K Number
K073381
Device Name
PLEXUS EBV IGM MULTI-ANALYTE DIAGNOSTICS, MODEL: MP0600M
Date Cleared
2008-08-04

(245 days)

Product Code
Regulation Number
866.3235
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Focus Diagnostics' Plexus™ EBV IgM Multi-Analyte Diagnostics test kit is intended for qualitatively detecting the presence or absence of human IgM class antibodies to viral capsid antigen (VCA), and heterophile antibodies in human sera. The test is indicated as an aid in the diagnosis of EBV infection and EBV-associated infectious mononucleosis. The performance of this assay has not been established for use in the diagnosis of nasopharyngeal carcinoma and Burkitt's lymphoma, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment. This assay has not been evaluated for donor screening.
Device Description
Multiplexed Immunoassay for the Qualitative Detection of Human IgM Antibodies to Epstein-Barr Virus The Focus Diagnostics Plexus™ EBV IgM uses an Antigen Bead suspension that contains two distinct EBV antigen bead types (VCA and Heterophile) and one process control bead type that fluoresce at different wavelengths and/or intensities. The Focus Diagnostics Plexus™ EBV IgM is a three step procedure. 1. Patient sera are diluted, and the diluted sera are incubated with Antigen Beads. If EBV antibodies are present, then the antibodies bind to the corresponding antigen beads. 2. Phycoerythrin-conjugated goat Anti-human IgM (Conjugate) is added, binds to the bound EBV antibody (if present), and forms a Conjugate-EBV antibody-antigen bead sandwich. 3. Fluorescence from each distinct EBV antigen bead type is measured and compared against a Cutoff Calibrator.
More Information

Osom® Mono Test, Diamedix EBV VCA IgM ELISA, Athena Multi-Lyte EBV VCA IgM Test System

Not Found

No
The summary describes a standard immunoassay technology and data analysis based on comparing fluorescence measurements to a cutoff calibrator. There is no mention of AI, ML, or any related concepts in the device description or performance studies.

No.
This device is a diagnostic test kit intended to aid in the diagnosis of EBV infection and EBV-associated infectious mononucleosis by detecting specific antibodies, not to treat or prevent a disease or condition.

Yes
The "Intended Use / Indications for Use" section explicitly states that the test kit "is intended as an aid in the diagnosis of EBV infection and EBV-associated infectious mononucleosis." This clearly indicates its purpose is to help identify a medical condition.

No

The device is a test kit that uses antigen beads and chemical reactions to detect antibodies in human sera, which are physical components and processes, not solely software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use explicitly states that the device is "intended for qualitatively detecting the presence or absence of human IgM class antibodies to viral capsid antigen (VCA), and heterophile antibodies in human sera." It is also indicated as "an aid in the diagnosis of EBV infection and EBV-associated infectious mononucleosis." This clearly describes a test performed on a biological sample (human sera) outside of the body to provide information for diagnosis.
  • Device Description: The description details a "Multiplexed Immunoassay for the Qualitative Detection of Human IgM Antibodies to Epstein-Barr Virus." Immunoassays performed on biological samples are a common type of IVD.
  • Methodology: The three-step procedure involves incubating patient sera with antigen beads, adding a conjugate, and measuring fluorescence. This is a typical laboratory-based diagnostic test performed on a biological specimen.
  • Performance Studies: The document describes performance studies comparing the device to predicate devices (which are also IVDs) using biological samples (sera) to evaluate its accuracy in detecting the target analytes.

All of these points align with the definition of an In Vitro Diagnostic device, which is a medical device intended for use in vitro for the examination of specimens derived from the human body solely or principally for the purpose of providing information concerning a physiological or pathological state, or concerning a congenital abnormality, or to monitor therapeutic measures.

N/A

Intended Use / Indications for Use

Focus Diagnostics' Plexus™ EBV IgM Multi-Analyte Diagnostics test kit is intended for qualitatively detecting the presence or absence of human IgM class antibodies to viral capsid antigen (VCA), and heterophile antibodies in human sera. The test is indicated as an aid in the diagnosis of EBV infection and EBV-associated infectious mononucleosis.

The performance of this assay has not been cstablished for use in the diagnosis of nasopharyngeal carcinoma and Burkit's lymphoma, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment. This assay has not been evaluated for donor screening.

Product codes (comma separated list FDA assigned to the subject device)

LJN

Device Description

Multiplexed Immunoassay for the Qualitative Detection of Human IgM Antibodies to Epstein-Barr Virus

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Prospective Population Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte (N = 723) Samples were collected and tested by the Northeast investigator (n = 350), Mid-West investigator (n=249) and Focus (n=124).

Prospective Population Samples: Plexus EBV vs. Heterophile Rapid Test for Heterophile IgM analyte (N = 723) Samples were collected and tested by the Northeast investigator (n = 350), Mid-West investigator (n=249) and Focus (n=124).

Presumed Acute Retrospective Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte Samples were collected and tested by Mid-West investigator (n=150).

Presumed Acute Retrospective Samples: Plexus EBV vs. Heterophile Rapid Test for Heterophile IgM analyte Samples were collected and tested by Mid-West investigator (n=150).

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Comparison Studies:
Performance of the Plexus EBV VCA IgM analyte was tested against a combination (hereafter referred to as 'consensus predicate') of a FDA-cleared commercially available ELISA, a FDA cleared commercially available immunofluroscent (IFA) test and a FDA cleared commercially available flow cytomcery based immunoassay. For each sample, a consensus based algorithm (2/3) was used to determine the predicate result for comparison with the Plexus VCA IgM result. The Plexus EBV Heterophile IgM analyte was tested against a FDA cleared heterophile rapid test. The studies were conducted at three United States testing sites: a hospital laboratory located in Northeast. a pediatric hospital laboratory located in the Mid-West, and Focus with serum samples in which EBV tests were ordered. The sera were sequentially submitted to the laboratory, archived, and masked, Samples were collected at three sites and include both prospective (n = 723) and retrospective (n = 150) specimens. The Plexus EBV IgM tests were run in conjunction with the Plexus EBV IgG tests for a complete antibody profile. The samples were then classified into EBV infection status using the Serological Status table above.

Prospective Population Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte (N = 723)
Table 4: EBV VCA IgM Results

Consensus PredicatenPositiveEquivocalNegative% Agreement
Positive88793685.9%(79/92), 95% CI:77.3-91.6%
Negative63012161797.8%(617/631), 95% CI:96. 3-98.7%
No consensus15104NA

Prospective Population Samples: Plexus EBV vs. Heterophile Rapid Test for Heterophile IgM analyte (N = 723)
Table 5: EBV Heterophile Results

Predicate Rapid TestnPositiveEquivocalNegative% Agreement
Positive756001580%(60/75), 95% CI:69.6-87.5%
Equivocal0000N/A
Negative6487064198.9%(641/648), 95% CI:97.8-99.5%

Presumed Acute Retrospective Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte

Consensus PredicatenPositiveEquivocalNegative% Agreement
Positive1431400397.2%(140/144), 95% CI:93.1-98.9%
Negative530233.3%(2/6), 95% CI:9.7-70.0%
No consensus12101NA

Presumed Acute Retrospective Samples: Plexus EBV vs. Heterophile Rapid Test for Heterophile IgM analyte
Table 8: EBV Heterophile Results

Predicate Rapid TestnPositiveEquivocalNegative% Agreement
Positive1129831187.5%(98/112), 95% CI:80.1-92.4%
Negative38303592.1%(35/38), 95% CI:79.2-97.3%

Inter-laboratory, Intra-assay and Inter-assay Reproducibility
The interfintra-assay reproducibility and the inter-laboratory reproducibility testing were performed at three laboratories. Each of the three laboratories tested twelve samples in triplicate on five different days.

Inter-Lot Reproducibility
The inter-lot reproducibility was evaluated with eleven (11) samples in triplicates on three (3) lots of Plexus EBV kit.

Cross-Reactivity
A cross-reactivity study was performed to determine if samples disease states and other potentially cross-reactivity factors interfere with test results when tested with the Plexus EBV IgM kit. A panel of (Antinuclear Antibody test (ANA) n=28, Cytomegalovirus (CMV) n=25, Herpes Simplex Virus-1 (HSV-1) and Herpes Simplex Virus-2 (HSV-2) n =2, Rheumatoid Factor (Rh) n=29, Rubella Virus n=5, and Varicella-Zoster Virus (VZV) n=42). samples for each cross reactant were evaluated for possible cross reactivity with the Plexus EBV IgM kit for each of the two (VCA and Heterophile) IgM analyte. The test samples were also evaluated on commercially available ELISA and heterophile rapid test.

Sample Storage and Handling
Fifteen (15) negative and positive for EBV IgM samples were used to assess the reactivity of unfrozen sample against samples that were frozen and thawed for up to five cycles. No effect was observed for any of the freezethaw cycling in either the positive or negative sample.

Interference
The test performance was evaluated with the presence of interfering substances. Four samples, two positive and two negative for EBV VCA IgM and heterophile antibody by Plexus EBV IgM were used in the study. Baseline ERV IgM levels were established for each sample before interferents were added. Two concentration levels of each interferent were used: triglycerides (1 mg/mL), albumin (6 mg/mL and 60 mg/mL), bilirubin (0.02 mg/mL and 0.2 mg/mL), and hemoglobin (22 mg/mL) The higher interferent concentrations represent levels that exceed the expected human range. All samples were evaluated to determine if these interfering substances affect the assay. No interference was observed for any of the substances in either the positive or negative sample.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

See tables in "Summary of Performance Studies" for positive and negative percent agreements and 95% CI.

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Osom® Mono Test, Diamedix EBV VCA IgM ELISA, Athena Multi-Lyte EBV VCA IgM Test System

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 866.3235 Epstein-Barr virus serological reagents.

(a)
Identification. Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).(b)
Classification. Class I (general controls).

0

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008 Page 1 of 13

| Applicant | Focus Diagnostics, Inc.
10703 Progress Way
Cypress, California 90630
USA | |
|--------------------------------|--------------------------------------------------------------------------------------------|--------------|
| Establishment Registration No. | 2023365 | AUG - 4 2008 |
| Contact Person | Constance Bridges
tel 714.220.1900
fax 714.995.6921
cbridges@focusdx.com | |
| Summary Date | June 27, 2008 | |
| Proprietary Name | Plexus EBV IgM Multi-Analyte Diagnostics | |
| Generic Name | Epstein-Barr Virus Serological Assays | |
| Classification | Class I | |
| Predicate Devices | Osom® Mono Test
Diamedix EBV VCA IgM ELISA
Athena Multi-Lyte EBV VCA IgM Test System | |

Device Description

Multiplexed Immunoassay for the Qualitative Detection of Human IgM Antibodies to Epstein-Barr Virus

Intended Use

Focus Diagnostics' Plexus™ EBV IgM Multi-Analyte Diagnostics test kit is intended for qualitatively detecting the presence or absence of human IgM class antibodies to viral capsid antigen (VCA), and heterophile antibodies in human sera. The test is indicated as an aid in the diagnosis of EBV infection and EBV-associated infectious mononucleosis.

Epstein-Barr Virus VCA (IgM) Recomb Immun Antibod

The performance of this assay has not been cstablished for use in the diagnosis of nasopharyngeal carcinoma and Burkit's lymphoma, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment. This assay has not been evaluated for donor screening.

The Focus Diagnostics Plexus™ EBV IgM uses an Antigen Bead suspension that contains two distinct EBV antigen bead types (VCA and Heterophile) and one process control bead type that fluoresce at different wavelengths and/or intensities.

The Focus Diagnostics Plexus™ EBV IgM is a three step procedure.

    1. Patient sera are diluted, and the diluted sera are incubated with Antigen Beads. If EBV antibodies are present, then the antibodies bind to the corresponding antigen beads.
    1. Phycoerythrin-conjugated goat Anti-human IgM (Conjugate) is added, binds to the bound EBV antibody (if present), and forms a Conjugate-EBV antibody-antigen bead sandwich.
    1. Fluorescence from each distinct EBV antigen bead type is measured and compared against a Cutoff Calibrator.

1

Image /page/1/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in a bold, sans-serif font, with a curved graphic element above and to the left of the word. Below "FOCUS" is the word "Diagnostics" in a smaller, sans-serif font, underlined with a thin line. The overall design is simple and professional.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008 Page 2 of 13

Comparison with Predicate Devices: VCA IgM analyte:

ItemDevicePredicates
NamePlexus™ EBV IgM
Multi-Analyte
DiagnosticsDiamedix VCA IgM,
ELISAAthena Multi-Lyte
EBV VCA IgM Test
SystemEpstein-Barr Virus
VCA (IgM) Recomb
Immun Antibod
(IFA)
Similarities between Device and Predicate
Intended useFocus Diagnostics'
Plexus™ EBV (VCA)
IgM Multi-Analyte
Diagnostics test kit is
intended for
qualitatively detecting
the presence or absence
of human IgM class
antibodies to viral capsid
antigen (VCA), and
heterophile antibodies in
human sera. The test is
indicated as an aid in the
diagnosis of EBV
infection and EBV-
associated infectious
mononucleosis.Diamedix Corp EBV
VCA IgM ELISA is
intended for the
qualitative and semi-
quantitative determination
of IgM antibodies to
Epstein-Barr Virus
(recombinant) viral capsid
antigen (EBV-VCA IgM)
in human serum by
indirect enzyme
immunoassay. The Is-
EBV-VCA IgM test kit
may be used in
combination with other
Epstein-Barr serologies,
Viral Capsid Antigen
(VCA) IgG, Epstein-Barr
Nuclear Antigen-1
(EBNA-1) IgG and IgM,
Early Antigen-Diffuse
(EA-D) IgG and IgM and
heterophile antibody, as
an aid in the diagnosis of
infectious mononucleosis
(IM).The Zeus Scientific,
Inc. AtheNA Multi-
Lyte® EBV VCA
IgM Test System is a
microparticle-based
immunoassay
intended for the
qualitative detection
of IgM class antibody
to Epstein-Barr virus,
viral capsid antigen in
human serum. The
test system is
intended to be used
for the laboratory
diagnosis of EBV-
associated infectious
mononucleosis and
provides
epidemiological
information on the
diseases caused by
Epstein-Barr virus.Focus Diagnostics'
Epstein-Barr Virus
Viral Capsid Antigens
(EBV VCA) IgG
Immunofluorescence
Antibody (IFA) test is
intended for the
detection and semi-
quantitation of human
IgG antibodies to the
viral capsid antigens
(VCA) of Epstein-Barr
virus in human serum
as an aid in the clinical
diagnosis of infectious
mononucleosis.
Indications for useThe test is indicated as
an aid in the diagnosis of
EBV infection and
EBV-associated
infectious
mononucleosis.The device is indicated for
use with patients with the
signs and symptoms of
infectious mononucleosis.The device is
indicated for
laboratory diagnosis
of EBV-associated
infectious
mononucleosis and to
provide
epidemiological
information on the
diseases caused by
Epstein-Barr virusThe test is indicated as
an aid in the clinical
diagnosis of infectious
mononucleosis.
Immunoglobulin
TypeIgMIgMIgMIgM
Sample matrixSerumSerumSerumSerum
CLIA complexityHighHighHighHigh
Difference between Device and Predicate
AntigenEBV-VCA: VCA gp
125, affinity purifiedEBV-VCA: Recombinant
47 kDa fusion half of p18EBV VCA pg125Recombinant VCA
antigen (rVCA)
ItemDevicePredicates
NamePlexus™ EBV IgM
Multi-Analyte
DiagnosticsDiamedix VCA IgM,
ELISAAthena Multi-Lyte
EBV VCA IgM Test
SystemEpstein-Barr Virus
VCA (IgM) Recomb
Immun Antibod
(IFA)
antigen
StrainN/A- purified proteinN/A- Recombinant proteinN/A native proteinN/A- Recombinant
protein
Host Cell LineEBV-VCA: Native
(P3H3 or P3HR-1)EBV-VCA: E. coli
(unknown)UnknownMammalian cell line
MethodologyMultiplex Microbead
Immunoassay (MMIA)
based on Luminex XMAP
technologyEnzyme Immunoassay
technology.Flow cytometry
immunoassayIndirect
Immunofluorescence
Antibody Test
Interpretation of
test resultsPerform automated
calculations using
Plexus software.Manual calculationAtheNA Multi-Lyte
instrumentManual

2

Image /page/2/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in bold, uppercase letters. A curved, swooping line extends from the left side of the "F", arcing over the word. Below "FOCUS" is the word "Diagnostics" in a smaller, less bold font, underlined with a thin line.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008

Page 3 of 13

Comparison with Predicate Devices: Heterophile IgM analyte:

ItemDevicePredicate
NamePlexus™ EBV IgM Multi-Analyte
DiagnosticsOsom® Mono Test, Heterophile
Similarity between Device and Predicate
Intended useFocus Diagnostics' Plexus™ EBV
(Heterophile) IgM Multi-Analyte Diagnostics
test kit is intended for qualitatively detecting
the presence or absence of human IgM class
antibodies to viral capsid antigen (VCA), and
heterophile antibodies in human sera. The test
is indicated as an aid in the diagnosis of EBV
infection and EBV-associated infectious
mononucleosis.The OSOM Mono Test is intended for the quantitative
detection infectious mononucleosis heterophile
antibodies in serum, plasma and whole blood as an aid
in the diagnosis of infectious mononucleosis.
Indications for useThe test is indicated as an aid in the diagnosis
of EBV infection and EBV-associated
infectious mononucleosis.Intended to make a serological diagnosis of EBV
infections
Immunoglobulin
TypeIgMIgM
Difference between Device and Predicate
Sample matrixSerumSerum / Plasma / Whole Blood
CLIA complexityHighModerate
MethodologyMultiplex Microbead Immunoassay (MMIA)
based on Luminex XMAP technology.Immunocheomatorgraphic Test

:

3

Image /page/3/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo consists of the word "FOCUS" in bold, sans-serif font, with a curved graphic element above and to the left of the word. Below the word "FOCUS" is the word "Diagnostics" in a smaller, less bold font, underlined by a thin line. The overall design is simple and professional.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M

Prepared Date: July 22, 2008

Page 4 of 13

ItemDevicePredicate
NamePlexus™ EBV IgM Multi-Analyte
DiagnosticsOsom® Mono Test, Heterophile
StrainNA purified proteinNA native protein
Host Cell LinePurified from bovine red blood cellsExtract of bovine erythrocytes
AntigenHeterophile: purified proteinHeterophile: native protein
Interpretation of
test resultsPerform automated calculations using Plexus
software.Visual evaluation

EXPECTED VALUES

Outside investigators assessed the device with prospective masked sequential samples that were submitted for routine EBV testing (n=723.) The prevalence of EBV will vary depending on age, geographic location, testing method used, and other factors. The comparator assay was performed by indirect enzyme immunoassay used in combination with other Spstein-Barr serologies (VCA) IgG and IgM, Epstein-Barr Nuclear Antigen-1 (EBNA-1) IgG, Early Antigen-Diffuse (EA-D) IgG and a heterophile rapid antibody test. The observed prevalence is in Table 1. Expected values are presented by age and gender in the tables below for serum sample from a Total Population (N= 723). For all analytes, index values of 1.10 are positives.

Table 1: Observed Prevalence – EBV Plexus
VCA IgMHeterophile
Positive12.7% (92/723)9.3% (67/723)
Equivocal0.6% (4/723)0% (0/723)
Negative86.7% (627/723)90.7% (656/723)
Table 2: EBV Plexus Results VCA IgM
PositiveEquivocalNegative
AgeGendern%n%n%Total
70F000071007
>70M111.100888.99
Total9212.740.662786.7723
Table 3: EBV Plexus Results Heterophile IgM
AgeGenderPositiveEquivocalNegativeTotal
n%n%n%
70F000071007
>70M000091009
Total679.30065690.7723

The table below summarizes the breakdown of the samples age and gender information. The distribution chart below exhibits the age distribution of all 873 samples included in the study.

Age Information:
Summary of Female SubjectsSummary of Male Subjects
n474n399
mean20.0mean18.3
median16.0median14.0
min1min1
max88max87

5

Image /page/5/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in bold, sans-serif font. Above and to the left of the word is a curved, swooping shape. Below the word "FOCUS" is the word "Diagnostics" in a smaller, less bold font, underlined with a thin line.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008 Page 6 of 13

Image /page/5/Figure/3 description: The image is a bar graph titled "Distribution of Patient Age". The x-axis is labeled "Age" and shows age ranges from 1-5 to 85-90. The y-axis is labeled "Frequency" and ranges from 0 to 300. The bar graph shows that the highest frequency of patients is in the 16-20 age range, with a frequency of approximately 240.

6

Image /page/6/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in bold, sans-serif font. Below "FOCUS" is the word "Diagnostics" in a smaller, sans-serif font with a line above it. To the left of the word "FOCUS" is a curved, black shape that resembles a crescent.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008

Page 7 of 13

PERFORMANCE CHARACTERISTICS

Typical Antibody Response Classification

The table below summarizes a generally accepted algoritum for classifying the EBV infection status via EBV serologic profiles.

| EBV Serological
Status | | EBNA-1 IgG | EBV VCA IgG | EBV EA-D IgG | EBV VCA IgM | Heterophile
Antibody | | |
|---------------------------|---------------|---------------------------------------|-------------|--------------|-------------|-------------------------|--|--|
| | | Negative | Negative | Negative | Positive | Negative | | |
| | | Negative | Negative | Negative | Positive | Positive | | |
| | | Negative | Positive | Negative | Positive | Positive | | |
| | Primary | Negative | Negative | Positive | Positive | Negative | | |
| | Acute | Negative | Negative | Positive | Positive | Positive | | |
| | | Negative | Positive | Positive | Positive | Negative | | |
| Acute | | Negative | Positive | Positive | Positive | Positive | | |
| | | Negative | Positive | Positive | Negative | Positive | | |
| | Late
Acute | Positive | Positive | Positive | Negative | Negative | | |
| | | Positive | Positive | Positive | Positive | Positive | | |
| | | Positive | Positive | Positive | Positive | Negative | | |
| | | Positive | Positive | Negative | Positive | Positive | | |
| | | Positive | Positive | Negative | Positive | Negative | | |
| | | Negative | Positive | Negative | Positive | Negative | | |
| Recovering | | Negative | Positive | Positive | Negative | Negative | | |
| Past Infection | | Negative | Positive | Negative | Negative | Negative | | |
| | | Positive | Positive | Negative | Negative | Negative | | |
| No Infection | | Negative | Negative | Negative | Negative | Negative | | |
| Indeterminant | | Combinations not listed above (n =18) | | | | | | |

Comparison Studies

Performance of the Plexus EBV VCA IgM analyte was tested against a combination (hereafter referred to as 'consensus predicate') of a FDA-cleared commercially available ELISA, a FDA cleared commercially available immunofluroscent (IFA) test and a FDA cleared commercially available flow cytomcery based immunoassay. For each sample, a consensus based algorithm (2/3) was used to determine the predicate result for comparison with the Plexus VCA IgM result. The Plexus EBV Heterophile IgM analyte was tested against a FDA cleared heterophile rapid test. The studies were conducted at three United States testing sites: a hospital laboratory located in Northeast. a pediatric hospital laboratory located in the Mid-West, and Focus with serum samples in which EBV tests were ordered. The sera were sequentially submitted to the laboratory, archived, and masked, Samples were collected at three sites and include both prospective (n = 723) and retrospective (n = 150) specimens. The Plexus EBV IgM tests were run in conjunction with the Plexus EBV IgG tests for a complete antibody profile. The samples were then classified into EBV infection status using the Serological Status table above.

Prospective Population Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte (N = 723) Samples were collected and tested by the Northeast investigator (n = 350), Mid-West investigator (n=249) and Focus (n=124).

The following table outlines the positive and negative percent agreements for prospective samples for VCA IgM analyte when the consensus predicate is used for VCA IgM analysis.

Table 4: EBV VCA IgM Results
Consensus PredicatenPositiveEquivocalNegative% Agreement
Positive88793685.9%(79/92), 95% CI:77.3-91.6%
Negative63012161797.8%(617/631), 95% CI:96. 3-98.7%

7

Image /page/7/Picture/1 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in large, bold, sans-serif font. Below the word "FOCUS" is a horizontal line, and below the line is the word "Diagnostics" in a smaller, sans-serif font. To the left of the word "FOCUS" is a curved, crescent-shaped graphic element.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M

Prepared Date: July 22, 2008

Page 8 of 13

Table 4: EBV VCA IgM Results
Consensus PredicatenPlexus
PositiveEquivocalNegative% Agreement
No consensus15104NA

1 No consensus results: the combination of three predicates could not yield a conclusive result for these samples – a 2/3 majority could not be obtained

Prospective Population Samples: Plexus EBV vs. Heterophile Rapid Test for Heterophile IgM analyte (N = 723) Samples were collected and tested by the Northeast investigator (n = 350), Mid-West investigator (n=249) and Focus (n=124).

The following table outlines the positive percent agreements for prospective samples for Heterophile IgM analyte when the Heterophile rapid test is used as predicate for Heterophile IgM analysis and consensus predicate is used for VCA IgM analysis.

Table 5: EBV Heterophile Results
Predicate Rapid TestnPositiveEquivocalNegative% Agreement
Positive756001580%(60/75), 95% CI:69.6-87.5%
Equivocal0000N/A
Negative6487064198.9%(641/648), 95% CI:97.8-99.5%

Prospective Population Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte (by Serological Status) (N = 723)

Samples were collected and tested by the Northeast investigator (n = 350), Mid-West investigator (n=249) and Focus (n=124).

The following table outlines the positive percent agreements across various serological classifications for prospective samples for VCA IgM analyte when the consensus predicate is used for VCA IgM analysis.

Table 6: EBV VCA IgM Results
Serostatus by PredicatesConsensus PredicatenPlexus% Agreement
PositiveEquivocalNegative
AcutePositive59562194.9%(56/59), 95% CI:86.1-98.3%
Negative1001100%(1/1), 95% CI:20.7-100%
No consensus0000NA
AcutePositive1491464.3%(9/14), 95% CI:38.8-83.7%
Negative58205696.6%(56/58), 95% CI:88.3-99%
No consensus0000NA
RecoveringPositive0000NA
RecoveringNegative1001100%(1/1), 95% CI:20.7-100%
RecoveringNo consensus0000NA
Previous InfectionPositive10010%(0/1), 95% CI:0-79.3%
Previous InfectionNegative2969128696.3%(286/297), 95% CI:93.5-97.9%
Previous InfectionNo consensus11100NA
No InfectionPositive110050%(1/2), 95% CI:9.5-90.5%

8

Image /page/8/Picture/1 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in bold, sans-serif font, with a curved graphic element above and to the left of the word. Below "FOCUS" is the word "Diagnostics" in a smaller, sans-serif font, underlined with a thin line. The overall design is simple and professional.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008

Page 9 of 13

Table 6: EBV VCA IgM Results
Consensus PredicatePlexus
Serostatus by
PredicatesnPositiveEquivocalNegative% Agreement
No InfectionNegative22500225100%(225/225), 95% CI:98.3-100%
No InfectionNo consensus11001NA
IndeterminatePositive13130081.3%(13/16), 95% CI:57.0-93.4%
IndeterminateNegative49104898%(48/49), 95% CI:89.3-99.6%
IndeterminateNo consensus13003NA

1 No consensus results: the combination of three predicates could not yied a conclusive result for these sumples - a 2/3 majority could not he obtained.

Prospective Population Samples: Plexus EBV vs. Heterophile IgM analyte (by Serological Status) (N = 723)

Samples were collected and tested by the Northeast investigator (n = 350), Mid-West investigator (n=249), and Focus (n=124).

The following table outlines the positive percent agreements across various scrological classifications for prospective samples for Heterophile IgM analyte when the Heterophile rapid test is used as predicate for Heterophile IgM analysis and consensus predicate is used for VCA IgM analysis.

| Predicate Heterophile

Rapid TestPlexusTable 7: EBV Heterophile IgM Results
Serological Status by
PredicatesPrimary AcutenPositiveEquivocalNegative% Agreement
Positive51480394.1%(48/51), 95% CI:84.1-98%
Primary
AcuteEquivocal0000NA
Negative910888.9%(8/9), 95% CI:56.5-98%
AcutePositive520340%(2/5), 95% CI:11.8-76.9%
Late
AcuteEquivocal0000NA
Negative67106698.5%(66/67), 95% CI:92-99.7%
RecoveringPositive0000NA
Equivocal0000NA
Negative1001100%(1/1), 95% CI:20.7-100%
Previous InfectionPositive0000NA
Equivocal0000NA
Negative2982029699.3%(296/298), 95% CI:97.6-99.8%
No InfectionPositive0000NA
Equivocal0000NA
Negative2273022498.7%(224/227), 95% CI:96.2-99.5%
Positive19100952.6%(10/19), 95% CI:31.7-72.7%
IndeterminateEquivocal0000NA
Negative460046100%(46/46), 95% CI:92.3-100%

9

Image /page/9/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo consists of the word "FOCUS" in bold, sans-serif font, with the word "Diagnostics" in a smaller font size underneath. A curved, black shape is positioned to the left of the word "FOCUS", resembling a stylized check mark or swoosh. A horizontal line separates the word "FOCUS" from the word "Diagnostics".

K073381 510(k) Summary of Safety and Effectiveness

Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008 Page 10 of 13

Presumed Acute Retrospective Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte Samples were collected and tested by Mid-West investigator (n=150).

The following table outlines the positive and negative percent agreements for retrospective samples for VCA IgM analyte when the consensus predicate is used for VCA IgM analysis.

Consensus PredicatenPlexus% Agreement
PositiveEquivocalNegative
Positive1431400397.2%(140/144), 95% CI:93.1-98.9%
Negative530233.3%(2/6), 95% CI:9.7-70.0%
No consensus12101NA

1 No consensus results: the combination of three predicates could not yield a conclusive result for these samples - a 2/3 majority could not be obtained.

Presumed Acute Retrospective Samples: Plexus EBV vs. Heterophile Rapid Test for Heterophile IgM analyte Samples were collected and tested by Mid-West investigator (n=150).

The following table outlines the positive and negative percents for retrospective samples for Heterophile IgM analyte when the Heterophile rapid test is used as predicate for Heterophile IgM analysis and consemasus predicate is used for VCA IgM analysis.

Table 8: EBV Heterophile Results
Predicate Rapid TestnPlexus% Agreement
PositiveEquivocalNegative
Positive1129831187.5%(98/112), 95% CI:80.1-92.4%
Negative38303592.1%(35/38), 95% CI:79.2-97.3%

Presumed Acute Retrospective Samples: Plexus EBV vs. Consensus Predicate for VCA IgM analyte (by Serological Status)

Samples were collected and tested by Mid-West investigator (n=150).

The following table outlines the positive percent agreements across various scrological classifications for retrospective samples for VCA IgM analyte when the consensus predicate is used for VCA IgM analysis,

Table 9: EBV VCA IgM Results
Serological Status by
PredicatesConsensus PredicatenPlexus% Agreement
PositiveEquivocalNegative
AcutePrimary
AcutePositive1041030199%(103/104), 95% CI:94.8-99.8%
Negative100150%(1/2), 95% CI:9.5-90.5%
No consensus11100NA
AcuteLate
AcutePositive870187.5%(7/8), 95% CI:52.9-97.8%
Negative0000NA
No consensus0000NA
No InfectionPositive0000NA
No InfectionNegative210150%(1/2), 95% CI:9.5-90.5%
No InfectionNo consensus0000NA
IndeterminatePositive31300193.8%(30/32), 95% CI:79.9-98.3%

10

Image /page/10/Picture/1 description: The image shows the logo for Focus Diagnostics. The logo consists of the word "FOCUS" in large, bold, sans-serif font, with the word "Diagnostics" in a smaller font size underneath. A curved, black shape is positioned to the left of the word "FOCUS", creating a visual element that complements the text.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008 Page 11 of 13

Table 9: EBV VCA IgM Results
Serological Status by
PredicatesConsensus PredicatenPositiveEquivocalNegative% Agreement
IndeterminateNegative22000%(0/2), 95% CI:0-65.8%
IndeterminateNo consensus11001NA

4 No consensus results: the contination of three predicates could not yield a conclusive result for these samples - a 2/3 majority could not be obtained.

Presumed Acute Retrospective Samples Plexus EBV vs. Heterophile Rapid Test for Heterophile IgM analyte (by Serological Status)

Samples were collected and tested by Mid-West investigator (n=150).

The following table outlines the positive percent agreements across various serological classifications for retrospective samples for Heterophile IgM analyte when the Heterophile rapid test is used as predicate for Heterophile IgM analysis and consensus predicate is used for VCA IgM analysis.

Table 10: EBV Heterophile IgM Results
Predicate Heterophile
Rapid TestPlexus
Serological Status by
PredicatesnPositiveEquivocalNegative% Agreement
AcutePrimary
AcutePositive877521086.2%(75/87), 95% CI:77.4-91.9%
Negative19201789.5%(17/19), 95% CI:68.6-97.1%
AcuteLate
AcutePositive320166.7%(2/3), 95% CI:20.8-93.9%
Negative5005100%(5/5), 95% CI:56.6-100%
No InfectionPositive0000NA
No InfectionNegative2002100%(2/2), 95% CI:34.2-100%
IndeterminatePositive22211095.5%(21/22), 95% CI:78.2-99.2%
IndeterminateNegative12101191.7%(11/12), 95% CI:64.6-98.5%

Inter-laboratory, Intra-assay and Inter-assay Reproducibility

The interfintra-assay reproducibility and the inter-laboratory reproducibility testing were performed at three laboratories. Each of the three laboratories tested twelve samples in triplicate on five different days. The results of the study are summarized in the table below.

Table 11: Inter-laboratory, Intra-assay and Inter-assay Reproducibility
Plexus VCA IgMPlexus Heterophile Antibody
IDIntra-assay & Inter-assay
%CVInter-LabIDIntra-assay & Inter-assay
%CVInter-Lab
Mean
IndexIntra-
assayInter-
assayMean
Index% CVMean
IndexIntra-
assayInter-
assayMean
Index% CV
64.482.2%13.7%4.488.8%66.123.0%16.7%6.1214.1%
52.674.2%15.0%2.6710.0%24.463.8%15.8%4.4610.2%
22.082.4%18.7%2.0816.3%42.195.0%13.8%2.191.9%
41.243.6%13.2%1.246.7%31.206.1%15.8%1.205.2%
121.124.4%25.3%1.129.0%130.746.4%14.0%0.747.2%
80.985.2%22.1%0.9816.0%80.555.7%12.8%0.553.3%
30.776.1%28.2%0.7724.3%50.278.3%47.2%0.2747.8%
130.596.1%46.8%0.5951.8%110.145.6%20.2%0.143.8%

11

Image /page/11/Picture/1 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in bold, sans-serif font, with a curved, swooping shape to the left of the word. Below the word "FOCUS" is a horizontal line, and below that line is the word "Diagnostics" in a smaller, sans-serif font.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M

Prepared Date: July 22, 2008

Page 12 of 13

| COLLECT CONSTITUTION THE THE CONTRACTOR COLLEGAL MOLECTION CONSULTIONAL CONSULTIONAL CONSULTIONAL CONSULTIONAL CONSULTIONAL CONSULTIONAL CONSULTIONAL CONSULTIONAL CONSULTIONA | . | | Call And Call of | MARK LE WARE & L MEL LEE & | and the program of the comments of the comments of the comments of the comments of the | | | | CALL CLAND STATUS A ST SERVICE PRODUCTION CONTROLLERS CONTRACTOR COLLECTION CONSULTER CONSULTER CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION C
11-11-11-11-2012-01-21-2-14-14-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1-1 | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-----|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------|--------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | | (0) | | 10/0 | | CHICA LANDRILLERIE & SUITE SHOW E MONDER BLOCK & L L POR SE & L L POST | NAME I | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
. MIN RE E RESE E E E R E R B UR R EN REMAIL LEWIN | .
10/0 |
| | 10/0
++ | 10/ | | ن جي معامل المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة المقارمة | | | -0- | 10/ | | 23.6% |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
11 | Carles of Children Company of Children Company Concession Company of Children
10/0 | | 10 | A0/
1114 | Free courts can and can can and can and a comments of the program | | 10 | | with county and considerer card a can a comprehens on | QU |

Inter-Lot Reproducibility

The inter-lot reproducibility was evaluated with eleven (11) samples in triplicates on three (3) lots of Plexus EBV kit. The results of the study are summarized in the table below.

Table 12: Inter-lot Reproducibility
Plexus VCA IgMPlexus Heterophile Antibody
IDMean Index%CVIDMean Index%CV
64.645.2%65.653.8%
52.837.6%24.424.5%
22.096.0%42.054.3%
41.135.0%31.125.6%
80.8716.9%130.787.2%
120.8417.5%80.5410.6%
30.7117.1%50.4437.9%
130.6024.4%110.138.4%
70.0519.4%70.0414.8%
110.0516.0%120.0420.4%
10.0419.3%10.0318.8%

Cross-Reactivity

A cross-reactivity study was performed to determine if samples disease states and other potentially cross-reactivity factors interfere with test results when tested with the Plexus EBV IgM kit. A panel of (Antinuclear Antibody test (ANA) n=28, Cytomegalovirus (CMV) n=25, Herpes Simplex Virus-1 (HSV-1) and Herpes Simplex Virus-2 (HSV-2) n =2, Rheumatoid Factor (Rh) n=29, Rubella Virus n=5, and Varicella-Zoster Virus (VZV) n=42). samples for each cross reactant were evaluated for possible cross reactivity with the Plexus EBV IgM kit for each of the two (VCA and Heterophile) IgM analyte. The test samples were also evaluated on commercially available ELISA and heterophile rapid test. The majority of all samples that elicit a negative result were also confirmed negative by the corresponding commercially available tests, showing that the Plexus EBV IgM kit did not have additional cross-reactivity

Table 13: Cross-Reactivity
Cross ReactivesNMethodEBV VCA IgMEBV Heterophile
PositiveEquivocalNegativePositiveEquivocalNegative
ANA28Plexus02260028
ELISA00280028
Discrepants220
Cytomegalovirus
(CMV)25Plexus421210024
ELISA10240025
Discrepants5114
HSV 1 & HSV 22Plexus002002
ELISA002002
Discrepants00
Rheumatoid Factor
(Rh)29Plexus40250029
29ELISA00290029

12

Image /page/12/Picture/0 description: The image shows the logo for Focus Diagnostics. The logo features the word "FOCUS" in bold, sans-serif font, with the word "Diagnostics" in a smaller font size underneath. A curved, swooping line is positioned to the left of the word "FOCUS", adding a dynamic element to the design.

510(k) Summary of Safety and Effectiveness Plexus EBV IgM Multi-Analyte Diagnostics Catalog No. MP0600M Prepared Date: July 22, 2008 Page 13 of 13

Discrepants40
Rubella5Plexus005005
ELISA005005
Discrepants00
Varicella-zoster
(VZV)42Plexus21392040
ELISA12392040
Discrepants330

1 One Equivocal Sample; 2 Two Equivocal Samples; "Three Equivocal Samples; 4 One Invalid Sample

Sample Storage and Handling

Fifteen (15) negative and positive for EBV IgM samples were used to assess the reactivity of unfrozen sample against samples that were frozen and thawed for up to five cycles. No effect was observed for any of the freezethaw cycling in either the positive or negative sample.

Interference

The test performance was evaluated with the presence of interfering substances. Four samples, two positive and two negative for EBV VCA IgM and heterophile antibody by Plexus EBV IgM were used in the study. Baseline ERV IgM levels were established for each sample before interferents were added. Two concentration levels of each interferent were used: triglycerides (1 mg/mL), albumin (6 mg/mL and 60 mg/mL), bilirubin (0.02 mg/mL and 0.2 mg/mL), and hemoglobin (22 mg/mL) The higher interferent concentrations represent levels that exceed the expected human range. All samples were evaluated to determine if these interfering substances affect the assay. No interference was observed for any of the substances in either the positive or negative sample.

13

Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG - 4 2008

Focus Diagnostics, Inc. C/O Deborah Morris 10703 Progress Way Cypress, California 90630

Re: K073381

Trade/Device Name: Plexus EBV IgM Multi-Analyte Diagnostics Regulation Number: 21 CFR 866.3235 Regulation Name: Epstein-Barr Virus Serological Reagents Regulatory Class: Class I Product Code: LJN Dated: November 30, 2007 Received: December 3, 2007

Dear Ms. Morris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

14

Page 2 - Focus Diagnostics, Inc.

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-frec number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Salez a Agre

Sally A. Hojvat. M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

15

510(k) Number (if known):

K073381

Device Name:

Plexus EBV IgM Multi-Analyte Diagnostics

Indications for Use:

Focus Diagnostics' Plexus™ EBV IgM Multi-Analyte Diagnostics test kit is intended for qualitatively detecting the presence or absence of human IgM class antibodies to viral capsid antigen (VCA), and heterophile antibodies in human sera. The test is indicated as an aid in the diagnosis of EBV infection and EBV-associated infectious mononucleosis.

The performance of this assay has not been established for use in the diagnosis of nasopharyngeal carcinoma and Burkitt's lymphoma, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment. This assay has not been evaluated for donor screening.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

the Schf
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of 1