K Number
K072777

Validate with FDA (Live)

Date Cleared
2008-01-24

(118 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CIBA Vision® (nelfilcon A) Spherical and Toric One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes.

CIBA Vision® (nelfilcon A) Progressives One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.

The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.

Device Description

The nelfilcon A lens material is 69% water and 31% nelfilcon A polymer (polyviny) alcohol partially acetalized with N-formylmethyl acrylamide). The lenses are tinted from edge to edge for visibility purposes with the color additive copper phthalocyaninks (UD).

Nelfilcon A lens designs currently include spherical, toric, and multifocal in the following parameter ranges:
Power Range: -20.00D to +20.00D
Center Thickness: 0.010 mm for -3.00D spherical (varies with power)

Lenses have the following properties:
Refractive index: 1.38
Light transmittance: approximately 96 %T
Water content : 69% by weight
Oxygen permeability . 26 barrer measured at 35°C (single point Dk-Polarographic method)

Lenses are supplied sterile in sealed blister-packs containing buffered saline.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Nelfilcon A Soft Contact Lenses for Daily Wear

1. Table of Acceptance Criteria and Reported Device Performance

The document describes "Substantial Equivalence" as the primary acceptance criterion, meaning the device must perform comparably to a predicate device. Specific quantitative acceptance criteria are not explicitly detailed in threshold values (e.g., "must achieve X% vision improvement"). Instead, the performance is described relative to the predicate device and established standards for contact lenses.

Acceptance Criteria (Demonstrates Substantial Equivalence To Predicate)Reported Device Performance
Non-clinical:
Lens physical and material properties are substantially equivalent.Demonstrated.
Lens material and extracts are substantially equivalent, non-toxic, and non-irritating.Demonstrated.
Clinical:
Similar overall performance in clinically relevant areas:Demonstrated.
- VisionSatisfactory.
- HealthSatisfactory.
- ComfortSatisfactory.
- FitSatisfactory.
Overall Equivalence:
Comparable to predicate device and other Group II soft hydrophilic contact lenses in:Demonstrated.
- Water content (69% water)(Matches predicate)
- Ionic characteristics (FDA Group II: high water, nonionic)(Matches predicate)
- Clinical performanceDemonstrated.
- Indications for useDemonstrated.
Any differences do not adversely affect safety and effectiveness.Demonstrated.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "a clinical study" and "an investigation" but does not provide the number of subjects or eyes included in the clinical test set.
  • Data Provenance: Not explicitly stated regarding country of origin. The study was a "daily disposable wear investigation." The document implies that the study was prospective, as it "investigated" the lens and was performed to "demonstrate" substantial equivalence.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the document. For contact lens studies demonstrating substantial equivalence, "ground truth" often refers to clinical assessments by ophthalmologists or optometrists, but the number and qualifications of such experts are not detailed here.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study, as typically understood in AI/medical imaging, was not performed. The study compared the device's clinical performance to an "FDA cleared and commercially available contact lens" (the control/predicate), but this is a direct clinical comparison, not a study of human readers improving with AI assistance. The device itself is a contact lens, not an AI system.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

N/A (Not Applicable). The device is a physical medical device (contact lens), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant. The "standalone" performance would be the lens's inherent physical and optical properties, which were part of the non-clinical testing.

7. The Type of Ground Truth Used

For the non-clinical tests, the "ground truth" would be established physical and material property measurements, and standard biocompatibility assays. For clinical testing, the "ground truth" was established through clinical assessments by medical professionals comparing the performance of the CIBA Vision® (nelfilcon A) lens to a control lens in terms of vision, health, comfort, and fit. This type of ground truth is based on expert clinical observation and evaluation, not pathology or outcomes data in the traditional sense for diagnostic devices.

8. The Sample Size for the Training Set

N/A (Not Applicable). The device is a physical contact lens, not an AI system or algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

N/A (Not Applicable), as there is no training set for a physical contact lens device.

{0}------------------------------------------------

CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097

21 September 2007 Page 1 of 3 v01

Nelfilcon A Soft Contact Lenses for Daily Wear 510(k) Summary of Safety and Substantial Equivalence

510(k) Summary

K072777

JAN

2 4 2006

1. Submitter Information:

Company:CIBA Vision Corporation11460 Johns Creek ParkwayDuluth, Georgia USA 30097
Contact Person:Martina Heim, PhD, RACSenior Regulatory Specialist, Global Regulatory Affairsmartina.heim@cibavision.com
Telephone:FAX:678-415-3565678-415-3454
Date Prepared:21 September 2007

2. Device Name:

  • . Common Name: Soft Contact Lens
  • CIBA Vision® (nelfilcon A) Trade/Proprietary Name: .
  • Classification Name: . Daily Wear Soft Contact Lens
  • Device Classification: Class II [21 CFR 886.5925 (b) (1)] .

3. Predicate Device:

CIBA Vision's Focus® DAILIES® (nelfilcon A) soft contact lenses have been selected as predicate devices for the modified CIBA Vision® (nelfilcon A) lenses. CIBA Vision obtained FDA 510(k) clearance for (nelfilcon A) lenses for daily wear on November 27, 1996 (K963487).

4. Description of Device:

The nelfilcon A lens material is 69% water and 31% nelfilcon A polymer (polyviny) alcohol partially acetalized with N-formylmethyl acrylamide). The lenses are tinted from edge to edge for visibility purposes with the color additive copper phthalocyaninks (UD).

Nelfilcon A lens designs currently include spherical, toric, and multifocal in the following parameter ranges:

Power Range:-20.00D to +20.00D
---------------------------------------
  • . Center Thickness: 0.010 mm for -3.00D spherical (varies with power)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image is a logo for CIBA Vision, a Novartis Company. The logo is black and white and features the words "CIBA Vision." in a bold, sans-serif font. The "i" in "Vision" is replaced with an eye symbol. Below the company name is the text "A Novartis Company" in a smaller font.

.

CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097

21 September 2007 Page 2 of 3 v01

Nelfilcon A Soft Contact Lenses for Daily Wear 510(k) Summary of Safety and Substantial Equivalence

Lenses have the following properties:

  • . Refractive index: 1.38
  • . Light transmittance: approximately 96 %T
    • Water content : 69% by weight
  • Oxygen permeability . 26 barrer

measured at 35°C (single point Dk-Polarographic method)

Lenses are supplied sterile in sealed blister-packs containing buffered saline. The compatibility and package integrity of the blister-pack packaging system has been demonstrated and successfully used for other CIBA Vision marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister-pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date.

5. Indications for Use:

CIBA Vision® (nelfilcon A) Spherical and Toric One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes.

CIBA Vision® (nelfilcon A) Progressives One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.

The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.

6. Description of Safety and Substantial Equivalence:

A series of non-clinical tests and a clinical study were performed to demonstrate the substantial equivalence of the device to the predicate device. All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses and in conformance to applicable device regulations. Results verify that the modified nelfilcon A lenses have material characteristics comparable to the predicate nelfilcon A soft contact lenses and are nontoxic and biocompatible. Clinically, the lens performed satisfactorily in a daily disposable wear investigation. Results from all tests demonstrate the substantial equivalence to previously FDA cleared predicate (control) lenses.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for CIBA Vision, a Novartis company. The logo features the word "CIBA" in large, bold, black letters on the top line. Below "CIBA" is the word "Vision" in a stylized font, with the "i" replaced by an eye symbol. Underneath the word "Vision" is the text "A Novartis Company" in a smaller font.

CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097

Nelfilcon A Soft Contact Lenses for Daily Wear 510(k) Summary of Safety and Substantial Equivalence

Non-clinical Testing:

A series of non-clinical testing was performed to verify equivalence of the device to the predicate device. Non-clinical biocompatibility testing was conducted in accordance with the GLP regulation (21 CFR Part 58).

The results of all non-clinical testing on (nelfilcon A) contact lens demonstrate:

  • Lens physical and material properties of the device are substantially equivalent to . the predicate lens.
  • . The lens material and extracts of the device are substantially equivalent to the predicate device and are non toxic and non-irritatina.

Clinical Testing:

The CIBA Vision® (nelfilcon A) contact lens was investigated in a clinical study for daily disposable wear. The clinical evaluation was conducted in accordance with current Good Clinical Practices and published regulations (21 CFR Parts 50, 56, 312, and 812). The study assessed the clinical performance of the lenses as compared to an FDA cleared and commercially available contact lens.

Clinical evaluation of the CIBA Vision® (nelfilcon A) lens demonstrated similar overall performance in the clinically relevant areas of vision, health, comfort and fit as compared to the control lens when used under daily disposable wear conditions.

Substantial Equivalence:

The CIBA Vision® (nelfilcon A) contact lens is substantially equivalent to the predicate lens and similar to other daily wear soft contact lenses in terms of water content (69% water) and ionic characteristics (FDA Group II: high water, nonionic), clinical performance, and indications for use.

Any differences which may exist between the CIBA Vision® (nelfilcon A) soft contact lens and other Group II soft hydrophilic contact lenses do not adversely affect the safety and effectiveness of the device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol is a representation of a human figure embracing a bird, which is meant to symbolize the department's mission of protecting the health of all Americans and providing essential human services.

JAN 2 4 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CIBA Vision Corporation c/o Martina Heim, Ph.D., RAC Senior Regulatory Specialist 11460 Johns Creek Parkway Duluth, GA 30097

Re: K072777

Trade/Device Name: CIBA Vision® (nelfilcon A) One-Day Soft Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: December 21, 2007 Received: December 26, 2007

Dear Dr. Heim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

M. B. Egelston, m.D.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Part III. Indications for Use Statement, v01, 21 September 2007, Page 1 of 1 Premarket Notification 510(k): (nelfilcon A) Soft Contact Lenses for Daily Wear

PART III. INDICATIONS FOR USE STATEMENT

510(k) Number:

KO72777

Device Name:

CIBA Vision® (nelfilcon A) Soft Contact Lenses

Indications For Use:

CIBA Vision® (nelfilcon A) Spherical and Toric One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes.

CIBA Vision® (nelfilcon A) Progressives One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.

The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Chandnamallika Ghosh

(Division Sign-Off)

Division of Ophthalmic Ear, Nose and Throat Devises

510(k) NumberK072777
------------------------
Prescription Use: or Over the Counter Use
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

N/A