K Number
K073118
Date Cleared
2008-03-05

(121 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Prodigy Voice Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus.

The alternative site testing in this system can be used only during steady-state blood glucose conditions.

This system contains a speaking functionality which provides step by step instructions to aid visually impaired persons.

Device Description

Prodigy Voice Blood Glucose Monitoring System consists of a meter and test strips. The system utilizes an electrochemical method-based meter and dry reagent biosensor (test strips) for blood glucose testing. The size of the current is proportional to the amount of glucose present in the sample, providing a quantitative measurement of glucose in fresh whole blood and control solutions.

This system contains speaker function which provides step by step instructions.

AI/ML Overview

The provided text offers minimal details regarding the acceptance criteria and the study that proves the device meets those criteria. Here's a breakdown of what can be extracted and what information is missing:

1. Table of Acceptance Criteria and Reported Device Performance:

The document explicitly states: "The studies demonstrated that the performance of this system meets its intended use." However, it does not provide specific acceptance criteria (e.g., accuracy percentages, precision metrics, bias limits) nor detailed performance results to populate such a table.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not specified.
  • Data Provenance: The studies were conducted "in the laboratory and in clinical settings by healthcare professionals and lay users." There is no mention of the country of origin or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not specified. The document only mentions "healthcare professionals" conducted studies.

4. Adjudication Method for the Test Set:

Not specified.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

Not mentioned or implied. The focus appears to be on the device's standalone performance rather than comparing human readers with and without AI assistance.

6. Standalone Performance (Algorithm Only without Human-in-the-loop Performance):

Yes, this was undoubtedly done. The "Performance Studies" section states, "The performance of Prodigy Voice Blood Glucose Monitoring System was studied in the laboratory and in clinical settings..." This implies the device was tested as a standalone system.

7. Type of Ground Truth Used:

While not explicitly stated, for a blood glucose monitoring system, the ground truth would typically be established by a reference method (e.g., a laboratory-grade blood glucose analyzer that is traceable to an international standard). The text mentions "quantitative measurement of glucose," which necessitates a highly accurate reference for comparison.

8. Sample Size for the Training Set:

Not specified.

9. How the Ground Truth for the Training Set Was Established:

Not specified.


Summary of Available Information from the Provided Text:

FeatureDetails from the Text
Acceptance Criteria & Reported Performance- Acceptance Criteria: NOT SPECIFIED. The document only states "The studies demonstrated that the performance of this system meets its intended use." - Reported Device Performance: NOT DETAILED. No specific metrics (e.g., accuracy, precision) or numerical results are provided.
Test Set Sample SizeNOT SPECIFIED.
Data Provenance (Test Set)"in the laboratory and in clinical settings by healthcare professionals and lay users." - Country of Origin: Not specified. - Retrospective/Prospective: Not specified.
Number & Qualifications of Experts (Ground Truth - Test Set)NOT SPECIFIED. Mentions "healthcare professionals" but no details on number, specialties, or experience.
Adjudication Method (Test Set)NOT SPECIFIED.
MRMC Comparative Effectiveness StudyNO. Not mentioned or implied.
Standalone PerformanceYES. Implied by "Performance Studies... in the laboratory and in clinical settings."
Type of Ground Truth UsedReference Method (e.g., laboratory-grade glucose analyzer). While not explicitly named, the nature of glucose measurement requires a highly accurate and precise reference method for comparative ground truth.
Training Set Sample SizeNOT SPECIFIED.
How Ground Truth for Training Set Was EstablishedNOT SPECIFIED.

Conclusion:

The provided 510(k) summary (a premarket notification for demonstrating substantial equivalence to a predicate device) is high-level. It confirms that performance studies were conducted and that the device met its intended use, which is sufficient for 510(k) clearance. However, it does not provide the detailed scientific and clinical evidence, including specific acceptance criteria, performance metrics, sample sizes, or ground truth methodologies that would typically be contained in the full submission to the FDA. The purpose of this summary is to briefly introduce the device and its regulatory status, not to serve as a comprehensive research paper.

{0}------------------------------------------------

MAR - 5 2008

510 (k) Summary

Page 1-of-3

Company nameDIAGNOSTIC DEVICES INC.
Contact personRick Admani Abulhaj
Address9300 Harris Corners ParkwaySuite 450, Charlotte, NC 28269
Phone1-704-285-6400
FAX1-704-285-6475
  1. Name of Device Trade Names

Common Names/Descriptions

Classification Names

Prodigy Voice Blood Glucose Monitoring System Blood Glucose Meter Blood Glucose Test Strips Class II devices (21 CFR Section 862.1345, Glucose Test System)

  1. Predicate Device Trade/Proprietary Name: Common/Usual Name:

Submitter 510 (k) Number Prodigy Blood Glucose Test System Blood Glucose Meter Blood Glucose Test Strips DIAGNOSTIC DEVICES INC. K060467

{1}------------------------------------------------

4. Device Description

Prodigy Voice Blood Glucose Monitoring System consists of a meter and test strips. The system utilizes an electrochemical method-based meter and dry reagent biosensor (test strips) for blood glucose testing. The size of the current is proportional to the amount of glucose present in the sample, providing a quantitative measurement of glucose in fresh whole blood and control solutions.

5. Intended Use

Prodigy Voice Blood Glucose Monitoring System is indicated for the quantitative measurement of glucose in fresh capillary whole blood taken from the finger and the alternative sites for self testing by persons with diabetes in the home or by healthcare professionals in healthcare facilities. Testing is done outside the body (in vitro diagnostic use).

The alternative site testing (the plam, the forearm, the upper arm, the calf and the thigh) in this system can be used only during steady-state blood glucose conditions.

This system contains speaker function which provides step by step instructions.

6. Comparison to Predicate Device

Prodigy Voice Blood Glucose Monitoring System has equivalent technological characteristics as the Prodigy Blood Glucose Test System (K060467). Prodigy Voice Blood Glucose Monitoring System also has the similar intended use as the Prodigy Blood Glucose Test System.

7. Performance Studies

The performance of Prodigy Voice Blood Glucose Monitoring System was studied in the laboratory and in clinical settings by healthcare professionals and lay users. The studies demonstrated that the performance of this system meets its intended use.

{2}------------------------------------------------

Page 3-of-3

8. Conclusion

.

Prodigy Voice Blood Glucose Monitoring System demonstrates satisfactory performance and is suitable for its intended use.

:

:

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three wing-like shapes. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle symbol.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Diagnostic Devices Inc. c/o Ms. Erica Li Management Representative 4F, No. 88, Sec. 1 Kwang-Fu Road San-Chung, Taipei County China (Taiwan) 241

MAR - 5 2008

Re: K073118

Trade/Device Name: Prodigy Voice Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system. Regulatory Class: Class II Product Code: NBW, CGA Dated: February 13, 2008 Received: February 14, 2008

Dear Ms. Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment A

Indications for Use

510(k) Number:

Image /page/5/Picture/3 description: The image shows a handwritten string of characters. The string appears to be a combination of letters and numbers. The characters are 'K073118'.

Device Name: Prodigy Voice Blood Glucose Monitoring System

Indications for Use:

The Prodigy Voice Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus.

The alternative site testing in this system can be used only during steady-state blood glucose conditions.

This system contains a speaking functionality which provides step by step instructions to aid visually impaired persons.

Prescription Use Over-The-Counter Use __ X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol Benson
Sign-Off

Page 1 of

Office of In Vitro Diagnostic Device
Evaluation and Safety

K073118

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.