K Number
K073030
Date Cleared
2008-07-21

(269 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AG-605 & AG-606 Blood Glucose Monitoring System is intended for in vitro diagnostic use. The system is intended to be used for the quantitative measurement of capillary whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

AG-605 & AG-606 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for both lay uses by people with diabetes and in a clinical setting by health care professionals, as an aid to monitoring levels in Diabetes Mellitus. Not for use on neonates.

Device Description

AG-605 & AG-606 Blood Glucose Monitoring System consists of a blood glucose meter, test strips, two levels of control solutions, lancets and lancing device.

AG-605 & AG-606 Blood Glucose Monitoring System is designed to provide an easy, accurate method for determining capillary blood glucose values. This analysis is based on amperometric technology using glucose oxidase that is specific for the blood glucose measurement. When the blood sample is applied to the test strip, electrons are formed by the reaction between glucose oxidase and blood glucose. The electrical current is measured by the meter and correlates with the concentration of glucose in the blood sample.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the AG-605 & AG-606 Blood Glucose Monitoring System, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary explicitly states that clinical tests were performed and complied with the accuracy requirements of ISO 15197. It also states, "The results meet or exceed the accuracy requirements of ISO 15197." However, the exact numerical acceptance criteria from ISO 15197 and the device's specific numerical performance are not provided in the summary.

To present this in a table, we can only state the standard referenced:

Acceptance CriteriaReported Device Performance
ISO 15197 Accuracy RequirementsMeets or exceeds ISO 15197 accuracy requirements

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated in the provided text. The document only mentions "Clinical Tests were performed."
  • Data Provenance: Not explicitly stated. It's implied to be from a clinical setting for "capillary whole blood from the fingertips," but the country of origin or whether it was retrospective/prospective is not mentioned.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the 510(k) summary. For a blood glucose monitoring system, the "ground truth" is typically established by comparative laboratory reference methods, not human expert consensus in the same way it would be for image-based diagnostic devices.

4. Adjudication Method for the Test Set

This information is not applicable/provided. Adjudication methods (like 2+1, 3+1) are typically used in studies involving human readers interpreting medical images or data. For a blood glucose meter, the primary ground truth is established through precise laboratory instrumentation.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was it done?: No. An MRMC study is not relevant for a blood glucose monitoring system. This type of study assesses how human readers' performance (e.g., diagnostic accuracy) changes with or without AI assistance, typically in image analysis.
  • Effect Size: Not applicable.

6. Standalone Performance Study (Algorithm Only)

  • Was it done?: Yes, implicitly. The performance of the AG-605 & AG-606 Blood Glucose Monitoring System, as described to meet ISO 15197, is inherently its standalone (algorithm only) performance, as it is a device designed to provide a direct measurement without human interpretation beyond reading the displayed result. The "clinical tests" reported would reflect this.

7. Type of Ground Truth Used

The ground truth for blood glucose monitoring systems is typically established using laboratory reference methods (e.g., YSI glucose analyzer) that are highly accurate and traceable to international standards. While the summary doesn't explicitly name the reference method, "ISO 15197" implicitly refers to a comparison against such a method to establish the accuracy of the device.

8. Sample Size for the Training Set

This information is not provided. Blood glucose meters are not typically "trained" in the same way machine learning algorithms are. The "device" is a biosensor-based system with a fixed enzymatic reaction and electrical current measurement, rather than a learning algorithm.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided for the same reasons as point 8. The "ground truth" (i.e., known glucose concentrations) would be used during the device's development, calibration, and manufacturing quality control, but not in the context of "training data" for a machine learning model.

{0}------------------------------------------------

510(k) Summary

JUL 2 1 2008

ldentification of the submitter:

Submitter:Andon Health Co., LTDNo 31, Changjiang Road, Nankai District, Tianjin,P.R. China, 300193
Telephone number:86-22-6052 6161
Fax number:86-22-6052 6162
Contact:Liu Yi
Date of Application:08/10/07

Identification of the product:

Trade name:AG-605 Blood Glucose Monitoring SystemAG-606 Blood Glucose Monitoring System
Common name:Glucose test System
Classification:Blood Glucose Meter and test strip are Class IIdevices (21 CFR 862.1345, Glucose Monitor)
Product code:CGA

Predicate Device:

One Touch Ultra Blood Glucose Monitoring system of Life Scan

K024194 510k number:

Device description:

AG-605 & AG-606 Blood Glucose Monitoring System consists of a blood glucose meter, test strips, two levels of control solutions, lancets and lancing device.

AG-605 & AG-606 Blood Glucose Monitoring System is designed to provide an easy, accurate method for determining capillary blood glucose values. This analysis is based on amperometric technology using glucose oxidase that is specific for the blood glucose measurement. When the blood sample is applied to the test strip, electrons are formed by the reaction between glucose oxidase and blood glucose. The electrical current is measured by the meter and correlates with the concentration of glucose in the blood sample.

{1}------------------------------------------------

Intended use:

AG-605 & AG-606 Blood Glucose Monitoring System is intended for in vitro diagnostic use. The system is intended to be used for the quantitative measurement of capillary whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Comparison to Predicate Devices(s):

:

CHARACTERISTICSNEW DEVICE:AG-605 & AG-606 BloodGlucose Monitoring SystemPREDICATE:One Touch Ultra BloodGlucose Monitoring system(K024194)
Type of MeterBiosensor (Electrode)Biosensor (Electrode)
Intended UseTo quantitatively measure glucosein fresh capillary whole blood.To quantitatively measure glucosein fresh capillary whole blood.
Sample SourceCapillary whole bloodCapillary whole blood
Sample ApplicationBlood sample is placed directly tothe test strip after finger is lanced.Blood sample is placed directly tothe test strip after finger is lanced.
Hematocrit Range30-55%30-55%
Operating Temperature Range10℃~40℃(50°-104°F)6℃~44℃(43°-111°F)
DimensionsAG-605: 50mmx 112mmx 22mmAG-606: 82 mm×59mm × 20mm3.12" X2.25" X0.85"
WeightAG-605: 60g (exclude batteries)AG-606: 55g (exclude batteries)1.5 ounces (43g) with battery
DisplayLCDLCD
Result Presentationmg/dL or mmol/Lmg/dL or mmol/L
Memory CapabilitiesAG-605: 160 times with time anddate displayingAG-606: 350 times with time anddate displaying150 blood glucose and controlsolution tests
Test StartAutomaticAutomatic
Test Time5 second5 second
Power SourceDC3V (2XAAA batteries)One replaceable 3.0v lithiumbattery
Battery LifeApprox. 1000 normal tests1000 tests
Measurement Range20mg/dL-600mg/dL(1.1mmol/L~33.3mmol/L)20mg/dL-600mg/dL(1.1mmol/L~33.3mmol/L)
Qualified Test StripAGS-600 Test StripONETOUCH Ultra Test Strip
Sample VolumeMinimum 1 micro literMinimum 1 micro liter

{2}------------------------------------------------

Summary:

The information provided in this pre-market notification demonstrates that AG-605 & AG-606 Blood Glucose Monitoring System is substantially equivalent to One Touch Ultra Blood Glucose Monitoring system. Substantial equivalent was demonstrated through comparison of intended use and physical properties to the commercially available and analytical predicate devices. The information supplied in this pre-market notification provides reasonable assurance that the AG-605 &AG-606 Blood Glucose Monitoring System is safe and effective for its stated intended use.

Clinical Tests:

Clinical tests were performed and complied the accuracy requirements of ISO 15197. The results meet or exceed the accuracy requirements of ISO 15197.

Non-clinical Tests:

All non-clinical tests coincide the following standards, including Product Safety test and Electromagnetic Compatibility test.

IEC 61010-1-2001

Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements

IEC 61010-2-101-2002

Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment

EN 61326

Electrical equipment for measurement, control and laboratory use -EMC requirements

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with its wings forming a curved shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Andon Health Co., Ltd. c/o Ms. Mona Technologi Department 7th Floor Hua Qiao Chuang Ye Plaza No. 10 JinPing Road, Ya An Road Nankai District, Tianjin, P.R. China 300193

JUL 2 1 2008

Re: K073030

Trade Name: AG-605 Blood Glucose Monitoring System, AG-606 Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Monitoring System Regulatory Class: Class II Product Codes: NBW, CGA, JJX Dated: July 9, 2008 Received: July 14, 2008

Dear Mona:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use

510(k) Number (if known): K073030

Device Name: AG-605 Blood Glucose Monitoring System AG-606 Blood Glucose Monitoring System

Indication For Use:

AG-605 & AG-606 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for both lay uses by people with diabetes and in a clinical setting by health care professionals, as an aid to monitoring levels in Diabetes Mellitus. Not for use on neonates.

Prescription Use (21 CFR Part 801 Subpart D) Over the Counter Use Yes (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

And/Or

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Sean Cooper
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K073036

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.