K Number
K073261

Validate with FDA (Live)

Device Name
VAR MONITOR
Date Cleared
2008-05-23

(185 days)

Product Code
Regulation Number
868.5925
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is to be used by properly trained personnel to deliver emergency, short term, constant flow - pressure cycled ventilatory support.

Device Description

The VAR-Monitor is a battery (9 VDC) operated device that is mounted onto the VAR to monitor cycling conditions during operation.

AI/ML Overview

The provided text describes a 510(k) summary for the "VAR with VAR-Monitor™" device, which is a powered emergency ventilator.

Here's an analysis based on the given information:

1. Acceptance Criteria and Reported Device Performance

The core acceptance criterion for the VAR with VAR-Monitor is its compliance with ASTM F 920-93 (Reapproved 1999), "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans."

Acceptance Criteria (Standard)Reported Device Performance
Mechanical Performance: Meets or exceeds ASTM Designation: F 920-93Passed: "The VAR with VAR-Monitor meets and exceeds 'Standard
(Reapproved 1999) "Standard Specification for Minimum Performance and SafetySpecification for Minimum Performance and Safety Requirements for Resuscitators
Requirements for Resuscitators Intended for Use with Humans."Intended for Use with Humans,' ASTM Designation: F 920-93 (Reapproved 1999)."
Alarm Functionality: Detection of non-cycling condition and activation ofFunctionality Confirmed: "VAR-Monitor detects a non-cycling condition of
visual and audible alarms within a preset time of eight (8) seconds.the VAR modulator Indicating that no pressure changes have occurred. When the
VAR-Monitor is powered on and ready, the device will check for the 'pressure
signal.' If the pressure stays unchanged for more than the preset time of eight
(8) seconds, both visual and audible alarms will be activated."
Clinical Efficacy (Indirect through equivalence): Substantially equivalentSubstantially Equivalent: "In conclusion, the modified VAR with VAR-Monitor
to legally marketed predicate devices in terms of intended use,is substantially equivalent (SE) to legally marketed predicate devices."
operational characteristics, and safety.
Absence of Adverse S&E Information: No new safety or effectiveness concerns.None: "Adverse S & E Information: None"
Overall Conclusion: Passed all required tests and met all predeterminedConclusion: "The VAR with VAR-Monitor passed all required tests and
acceptance criteria for applicable standards.demonstrated that it meets all predetermined acceptance criteria for applicable
standards."

2. Sample Size Used for the Test Set and Data Provenance

The document states "Clinical Tests: None" and does not refer to a "test set" in the context of clinical data. The evaluation appears to be based on engineering tests against the ASTM standard and comparison to predicate devices, rather than a clinical study with patients or a specific dataset. Therefore, there is no mention of sample size or data provenance in this context.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

As "Clinical Tests: None" is stated, there is no indication of a test set requiring ground truth established by experts. The evaluation primarily relies on compliance with the ASTM standard and comparison to predicate devices.

4. Adjudication Method for the Test Set

Given that no clinical test set is described, there is no mention of an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. The document explicitly states "Clinical Tests: None." This type of study would fall under clinical testing.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

The device is a physical medical device (ventilator) with a monitor, not an algorithm in the typical AI sense. The performance evaluation focuses on the mechanical and alarm functions of the device itself, which are inherently "standalone" in their operation rather than involving a human-in-the-loop for interpreting algorithm outputs.

7. The Type of Ground Truth Used

The "ground truth" for the device's acceptable performance is primarily derived from:

  • Engineering Standards: Compliance with ASTM F 920-93 (Reapproved 1999) for performance and safety.
  • Operational Characteristics: Whether the device successfully detects a non-cycling condition and triggers alarms within the specified time.
  • Predicate Device Comparison: The established safety and effectiveness of the existing predicate devices (RespirTech PRO, RespirTech PRO O₂C, VAR-Plus Model PCM).

There is no mention of expert consensus, pathology, or outcomes data as "ground truth" for a specific clinical dataset.

8. The Sample Size for the Training Set

The document does not describe the development of an AI algorithm or a "training set." The device is a ventilator with an alarm monitor, evaluated against engineering standards and substantial equivalence to predicate devices, not through machine learning.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of a training set or AI algorithm.

{0}------------------------------------------------

1

510(k) Summary

K07326
1. SubmitterInformationVORTRAN Medical Technology 1, Inc.21 Golden Land Court, Sacramento, CA 95834MAY 23 2008
2. ContactInformationJames Lee, Senior Vice PresidentTEL: (800) 434-4034 FAX: (916) 648-9751
3. Trade NameVAR with VAR-Monitor™
4. Common NameVentilator, Emergency, Powered (Resuscitator)
5. Device ClassClass II
6. Product Code73 BTL
7. ProductClassificationPer CFR section 868.5925
8. ClassificationPanelAnesthesiology
9. PredicateDeviceRespirTech PRORespirTech PRO O₂CVAR-Plus Model PCM510(k) No.: K973975510(k) No.: K001430510(k) No.: K041473
10. DeviceDescriptionThe VAR-Monitor is a battery (9 VDC) operated device that ismounted onto the VAR to monitor cycling conditions duringoperation.
11. Intended UseThe device is to be used by properly trained personnel todeliver emergency, short term, constant flow - pressure cycledventilatory support.
12. SubstantialEquivalencyEvaluationThe VAR with VAR-Monitor meets and exceeds "StandardSpecification for Minimum Performance and SafetyRequirements for Resuscitators Intended for Use withHumans," ASTM Designation: F 920-93 (Reapproved 1999).
13. OperationalCharacteristicsVAR-Monitor detects a non-cycling condition of the VARmodulator Indicating that no pressure changes have occurred.When the VAR-Monitor is powered on and ready, the devicewill check for the "pressure signal." If the pressure staysunchanged for more than the preset time of eight (8) seconds,both visual and audible alarms will be activated.

{1}------------------------------------------------

510(k) Summary

  • The VAR with VAR-Monitor provides short term, pressure 14.Clinical cycled, constant flow ventilatory support using either pressure Application control or pressure support. The device is not an ICU stand alone ventilator. Set up and use of the VAR with VAR-Monitor is simple. Set desired flow and adjust pressure dial to obtain desired I-time and/or tidal volume (see tidal volume chart in instructions), and adjust rate dial to obtain desired rate and I to E ratio for the VAR. Power on the VAR-Monitor to monitor the VAR cycling and detect any VAR non-cycling condition. This non-cycling condition will trigger the alarm (audible and visual) to notify attending personnel of an abnormal condition. Possible changing conditions could include but are not limited to: (i) increases or decreases in supply gas flow or no gas flow because tank is empty or has been turned off; (ii) patlent disconnect or airway leak; or (iil) a change in patient All these conditions may cause the VAR to stop condition. cycling or stay in a spontaneous "pressure assisted" mode, waiting for the patient to trigger the next breath.
    1. Clinical Tests None
  • 16.Adverse S & E None Information
  • The VAR with VAR-Monitor passed all required tests and 17. Conclusion demonstrated that it meets all predetermined acceptance criteria for applicable standards. In conclusion, the modified VAR with VAR-Monitor is substantially equivalent (SE) to legally marketed predicate devices.

[Signature]

James Lee

[Typed Name]

November 16, 2007

[Dated]

Senior Vice President

[Title]

VAR-P 510k Summary

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird with outstretched wings. The bird is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 3 2008

Mr. James Lee Senior Vice President VORTRAN Medical Technology 1, Incorporated 21 Golden Land Court Sacramento, California 95834

Re: K073261 Trade/Device Name: VAR with VAR-Monitor Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: II Product Code: BTL Dated: April 23, 2008 Received: April 28, 2008

Dear Mr. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Lee

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Svetta y Michael ms

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K073261

Device Name: VAR with VAR-Monitor

Indications For Use:

The device is to be used by properly trained personnel to deliver emergency, short term, constant flow - pressure cycled ventilatory support.

Prescription Use 2 (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

i Vhm

Page 1 of 1

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: 1607326

§ 868.5925 Powered emergency ventilator.

(a)
Identification. A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.(b)
Classification. Class II (performance standards).