K Number
K072157
Manufacturer
Date Cleared
2007-08-27

(24 days)

Product Code
Regulation Number
880.5130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SPO2 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.
For professional use only, by trained clinicians.

Device Description

The Ohmeda Medical Giraffe and Panda Warmers are devices with a radiant heating source intended to maintain the thermal balance of an infant patient by direct radiation of energy in the infrared region of the electromagnetic spectrum.
The warmers operate similarly to warmers currently in use in hospitals. Radiant heat from an infrared heat source is focused onto the bed to warm the patient. The operator may select either the heater power or skin temperature control method. Depending on the control method selected, the heater is either regulated at the operator selected power level or the heater output is modulated to maintain the patient's temperature at the value selected by the operator.
Infant radiant warmers are also used to provide thermal support during surgical procedures and during procedures such as extracorporeal membrane oxygenation, resuscitation, or other procedures requiring open access and thermal support.
Both units also feature optional intearated SpO2 and Resuscitation Modules. The Resuscitation Module may feature either a traditional bag-and-mask technology or a T-piece technology. Both the SpO2 module and the Resuscitation Modules use existing technology.
Infant radiant warmers may incorporate other features, such as tilting of the bed, elevating base, and data output to remote monitors or nurse call systems. Infant radiant warmers may also allow use with or attachment of an independent phototherapy device or other accessories.
The Giraffe and Panda UPS is a stand-by power supply that will serve the purpose of temporarily maintaining the operation of the parent devices, the Giraffe and Panda Warmers.

AI/ML Overview

The provided text describes the acceptance criteria and a study demonstrating the device's performance, but it focuses on performance testing rather than an AI/ML-driven diagnostic device. Therefore, several requested fields, such as MRMC studies, details about expert ground truth, and training set information, are not applicable.

Here's the available information presented in the requested format:

Acceptance Criteria and Device Performance Study

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Performance Specification)Reported Device Performance
Compliance with consensus standards for infant radiant warmers.Conformance to performance specifications and multiple recognized performance standards for infant radiant warmers (established through bench testing).
UPS Battery Life (nominal operating conditions, 60% heater power)20 minutes
UPS Battery Life (worst case operating condition, 100% heater power)15 minutes
UPS Battery Life (worst case operating condition, 100% heater power, with phototherapy accessory and all lights on)10 minutes
UPS Recharge Time6 hours
Thermal balance maintenance for neonates (by direct radiation of energy in the infrared region).The device operates similarly to warmers currently in use in hospitals, providing radiant heat from an infrared source focused onto the bed to warm the patient. The heater is regulated either at an operator-selected power level or modulated to maintain the patient's selected temperature.
Provision of thermal support during surgical procedures, extracorporeal membrane oxygenation, resuscitation, or other procedures requiring open access and thermal support.Infant radiant warmers are explicitly stated to be used for these purposes.
Continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (with optional integrated SpO2 monitoring feature).The optional integrated SpO2 module uses existing technology.
Provision of basic equipment required for pulmonary resuscitation of infants through an optional integrated resuscitation system (including establishing clear airway, providing oxygen/air/oxygen mixtures, and/or manual ventilation).The optional integrated Resuscitation Module may feature either traditional bag-and-mask technology or T-piece technology, both using existing technology.
Stand-by power supply function for temporary operation maintenance during power interruptions (e.g., power failure, unintentional disconnect, short-duration intra-hospital transport).The Giraffe and Panda UPS (medical grade battery and shelf) serves this purpose. "When disconnected from wall power, the parent device continues receiving the power that it needs to operate, but the energy is supplied by the UPS rather than the distribution line."

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated. The study relies on "bench testing" against product specifications and recognized consensus standards.
  • Data Provenance: Not applicable, as no patient data was used. The testing was conducted in a laboratory setting ("bench testing").

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Ground truth was established by conformance to engineering specifications and recognized consensus standards for infant radiant warmers, not by expert interpretation of data.

4. Adjudication method for the test set:

  • Not applicable. The study involved bench testing against predefined performance criteria and standards, not clinical adjudication of diagnostic findings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is a medical device (infant warmer) without AI or diagnostic functionality, so an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. The device is not an algorithm, but a physical medical device. Performance was assessed through bench testing of its physical and electrical functions.

7. The type of ground truth used:

  • Performance Specifications and Recognized Consensus Standards: The ground truth was based on the device meeting its predefined engineering performance specifications (e.g., battery life, recharge time) and conforming to established, recognized consensus standards for infant radiant warmers.

8. The sample size for the training set:

  • Not applicable. This device does not involve machine learning and therefore has no "training set."

9. How the ground truth for the training set was established:

  • Not applicable. This device does not involve machine learning and therefore has no "training set" or associated ground truth establishment process for training.

{0}------------------------------------------------

K072/57 p. lot 3

510(k) Summary

AUG 2 7 2007

Submitter Information and Date Prepared

Agata Smieja GE Healthcare 8880 Gorman Road Laurel, MD 20723 USA

Phone: 410 888 5218 Mobile: 410 456 0329

Prepared August 24, 2007

Device Identification

Proprietary Name:Ohmeda Medical Giraffe and Panda Warmers
Common Name:Giraffe and Panda Warmers

Predicate Device Information

Predicate Device510(k) Number
Giraffe and Panda WarmersK070377

Intended Use Statement

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-requlate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated Sp02 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of

{1}------------------------------------------------

072157 p. 2 of 3

infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Functional Description and Technological Characteristics

The Ohmeda Medical Giraffe and Panda Warmers are devices with a radiant heating source intended to maintain the thermal balance of an infant patient by direct radiation of energy in the infrared region of the electromagnetic spectrum.

The warmers operate similarly to warmers currently in use in hospitals. Radiant heat from an infrared heat source is focused onto the bed to warm the patient. The operator may select either the heater power or skin temperature control method. Depending on the control method selected, the heater is either regulated at the operator selected power level or the heater output is modulated to maintain the patient's temperature at the value selected by the operator.

Infant radiant warmers are also used to provide thermal support during surgical procedures and during procedures such as extracorporeal membrane oxygenation, resuscitation, or other procedures requiring open access and thermal support.

Both units also feature optional intearated SpO2 and Resuscitation Modules. The Resuscitation Module may feature either a traditional bag-and-mask technology or a T-piece technology. Both the SpO2 module and the Resuscitation Modules use existing technology.

Infant radiant warmers may incorporate other features, such as tilting of the bed, elevating base, and data output to remote monitors or nurse call systems. Infant radiant warmers may also allow use with or attachment of an independent phototherapy device or other accessories.

The Giraffe and Panda UPS is a stand-by power supply that will serve the purpose of temporarily maintaining the operation of the parent devices, the Giraffe and Panda Warmers. Instances where normal line supply is interrupted include but are not limited to:

  1. power failure due to brown/black out scenario

  2. unintentional disconnect of parent device power cord from wall receptacle 3) purposeful disconnect from wall receptacle of the parent device to affect short duration, intra-hospital transport of the parent device

{2}------------------------------------------------

072157 p. 3 of-3

The power cord of the Giraffe and Panda Warmer devices is plugged into Giraffe and Panda UPS, and, in turn, the power cord of the UPS is then connected to the wall receptacle. When disconnected from wall power, the parent device continues receiving the power that it needs to operate, but the energy is supplied by the UPS rather than the distribution line.

The UPS comprises a medical grade battery and a shelf.

Performance Data

Because care of infants in infant radiant warmers is a well established clinical practice and because the Giraffe and Panda Warmer will comply with all applicable clauses of the recognized consensus standard for infant radiant warmers, animal or clinical testing to support safety and effectiveness is not necessary. The conformance of the Giraffe and Panda Warmers to performance specifications and to multiple recognized performance standards is being established through bench testing.

The UPS battery has a life of 20 minutes in nominal operating conditions (60% heater power), 15 minutes at the worst case operation condition (100% heater power) and 10 minutes at the worst case operating condition with the addition of the phototherapy accessory and with all available lights on (at 100% heater power). The recharge time is 6 hours.

Performance of the Giraffe and Panda Warmers with the addition of UPS has been established by bench testing against product specifications and recognized consensus standards.

Prepared by: Date 08/24/07

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the circumference. Inside the circle is a stylized symbol resembling an abstract caduceus or a series of flowing lines, possibly representing a human form or a symbol of health and well-being.

Public Health Service

AUG 2 7 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Agata Smieja Global Compliance Leader Ohmeda Medical, a Division of Datex Ohmeda, a GE Healthcare Company 8880 Gorman Road Laurel, Maryland 20723

Re: K072157

Trade/Device Name: Giraffe and Panda Warmer Regulation Number: 21 CFR 880.5130 Regulation Name: Infant Radiant Warmer Regulatory Class: II Product Code: FMT Dated: August 2, 2007 Received: August 3, 2007

Dear Ms. Smieja:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Smieja

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K072157

Device Name: Giraffe and Panda Warmer

Indications For Use:

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SPO2 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

For professional use only, by trained clinicians.

X__ Prescription Use

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 880.5130 Infant radiant warmer.

(a)
Identification. The infant radiant warmer is a device consisting of an infrared heating element intended to be placed over an infant to maintain the infant's body temperature by means of radiant heat. The device may also contain a temperature monitoring sensor, a heat output control mechanism, and an alarm system (infant temperature, manual mode if present, and failure alarms) to alert operators of a temperature condition over or under the set temperature, manual mode time limits, and device component failure, respectively. The device may be placed over a pediatric hospital bed or it may be built into the bed as a complete unit.(b)
Classification. Class II (Special Controls):(1) The Association for the Advancement of Medical Instrumentation (AAMI) Voluntary Standard for the Infant Radiant Warmer;
(2) A prescription statement in accordance with § 801.109 of this chapter (restricted to use by or upon the order of qualified practitioners as determined by the States); and
(3) Labeling for use only in health care facilities and only by persons with specific training and experience in the use of the device.