(16 days)
Not Found
No
The document describes a standard blood pressure monitor using the auscultatory method and does not mention any AI or ML terms, image processing, or data training/testing sets typically associated with AI/ML applications.
No
The device is a non-invasive blood pressure monitor, which is used for measurement and monitoring, not for therapeutic purposes.
Yes
Explanation: The device is a "Non-Invasive Blood Pressure Monitor" that measures "blood pressure (systolic and diastolic)" and "heart rate." These measurements are used to assess a patient's health status, which falls under the definition of a diagnostic device.
No
The device description explicitly states it is a "Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the ausculatory method." This method typically involves hardware components like a cuff, stethoscope, and pressure sensor, and the listed performance standards (IEC/EN 60601-1, ANSI/AAMI SP10, EN1060) are for physical medical devices, not solely software.
Based on the provided information, the BPCard is not an IVD (In Vitro Diagnostic) device.
Here's why:
- IVD Definition: In vitro diagnostic devices are used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information about a person's health.
- BPCard Function: The BPCard is described as a "Non-Invasive Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the auscultatory method." This is a direct measurement of a physiological parameter from the body, not an analysis of a specimen taken from the body.
- Lack of Specimen Analysis: The description does not mention any analysis of biological samples.
Therefore, the BPCard falls under the category of a non-invasive medical device used for monitoring physiological signs, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The BPCard is a Non-Invasive Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the auscultatory method in both male and female, adult subjects. The BPCard also measures heart rate.
- Patient Population: Male/Female, Adults .
- Environment of Use: Hospital, Clinic and Home Use .
- Prescription Device by a Physician .
Product codes (comma separated list FDA assigned to the subject device)
DXN
Device Description
Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the ausculatory method. The BPCard also measures heart rate.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Adults
Intended User / Care Setting
Hospital, Clinic and Home Use
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The modified BPCard has been tested or found otherwise to comply with the following standards:
- IEC/EN 60601-1, Medical electrical equipment Part 1: General requirements . for basic safety and essential performance
- IEC/EN/UL 60601–1-2, Medical Electrical Equipment Part 1-2: General . requirements for safety - collateral standard: Electromagnetic compatibility -Requirements and Test
- ISO 10993-1, Biological Evaluation Of Medical Devices Part 1: Evaluation . And Testing
- . ANSI/AAMI SP10:2002, Manual, Electronic or Automated Sphygmomanometers
- EN1060-1 "Non-Invasive Sphygmomanometers General Requirements" .
- EN1060-3 "Non-Invasive Sphygmomanometers Supplementary . Requirements"
Conclusions: The results of the tests discussed above, indicate that the modified QRS Diagnostic BPCard is as safe, as effective, and performs as well as or better than the non modified device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
510(k) Summary of Safety and Effectiveness
Date:12/14/07
Submitter:
QRS Diagnostic, LLC Street Address: 14755 27th Ave N. City: Plymouth State: MN Zip Code: 55447 Telephone: 763-559-8492 Facsimile: 763-559-2961
JAN 1 1 2008
Contact: Mary Kay Jensen Phone: 763-559-8492 Ext 958 Facsimile: 763-559-2961 e-mail: mkjensen@grsdiagnostic.com
Device Name: Trade Name: BPCard Common Name: Blood Pressure Monitor Classification Name: System, Measurement, Blood-Pressure, Non-invasive Classification: Class II Panel Code: DXN Regulation Number: 21CFR. 870.1130 Noninvasive blood pressure measurement system.
Identification of Legally Marketed (Unmodified) Device (Predicate Device):
| Name of
Predicate | Manufacturer | Use | 510K) | Date
Cleared |
|----------------------|---------------------------|---------------------------|---------|-----------------|
| BPCard™ | QRS
Diagnostic,
LLC | Blood Pressure
Monitor | K031964 | 9/5/03 |
Device Description:
Indications for Use
Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the ausculatory method. The BPCard also measures heart rate.
- . Patient Population: Male/Female Adults
- . Environment of Use: Hospital, Clinic and Home Use
- . Prescription Device by a Physician
1
Technological Comparison to (Unmodified) Predicate Device:
The following summary table of comparisons compares the Modified BPCard Device to the Previously Cleared BPCard (Predicate) Device.
# | Area | Modified Device | Previously Cleared Device | Same | Different |
---|---|---|---|---|---|
Indications for Use | |||||
1 | Patient Population | Male/Female | |||
Adult | Male/Female | ||||
Adult | X | ||||
2 | Environment of Use | Hospital, Clinic, Home Use | Hospital, Clinic, Home Use | X | |
Fundamental Scientific Technology | |||||
3 | Measurements taken by Device | • Systolic Pressure | |||
• Diastolic Pressure | |||||
• Pulse Rate | • Systolic Pressure | ||||
• Diastolic Pressure | |||||
• Pulse Rate | X | ||||
4 | Measurement Method | Auscultatory | Auscultatory | X | |
5 | Device Components | • BPCard (PCMCIA) | |||
• Microphone Assembly | |||||
• Blood Pressure Cuff | |||||
• Software | • BPCard (PCMCIA) | ||||
• Microphone Assembly | |||||
• Blood Pressure Cuff | |||||
• Software | X | ||||
Contraindications | |||||
6 | Contraindications | No contraindications | Contraindicated in patients who are not ambulatory, have heart failure, arrhythmia or cardiac valve abnormality. | X | |
Sterility/Expiration Dating | |||||
7 | Sterile/Non Sterile | Non-Sterile/No Expiration Date | Non-Sterile/No Expiration Date | X | |
Energy Type | |||||
8 | AC/DC | Both | Both | X | |
Environmental Specifications | |||||
9 | Temperature/ | ||||
Humidity/EMC | |||||
Requirements | Same | Same | X | ||
Performance Standards | |||||
10 | AAMI SP10 | Meets AAMI | Meets AAMI | X | |
# | Area | Modified Device | Previously Cleared | ||
Device | Same | Different | |||
SP10:2002 | SP10:1992 | ||||
Device Hardware | |||||
11 | Dimensional and | ||||
Physical | |||||
Specifications | Same | Same | X | ||
12 | PCB Board | Same | Same | X | |
13 | Air Coupling | Bayonet Style | |||
Connector | Luer Slip | X | |||
14 | Deflation Valve | Lee Valve | Pneutronic Valve | X | |
15 | EMC Protection | Added measures for | |||
EMC protection to | |||||
meet IEC/EN | |||||
60601-1-2:2001 as a | |||||
result of other | |||||
changes | X | ||||
16 | Microphone Gain | ||||
(Sensitivity) | X40 | X10 | X | ||
17 | Microcontroller | ||||
Firmware | V 1.05 | V 0.03 | X | ||
Materials/Components | |||||
18 | Blood Pressure | ||||
Cuff | Trimline Kuff Link. | ||||
Pre-amendment | |||||
device. | Pymah Corporation | ||||
K884421 | X | ||||
19 | Microphone | ||||
Material | Medical Grade ABS | ||||
Plastic | ABS Plastic | X | |||
20 | Microphone Cable | ||||
Assembly | Dual Lumen Tubing | ||||
enclosing | |||||
microphone cable | No tubing enclosing | ||||
microphone cable | X | ||||
Algorithms | |||||
21 | Automatic | ||||
Microphone Gain | |||||
Control | Available | Not available | X | ||
22 | Level Based Pulse | ||||
Detection | Dynamic | Static | X | ||
23 | Pulse Proximity | ||||
Evaluation | Available | Not available | X | ||
24 | Pulse Rate | ||||
Calculation | Calculated from | ||||
median pulse | |||||
intervals. | Calculated from | ||||
minimum pulse | |||||
intervals. | X | ||||
Software | |||||
25 | Print Test Feature | Available | Not Available | X | |
26 | Save and Review | ||||
Tests Feature | Available | Not Available | X | ||
27 | Accept/Reject or | Available | Not Available | X | |
# | Area | Modified Device | Previously Cleared | ||
Device | Same | Different | |||
Device | |||||
Edit Test Feature | |||||
28 | Operating System | Windows | Windows | X | |
29 | Software Platform | Integrated with QRS | |||
Office Medic | Stand alone software | X | |||
30 | Deflation Rate | 2.5 mmHg and 5.0 | |||
mmHg | 2.5 mmHg | X | |||
31 | Deflation Start | Starts when pressure | |||
and pulse detection | |||||
criteria met | Starts when start | ||||
button pushed | X | ||||
32 | Maximum Pressure | ||||
Exceeded | Test will stop and | ||||
cuff will deflate | Software provides | ||||
warning | X | ||||
33 | Zero Function and | ||||
Open and Close | |||||
Valve function | Performed | ||||
Automatically at | |||||
start and end of test | User selects Zeroing | ||||
Button and End Test | |||||
Button | X | ||||
34 | Calibration | User can check | |||
pressure calibration | User can check | ||||
pressure calibration | X | ||||
35 | Pulse Detection | ||||
Beep and Display | System does not | ||||
provide audible or | |||||
display indication | |||||
when pulse is | |||||
detected | System provides | ||||
audible indication | |||||
when pulse is | |||||
detected and displays | |||||
number of pulses | X | ||||
36 | Raw Data | ||||
Processing | Raw Data file can | ||||
only be reprocessed | |||||
with QRS | |||||
development | |||||
software. | Raw Data File can be | ||||
reprocessed | X | ||||
37 | Pressure Display | Pressure displayed | |||
after test has begun | Pressure | ||||
continuously | |||||
displayed | X | ||||
38 | Start of Test | Test begins after | |||
auto-zeroing | |||||
completed and user | |||||
begins manual | |||||
inflation | Test begins after user | ||||
completed manual | |||||
inflation | X | ||||
39 | Battery Test | OS/Computer | |||
Battery warning | BPCard software | ||||
provides low battery | |||||
warning | X |
2
3
4
Summary of Performance Testing:
The modified BPCard has been tested or found otherwise to comply with the following standards:
- IEC/EN 60601-1, Medical electrical equipment Part 1: General requirements . for basic safety and essential performance
- IEC/EN/UL 60601–1-2, Medical Electrical Equipment Part 1-2: General . requirements for safety - collateral standard: Electromagnetic compatibility -Requirements and Test
- ISO 10993-1, Biological Evaluation Of Medical Devices Part 1: Evaluation . And Testing
- . ANSI/AAMI SP10:2002, Manual, Electronic or Automated Sphygmomanometers
- EN1060-1 "Non-Invasive Sphygmomanometers General Requirements" .
- EN1060-3 "Non-Invasive Sphygmomanometers Supplementary . Requirements"
Conclusions:
The results of the tests discussed above, indicate that the modified QRS Diagnostic BPCard is as safe, as effective, and performs as well as or better than the non modified device.
5
Public Health Service
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is smaller than the caduceus symbol.
JAN 1 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
QRS Diagnostic, LLC c/o Ms. Mary Kay Jensen QA/RA Manager 14755 27th Ave. N. Plymouth, MN 55447
Re: K073630
BPCard, Model Z-7000-0700 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 14, 2007 Received: December 26, 2007
Dear Ms. Jensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
6
Page 2 -- Ms. Mary Kay Jensen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements nft he Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odrh/industry/support/index.html.
Sincerely yours.
B. Bemmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
Indications for Use
510(k) Number (if known): To be determined
Device Name: BPCard
Indications for Use:
The BPCard is a Non-Invasive Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the auscultatory method in both male and female, adult subjects. The BPCard also measures heart rate.
- Patient Population: Male/Female, Adults .
- Environment of Use: Hospital, Clinic and Home Use .
- Prescription Device by a Physician .
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhumima
Page 1 of 1
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K073630