
K080254

Validate with FDA (Live)

Date Cleared
2008-10-31

(274 days)

Product Code
Regulation Number
864.1860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ScanScope® System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.

The ScanScope® system is intended for use as an aid to the pathologist in the detection and quantitative measurement of PR (Progesterone Receptor) by manual examination of the digital slide of formalin-fixed, paraffin-embedded normal and neoplastic tissue immunohistochemically stained for PR on a computer monitor.

It is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer.

Device Description

The system comprises a ScanScope® XT digital slide scanner instrument and a computer system executing Spectrum " software. The system capabilitics include digitizing microscope slides at diagnostic resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing facilities for annotating digital slides and entering and editing metadata associated with digital slides, and facilities for image analysis of digital slides, including the ability to quantify characteristics useful to Pathologists, such as measuring and scoring immunohistochemical stains applied to histology specimens, such as Dako PR, which reveal the presence of PR (Progesterone Receptor) protein expression, which may be used to determine patient treatment for breast cancer.

AI/ML Overview

The provided document, K080254, describes the Aperio ScanScope® XT System, intended for in vitro diagnostic use as an aid to pathologists in the display, detection, counting, and classification of tissues and cells of clinical interest. Specifically, the document focuses on its use for the detection and quantitative measurement of Progesterone Receptor (PR) by manual examination of digital slides.

Here's an analysis based on the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly based on the agreement between manual microscopy and manual reading of digital slides, and the reproducibility of the digital slide system. The document does not explicitly state pre-defined numerical acceptance criteria (e.g., "agreement must be >X%"). Instead, it reports the range of observed agreements and precision metrics.

MetricAcceptance Criteria (Implicit)Reported Device Performance (ScanScope® XT System)
Clinical Performance (Pathologist Agreement)
Inter-Pathologist Agreement (Digital Slides) - % Positive NucleiN/A (Comparison study to conventional microscopy for substantial equivalence)76.3% - 98.0%
Inter-Pathologist Agreement (Manual Microscopy) - % Positive NucleiN/A83.8% - 99.0%
Agreement between Manual Microscopy & Digital Slides - % Positive NucleiN/A (Demonstrate substantial equivalence)78.8% - 100.0%
Inter-Pathologist Agreement (Digital Slides) - Intensity ScoreN/A58.8% - 78.0%
Inter-Pathologist Agreement (Manual Microscopy) - Intensity ScoreN/A58.8% - 88.0%
Agreement between Manual Microscopy & Digital Slides - Intensity ScoreN/A62.5% - 96.0%
Analytical Performance (Reproducibility via Image Analysis Algorithm)
Intra-system (10 scans) - % Positive Nuclei Standard DeviationN/A (Demonstrate precision/reproducibility)Overall SD: 0.54% (max 1.47%), Average Range: 1.06% (max 4.78%)
Intra-system (10 scans) - Intensity Values Standard DeviationN/AOverall SD: 0.9 (max 1.60), Average Range: 2.48 (max 4.27)
Inter-day/Intra-system (20 scans) - % Positive Nuclei Standard DeviationN/AOverall SD: 0.54% (max 1.09%), Average Range: 1.52% (max 3.90%)
Inter-day/Intra-system (20 scans) - Intensity Values Standard DeviationN/AOverall SD: 1.44 (max 2.43), Average Range: 5.29 (max 11.39)
Inter-system (3 systems, 10 scans each) - % Positive Nuclei Standard DeviationN/AIndividual System Average SD: 0.54%, 0.53%, 0.75% (max 1.47%, 1.23%, 2.05%) Combined Systems Overall Average SD: 0.87% (max 1.57%)
Inter-system (3 systems, 10 scans each) - Intensity Values Standard DeviationN/AIndividual System Average SD: 0.9%, 1.01%, 0.93% (max 1.60%, 1.48%) Combined Systems Overall Average SD: 1.35% (max 2.03%)
Intra-pathologist (5 reads) - % Positive Nuclei Standard Deviation (Manual Microscopy)N/AOverall Average SD: 6.73% (max 16.73%), Average Range: 9.8% (max 40%)
Intra-pathologist (5 reads) - % Positive Nuclei Standard Deviation (Digital Slides)N/AOverall Average SD: 11.81% (max 28.72%), Average Range: 16.2% (max 75%)
Intra-pathologist (5 reads) - Intensity Score Outliers (Manual Microscopy)N/A8 outliers out of 50 scores (16%)
Intra-pathologist (5 reads) - Intensity Score Outliers (Digital Slides)N/A9 outliers out of 50 scores (18%)
Inter-pathologist - % Positive Nuclei Standard Deviation (Manual Microscopy)N/AOverall Average SD: 13.30% (max 32.15%), Average Range: 17.2% (max 60%)
Inter-pathologist - % Positive Nuclei Standard Deviation (Digital Slides)N/AOverall Average SD: 11.3% (max 20.82%), Average Range: 16.0% (max 40%)
Inter-pathologist - Intensity Score Outliers (Manual Microscopy)N/A7 outliers out of 30 scores (23%)
Inter-pathologist - Intensity Score Outliers (Digital Slides)N/A7 outliers out of 30 scores (23%)

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: 180 formalin-fixed, paraffin-embedded breast tissue specimens.
    • Clinical Site 1: 80 slides.
    • Clinical Site 2: 100 slides.
  • Data Provenance: The study was conducted at two Clinical Laboratory Improvement Amendments (CLIA) qualified clinical sites. The specimens were immunohistochemically stained at these clinical sites using Dako IVD FDA cleared reagents. Glass slides were prepared in the sites' clinical laboratories.
    • Clinical Site 1: Specimens selected based on existing clinical scores to provide an equal distribution of PR slides across different percentage positive nuclei ranges (0%, 1-9%, 10-49%, 50-100%).
    • Clinical Site 2: Routine specimens taken from clinical operations, representing a typical clinical setting.
  • Retrospective/Prospective: Not explicitly stated as strictly retrospective or prospective. The specimens were "from both clinical sites" and for Site 1, "selected based on their clinical scores on file," suggesting a retrospective selection of cases. For Site 2, "routine specimens taken from their clinical operation" could imply concurrent collection or a recent retrospective selection. The reading of these slides by pathologists for the study itself was a prospective activity within the study design.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Three different board-certified pathologists at each of the two clinical sites were used. This totals 6 pathologists for the initial manual reads (3 at site 1, 3 at site 2). Later, for the digital reads, the "same three Pathologists at each clinical site" performed the digital read. It's unclear if the same 6 pathologists or 3 total pathologists across both sites participated in all stages. However, for ground truth generation, it was 3 pathologists per site.
  • Qualifications of Experts: "Board-certified staff pathologists."

4. Adjudication Method for the Test Set

The "ground truth" for the comparison study was established by the consensus or average of the three pathologists' manual microscopy reads. The statistical analyses were presented for each of the scores (percentage of positive nuclei and intensity scores) and "comparatively between the two methods for the clinical sites with their different three pathologists." This implies a form of expert consensus, where the average/distribution of their reads served as the reference for comparison, rather than a formal adjudication to a single "correct" answer by an independent panel.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? Yes, a comparative study was conducted where three pathologists at each of two clinical sites read 180 cases by traditional manual microscopy and then later (after a wash-out period and randomization) read the digital slides of the same cases on a computer monitor. This fits the description of a multi-reader, multi-case comparison study.
  • Effect size of human readers improve with AI vs without AI assistance: This study does not describe AI assistance for human readers. The ScanScope® XT System is described as a digital slide creation, management, viewing, and analysis system intended as an aid to the pathologist by manual examination of digital slides. The analytical performance section mentions an "image analysis algorithm," but this algorithm was used for precision/reproducibility studies of the system itself, not to assist pathologists in their interpretation of diagnostic cases. The clinical comparison study directly compares manual microscopy performance to human pathologists reading digital slides visually. Therefore, an effect size of human readers with AI vs without AI assistance is not reported because the clinical study did not involve AI assistance for the pathologists.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Was a standalone study done? No, a standalone study demonstrating the algorithm's diagnostic performance without human-in-the-loop was not performed or reported for its intended diagnostic use.
    • An "image analysis algorithm" was used in the analytical performance (precision/reproducibility) section to quantify cell features and scoring schemes objectively. This algorithm reported percentage of positive nuclei and intensity scores for system assessment, not for diagnostic claims for the algorithm itself. It was used to understand scanning variability, not to replace or assist a pathologist's diagnosis.

7. Type of Ground Truth Used

The ground truth for the comparison study (clinical performance) was established by expert consensus based on the average/distribution of manual microscopy readings from three board-certified pathologists for each slide. The Dako IVD FDA cleared Monoclonal Mouse Anti-Human Progesterone Receptor (Clone PgR 636) (K020023) was used for immunohistochemical staining, providing a standardized basis for the pathologists' assessment.

8. Sample Size for the Training Set

The document does not mention a training set for the ScanScope® XT System itself, as it is a digital slide scanner and management system, with the focus of the clinical study being on the equivalence of manual pathologist interpretation of digital slides compared to glass slides. The "image analysis algorithm" used in the analytical performance section is not presented as a component needing a separate training set for diagnostic purposes described here.

9. How the Ground Truth for the Training Set Was Established

As no training set is described for a diagnostic algorithm, the method for establishing its ground truth is not applicable in this context. The clinical study focuses on establishing the equivalence of the digital viewing method to conventional microscopy, with the human pathologist remaining the primary interpreter.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" in black, with a white line running through it. To the right of the symbol is the word "aperio" in a sans-serif font. The word is in a light gray color.

K080254

510(k) Summary of Substantial Equivalence Aperio Technologies, Inc. (ScanScope® XT System)

OCT 3 1 2008

21 CFR 807.92(a):

21 CFR 807.92(a) (1):

Submitter's name and address:

Aperio Technologies, Inc. 1360 Park Center Drive Vista, CA 92081

Submitter's telephone and fax numbers:

Phone:(760) 539-1114
Fax:(760) 539-1116

Contact person:

Jeff Ryberg Director of Quality, Regulatory and Clinical Aperio Technologies, Inc. 1360 Park Center Drive Vista, CA 92081 jryberg@apcrio.com

Date this 510(k) summary was prepared:

October 23, 2008

21 CFR 807.92(a)(2):

Trade Name of Device:ScanScope® XT System
Regulatory Section:21 CFR 864.1860 Immunhistochemistery reagents and kits
Classification:Class II
Product Code:OEO (automated digital image manual interpretation microscope)

510k Summary - ScanScope® System - Rev23Oct19Sep08

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" in black, followed by the word "aperio" in a light gray, sans-serif font. The "O" is an oval shape with a white line running through it.

21 CFR 807.92(a)(3): Legally marketed predicate device to which substantial equivalence is claimed:

Predicate Device:ScanScope® XT System
Manufacturer:Aperio Technologies, Inc.
Predicate Device k#:K071671
21 CFR 807.92(a)(4):Description of the device that is the subject of this premarket

notification :

System: The system comprises a ScanScope® XT digital slide scanner instrument and a computer system executing Spectrum " software. The system capabilitics include digitizing microscope slides at diagnostic resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing facilities for annotating digital slides and entering and editing metadata associated with digital slides, and facilities for image analysis of digital slides, including the ability to quantify characteristics useful to Pathologists, such as measuring and scoring immunohistochemical stains applied to histology specimens, such as Dako PR, which reveal the presence of PR (Progesterone Receptor) protein expression, which may be used to determine patient treatment for breast cancer.

Hardware Operation: The ScanScope XT digital slide scanner creates seamless true color digital slide images of entire glass slides in a matter of minutes. A high numeric aperture 20x, as found on conventional microscopes, is used to produce high-quality images. (When the 2X magnification changer is inserted, the effective magnification of the images is 40X.) The ScanScope XT employs a linear-array scanning technique that generates images free from optical aberrations along the scanning axis. The result is digital slide images that have no tiling artifacts and are seamless.

Software Operation: The Spectrum software is a full-featured digital pathology management system. The software runs on a server computer called a Digital Slide Repository (DSR), which stores digital slide images on disk storage such as a RAID array, and which hosts an SQL database that contains digital slide metadata. Spectrum includes a web application and services which encapsulate database and digital slide image access for other computers. The Spectrum server supports the capability of running a variety of image analysis algorithms on digital slides, and storing the results of analysis into the database. Spectrum also includes support for locally or remotely connected image workstation computers, which run digital slide vicwing and analysis software provided as part of Spectrum.

Overview of System Operation: The laboratory technician or operator loads glass microscope slides into a specially designed slide carrier with a capacity of up to 120 slides. The scanning process begins when the operator starts the ScanScope scanner and finishes when the scamer has completed scaming of all loaded slides. As each glass slide is processed, the system automatically stores individual "striped" images of the tissue contained on the glass slide and integrates the striped images into a single digital

510k Summary - ScanScope® System - Rev23Oct19Sep08

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" in black, followed by the word "aperio" in a lighter color. The letter "O" is an oval shape with a white line running through it.

slide image, which represents a histological reconstruction of the entire tissue section. After scanning is completed, the operator is able to view and perform certain analytical tests on the digital slides.

21 CFR 807.92(a)(5): Intended use and labeled indications for use:

The ScanScope® System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.

The IHC PR Manual Read of Digital Slides application is intended for use as an aid to the pathologist in the detection and quantitative measurement of PR (Progesterone Receptor) by manual examination of the digital slide of formalin-fixed, paraffin-embedded normal and neoplastic tissue immunohistochemically stained for PR on a computer monitor.

It is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of brcast cancer.

21 CFR 807.92(a)(6): Technological characteristics:

The design, construction, energy source and other characteristics of the ScanScope System candidate device are considered to be substantially equivalent to the relevant features of the predicate device. A summary of the technological characteristics of the ScanScope System candidate device in comparison to the predicate device follows:

Method of cell detection. The method of cell detection is by colorimetric and morphometric pattern recognition by microscopic examination of prepared cells by size, shape, color, and intensity as observed by a computer-automated, microscopic digital slide scanner system and/or by visual observation by a health care professional.

System Components. The system components comprising the ScanScope System candidate device are substantially equivalent to those in the predicate device; i.e., a computer-automated digital microscope slide scanner, computer, color monitor, and keyboard.

Energy Source. The electrical service is 100vAC -- 240vAC. 50Hz/60 Hz. 2 amps which is similar to the predicate device electrical service requirements.

21 CFR 807.92(b): 510(k) summaries for those premarket submissions in which determination of substantial equivalence is also based on an assessment of performance data shall contain the following information:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" in black, with a white line running through it. To the right of the logo is the word "aperio" in a sans-serif font. The word is in a lighter color than the logo, possibly gray or a faded black.

21 CFR 807.92(b)(1): Brief discussion of non-clinical tests submitted, referenced or relied on in this premarket notification:

There are no non-clinical tests submitted, referenced or relied on in this submission.

21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:

Comparison studies:

a. Method comparison with predicate device:

The substantial equivalence study was based on comparison of reading digital slides on a computer monitor to reading glass slides using conventional microscopy. Manual microscopy and the manual reading of digital slides on a computer monitor were performed in accordance with the reagent vendor's instructions for use.

Two Clinical Laboratory Improvement Amendments (CLIA) qualified clinical sites participated in the study. Prior to their participation in the study each clinical site obtained exemption status from an Institutional Review Board (IRB).

A total set of 180 formalin-fixed, paraffin-embedded breast tissue specimens from both clinical sites were used for the PR study; 80 slides from the first clinical site and 100 slides from the second clinical site.

The specimens at the first clinical site were selected based on their clinical scores on file to provide an equal distribution of PR slides in the percentage of positive nuclei ranges 0%, 1% to 9%, 10% to 49%, and 50% to 100%. The specimens at the second clinical site were routine specimens taken from their clinical operation, representing the true target population of cases in a typical clinical setting.

All specimens for the PR study were immunohistochemically stained at the clinical sites using Dako in vitro diagnostic (IVD) FDA cleared Monoclonal Mouse Anti-Human Progesterone Receptor (Clone PgR 636) (K020023).

The study was performed primarily at the participating clinical sites and all parts except the scanning of glass slides were performed at their facilities using their typical workflow. The glass slides were propared in the sites' clinical laboratories and read by board certified staff pathologists. For the scanning of glass slides ScanScope XT instruments were operated in a simulated clinical setting at Aperio (designed to be representative of a typical lab environment).

All ScanScope XT instruments used in the study were production units and were delivered, installed, and maintained in accordance with the approved procedures, per Aperio's OSPs (Ouality Systems Procedures), and as described in product documentation and labeling.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" that is black and bolded, with a white line running through it. To the right of the symbol is the word "aperio" in a smaller, sans-serif font. The word is in a lighter color, possibly gray or a faded black.

Three different board-certified pathologists at each clinical site performed a blinded manual review of each glass slide using a conventional light microscope. The pathologists reported the percentage of positive nuclei [0%, 1%, ... 100%] and average intensity score of 0, 1+, 2+ or 3+ for each of the reviewed glass slides.

Based on the manual microscopy average percentages of positive nuclei from the three pathologists, the glass slides used for the PR study provided the following percentages of positive nuclei distribution.

PercentageClinical Site 1Clinical Site 2Total
0%293362
[ 1%- 5%)12618
[ 5%-10%)8311
[10%-50%)151126
[50%-100%]164763
Total80100180

PR Percentage of Positive Nuclei Distributions.

Based on the manual microscopy average intensity scores from the three pathologists, the glass slides used for the PR study provided the following average intensity score distribution.

Intensity ScoreClinical Site 1Clinical Site 2Total
0263157
1+14317
2+201232
3+205474
Total80100180

PR Average Intensity Score Distributions.

As it can be seen from the PR percentage of positive nuclei distribution, it was not possible to obtain an equal distribution of the percentage of positive nuclei in the range from 1% to 10%. This difficulty was founded in the limited representation of this percentage range in the true target population of cases.

All glass slides were scanned using a different ScanScope XT instrument for each clinical site,

After a wash-out period of over one week and subsequent randomization of the same three Pathologists at each clinical site performed a blinded manual read of each digital slide displayed on a computer monitor using the ScanScope XT System's remote viewing capability. The pathologists reported the percentage of positive nuclei [0%, 1%, ... 100%] and average intensity score of 0, 1+, 2+ or 3+ for each of the reviewed digital slides.

The statistical analyses are presented for each of the scores: percentage of positive nuclei and intensity scores. The statistical analyses are presented across all slides for manual microscopy and manual reading of digital slides on a computer monitor, and comparatively between the two methods for the clinical sites with their different three pathologists.

510k Summary - ScanScope® System - Rev23Oct19Sep08

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized oval shape with a line running through it, followed by the word "aperio" in lowercase letters. The oval shape is black, while the word "aperio" is in a lighter color, possibly gray or a faded black.

Percentage of Positive Nuclei - Progesterone Receptor (PR)

The inter-pathologist agreements for reading digital slides were in the range of 76.3%-98.0% and the inter-pathologist agreements for manual microscopy were in the range of 83.8%-99.0%.

The agreements between the pathologists' manual microscopy and reading digital slides were in the range of 78.8%-100.0% and the inter-pathologist agreements for manual microscopy were in the range of 83.8%-99.0%

Intensity Score - Progesterone Receptor (PR)

The inter-pathologist agreements for reading digital slides were in the range of 58.8%-78.0% and the inter-pathologist agreements for manual microscopy were in the range of 58.8%-88.0%.

The agreements between the pathologists' manual microscopy and reading digital slides were in the range of 62.5%-96.0% and the inter-pathologists agreements for manual microscopy were in the range of 58.8%-88.0%.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized, dark oval shape with a white, curved line running through it, resembling a stylized letter 'O'. To the right of the oval is the word "aperio" in lowercase letters, with a slightly faded or textured appearance.

Analytical Performance:

a. Precision/Reproducibility:

The precision of the ScanScope XT System was determined in a suite of intra-system, inter-day/intrasystem, inter-system, intra-pathologist and inter-pathologist studies.

To measure the precision and reproducibility of the system for the manual reading of digital slides, an image analysis algorithm was used that detects and quantifies the same cell features and uses the same scoring scheme as the pathologists. This method allowed the quality of the digital slides to be quantified objectively in terms of their interpretability for PR scoring, and enabled detection of small variations of the system humans would not be able to detect.

10 PR slides from the comparison study were used for this study. The PR slides consisted of formalinfixed, paraffin-embedded breast tissue specimens immunohistochemically stained using Dako in vitro diagnostic (IVD) FDA cleared Monoclonal Mouse Anti-Human Progesterone Receptor (Clone PgR 636) (K020023).

The pathologists' selection of tumor regions for image analysis introduces some variability to the system. To properly assess the true variability of the system the influence of the pathologists' selections in the intra- system and inter-systems studies was eliminated by using the same tumor regions for image analysis of all scans of the same slide.

The image analysis algorithm reported the percentage of positive nuclei [0.0%, ... 100.0%] and average intensity score of 0, 1+, 2+, or 3+ as well as the underlying average intensity on a scale from 0 to 255.

The statistical analyses are presented for the percentage of positive nuclei and intensity values.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for "aperio". The logo consists of a stylized letter "O" in black, with a horizontal line cutting through the top portion of the oval shape. To the right of the symbol is the word "aperio" in a lowercase, sans-serif font. The text is in a lighter shade, giving it a softer appearance compared to the bold, black symbol.

Intra- system: The slide scores provided by image analysis over 10 consecutive scans were analyzed for all 10 PR slides.

Percentage of Positive Nuclei

The image analysis results show an overall standard deviation of 0.54% (maximum 1.47%) and average range (maximum – minimum) of 1.06% (maximum 4.78%) for the percentage of positive nuclei [0.0-100.0%] across all runs.

Intensity Values

The image analysis results show an overall standard deviation of 0.9 (maximum 1.60) and average range (maximum - minimum) of 2.48 (maximum 4.27) for the intensity values [0-255] across all runs.

Inter-day/intra-system: The slide scores provided by image analysis over 20 scans on different days were analyzed for all 10 PR slides.

Percentage of Positive Nuclei

The image analysis results show an overall standard deviation of 0.54% (maximum 1.09%) and average range (maximum - minimum) of 1.52% (maximum 3.90%) for the percentage of positive nuclei [0.0-100.0%] across all runs.

Intensity Values

The image analysis results show an overall standard deviation of 1.44 (maximum 2.43) and average range (maximum – minimum) of 5.29 (maximum 11.39) for the intensity values [0-255] across all runs

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" in black, with a horizontal line cutting through the top portion of the "O". To the right of the symbol is the word "aperio" in a lowercase font. The word is in a light gray color.

Inter-system: The side scores provided by image analysis over 10 consecutive scans on three different ScanScope XT instruments were analyzed for all 10 PR slides.

Percentage of Positive Nuclei

The image analysis results on each of the three ScanScope systems show an overall average standard deviation of 0.54%, 0.53% and 0.75% (maximum 1.47%, 1.23%, 2.05%) and average range of 1.06%, 1,23%, and 1.50% (maximum 4.78%, 4.17%, 7.20%) for the percentage of positive nuclei [0.0-100.0%] across all runs.

The image analysis results of the three ScanScope systems combined show an overall average standard deviation of 0.87% (maximum 1.57%) and average range of 2.54% (maximum 8.13%) for the percentage of positive nuclei [0.0-100.0%] across all runs.

The image analysis results show minimal variation from one ScanScope system to another as shown in the following table that shows the mean over all runs of the reported percentage of positive nuclei [0.0-100.0%] for the 10 PR slides (#S) for the three ScanScope systems.

and announce of the demands of the children the desires of the closeChicago Corporation Controllers of CharlesCharles of the Research andCHERRE REERS COUNTERS FREE CHARRETH CONNECT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CARDER CONTRACT CARDERCONTRACTORICAL PRODUCTION COLLEGION COLLECTION CONTINUES COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR CONTRACTOR COLLECTION CONTRAC********* ***- Bolly Blood Planel PL Lecturities Production AMARSCall College
AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA1 1 2ANNUAL----------Children
Car Leasurements of*******-------RESERVER DOM R. F.on passes were passes manufact a aANNER LEE ENURRES ERRESHOUSERS RESEMBERS FREE A******КАРАДАННОСОВАЯ В РАНИЦИВАЕВ ГУРАПО-----------------------------------------.и в вла и полникая македонA MILLER - CALL PROPER COLLEGION CONNECT CONTRACTORCARRER CHRINNERAL CreationАКАКАВИНИНИЯ АКИНИИ КАКСАНИЕ ТАКИ КCE EREASERIES: BERRENIE E EAR WHENстание продукции институции и представление интеренности подательности в поставления по седения по седения по седенHARRENARIONER FR IN IN IN IN ERRESERATIONи и в вед и полнедаванными в в- RENER SEA & M LEASE ERESERVER******ERRESEMBERRESERESSESSMENT. FEE Boom . FAREEEEE Is an assacted****

Intensity Values

The image analysis results on each of the three ScanScope systems show an overall average standard deviation of 0.9%, 1.01%, and 0.93% (maximum 1.60%, 1.48%) and average range of 2.48%, 2.62%, and 2.60% (maximum 4.27%, 5.09%, 4.85%) for the intensity values [0-255] across all runs.

The image analysis results of the three ScanScope systems combined show an overall average standard deviation of 1.35% (maximum 2.03%) and average range of 4.55% (maximum 6.86%) for the intensity values [0-255] across all runs.

The image analysis results show minimal variation from one ScanScope system to another as shown in the following table that shows the mean over all runs of the reported percentage of positive nuclei [0.0-100.0%} for the 10 PR slides (#S) for the three ScanScope systems.

NA NA SATAI assemme------------------------------------------------------------------------------------------------------------------------------------------------------------------------------65 191 34 186.Children-1 14867
NA INA S160.00 204 184.05 184.07 175.62 49.214 # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # #
ירונג ידANGIAH1999 1999 100 172 11 100 172 1990 100,430191.68 185 4----------------------------------------------------一度的心理。 上一篇:1 100 7 40 1 7 2 1

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" in black, followed by the word "aperio" in a light gray color. The "O" is an oval shape with a white line running through it, creating a sense of depth. The word "aperio" is written in a simple, sans-serif font.

510k Summary – ScanScope® System – Rev23Oct19Sep08

.

.

.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Aperio. The logo consists of a stylized letter "O" in black, followed by the word "aperio" in a smaller, sans-serif font. The letter "O" is designed as an oval shape with a curved line running through it.

Intra-Pathologist: One pathologist read the same 10 PR slides 5 times using manual microscopy and 5 times using a manual read of digital slides on a computer monitor.

Percentage of Positive Nuclei

The manual microscopy results show an overall average standard deviation of 6.73% (maximum 16.73%) and average range of 9.8% (maximum 40%) and the manual read of digital slides results show an overall average standard deviation of 11.81% (maximum 28.72%) and average range of 16.2% (maximum 75%).

Intensity Scores

The manual microscopy results show 8 outliers out of 50 scores (16%) and the manual read of digital slides results show 9 outliers out of 50 scores (18%). Outliers are defined as scores that are different from the median values of the scores provided by the pathologist over 5 runs of the method.

Inter-pathologist: Three pathologists read the same 10PR slides using manual microscopy and using a manual read of digital slides on a computer monitor (based on the clinical comparison to manual microscopy study).

Percentage of Positive Nuclei

The manual microscopy results show an overall average standard deviation of 13.30% (maximum 32.15%) and average range of 17.2% (maximum 60%) and the manual read of digital slides results show an overall average standard deviation of 11.3% (maximum 20.82%) and average range of 16.0% (maximum 40%).

Intensity Scores

The manual microscopy results show 7 outliers out of 30 scores (23%) and the manual read of digital slides results show 7 outliers out of 30 scores (23%). Outliers are defined as scores that are different from the median values of the scores provided by the three pathologists.

21 CFR 807.92(b)(3): Conclusions drawn from the non-clinical and clinical tests:

Based on the results of the clinical studies described in this 510(k) submission, it is concluded that the ScanScope System device is as safe and cffective (therefore substantially equivalent) as the predicate device as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer.

... End of 510(k) Summary ....

510k Summary -- ScanScope® System - Rev23Oct19Scp08

{11}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the seal is a stylized graphic of a person embracing or protecting another person, symbolizing the department's mission to protect the health of all Americans.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 31 2008

Re: K080254

c/o Kim Bloom

Aperio Technologies, Inc.

Vista, California 92081

1430 Vantage Court, Suite 106

Trade/Device Name: ScanScope® XT System Regulation Number: 21 CFR 864.1860 Regulation Name: Immunohistochemistry Reagents & Kits Regulatory Class: Class II Product Code: OEO Dated: October 02, 2008 Received: October 03, 2008

Dear Ms Bloom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed

{12}------------------------------------------------

Page 2 - Ms. Bloom

predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Marian Chan, PhD

Acting Division Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Aperio. The logo consists of a stylized letter "O" that is tilted to the left and the word "aperio" written in lowercase letters to the right of the "O". The "O" is black, while the word "aperio" is a lighter color, possibly gray or a light shade of black.

Indications for Use

2018/06/2017 11:52:52 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11:50 PM IST 11

510(k) Number (if known): K080254

Device Name: ScanScope® XT System

Indications for Use:

The ScanScope® System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.

The ScanScope® system is intended for use as an aid to the pathologist in the detection and quantitative measurement of PR (Progesterone Receptor) by manual examination of the digital slide of formalin-fixed, paraffin-embedded normal and neoplastic tissue immunohistochemically stained for PR on a computer monitor.

It is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer.

Prescription Use (Part 21CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign Off
Office of In Vitro Diagnostic Device
Evaluation and Safety 510(k) K080254

§ 864.1860 Immunohistochemistry reagents and kits.

(a)
Identification. Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.(b)
Classification of immunohistochemistry devices. (1) Class I (general controls). Except as described in paragraphs (b)(2) and (b)(3) of this section, these devices are exempt from the premarket notification requirements in part 807, subpart E of this chapter. This exemption applies to IHC's that provide the pathologist with adjunctive diagnostic information that may be incorporated into the pathologist's report, but that is not ordinarily reported to the clinician as an independent finding. These IHC's are used after the primary diagnosis of tumor (neoplasm) has been made by conventional histopathology using nonimmunologic histochemical stains, such as hematoxylin and eosin. Examples of class I IHC's are differentiation markers that are used as adjunctive tests to subclassify tumors, such as keratin.(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
(3) Class III (premarket approval). IHC's intended for any use not described in paragraphs (b)(1) or (b)(2) of this section.
(c)
Date of PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the Federal Food, Drug, and Cosmetic Act is required for any device described in paragraph (b)(3) of this section before this device may be commercially distributed. See § 864.3.