K Number
K073565
Date Cleared
2009-02-06

(415 days)

Product Code
Regulation Number
890.3110
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Patients in wheelchairs present problems to clinicians who need to treat them and the patients themselves can be seriously disadvantaged by their situation. This is particularly true of dental treatment. To be treated the dentist would need to bend around the patient in the wheelchair, that cusses back pain and sometimes injury. An alternate approach is to transfer the patient to a conventional chair by slings and hoists or other approaches, all of which hold significant health and safety risks. The wheelchair platform offers a safe environment for both clinician and patient. The wheelchair is moved onto the platform up against a back wall support and the wheelchair and patient are then supported by a backrest and a head support. The platform is reclined, and possibly raised and lowered, to present the patient in the best position for treatment.

Device Description

The device accommodates wheelchairs in a secure way, to enable them to be tilted backwards. Patient support is by an adjustable backrest with head support using an adjustable head cushion. The primary use is with patients who cannot be easily transferred from their chairs for such procedures as dentistry, Podiatry, Ear, nose and throat, maxio-facial work. Basic Platform shape: Steel fabrication. Drive mechanism: Electrical Linear Drive. Power source: Battery 24v and Power supply delivering 24v. Backrest support: Pivots in and out adjustable by worm and wheel. Headrest: Adjustable using a slotted arm with an adjustable headrest cushion for neck and head support. Motions command input: Tilt only Fixed switches and radio remote control. Foot crush protection: Pressure sensitive mat and guarding. Max. angle of tilt: 45deg. Rated capacity: 594lb (BSEN1570:1999). Proposed uses: Dental and medical applications. Static or mobile: Mobile on air cushion pads or ball castors.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the Wheelchair Platform, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Wheelchair Platform are primarily derived from the standard BSEN1570:1999. The submission confirms that the device meets these requirements.

Acceptance Criteria (from BSEN1570:1999)Reported Device Performance (Adherence)
Allowable stresses at design load: 60% of yield stressMet: Implied by the statement "Testing is carried out under the requirements as set out in BSEN1570:1999," which specifies this stress limit.
Full dynamic test at rated load recording operation timesMet: Implied by the statement "Testing is carried out under the requirements as set out in BSEN1570:1999," which lists this as a test.
10 min delay under load to determine sinkMet: Implied by the statement "Testing is carried out under the requirements as set out in BSEN1570:1999," which lists this as a test.
Load increase by 10% and full dynamic testMet: Implied by the statement "Testing is carried out under the requirements as set out in BSEN1570:1999," which lists this as a test.
Static test to 25% increase loadMet: Implied by the statement "Testing is carried out under the requirements as set out in BSEN1570:1999," which lists this as a test.
Rated capacity594lb (269.4 kg): Explicitly stated as per BSEN1570:1999.

Note: The provided text states that testing was carried out under BSEN1570:1999 and that this standard specifies the allowable stresses and types of tests. It then explicitly states the rated capacity of 594lb as per this standard. This implies that the device successfully met all the specified criteria within that standard.

2. Sample Size and Data Provenance

  • Sample Size for Test Set: Not specified in the provided text. The testing seems to be focused on a single device prototype or production model, as is typical for mechanical device performance testing against a standard. It's not a study involving a "test set" in the sense of a population of patients or data points.
  • Data Provenance: Not explicitly stated, however, the standard BSEN1570:1999 is a European standard, suggesting testing might have been done in a location adhering to these standards, possibly the UK given the "Ltd" in "Design Specific Ltd". The submission date is October 21, 2008, indicating retrospective data collection relative to the submission.

3. Number of Experts and Qualifications for Ground Truth

Not applicable. This is not a study requiring expert-established ground truth for a test set. The performance is assessed against engineering and safety standards.

4. Adjudication Method

Not applicable. This is not a study requiring adjudication of expert opinions.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is a mechanical device, not an AI or imaging diagnostic device.

6. Standalone (Algorithm Only) Performance Study

Not applicable. This is a mechanical device, not an algorithm.

7. Type of Ground Truth Used

The ground truth used for proving device performance is defined by the engineering and safety standards outlined in BSEN1570:1999. This standard specifies objective, measurable criteria related to stress, load capacity, and operational tests.

8. Sample Size for the Training Set

Not applicable. This is not a machine learning model, so there is no "training set."

9. How Ground Truth for the Training Set was Established

Not applicable. As there is no training set for an algorithm, there's no ground truth established for it.

{0}------------------------------------------------

design & specific

Executive Summary 510(k) Submission

Wheelchair Platform

Description

The device accommodates wheelchairs in a secure way, to enable them to be tilted backwards. Patient support is by an adjustable backrest with head support using an adjustable head cushion. The primary use is with patients who cannot be easily transferred from their chairs for such procedures as dentistry, Podiatry, Ear, nose and throat, maxio-facial work

Performance testing

$\nabla$

Testing is carried out under the requirements as set out in BSEN1570:1999. This specifies the allowable stresses as 60% of yield stress at the design load and states that:

  • . Full dynamic test at rated load recording operation times
  • . 10 min delay under load to determine sink
  • Load increase by 10% and full dynamic test .

Static test to 25% increase load

R. Fletcher 18.10.07

{1}------------------------------------------------

design → specific

Substantial Equivalence Discussion 510(k) Submission

Wheelchair Platform

Predicate Devices

Convertible I-series Positioning and Transfer Chair Barton Medical Corp. 5725 Hway 290 West #103 Austin TX 78735 8701 510(k) No. K071793

Electronic Positioning Chair Medline Industries Inc One Medline Place 11 60060 510(k) No. K001937

Dental Tilting Ramp TRICO Metal Products Wyandotte Rd. Willow Grove Pennsylvania 19060 Manufacturers claim distribution prior to 1976

Rehabilitation Hospital Philadelphia Introduction 1968, several sold. Currently in use manufacturer unknown

Substantial equivalence is sought within the category 'Electronic Positioning Chair' (INO)

The Barton and Medline equipment under the heading is intended for use by patients who are NOT in wheelchairs. The Trico product is designed for the specific treatment of wheelchair patients. Equivalence is therefore sought, not on the grounds of visual appearance only, but on functionality. Patients in wheelchairs must be considered as being integral with their chairs, for it is not safe or beneficial for them to be removed

The Philadelphia Rehabilitation Hospital unit clearly predates the May 28th 1976 date and is the earliest version of equipment found so far. It has all the features and functions of a modern wheelchair recliner but uses the technology available at the period. At least two of these units are in current use and they were made available for sale.

{2}------------------------------------------------

The table, attached, shows that the Design Specific Standard Wheelchair Platform is equivalent in both function and technology to both the Medline and Barton positioning chairs. It shows that the Trico Wheelchair Tilting Ramp is similar in form and designed for exactly the same medical procedures. The Philadelphia Recliner has the same function and intended purpose as the Design Specific model, but takes a different design solution. The technology is appropriate to the period of the design.

Flowchart Analysis

Is the device new to the market?Yes
Is new device to be compared to an existing deviceYes
Does new device have same indicated use statement?Yes
Does the new device have the same technological characteristics(See table above)Yes
Are descriptive characteristics enough to ensure equivalenceYes
Substantial Equivalence been shown.

R. Fletcher 19.3.08

·

{3}------------------------------------------------

CharacteristicMedline Ind. Inc.Barton Medical CorpsTrico Metal Ind.Rehab. Hospital Philadelphia 1968Design Specific Wheelchair Platform
Basic Platform shapeSteel fabricationSteel fabricationSteel with cast iron baseSteel assemblySteel fabrication
Drive mechanismElectrical Linear DriveElectrical Linear DriveElectrical Linear DriveHydraulicElectrical Linear Drive
Power sourceBattery 12vBattery 12v110v supply acManualBattery 24v and Power supply delivering 24v
Backrest supportNormal chair backNormal chair backUses wheelchair back with wheelchair locked in placeUses wheelchair back with wheelchair locked in placePivots in and out adjustable by worm and wheel
HeadrestNo separate headrestVertically adjustable headrestFixed headrestHeadrest attached to the wheelchairAdjustable using a slotted arm with an adjustable headrest cushion for neck and head support.
Motions command inputTilt onlyFixed switchesTilt and rise and fallHand held controllerTilt and riseFixed stalk with switch boxTilt and riseTilt onlyFixed switches and radio remote control.
Foot crush protectionNoneNoneNoneNonePressure sensitive mat and guarding
Max. angle of tiltHorizontalHorizontalApprox 25deg.35deg. approx45deg
Rated capacity850lb1000lbNot knownNot known594lb (BSEN1570:1999)
Proposed uses"...medical purposes.""...used to allow postural position."Wheelchair dentistryWheelchair DentistryDental and medical applications
Static or mobileMobile on wheelsMobile on wheelsMobile on air cushion baseStatic fixed baseMobile on air cushion pads or ball castors

:

、、

:

.

·

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Design Specific Ltd % Ms. Dawn Edwards 1310 Templar Boulevard Norfolk, Virginia 23518

Re: K073565

Trade Name: Standard Wheelchair Platform Regulation Number: 21 CFR 890.3110 Regulation Names: Electric positioning chair Regulatory Class: II Product Code: INO Dated: October 21, 2008 Received: October 31, 2008

FEB - 6 2009

Dear Ms. Edwards:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely vours,

Mark M. Mullerian

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for use

510(k) Number (if known): K073565

Device Name: standard Wheelchair Platform

Indications For Use:

Patients in wheelchairs present problems to clinicians who need to treat them and the patients themselves can be seriously disadvantaged by their situation. This is particularly true of dental treatment. To be treated the dentist would need to bend around the patient in the wheelchair, that cusses back pain and sometimes injury. An alternate approach is to transfer the patient to a conventional chair by slings and hoists or other approaches, all of which hold significant health and safety risks.

The wheelchair platform offers a safe environment for both clinician and patient. The wheelchair is moved onto the platform up against a back wall support and the wheelchair and patient are then supported by a backrest and a head support. The platform is reclined, and possibly raised and lowered, to present the patient in the best position for treatment.

Prescription Use 7 (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS) LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division S

Division of General, Restorative. and Neurological Devices

§ 890.3110 Electric positioning chair.

(a)
Identification. An electric positioning chair is a device with a motorized positioning control that is intended for medical purposes and that can be adjusted to various positions. The device is used to provide stability for patients with athetosis (involuntary spasms) and to alter postural positions.(b)
Classification. Class II. The electric positioning chair is exempt from premarket notification procedures in subpart E of part 807 of this chapter, subject to § 890.9 and the following conditions for exemption:(1) Appropriate analysis and non-clinical testing must demonstrate that the safety controls are adequate to ensure safe use of the device and prevent user falls from the device in the event of a device failure;
(2) Appropriate analysis and non-clinical testing must demonstrate the ability of the device to withstand the rated user weight load with an appropriate factor of safety;
(3) Appropriate analysis and non-clinical testing must demonstrate the longevity of the device to withstand external forces applied to the device and provide the user with an expected service life of the device;
(4) Appropriate analysis and non-clinical testing must demonstrate proper environments of use and storage of the device to maximize the longevity of the device;
(5) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ES60601-1, “Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and Essential Performance,” and ANSI/AAMI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Basic Safety and Essential Performance—Collateral Standard: Electromagnetic Disturbances—Requirements and Tests”) must validate electromagnetic compatibility and electrical safety;
(6) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ISO 10993-1, “Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process,” ANSI/AAMI/ISO 10993-5, “Biological Evaluation of Medical Devices—Part 5: Tests for In Vitro Cytotoxicity,” and ANSI/AAMI/ISO 10993-10, “Biological Evaluation of Medical Devices—Part 10: Tests for Irritation and Skin Sensitization”) must validate that the skin-contacting components of the device are biocompatible;
(7) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of IEC 62304, “Medical Device Software—Software Life Cycle Processes”) must validate the software life cycle and that all processes, activities, and tasks are implemented and documented;
(8) Appropriate analysis and non-clinical testing must validate that the device components are found to be non-flammable;
(9) Appropriate analysis and non-clinical testing must validate that the battery in the device (if applicable) performs as intended over the anticipated service life of the device; and
(10) Adequate patient labeling is provided to the user to document proper use and maintenance of the device to ensure safe use of the device by the patient in the intended use environment.