(26 days)
The D131 Dideco Kids with 40 micron screen phosphorilcholine (Ph.I.S.I.O.) coated is recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filter is used to trap and remove gaseous emboli as well as particulate debris that maybe introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
The D131 Ph.I.S.I.O. is a sterile, non-pyrogenic disposable filter for use in arterial line of the cardiopulmonary bypass circuit with the flow rate not exceeding 2.5 liters/minute. The D131 Ph.I.S.I.O. is a Infant Arterial Filter with a 40 micron filter screen designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris greater than 40 microns from the arterial line agground blood been reduced in the modified in the modified version of the D736 Ph.I.S.I.O. unmodified device resulting in decreased priming volume and surface area of the filtering material. The internal modifications which mainly involve the filter screen geometry, make the new filter essentially a downscaled and simplified version of the D736 Ph.I.S.I.O.. The modifications to some of the external features result in enhanced ergonomics.
The provided 510(k) summary (K072308) describes the D131 Ph.I.S.I.O. Dideco Kids Infant Arterial Filter. This document primarily focuses on demonstrating substantial equivalence to a predicate device through in vitro testing, not a clinical study involving human readers or AI. Therefore, several of the requested sections related to AI or human reader studies are not applicable.
Here's the breakdown of the information that can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria (Established Specifications) | Reported Device Performance (D131 Ph.I.S.I.O. vs. D736 Ph.I.S.I.O.) |
|---|---|---|
| Operating Blood Volume | Met established specifications (Implied to be reduced compared to predicate) | Reduced priming volume compared to D736 Ph.I.S.I.O. |
| Structural Integrity | Met established specifications | Comparable to D736 Ph.I.S.I.O.; demonstrated integrity when pressurized over the maximum recommended pressure. |
| Pressure Integrity | Met established specifications | Part of the structural integrity assessment, demonstrated integrity when pressurized over the maximum recommended pressure. |
| Pressure Drop | Met established specifications (Implied to be reduced compared to predicate at same max blood flow) | Reduced pressure drop at all blood flow rates compared to D736 Ph.I.S.I.O. |
| Filter Flow Rate Capacity | Met established specifications | Effective filter flow rate capacity at all blood flow rates as compared to D736 Ph.I.S.I.O. |
| In vitro Hemolysis/Cell Depletion | Met established specifications | Comparable to D736 Ph.I.S.I.O. at maximum blood flow. |
| Filtration Efficiency | Met established specifications | Comparable to D736 Ph.I.S.I.O.; overall mean percent removal greater than 80% for particles equal to the nominal pore size of the filter (40µm). |
| Leaching | Met established specifications (No evidence of leaching) | No analytical evidence of the possibility of leaching of phosphorylcholine coating from the D131 Ph.I.S.I.O. was evidenced during the leaching test. |
| Air Handling Characteristics | Met established specifications | Similar removal capability for microbubbles as well as air bolus introduced in the circuit at minimum, mean, and maximum blood flow tested compared to D736 Ph.I.S.I.O. |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document mentions "aged D130 Ph.I.S.I.O. samples (K063255)" for biocompatibility and "sterilized aged devices comparing the D131 Ph.I.S.I.O. vs. the D736 Ph.I.S.I.O." for in vitro tests. Specific numerical sample sizes for these tests are not provided.
- Data Provenance: The studies were in vitro tests conducted by the manufacturer, Sorin Group Italia S.r.l. in Mirandola, Italy. The tests were performed on aged devices, including one additional year of aging for a "truly worst case." No human or clinical data is presented, so country of origin for patients/data is not applicable. The studies are retrospective of the manufacturing process and device characteristics.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This was an in vitro device performance study, not a study requiring expert clinical assessment for ground truth. The "ground truth" was established by the physical and chemical properties of the materials and the performance measurements using standardized test methods.
4. Adjudication method for the test set:
- Not applicable. This was an in vitro device performance study; there was no subjective assessment or need for adjudication between observers.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an arterial filter for cardiopulmonary bypass circuits, not an AI-powered diagnostic tool requiring human reader studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device does not involve an algorithm or AI. It is a mechanical filter.
7. The type of ground truth used:
- For the in vitro performance tests, the "ground truth" was established by objective measurements and physical/chemical properties of the device under various simulated physiological conditions (e.g., flow rates, pressure). These measurements were compared against established specifications derived from relevant guidance documents (e.g., "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" Final Guidance for Industry, dated November 29, 2000) and the performance of the predicate device (D736 Ph.I.S.I.O.).
- For biocompatibility, the ground truth was based on the biological response to the materials as assessed through a battery of tests compliant with ISO 10993-1:2002 and FDA guidance, compared against established specifications for biological safety.
8. The sample size for the training set:
- Not applicable. This is not an AI/machine learning device that requires a training set. The device itself is the "product" being tested, not an algorithm.
9. How the ground truth for the training set was established:
- Not applicable. As noted above, there is no training set for this type of device.
{0}------------------------------------------------
510(k) SUMMARY
| SUBMITTER: | Sorin Group Italia S.r.l.86, Via Statale 12 Nord41037 Mirandola (MO) Italy | SEP 12 2007 |
|---|---|---|
| CONTACT PERSON: | Luigi VecchiPhone: 011 39 0535 29811Fax: 011 39 0535 25229 | |
| DATE PREPARED: | August 16, 2007 | |
| DEVICE TRADE NAME: | D131 Dideco Kids Infant Arterial Filter with 40 micron screenphosphorylcholine (Ph.I.S.I.O.) coated (hereafter referred toas D131 Ph.I.S.I.O.) | |
| COMMON NAME: | Arterial Filter | |
| CLASSIFICATION NAME: | Cardiopulmonary Bypass Arterial Line Blood Filter | |
| LEGALLY MARKETED UNMODIFIED DEVICE | D736 MICRO 40 Ph.I.S.I.O.: Dideco Newborn-Infant ArterialFilter with Ph.I.S.I.O. coating (Phosphorilcholine coating) asdescribed in K002493 and modified in K033987 | |
| PREDICATE DEVICE: | D130 Dideco Kids Newborn Arterial Filter with 40 micronscreen phosphorylcholine (Ph.I.S.I.O.) coated (hereinafterreferred to as D130 Ph.I.S.I.O.) originally cleared for use witha maximum blood flow rate of 0.7 l/min under K063255 ismanufactured by Sorin group Italia S.r.l. |
DEVICE DESCRIPTION:
The D131 Ph.I.S.I.O. is a sterile, non-pyrogenic disposable filter for use in arterial line of the cardiopulmonary bypass circuit with the flow rate not exceeding 2.5 liters/minute. The D131 Ph.I.S.I.O. is a Infant Arterial Filter with a 40 micron filter screen designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris greater than 40 microns from the arterial line agground blood been reduced in the modified in the modified version of the D736 Ph.I.S.I.O. unmodified device resulting in decreased priming volume and surface area of the filtering material. The internal modifications which mainly involve the filter screen geometry, make the new filter essentially a downscaled and simplified version of the D736 Ph.I.S.I.O.. The modifications to some of the external features result in enhanced ergonomics.
INDICATION FOR USE:
The D131 Dideco Kids with 40 micron screen phosphorilcholine (Ph.I.S.I.O.) coated is recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filter is used to trap and remove gaseous emboli as well as particulate debris that maybe introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
TECHNOLOGICAL CHARACTERISTICS:
The D131 Ph.I.S.I.O. has the same operating principles and control mechanisms when compared to the D736 Ph.I.S.I.O. unmodified device and D130 Ph.I.S.I.O. predicate device. The D131 Ph.I.S.I.O. utilizes the
{1}------------------------------------------------
same materials, the same filtering media, the same main blood flow path and the same filtering pore size (40 micron) as both the unmodified and predicate devices.
million) as both the unitiounce and products and seen updated with respect to those of the current D736 The Gesign leatures of the D101 Ph.I.S.I.O. shares essentially the D131 Ph.I.S.I.O. shares essentially the same PIT.C.S.C. Newbon-mail diffindined act not not hose in actuate device. No change of the intended basic design philosophy and goornely with the fevice. Both devices share the identical manufacturing use has been made for the modified torolor or the active and has a nonpyrogenic fluid path. It is for single use only.
BIOCOMPATIBILITY TEST RESULTS:
As no new materials are used into the D131 Ph..S.I.O. infant arterial filter as compared to the D130 Ph.I.S.I.O. predicate device, data collected on aged D130 Ph.I.S.I.O. samples (K063255) are considered applicable also to the D131 Ph.I.S.I.O. modified device. A complete battery of tests were conducted in applicance with the requirements of ISO 10993-12002 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on the raw materials. Testing was performed on the D 130 Ph.I.S.I.O. (accelerated aging). The results of the testing met established specifications.
IN VITRO TEST RESULTS:
In vitro testing was carried out in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" Final Guidance for Industry, dated November 29, 2000 for providing the data necessary to demonstrate both substantial equivalence with the unmodified device and show that the device is compliant with safety and effectiveness requirements. The device was aged up to 3 years (+ 1 additional year of aging in order to test a truly worst case) and tested for operating blood volume, structural integrity test, pressure integrity test, pressure drop, filter flow rate capacity, in vitro hemolysis/cell depletion, filtration efficiency, leaching and air handling characteristics. For comparative purposes all tests, when applicable, were performed on sterilized aged devices comparing the D131 Ph.I.S.I.O. vs. the D736 Ph.I.S.I.O. operated at same max blood flow. The results of these tests met established specifications..
CONCLUSIONS:
The results of in vitro studies demonstrate that the design modifications result in reduced priming volume and pressure drop with an effective filter flow rate capacity at all blood flow rates as compared to the D736 Ph.I.S.I.O. unmodified device. Both filters demonstrated a comparable structural integrity, hemolysis/cell depletion at the maximum blood flow. Both devices share similar removal capability for microbubbles as well as air bolus introduced in the circuit at minimum, mean and maximum blood flow tested. Likewise both filters demonstrated comparable filtration efficiency with an overall mean percent removal greater than 80% for particles equal to the nominal pore size of the filter (40p). The D131 Ph.l.S.I.O. filter demonstrated integrity when pressurized over the maximum recommended pressure and no analytical evidence of the possibility of leaching of phosphoriylcholine coating from the D131 Ph.l.S.I.O. was evidenced during the leaching test. The results are in line with expectations because the D131 Ph.I.S.I.O. is smaller in overall size, has a more compact design, and contains a different filter screen design as compared to the unmodified device. The smaller size offers theoretical advantages in terms of reduced priming volume and consequently less hemodilution. A lower priming volume is desirable as it results in advantageous patient hemodynamics, reduced exposure of the blood cells and plasma proteins to large surface areas.
Biocompatibility tests demonstrate that its performance is equivalent to the D736 Ph.I.S.I.O. unmodified device, according to its intended use. Additional testing has demonstrated the effectiveness of production techniques assuring that the newborn arterial filter is sterile and non-pyrogenic
In conclusion test result of this study suggests the D131 Ph.I.S.I.O. arterial filter is equivalent to the D736 Ph.I.S.I.O. arterial filter with respect to device function.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and head. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
ood and Drug Administration 200 Corporate Boulevard Rockville MD 20850
Sorin Group Italia S.R.L c/o Mr. Barry Sall Principal Consultant Parexel Consulting 200 West Street Waltham. MA 02451
SEP 1 2 2007
K072308 Re:
D131 Ph.I.S.I.O. Dideco Kids Infant Arterial Filter Regulation Number: 21 CFR 870.4260 Regulation Name: Cardiopulmonary bypass arterial line blood filter Regulatory Class: Class II Product Code: DTM Dated: August 16, 2007 Received: August 17, 2007
Dear Mr. Sall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Barry Sall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
SORIN GROUP
AT THE HEART OF MEDICAL TECHNOLOGY
Sorin Group Italia S.r.l.
510(k) Number (if known):
14072308
Device Name: D131 Dideco Kids Infant Arterial Filter with 40 micron screen phosphorycholine (Ph.J.S.I.O.) coated
Indications For Use:
D131 Dideco Kids with 40 micron screen phosphorylcholine (Ph.I.S.I.O.) coated is recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filter is used to trap and remove gaseous emboli as particulate debris that maybe introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
Prescription Use X (Part 21 CFR 801 Subpart D) and/or Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Bro
§ 870.4260 Cardiopulmonary bypass arterial line blood filter.
(a)
Identification. A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”