K Number
K073272

Validate with FDA (Live)

Date Cleared
2007-12-11

(20 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The WorkstationOne™ Breast Imaging Workstation is intended for use with a regionally approved digital mammography system. The workstation displays images from multiple modalities, which include X-ray mammography MG, breast US and breast MRI. The workstation allows selection, display, manipulation, quantification, markup, print composition and media exchange of breast images. Here, quantification refers to measurements (such as area and distance) within a region of interest that the radiologist manually draws on the images. Similarly, markup refers to graphics that are manually drawn by the radiologist to indicate a region of interest. Note that the region of interest is not automatically generated by the computer.

The WorkstationOne™ Breast Imaging Workstation is intended for softcopy reading and interpretation of digital mammography images by Radiologists.

The WorkstationOne™ Breast Imaging Workstation when used for interpretation of images acquired using a FFDM image acquisition system shall display the images only on FDA-cleared high-resolution monitors. The images used for primary diagnostic reading must be in a lossless format, unless lossy formats are approved for use in digital mammography.

Device Description

The WorkstationOne™ is a diagnostic breast imaging workstation which consists of a software system that obtains breast imaging screening or diagnosis exams from PACS or Modalities; displays and manipulates the images for radiologists to perform interpretation task. The workstation also supports media exchange and film printing. Optionally, the workstation can interface with a reporting system to generate an interpretation report.

The enterprise workflow of the workstation follows IHE integration profiles, specifically, MAMMO (Mammography Image Profile) and RWP (Reporting Workflow Profile). The workstation can be configured to use mammographic specific hanging protocol and reading workflow. The workstation obtains the source images and CAD reports either as the recipient of the push of that data, or by querying and retrieving them from a PACS archive. In both models, DICOM is used. The images can only be the lossless compressed or non-compressed DICOM images.

The WorkstationOne™ is a software system that can be installed on an off-the-shelf general-purpose computer with one or two aray-scale high-resolution monitors and one color monitor. The high-resolution monitors are used to display digital mammography images and associated overlays for the purpose of primary interpretation by radiologists. The color monitor is used for selecting studies, displaying color images, navigating workflow, and other user interface elements. Optionally, a dedicated keypad or touchpad is included for ergonomic reasons.

AI/ML Overview

The provided 510(k) summary for the WorkstationOne™ Breast Imaging Workstation (K073272) does not include specific acceptance criteria or a detailed study proving the device meets acceptance criteria in the format typically associated with performance claims for AI/CAD devices. This submission predates the extensive regulatory guidance on clinical performance studies for AI/ML devices.

However, based on the information provided, we can infer the approach taken:

1. Acceptance Criteria and Reported Device Performance:

The submission explicitly states:

  • "The software testing procedure has been developed. The procedure with pass/fail criteria has been run to ensure that the product meets all the specified requirements."

This indicates that internal acceptance criteria were established, likely focusing on technical specifications, functionality, and safety rather than a clinical performance metric like sensitivity or specificity. Given that it's a diagnostic workstation for displaying and manipulating images, the "performance" would be related to:

  • Image Display Quality: Ensuring accurate and lossless display of various breast imaging modalities (MG, US, MRI) on FDA-cleared high-resolution monitors.
  • Functionality: Correct operation of selection, manipulation, quantification, markup, print composition, media exchange, and integration with PACS/modalities/reporting systems.
  • Safety: Adherence to a Risk Management Plan.

Since no specific quantitative performance metrics (e.g., sensitivity, specificity, AUC) are reported, a table comparing those to acceptance criteria cannot be generated as it would for an AI algorithm making a diagnostic claim. The "device performance" in this context refers to its ability to meet its functional and technical specifications as a workstation.

No quantitative performance criteria or results are provided in the document for comparison.

2. Sample Size for Test Set and Data Provenance:

The document describes software testing procedures, but does not specify a sample size for a "test set" of medical images or its provenance. The testing appears to be internal software validation rather than a clinical performance study using patient data.

3. Number of Experts and Qualifications for Ground Truth:

The document does not mention the use of experts to establish ground truth for a test set, nor does it specify their number or qualifications. This is likely because the device is a workstation for display and manipulation, not an AI algorithm making a diagnostic interpretation that requires expert ground truth for performance evaluation.

4. Adjudication Method for Test Set:

Since no test set with expert ground truth is described, no adjudication method is mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study is mentioned. The device is a workstation for interpretation, not an AI assisting human readers to improve their diagnostic accuracy. Its purpose is to provide the tools for interpretation.

6. Standalone (Algorithm Only) Performance Study:

No standalone performance study for an algorithm is mentioned. This device is a "Picture archiving and communications system" and a "diagnostic breast imaging workstation," meaning its primary function is to display images and provide tools for radiologists. It does not contain an AI algorithm that makes a diagnostic interpretation on its own. The mention of CAD reports being displayed indicates it can integrate with CAD systems, but it is not a CAD system itself.

7. Type of Ground Truth Used:

As established, the submission does not describe a study involving medical images requiring clinical ground truth (e.g., expert consensus, pathology, outcomes data). The "ground truth" for the software testing would have been against defined functional requirements and technical specifications.

8. Sample Size for Training Set:

No training set is mentioned or applicable as this device is not an AI/ML algorithm that learns from data.

9. How Ground Truth for Training Set Was Established:

Not applicable as there is no training set for an AI/ML algorithm.

Summary of Study (Based on 510(k) Text):

The "study" described in the 510(k) is an internal Non-clinical Test that involved:

  • Risk Management Plan: Identifying and controlling potential hazards.
  • Software Testing Procedure: Developing and executing a procedure with pass/fail criteria to ensure the product meets all specified requirements.

This type of testing focuses on:

  • Functional Verification: Ensuring all features (selection, display, manipulation, quantification, markup, printing, media exchange, PACS/modality interface) work as intended.
  • Technical Compliance: Adherence to standards like DICOM and IHE profiles.
  • Safety and Reliability: Confirming the software's stability and absence of critical bugs.

The conclusion drawn from this "study" was that "The materials provided in this 510(k) submission have demonstrated the device is as safe, as effective, and performs as well as the predicate devices." This substantial equivalence argument is based on similar technological characteristics and the non-clinical testing performed to ensure its functionality and safety met internal criteria. It's crucial to understand that this is a 2007 submission for a workstation, not an AI diagnostic algorithm, hence the absence of clinical performance metrics and studies typical for AI devices today.

{0}------------------------------------------------

K073272

Image /page/0/Picture/1 description: The image is mostly black with some white markings. In the upper portion of the image, there are some white markings that appear to be text or symbols. The white markings are small and difficult to read. The rest of the image is mostly black, with a few small white dots scattered throughout.

Three Palm Software, LLC

5. 510(k) Summary (as required by section 807.92(c))

5.1. 510(k) Submitter

510(k) owner's name: Three Palm Software, LLC Address: 367 Penn Way, Los Gatos, CA 95032 USA Phone: 408-356-3240 Name of contact person: Heidi Daoxian Zhang Date the summary was prepared: November 19, 2007

DEC 1 1 2007

5.2. Device Name

Trade name: WorkstationOne™ Breast Imaging Workstation

Common name: Picture archiving and communications system

Classification name: System, Image Processing, Radiological (21 CFR 892.2050 Product Code LLZ)

5.3. Predicate Devices

WorkstationOne™ Breast Imaging Workstation is substantially equivalent to the devices listed below:

Trade NameManufacturer510(k) Number
Seno AdvantageGeneral Electric MedicalSystemsK033400
CedaraI-ReadMammoCedara Software Corp.K040468
Secta WorkstationSectra Imtec ABK063093

5.4. Device Description

The WorkstationOne™ is a diagnostic breast imaging workstation which consists of a software system that obtains breast imaging screening or diagnosis exams from PACS or Modalities; displays and manipulates the images for radiologists to perform interpretation task. The workstation also supports media exchange and film printing. Optionally, the workstation can interface with a reporting system to generate an interpretation report.

The enterprise workflow of the workstation follows IHE integration profiles, specifically, MAMMO (Mammography Image Profile) and RWP (Reporting Workflow Profile). The workstation can be configured to use mammographic specific hanging protocol and reading workflow. The workstation obtains the source images and CAD reports either as the recipient of the push of that data, or by querying and retrieving them from a PACS archive. In both models, DICOM is used. The images can only be the lossless compressed or non-compressed DICOM images.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image is a black and white photograph. The image is mostly black, with some white speckles scattered throughout. The white speckles are concentrated in the center and on the right side of the image.

Three Palm Software, LLC

The WorkstationOne™ is a software system that can be installed on an off-the-shelf general-purpose computer with one or two aray-scale high-resolution monitors and one color monitor. The high-resolution monitors are used to display digital mammography images and associated overlays for the purpose of primary interpretation by radiologists. The color monitor is used for selecting studies, displaying color images, navigating workflow, and other user interface elements. Optionally, a dedicated keypad or touchpad is included for ergonomic reasons.

5.5. Indications for Use

The WorkstationOne™ Breast Imaging Workstation is intended for use with a regionally approved digital mammography system. The workstation displays images from multiple modalities, which include X-ray mammography MG, breast US and breast MRI. The workstation allows selection, display, manipulation, quantification, markup, print composition and media exchange of breast images. Here, quantification refers to measurements (such as area and distance) within a region of interest that the radiologist manually draws on the images. Similarly, markup refers to graphics that are manually drawn by the radiologist to indicate a region of interest. Note that the region of interest is not automatically generated by the computer.

The WorkstationOne™ Breast Imaging Workstation is intended for softcopy reading and interpretation of digital mammography images by Radiologists.

The WorkstationOne™ Breast Imaging Workstation when used for interpretation of images acquired using a FFDM image acquisition system shall display the images only on FDA-cleared high-resolution monitors. The images used for primary diagnostic reading must be in a lossless format, unless lossy formats are approved for use in digital mammography.

5.6. Technological Characteristics Comparison

WorkstationOne™ Breast Imaging Workstation has the similar technological characteristics to the equivalent devices as all support radiology workflow integration, obtaining, displaying, and manipulating of MG, US or MR images. Both the WorkstationOne™ Breast Imaging Workstation and the Sectra Workstation allow displaying of CAD reports.

5.7. Non-clinical Test

The potential hazards have been studied and controlled by a Risk Management Plan. The software testing procedure has been developed. The procedure with pass/fail criteria has been run to ensure that the product meets all the specified requirements.

5.8. Conclusions

The materials provided in this 510(k) submission have demonstrated the device is as safe, as effective, and performs as well as the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is a stylized graphic of three wavy lines, which are meant to represent the flow of people and services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 1 2007

Ms. Heidi Daoxian Zhang Vice President, Clinical Affairs Three Palm Software, LLC 367 Penn Way LOS GATOS CA 95032

Re: K073272

Trade/Device Name: WorkstationOne™ Breast Imaging Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 19, 2007 Received: November 21, 2007

Dear Ms. Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is very dark and difficult to make out any details. It appears to be a low-resolution image with a lot of noise. There are some brighter areas, but it is hard to tell what they are.

4. Indications for Use Statement

510(k) Number (if known): K073272

Device Name: WorkstationOne™ Breast Imaging Workstation

Indications for Use:

The WorkstationOne™ Breast Imaging Workstation is intended for use with a regionally approved digital mammography system. The workstation displays images from multiple modalities, which include X-ray mammography MG, breast US and breast MRI. The workstation allows selection, display, manipulation, quantification, markup, print composition and media exchange of breast images. Here, quantification refers to measurements (such as area and distance) within a region of interest that the radiologist manually draws on the images. Similarly, markup refers to graphics that are manually drawn by the radiologist to indicate a region of interest. Note that the region of interest is not automatically generated by the computer.

The WorkstationOne™ Breast Imaging Workstation is intended for softcopy reading and interpretation of digital mammography images by Radiologists.

The WorkstationOne™ Breast Imaging Workstation when used for interpretation of images acquired using a FFDM image acquisition system shall display the images only on FDA-cleared high-resolution monitors. The images used for primary diagnostic reading must be in a lossless format, unless lossy formats are approved for use in digital mammography.

Prescription Use マ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use
(21 CER 801 Subnart C (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

hontmkhang

(Division Division of Reproductive, Abdominal an Page 6 of 48 Radiological Devices 510(k) Number _

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).