(268 days)
Not Found
No
The device description details a turbidimetric assay based on antibody-antigen reactions and absorbance measurements, which is a standard laboratory technique and does not mention or imply the use of AI or ML. The "Mentions AI, DNN, or ML" field is explicitly marked as "Not Found".
No.
This device is an in vitro diagnostic assay used to quantify apolipoprotein A-I, which can aid in assessing the risk of coronary artery disease, but it does not treat or prevent a disease.
Yes
The 'Intended Use / Indications for Use' section explicitly states, "For in vitro Diagnostic use." and "It can be used as an aid for assessing the risk of coronary artery disease," which indicates its role in disease diagnosis or risk assessment.
No
The device description clearly indicates a method based on a chemical reaction and turbidimetric measurement, which requires physical reagents and an instrument (spectrophotometer or similar) to perform the assay. This is an in vitro diagnostic device with hardware components, not a software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Explicit Statement: The "Intended Use / Indications for Use" section clearly states: "For in vitro Diagnostic use." This is the most direct indicator.
- Nature of the Assay: The assay measures a substance (apolipoprotein A-I) in a biological sample (serum) to provide information for assessing a health condition (risk of coronary artery disease). This is the fundamental purpose of an in vitro diagnostic device.
- Components: The description includes components like calibrators and controls, which are standard for IVD assays to ensure accurate and reliable results.
- Predicate Device: The mention of a predicate device (K993345; K-Assay Apo AI Assay) which is also an Apo AI assay, further supports its classification as an IVD, as predicate devices are used for comparison in regulatory submissions for new IVDs.
N/A
Intended Use / Indications for Use
The Diazyme Apolipoprotein A-I Assay is intended for the in vitro quantitative determination of apolipoprotein A-I (apo A-I) in serum. It can be used as an aid for assessing the risk of coronary artery disease. For in vitro Diagnostic use.
Calibrator: For calibration of the Diazyme Apolipoprotein A-I Assay in serum. For in vitro Diagnostic Use.
Controls: To monitor the performance of Diazyme Apolipoprotein A-I Assay in serum. For in vitro Diagnostic Use.
Product codes (comma separated list FDA assigned to the subject device)
MSJ, JIT, JJX
Device Description
Not Found
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance:
Reportable Range:
Serum: 20- 228 mg/dL
Precision/Serum:
Within Run: 0.6% -0.9%
Total: 1.8%-3.3%
Accuracy/Serum:
Correlation Coefficient: 0.9789
Slope/Intercept:
y = 0.91x + 2.813
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1475 Lipoprotein test system.
(a)
Identification. A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.
0
JUL 1 6 2008
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807,92
Submitter's name: | Diazyme Laboratories |
---|---|
Submitter's address: | 12889 Gregg Court |
Poway, CA 92064 | |
USA | |
Name of Contact Person: | Dr. Abhijit Datta |
Diazyme Laboratories | |
12889 Gregg Court | |
Poway, CA 92064 | |
Phone: 858-455-4762 | |
Fax: 858-455-2120 | |
Date the Summary was Prepared: | September 24, 2007, June 5, 2008, and July 8, 2008 |
Name of the Device | Diazyme Apolipoprotein A-I Assay |
Trade Name: | Diazyme Apolipoprotein A-I Assay |
Common/Usual Name | Apolipoprotein A-I Test System |
Device Classification Name | Lipoprotein test system |
Product code: | MSJ (reagent), JIT (calibrator), JJX (control) |
Submission Type | |
============================================================================================================================================================================== | 510k |
Regulation Number | 862.1475(reagent), 862.1150(calibrator), 862.1660(control) |
Device Class | II (reagent), II (calibrator), I Reserved (control) |
Predicate Device: | For the Alpha-1-lipoprotein Immunological Test System |
Lipoprotein test system, we are claiming equivalence | |
[807.92(a) (3) to K-ASSAY APO AI ASSAY (K993345) | |
manufactured by Kamiya Biomedical Company |
1
Substantial Equivalence Information
l. Predicate device name(s):
K-Assay Apo AI Assay
2. Predicate 510(k) number(s): K993345
3. Comparison with predicate:
Indications for Use
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay | Equivalency |
---|---|---|
The Diazyme Apolipoprotein A-I Assay is intended for the in vitro quantitative determination of apolipoprotein A-I (apo A-I) in serum. It can be used as an aid for assessing the risk of coronary artery disease. | For the quantitative determination of human Apolipoprotein AI (Apo AI) in serum by immunoturbidimetric assay. | Same |
Principle
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay | Equivalency |
---|---|---|
This method is based on the reaction of a sam- | ||
ple containing human Apo A-I and specific | ||
antiserum to form an insoluble complex which | ||
can be measured turbidimetrically at 340nm. | ||
By constructing a standard curve from the ab- | ||
sorbance of standards the concentration of Apo | ||
A-1 can be determined. | This method quantifies Apolipoprotein AI | |
based on immunoturbidimetric assay. The re- | ||
agent uses a goat polyclonal antibody specific | ||
for human Apolipoprotein AI. The antibody | ||
binds to the Apo AI in the serum forming | ||
light scattering immune complexes, which | ||
increase the turbidity of the sample. Since the | ||
turbidity is proportional to the amount of Apo | ||
AI in the sample, the Apolipoprotein AI con- | ||
centration can be determined by measuring | ||
this increase in turbidity. The increase in tur- | ||
bidity is measured at 800 nm. Apolipoprotein | ||
AI in the sample is quantitatively determined. | Similar |
Test Objective
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay | Equivalency |
---|---|---|
The Diazyme LDL-Cholesterol Assay is in- | ||
tended for the in vitro quantitative determina- | ||
tion of Low Density Lipoprotein Cholesterol | ||
in human serum or plasma. | For the quantitative determination of human | |
Apolipoprotein AI (Apo AI) in serum by im- | ||
munoturbidimetric assay. | Same | |
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay | Equivalency |
Quantitative | Quantitative | Same |
:
2
Specimen Type
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay | Equivalency |
---|---|---|
Human serum | Human serum | Same |
Product Type
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay | Equivalency |
---|---|---|
Calibrator, Reagent, Instrument | Calibrator, Reagent, Instrument | Same |
Performance
Performance | |
---|---|
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay |
Reportable Range: | Reportable Range: |
Serum: 20- 228 mg/dL | Serum: 20 – 300 mg/dL |
Precision/Serum: | Precision/Serum: |
Within Run: 0.6% -0.9% | Within Run: 1.51% -1.87% |
Total: 1.8%-3.3% | Total: 1.05%-1.57% |
Accuracy/Serum: | Accuracy/Serum: |
Correlation Coefficient: 0.9789 | Correlation Coefficient: 0.970 |
Slope/Intercept: | Slope/Intercept: |
$y = 0.91x + 2.813$ | $y = 0.980x + 4.776$ mg/dL |
Calibrator Comparison
Calibrator Comparison | ||
---|---|---|
Diazyme Apolipoprotein A-I Assay | K-Assay Apo AI Assay | Equivalency |
Lyophilized form | Lyophilized form | Same |
Diazyme Apolipoprotein A-I calibrator value is | ||
traceable to the WHO International Reference | ||
Material for Apo AI, Sp1-01. | K-Assay Apo AI calibrator value is traceable | |
to the WHO International Reference | ||
Material for Apo AI, Sp1-01. | Same |
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
General Atomics Diazyme Laboratories c/o Mr. Charles Yu. Quality System Manager 12889 Gregg Court Poway, California 92064
K072977 Trade Name: Diazyme Apolipoprotein A-I Assay; Diazyme Apolipoprotein A-I Calibrator, and Diazyme Apolipoprotein A-I Control, Regulation Number: 21 CFR 862.1475 Regulation Name: Lipoprotein test system Regulatory Class: Class I, meets the limitation to exemption in 862.9(c)(4) Product Code: MSJ, JIT, JJX Dated: June 5, 2008 Received: June 17, 2008
JUL 1 6 2008
Dear Mr. Yur.
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
4
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.v.M.
Yéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number: | K072977 |
---|---|
Device Name: | Diazyme Apolipoprotein A-I Assay |
Indications for Use: | The Diazyme Apolipoprotein A-I Assay is intended for the in vitro quantitative |
determination of apolipoprotein A-I (apo A-I) in serum. It can be used as an | |
aid for assessing the risk of coronary artery disease. For in vitro Diagnostic | |
use. | |
Calibrator: | For calibration of the Diazyme Apolipoprotein A-I Assay in serum. For in vitro |
Diagnostic Use. | |
Controls: | To monitor the performance of Diazyme Apolipoprotein A-I Assay in serum. |
For in vitro Diagnostic Use. |
7
Division/Sign-Off
Office of In Vir o Dingnostic Device Evaluation and Unitedy
510(k) K072977
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)