(267 days)
VIDAS® AFP is an automated quantitative test for use on the VIDAS instruments for the quantitative measurement of alpha-fetoprotein (AFP) in human serum using the ELFA technique (Enzyme Linked Fluorescent Assay). The VIDAS AFP assay is indicated for the quantitative measurement of Alpha-Fetoprotein (AFP) in serum to aid in the management of patients with nonseminomatous testicular carcinoma.
The assay principle combines a one-step immunoassay sandwich method with a final fluorescent detection (ELFA). The Solid Phase Receptacle (SPR), a pipette tip-like device, serves as the solid phase as well as the pipetting device for the assay. The assay reagents are ready-to-use and pre-dispensed in the sealed reagent strips (STRs). All of the assay steps are performed automatically by the VIDAS instrument. The sample is transferred into the well containing AFP antibody (conjugate) labeled with alkaline phosphatase. The sample/conjugate mixture is cycled in and out of the SPR several times. This operation enables the antigen to bind with the immunoglobulins fixed to the interior wall of the SPR and to the conjugate to form a sandwich. Unbound compounds are eliminated during washing steps. Two detection steps are performed successively. During each step, the substrate (4-Methylumbeliferyl phosphate) is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methyl-umbelliferone) the fluorescence of which is measured at 450 nm. The intensity of the fluorescence is proportional to the concentration of antigen present in the sample. At the end of the assay, results are automatically calculated by the VIDAS instrument in relation to two calibration curves corresponding to the two detection steps stored in memory, and then printed out.
Here's a breakdown of the acceptance criteria and study information for the VIDAS® AFP Assay, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria Category | Acceptance Criteria (Implicit from Study Findings) | Reported Device Performance |
|---|---|---|
| Precision | Total variability (%CV) < 6.77% (across sites and samples) | < 6.77% |
| Limit of Blank (LoB) | Value for LoB established and reported | 0.178 U/mL |
| Limit of Detection (LoD) | Value for LoD established and reported | 0.444 U/mL |
| Limit of Quantitation (LoQ) | Value for LoQ established and reported | 0.444 ng/mL |
| Interference | No significant interference from tested substances | No interfering effect detected from hemoglobin, triglycerides, bilirubin, human albumin, RF, HAMA, and potential drug interferents. |
| Hook Effect | No hook effect up to a specified concentration | No hook effect observed up to 242,000 IU/mL |
| Dilution/Linearity | Samples can be diluted, and linearity maintained within the assay range. | May be used to dilute out-of-range samples (>400 IU/mL) up to 1/20. Linearity is 0.500 - 400.00 IU/mL. |
| Clinical Equivalence to Predicate | Clinical trial results demonstrate agreement between the new device and the predicate device, as evidenced by regression analysis. | Passing and Bablok regression analysis of 253 samples showed a slope = 1.128 (95% CI: 1.115 to 1.143) and an intercept = -0.530 (95% CI: -0.658 to -0.416). |
Study Information:
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Clinical Trial Test Set: 257 samples were tested to compare the VIDAS® AFP and the TOSOH ST AIA-PACK AFP assays. 253 samples were used for the Passing and Bablok regression analysis.
- Data Provenance: Not explicitly stated. The nature of the samples for the clinical trial (e.g., patient samples) suggests retrospective or prospective testing, but the specific origin (e.g., country) is not mentioned.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. For this in-vitro diagnostic device, ground truth for the clinical trial was established by comparison to a legally marketed predicate device (TOSOH ST AIA-PACK AFP assay), which itself serves as the reference standard in this context. There were no human expert readers establishing ground truth in the traditional sense for image interpretation or similar applications.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As the ground truth was established by comparison to a predicate device's quantitative measurement, adjudication by multiple human experts is not relevant to this type of study.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This was not a MRMC comparative effectiveness study involving human readers and AI assistance. It is an in-vitro diagnostic device measuring a biomarker.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, in essence. The VIDAS® AFP assay is an automated quantitative test. Its performance characteristics (precision, limits of detection, linearity, interference, hook effect) are evaluated as a standalone system. The "clinical trial" section compares its performance directly to another automated in-vitro diagnostic device (the predicate), essentially evaluating the algorithm/device's output without direct human interpretation influencing the final quantified result.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the clinical trial, the ground truth was established by comparison to a legally marketed predicate device (TOSOH ST AIA-PACK AFP assay). The predicate device's measurements served as the reference standard for evaluating the new device's quantitative agreement.
- For the precision, limits of detection, linearity, interference, and hook effect studies, the ground truth was based on the inherent characteristics of the samples used (e.g., known concentrations of AFP, spike/recovery studies), measured against the expected analytical performance.
-
The sample size for the training set
- Not Applicable/Not provided. This submission describes an in-vitro diagnostic assay rather than a machine learning algorithm that typically undergoes distinct "training" with a labeled dataset. The device's calibration curves are "stored in memory," and for each kit lot and calibrator lot, there's a "master curve." This suggests a calibration process, not a machine learning training phase in the conventional sense.
-
How the ground truth for the training set was established
- Not Applicable/Not provided. Similar to the above, there isn't a "training set" with ground truth in the context of a machine learning algorithm. The assay uses two calibration curves related to two detection steps, which are stored in memory. It also mentions a "Master curve for each kit lot and each calibrator lot are traceable to 1st IS 72/225 International Reference Preparation (IRP)". This traceability to an International Reference Preparation is how the assay's quantitative accuracy is established and maintained.
{0}------------------------------------------------
510(k) SUMMARY
VIDAS® AFP Assay
A. Submitter Information
bioMérieux, Inc. Submitter's Name: 595 Anglum Road Address: Hazelwood, MO 63042 Sandra Perreand Contact Person: 314-731-8594 Phone Number: 314-731-8689 Fax Number: Date of Preparation: September 18, 2008 B. Device Name VIDAS® AFP Assay Trade Name: Kit, Test, Alpha-Fetoprotein for Testicular Cancer Common Name: Kit, Test, Alpha-Fetoprotein for Testicular Cancer Classification Name:
C. Predicate Device Name
Tosoh Medical, Inc. ST AIA Pack AFP Enzyme Immunoassay
D. Device Description
Trade Name:
The assay principle combines a one-step immunoassay sandwich method with a final fluorescent detection (ELFA). The Solid Phase Receptacle (SPR), a pipette tip-like device, serves as the solid phase as well as the pipetting device for the assay. The assay reagents are ready-to-use and pre-dispensed in the sealed reagent strips (STRs).
All of the assay steps are performed automatically by the VIDAS instrument. The sample is transferred into the well containing AFP antibody (conjugate) labeled with alkaline phosphatase. The sample/conjugate mixture is cycled in and out of the SPR several times. This operation enables the antigen to bind with the immunoglobulins fixed to the interior wall of the SPR and to the conjugate to form a sandwich. Unbound compounds are eliminated during washing steps.
Two detection steps are performed successively. During each step, the substrate (4-Methylumbeliferyl phosphate) is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methyl-umbelliferone) the fluorescence of which is measured at 450 nm. The intensity of the fluorescence is proportional to the concentration of antigen present in the sample.
{1}------------------------------------------------
At the end of the assay, results are automatically calculated by the VIDAS instrument in relation to two calibration curves corresponding to the two detection steps stored in memory, and then printed out.
E. Intended Use
VIDAS® AFP is an automated quantitative test for use on the VIDAS instruments for the quantitative measurement of alpha-fetoprotein (AFP) in human serum using the ELFA technique (Enzyme Linked Fluorescent Assay). The VIDAS AFP assay is indicated for the quantitative measurement of Alpha-Fetoprotein (AFP) in serum to aid in the management of patients with nonseminomatous testicular.
F. Technological Characteristics Summary
A general comparison of the similarities and differences of the VIDAS AFP assay to the predicate device is presented in the table below.
| Item | VIDAS® AFP Assay | TOSOH ST AIA-PACK AFP(K023894) |
|---|---|---|
| Intended Use | VIDAS® AFP is an automated quantitative test for use on the VIDAS instruments for the quantitative measurement of alpha-fetoprotein (AFP) in human serum using the ELFA technique (Enzyme Linked Fluorescent Assay). The VIDAS AFP assay is indicated for the quantitative measurement of Alpha-Fetoprotein (AFP) in serum to aid in the management of patients with nonseminomatous testicular carcinoma. | ST AIA-PACK AFP is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of Alpha-Fetoprotein (AFP) in serum to aid in the management of patients with nonseminomatous testicular carcinoma. |
| Specimen | Serum | Serum |
| Antibody | Two mouse monoclonal AFP antibodies | Two mouse monoclonal AFP antibodies |
| Assay Principle | Two antibody "sandwich" binding of AFP. One antibody is bound to a solid phase and the second antibody is in liquid form and is labeled with fluorescent compound | Two antibody "sandwich" binding of AFP. One antibody is bound to a solid phase and the second antibody is in liquid form and is labeled with fluorescent compound |
| Automated | Yes | Yes |
| Assay Technique | Enzyme-linked fluorescent assay (ELFA) | Two-site immunoenzymometric assay |
| Sample Volume | 100 µL | 25 µL |
| Traceability/ | Master curve for each kit lot and each | Each calibrator lot are traceable |
{2}------------------------------------------------
| Item | VIDAS® AFP Assay | TOSOH ST AIA-PACK AFP(K023894) |
|---|---|---|
| Standardization | calibrator lot are traceable to 1st IS 72/225 International Reference Preparation (IRP) | to 1st IS 72/225 International Reference Preparation (IRP) |
| Measurement range | 0.500 – 400.00 IU/mLMaster curve highest calibration point = 400.0 IU/mL | 1 – 400 ng/mL(0.83 – 330.6 IU/mL)Highest calibrator = 200 ng/mL(165.3 IU/mL) |
G. Performance Data
A summary of the non-clinical and clinical test results is presented below.
Precision
A panel of three samples was assayed in duplicate in 40 different runs (2 runs per day) with two reagent lots using the same VIDAS instrument at three sites. The total variability (%CV) across all sites and across all samples was < 6.77%.
Limits of Detection
The Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) were determined using two kit lots on two VIDAS instrument per kit lot). The results are: LoB = 0.178U/mL, LoD = 0.444/mL, and LoQ = 0.444 ng/ml.
Interfering Substances/Drug Interference
In the study, hemoglobin, triglcerides, bilirubin, human albumin, RF, HAMA, and potential drug interferents were added to a human serum pool containing a measured concentration of AFP. No interfering effect was detected for any of the substances tested.
Hook Effect
No hook effect was observed for AFP concentrations up to 242,000 IU/mL.
Dilution/Linearity
Samples with very high AFP results were diluted with VIDAS AFP kit Diluent to evaluate the appropriateness of the diluent and the recommended maximum dilution. The VIDAS AFP diluent may be used to dilute out-of-range samples (>400 IU/mL) up to 1/20 prior to retesting.
Two high titer natural sera and one low titer natural serum are mixed in variable proportions to yield dilutions that cover the assay's full measurement range and the medical decision level (10 IU/mL). The linearity of the assay is 0.500 - 400.00 IU/mL.
Clinical Trial
A clinical trial was conducted to compare the VIDAS® AFP and the TOSOH ST AIA-PACK AFP assays by testing 257 samples with both assays. A Passing and Bablok regression analysis of the 253 samples gave a slope = 1.128 (95% confidence interval = 1.115 to 1.143) and an intercept = - 0.530 (95% confidence interval = -0.658 to -0.416).
{3}------------------------------------------------
H. Conclusion
:
The VIDAS® AFP Assay is substantially equivalent to the Tosoh Medical, Inc. ST AIA Pack AFP Assay.
:
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an eagle with its wings spread, symbolizing protection and strength. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
SEP 2 6 2008
bioMèricux, Inc. c/o Ms. Sandra Perreand Senior Director, North American Regulatory Affairs 595 Anglum Road Hazelwood, MO 63042
Re: K080017
Trade/Device Name: Vidas® AFP Assay Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor associated antigen immunological test systems Regulatory Class: Class II Product Code: LOJ Dated: August 28, 2008 Received: September 2, 2008
Dear Ms. Perreand:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter
{5}------------------------------------------------
Page 2 - bioMèrieux, Inc.
will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
ia Mc Chan
Maria M. Chan, Ph.D. Acting Division Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety
Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): K080017
Device Name: VIDAS® AFP
Indications For Use:
VIDAS® AFP is an automated quantitative test for use on the VIDAS instruments for the quantitative measurement of alpha-fetoprotein (AFP) in human serum using the ELFA technique (Enzyme Linked Fluorescent Assay). The VIDAS AFP assay is indicated for the quantitative measurement of Alpha-Fetoprotein (AFP) in serum to aid in the management of patients with nonseminomatous testicular carcinoma.
Prescription Use _ X ____________________________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use_ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Qeena Philip
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) - K080017
§ 866.6010 Tumor-associated antigen immunological test system.
(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.