K Number
K073523
Manufacturer
Date Cleared
2008-08-01

(231 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Mcdtronic Interceptor® PLUS Coronary Filter System is intended for use in saphenous vein bypass grafts, with a reference vessel diameter between 2.5 mm and 5.25mm, in conjunction with percutaneous transluminal coronary intervention (PTCI) for embolic particulate capture. The Interceptor® PLUS Coronary Filter System is intended for temporary use in conjunction with other therapeutic devices and is intended to be removed at the completion of the interventional procedure.

The safety and effectiveness of this device as an embolic protection system has not been established in the cerebral, carotid, or peripheral vasculature

Device Description

The Interceptor® PLUS Coronary Filter System (the Device) consists of six components; an 180cm extendable Steerable Vascular Filter, an Actuator Handle, a Torque Handle, a Guidewire Introducer, Peel Away Introducer and a RX Facilitator Catheter . The Steerable Vascular Filter is comprised of a radiopaque nitinol filter mounted near the distal tip of a steerable 0.014 inch nitinol hypotube guidewire. The guidewire extends 3.5cm beyond the distal end of the filter in the collapsed state to facilitate distal tracking and steering through tortuous vessels and tight lesions. The proximal end of the steerable vascular filter contains four large openings (Approximately 1400 microns for the 3.5mm filter, >1500 microns for the 4.5mm filter size and >1800 microns for the 5.5mm filter size) that allow embolic particulate to enter the filter.

In addition to the Interceptor® PLUS Coronary Filter System, there is an optional Angled RX Facilitator Catheter when an alternative tip configuration is desired for filter withdrawal.

AI/ML Overview

The Medtronic Interceptor® PLUS Coronary Filter System's acceptance criteria and the study proving its performance are detailed below:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Non-Inferiority Margin)Reported Device Performance (MACE Rate Difference)Result
Delta of 4.5% (non-inferiority margin for MACE)Upper 95% CI of the difference: 4.0%Met

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Original Analysis: 800 saphenous vein graft patients total (501 in Treatment arm, 247 in Control arm).
    • Adjusted Analysis (due to missing CK-MB data):
      • Interceptor PLUS (Treatment arm): 139 patients
      • Control arm: 82 patients
      • Additional data for estimating non-Q wave MI: 158 patients from SAFER study (GuardWire arm) and 178 patients from Amethyst study (Control arm) with non-missing CK-MB.
  • Data Provenance: The study was a randomized clinical trial, implying prospective data collection. The document does not specify the country of origin, but given the FDA submission, it's likely a multi-center trial with sites possibly in the US.

3. Number of Experts and Qualifications for Ground Truth

The document does not provide information about the number of experts or their qualifications used to establish the ground truth for the test set. The ground truth (MACE) appears to be derived from clinical outcomes and laboratory results (CK-MB levels) which are typically assessed by physicians and laboratory personnel, rather than explicitly stated "experts" creating a ground truth set for an AI model.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method for the test set in the context of expert review. The primary endpoint, Major Adverse Cardiac Events (MACE), is a composite clinical outcome determined by predefined criteria (death, Q-wave or non-Q wave MI, emergent CABG, target vessel revascularization). The assessment of these components would follow standard clinical practice and laboratory interpretations, rather than a specific adjudication process for a test set in the way one would for an imaging AI study.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This study is a clinical trial assessing the effectiveness of a medical device (filter system) in human patients, not a study evaluating human readers' performance with or without AI assistance. Therefore, an MRMC comparative effectiveness study was not conducted, and no effect size for human readers' improvement with AI is reported.

6. Standalone (Algorithm Only) Performance Study

Not applicable. The Interceptor® PLUS Coronary Filter System is a physical medical device (embolic protection filter), not an algorithm or AI model. Therefore, a standalone (algorithm only) performance study was not conducted. The study evaluated the device's performance in preventing adverse cardiac events in patients.

7. Type of Ground Truth Used

The ground truth used was clinical outcomes data, specifically Major Adverse Cardiac Events (MACE), which is a composite endpoint defined by:

  • Death
  • Q-wave or non-Q wave MI (Myocardial Infarction), assessed via CK-MB results
  • Emergent Coronary Artery Bypass Surgery (CABG)
  • Target Vessel Revascularization (TVR) within 30 days.

8. Sample Size for the Training Set

Not applicable. This is a study of a physical medical device in a clinical trial, not an AI model. Therefore, there is no "training set" in the context of machine learning.

9. How Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for an AI model.

{0}------------------------------------------------

K073523

AUG - 1 2008

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Pursuant to Section 12, Safe Medical Devices Act of 1990)

    1. Identifying Information:
Submitter:Medtronic Vascular37A Cherry Hill DriveDanvers, MA 01923
Contact Person:Colleen MullinsSr. Regulatory Affairs Specialist(978) 739-3267
    1. Classification Name: Device, Coronary Saphenous Vein Bypass Graft, Temporary, For Embolization Protection (21 CFR Part 870.1250) Product Code NFA
  • Proprietary Name: 3. Interceptor® PLUS Coronary Filter System

4. Name of Predicate Devices:

FilterWire EZ TM Embolic Protection System (2.25mm-3.5mm)K061332
FilterWire EZTM Embolic Protection SystemK052280
SpiderFXTM Embolic Protection DeviceK063785
PercuSurge GuardWire Temporary Occlusion and Aspiration SystemK013913
GuardWire 3-6 Temporary Occlusion and Aspiration SystemK023878

{1}------------------------------------------------

5. Description:

The Interceptor® PLUS Coronary Filter System (the Device) consists of six components; an 180cm extendable Steerable Vascular Filter, an Actuator Handle, a Torque Handle, a Guidewire Introducer, Peel Away Introducer and a RX Facilitator Catheter . The Steerable Vascular Filter is comprised of a radiopaque nitinol filter mounted near the distal tip of a steerable 0.014 inch nitinol hypotube guidewire. The guidewire extends 3.5cm beyond the distal end of the filter in the collapsed state to facilitate distal tracking and steering through tortuous vessels and tight lesions. The proximal end of the steerable vascular filter contains four large openings (Approximately 1400 microns for the 3.5mm filter, >1500 microns for the 4.5mm filter size and >1800 microns for the 5.5mm filter size) that allow embolic particulate to enter the filter.

In addition to the Interceptor® PLUS Coronary Filter System, there is an optional Angled RX Facilitator Catheter when an alternative tip configuration is desired for filter withdrawal.

Intended-Üse: 6.

The Interceptor® PLUS Coronary Filter System is intended for use in saphenous vein bypass grafts, with a reference vessel diameter between 2.5 mm and 5.25mm, in conjunction with percutaneous transluminal coronary intervention (PTCI) for embolic particulate capture. The Interceptor® PLUS Coronary Filter System is intended for temporary use in conjunction with other therapeutic devices and is intended to be removed at the completion of the interventional procedure.

156

{2}------------------------------------------------

7. Summary Of Substantial Equivalence:

The Interceptor® PI.US Coronary Filter System for use in saphenous vein bypass grafts, is equivalent to the following predicate devices in that the intended use, principles of operation, design and materials are similar:

FilterWire EZ™ Embolic Protection System (2.25mm-3.5mm)K061332
FilterWire EZ™ Embolic Protection SystemK052280
SpiderFX™ Embolic Protection DeviceK063785
PercuSurge GuardWire Temporary Occlusion and Aspiration SystemK013913
GuardWire 3-6 Temporary Occlusion and Aspiration SystemK023878

Summary of Safety and Performance: 8.

A Clinical evaluation, AMEthyst study, was conducted in a randomized clinical trial involving a total of 800 saphenous vein graft patients. Patients were enrolled and randomized to one of two groups (Treatment )Interceptor or Interceptor PLUS Coronary Filter System] or the Control [GuardWire® Temporary Occlusion Balloon and Aspiration System or FilterWire EZTM Embolic Protection System]) at an unbalanced 2:1 ratio. The primary endpoint of this trial was Major Adverse Cardiac Events (MACE) related to the target vessel at 30 days: defined as death, Q-wave or non-Q wave MI, emergent coronary artery bypass surgery, or target vessel revascularization within 30 days of the index procedure. The primary endpoint analysis resulted in a MACE rate of 8.0% (40/501) in the Treatment arm and 7.3% (18/247) in the Control arm which demonstrates non-inferiority when tested to a delta of 4.5%. The upper 95% confidence interval of the difference between the two groups was 4.0%.

{3}------------------------------------------------

Notc: During the conduct of the AMFithyst study there were 163 Treatment (Interceptor and Interceptor PLUS) patients and 82 Control (GuardWire and FilterWire) patients who had at least one missing CK-MB result. To address this issue of missing CK-MB data an additional analysis on MACE was performed to determine a statistical estimate regarding the number of patients who may have experienced a non-Q wave MI. For the purposes of the analysis Interceptor subjects were excluded from the Treatment group and only Interceptor PLUS subjects were analyzed resulting in 139 patients. The number of Control patients remained 82. Data from 158 patients in the GuardWire arm of the SAFER study and 178 patients in the Control arm form the Amethyst study with non-missing CK-MB at all three time points (6-8, 12-16, and 18-24) were used in conjunction with conditional probability theory to estimate the number of non-Q wave MI that could have resulted for each of the various missing data patterns. The probabilities of a non-Q Wave MI for the Interceptor PLUS arm was based on the probabilities calculated for the SAFER study. The probabilities of this event in the Control arm were based on the Control Arm of the Amethyst study. For the Interceptor PLUS group, the total expected number of additional patients with non-Q wave MI was 2.392. For the Control group, the total expected number of additional patients with non-Q wave MI was 2.077. Taking a conservative approach and assuming 3 additional Interceptor PLUS patients experienced 30day MACE, versus only 2 in the control group, the rate of MACE at 30 days would increase from 8.4% to 9.1% for Interceptor PLUS arm and from 7.3% to 8.1% for the Control arm. The resulting non-inferiority p-value on the Interceptor PLUS vs. Control risk difference, using an absolute non-inferiority margin of 4.5%, is 0.050, allowing non-inferiority of Interceptor Plus to Control to be claimed.

Based on acute procedural and 30-day clinical results, the Interceptor PLUS Coronary Filter System has demonstrated non-inferior safety and efficacy compared with the GuardWire Temporary Occlusion Balloon and Aspiration System and the FilterWire EZ Embolic Protection System.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 1 2008

Medtronic Vascular c/o Ms. Colleen Mullins Senior Regulatory Affairs Specialist 37 A Cherry Hill Drive Danvers, MA 01923

Re: K073523

Trade/Device Name: Interceptor® PLUS Coronary Filter System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NFA Dated: June 30, 2008 Received: July 1, 2008

Dear Ms. Mullins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality

{5}------------------------------------------------

Page 2 - Ms. Colleen Mullins

systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entifled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Dona R. Vachner

A Bram D. Zuckerman, M.D. - Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number: K073523

Device Name: Interceptor® PLUS Coronary Filter System

Indications for Use:

The Mcdtronic Interceptor® PLUS Coronary Filter System is intended for use in saphenous vein bypass grafts, with a reference vessel diameter between 2.5 mm and 5.25mm, in conjunction with percutaneous transluminal coronary intervention (PTCI) for embolic particulate capture. The Interceptor® PLUS Coronary Filter System is intended for temporary use in conjunction with other therapeutic devices and is intended to be removed at the completion of the interventional procedure.

The safety and effectiveness of this device as an embolic protection system has not been established in the cerebral, carotid, or peripheral vasculature

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duane R. Barnes

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_Ko73523

Page 1 of 1

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).