K Number
K073706
Date Cleared
2008-03-31

(91 days)

Product Code
Regulation Number
868.5450
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fisher & Paykel Healthcare MR850 humidifier is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation or positive pressure breathing assistance or general medical gases.

The heated-wire breathing circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of humidified inspired gas, to reduce condensation. They are accessories for the Fisher & Paykel Healthcare MR850 Respiratory Gas Humidifier. The RT130 is used for flow rates between 0.3 and 4 L/min, and the RT131 is for flow rates greater than 4 L/min, for neonatal patients.

Device Description

The MR850 Respiratory Humidifier is designed to condition the ventilatory gases for patients requiring assisted breathing, by raising the water vapor content and temperature of the gases delivered to patients.

It consists of an electrically powered heat controller, utilizing a microprocessor with embedded software, to control a heating element that transfers heat to the water in a chamber. Breathing tubes enable the humidified gas to be transported to the patient. Depending on the chosen configuration, these tubes may be electrically heated, by means of a heater-wire placed internally to the tubes, to minimize the loss of humidity. An electrical adaptor provides electrical energy from the humidifier to the heater-wire in the breathing circuit and incorporates protection circuitry to prevent voltage and current transients on the heater-wire.

Temperature probes in the gas path provide feedback on temperature and flow of the gas to regulate temperature and humidity to the patient.

AI/ML Overview

The provided text describes a 510(k) summary for the Fisher & Paykel Healthcare MR850 Respiratory Humidifier. The focus of the submission is on a modification to the electrical adaptor for the heated breathing circuit, specifically the incorporation of surge and over-current protection circuitry. The main objective of the regulatory submission is to demonstrate substantial equivalence to the predicate device.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria or detailed performance metrics in a tabular format as you might expect for a new device's primary performance claims. Instead, it relies on the concept of substantial equivalence to a predicate device.

Acceptance Criteria (Inferred from Substantial Equivalence Claim)Reported Device Performance (as described in the text)
Performance and Safety equivalent to predicate device"Testing of the MR850 Respiratory Humidifier was compared to the MR850 (predicate) humidifier for performance and safety. These tests show that the MR850 Respiratory Humidifier has substantial equivalence to the MR850 (predicate) humidifier."Specifics on the modified component:- Surge Condition: Adaptor disconnects heater-wire for a period (up to 4 seconds) dependent on transient magnitude.- Detection Logic: If heater-wire remains disconnected during a detection cycle, humidifier illuminates warning indicators and audibly alarms.- Reconnection: Normal operation resumes once heater-wire is reconnected.- Over-current Condition: Heater-wire disconnected for a fixed 4 seconds.- Persistent Over-current: If over-current persists after reconnection, heater-wire is disconnected again, and humidifier continues to alarm.- Risk Minimization: These actions aim to minimize risks associated with fault conditions.
Maintains intended use and technological characteristics"The modified MR850 Respiratory Humidifier is substantially equivalent to the predicate MR850 humidifier and uses the same method of control and delivery of humidity.""The comparison of features, performance, and intended use demonstrate that the MR850 Respiratory Humidifier is at least as safe and effective for its intended purpose."

Explanation of "Acceptance Criteria" in this context: For a 510(k) submission, the primary acceptance criterion is often demonstrating "substantial equivalence" to a legally marketed predicate device. This means the new device is as safe and effective as the predicate, either by having the same technological characteristics and intended use, or by having different technological characteristics but no new questions of safety and effectiveness, and demonstrating that it is as safe and effective as the predicate device.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not specify a "sample size" in terms of patient data or clinical cases for the tests. The testing described is non-clinical and focuses on the device's electrical and functional performance as compared to the predicate.
  • Data Provenance: The testing was conducted by Fisher & Paykel Healthcare Limited. Given the company's location (New Zealand) and the regulatory submission to the FDA, it is likely the testing was conducted in New Zealand or under their direct supervision. The tests are described as non-clinical tests, which typically involve laboratory or bench testing rather than human subject data. The data is retrospective in the sense that it compares the new device's performance against an existing predicate device's established performance, rather than a new prospective clinical trial.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • This information is not provided in the document. For non-clinical, bench-level performance, the "ground truth" would likely be defined by engineering specifications, safety standards, and the performance characteristics of the predicate device. Expertise would come from the device's design and testing engineers, rather than medical experts establishing a diagnostic ground truth.

4. Adjudication Method for the Test Set:

  • This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of medical images or patient outcomes, where there might be disagreement among reviewers regarding a diagnosis or outcome. Since the described tests are non-clinical and focus on electrical and functional performance, such adjudication methods are not relevant.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not done. The document describes non-clinical testing comparing the modified device to its predicate for performance and safety. It does not involve human readers interpreting data or a study to determine how human readers improve with or without AI assistance. The device is a respiratory humidifier, not an AI-powered diagnostic tool.

6. Standalone (Algorithm Only) Performance Study:

  • No, a standalone (algorithm only) performance study was not done. This concept primarily applies to AI/ML algorithms. The MR850 Respiratory Humidifier is a physical medical device with embedded software for control, but it is not an AI diagnostic algorithm in the sense that would require a standalone performance study. The performance described relates to the device's functional operation (e.g., surge protection, alarming mechanisms).

7. Type of Ground Truth Used:

  • The "ground truth" for the non-clinical tests would have been established by engineering specifications, safety standards (e.g., electrical safety standards, medical device specific standards), and the known performance characteristics of the predicate device. For example, the ground truth for surge protection would be successful disconnection of power within specified parameters during a simulated surge event, as defined by engineering design and relevant electrical safety standards.

8. Sample Size for the Training Set:

  • This information is not applicable and not provided. "Training set" refers to data used to train AI/ML models. The MR850 is a hardware device with embedded control software; it does not involve machine learning in the conventional sense that would require a training set. The embedded software's logic is typically programmed based on design specifications and engineering principles, not "trained" on data.

9. How the Ground Truth for the Training Set was Established:

  • This information is not applicable and not provided for the same reasons as (8). The functional parameters and control logic for the humidifier's software are based on engineering design, medical requirements for humidification, and safety considerations, rather than ground truth established for an AI training set.

{0}------------------------------------------------

510(K) SUMMARY 5

Image /page/0/Picture/4 description: The image contains a sequence of handwritten digits and a letter. The sequence starts with the letter 'K', followed by the digits '073706'. The digits are written in a clear, legible manner, and the image appears to be a close-up of the sequence.

MAR 3 1 2008

Image /page/0/Picture/6 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are on the top line, with a large ampersand between the two words. The word "HEALTHCARE" is on the second line, and it is underlined.

Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place, East Tamaki P O Box 14 348, Panmure Telephone: +64 9 574 0100 Facsimile: +64 9 574 0158 Website: www.fphcare.com

Contact personAdele Bindon
Date prepared21 December 2007
Trade nameMR850 Respiratory Humidifier
Common nameRespiratory Gas Humidifier
Classification nameRespiratory Gas HumidifierII (21 CFR § 868.5450), product code BTT
Predicate deviceK033710 Fisher & Paykel Healthcare MR850 Respiratory Humidifier

{1}------------------------------------------------

Description 5.1

The MR850 Respiratory Humidifier is designed to condition the ventilatory gases for patients requiring assisted breathing, by raising the water vapor content and temperature of the gases delivered to patients.

It consists of an electrically powered heat controller, utilizing a microprocessor with embedded software, to control a heating element that transfers heat to the water in a chamber. Breathing tubes enable the humidified gas to be transported to the patient. Depending on the chosen configuration, these tubes may be electrically heated, by means of a heater-wire placed internally to the tubes, to minimize the loss of humidity. An electrical adaptor provides electrical energy from the humidifier to the heater-wire in the breathing circuit and incorporates protection circuitry to prevent voltage and current transients on the heater-wire.

Temperature probes in the gas path provide feedback on temperature and flow of the gas to regulate temperature and humidity to the patient.

5.2 Intended use

Gases available for medical use do not contain sufficient moisture and heat, which may damage or irritate the respiratory tract by desiccation in patients whose supraglottic airways have been bypassed.

Heated humidification, as provided by the MR850 Respiratory Humidifier, may be used for patients requiring mechanical ventilation, positive pressure breathing assistance, or general The MR850 humidifier is to be used on patients who are in the medical gases. hospital/institutional environment.

5.3 Technological characteristics comparison

The modified MR850 Respiratory Humidifier is substantially equivalent to the predicate MR850 humidifier and uses the same method of control and delivery of humidity.

The MR850 Respiratory Humidifier differs from the predicate device in the electrical adaptor for the heated breathing circuit. In the predicate system, the heater-wire adaptor is simply an electrical connection to provide power to the heater-wires in the breathing tubes. In the modified MR850 system, the electrical heater-wire adaptor incorporates surge and over-current protection circuitry.

In the event of a surge condition, the adaptor disconnects the heater-wire for a period of time which is dependent on the transient magnitude (for up to 4 seconds), to minimize the risks associated with such fault conditions. As the MR850 humidifier detects the presence of a heaterwire, if the heater-wire remains disconnected during a detection cycle, then the humidifier will illuminate the warning indicators and audibly alarm for the disconnection, thus alerting the user to abnormal operating conditions. Once the heater-wire is reconnected, normal operation resumes.

In the event of an over-current condition (such as an incompatible heated circuit or a fault condition such as a short-circuit), the heater-wire is disconnected for a fixed period of 4 seconds. When the heater-wire is reconnected (and should the over-current condition persist) the heaterwire is disconnected again, and the humidifier continues to alarm, thus minimizing the risk associated with such a fault condition.

{2}------------------------------------------------

Non-clinical tests 5.4

Testing of the MR850 Respiratory Humidifier was compared to the MR850 (predicate) humidifier for performance and safety. These tests show that the MR850 Respiratory Humidifier has substantial equivalence to the MR850 (predicate) humidifier.

5.5 Conclusion

The MR850 Respiratory Humidifier is substantially equivalent to the MR850 (predicate) humidifier. The comparison of features, performance, and intended use demonstrate that the MR850 Respiratory Humidifier is at least as safe and effective for its intended purpose.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 3 1 2008

Ms. Adele Bindon Regulatory Affairs Manager - RH Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place, East Tamaki P.O. Box 14-348, Panmure Auckland, NEW ZEALAND

Re: K073706

Trade/Device Name: Fisher & Paykel Healthcare MR850 Respiratory Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: February 27, 2008 Received: March 3, 2008

Dear Ms. Bindon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Bindon

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Syltte Y. Michaud
md.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and · Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT 4

510(k) Number

Device Name

Fisher & Paykel Healthcare MR850 Respiratory Humidifier

The Fisher & Paykel Healthcare MR850 humidifier is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation or positive pressure breathing assistance or general medical gases.

The heated-wire breathing circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of humidified inspired gas, to reduce condensation. They are accessories for the Fisher & Paykel Healthcare MR850 Respiratory Gas Humidifier. The RT130 is used for flow rates between 0.3 and 4 L/min, and the RT131 is for flow rates greater than 4 L/min, for neonatal patients.

V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

Please do not write below this line - continue on another page if needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Minh Khai

(Division Sign-Off) Tivision of Anesthesiology General Hospital nfection Control, Dental Devices

:10(k) Number: K073706

Page 1 of 1

§ 868.5450 Respiratory gas humidifier.

(a)
Identification. A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.(b)
Classification. Class II (performance standards).