K Number
K072134
Date Cleared
2007-11-02

(92 days)

Product Code
Regulation Number
872.1800
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Computed Oral Radiology System is intended for intra-oral x-rav examinations and indicated for dental patients. It produces instant, digital, intraoral x-ray images of a patient's mouth while reducing the necessary x-rav dosage. This device modification in no way alters the indications for use of this machine beyond what was most recently cleared in K041385.

Device Description

The device and its predicates are small digital imaging receptors that may be used in place of dental x-ray film. A new control mechanism differs from the predicate in that image acquisition is initiated via a new hardware trigger residing on the CMOS ASIC. Direct triggering via an x-ray tube continues to be supported as was cleared in K041385. The modification offers an improvement in system-level support. The modification in no way alters the fundamental technology nor intended use.

AI/ML Overview

The provided text is a 510(k) summary for a dental imaging device (Computed Oral Radiology System) and primarily focuses on demonstrating substantial equivalence to a predicate device after a modification. As such, it does not contain the detailed information typically found in a study report proving a device meets specific performance acceptance criteria.

Specifically, the document states: "All validation activities have demonstrated that the predetermined acceptance criteria were met." However, it does not explicitly define these acceptance criteria, nor does it describe the specific studies and their results in enough detail to populate the requested table and answer many of the questions.

Here's a breakdown of what can be inferred and what is missing, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Not explicit: The document states "predetermined acceptance criteria were met" but does not define what those criteria were (e.g., specific image quality metrics, diagnostic accuracy thresholds).Not explicit: The document does not provide specific performance data (e.g., sensitivity, specificity, resolution metrics, contrast ratios) from any studies. It only generally states that "All validation activities have demonstrated that the predetermined acceptance criteria were met."
Principal risk: Unintended x-ray exposure (inferred as an acceptance criterion for safety)Evaluated through "imaging, software validation, and third-party safety testing." (No specific results provided).

2. Sample size used for the test set and the data provenance:

  • Not specified. The document mentions "validation studies" but does not provide details about sample sizes or data provenance (country of origin, retrospective/prospective). Given that the modification is related to a control mechanism and not a fundamental change in imaging technology or intended use, it's possible that extensive clinical test sets were not deemed necessary for this specific 510(k) rather than for the initial clearance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not specified. Ground truth establishment methods, number of experts, and their qualifications are not mentioned.

4. Adjudication method for the test set:

  • Not specified.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. The device is an imaging sensor, not an AI-assisted diagnostic tool. There is no mention of AI or any MRMC studies involving human readers and AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not explicitly stated how "standalone" performance translates to this device. The device is a digital imaging receptor. Performance would relate to image quality, dose reduction, and overall system function, rather than an "algorithm only" performance separate from the human operator. While "imaging, software validation" is mentioned, no specific standalone performance metrics are provided.

7. The type of ground truth used:

  • Not specified.

8. The sample size for the training set:

  • Not applicable/Not specified. As there's no mention of AI or deep learning, there isn't a "training set" in the typical sense for an algorithm. If "training set" refers to data used for system development and calibration, it is not detailed.

9. How the ground truth for the training set was established:

  • Not applicable/Not specified.

In summary: The provided 510(k) pertains to a modification of an already cleared device, specifically "a new control mechanism" for image acquisition. The document emphasizes that the modification "in no way alters the fundamental technology nor intended use." This suggests that the validation focused on ensuring the new control mechanism's safety and functionality, and its compatibility with the existing system, rather than re-proving the core diagnostic performance of the imaging technology. Therefore, the detailed performance study information typically required for a novel diagnostic device is not present in this summary.

{0}------------------------------------------------

K072134
Page 1 of 2

Exhibit IV: 510(k) Summary

Schick Computed Oral Radiology System ●

NOV - 1 2007

Common/Classification Name: Solid State X-ray Imager 21 CFR892.1650

Schick Technologies, Inc. 30-30 47th Avenue Long Island City. NY 11101 718-937-5765, 718-937-5962 (FAX) Contact: Daniel Michaeli, Prepared: October 12, 2007

A. Legally Marketed Predicate Devices

The Computed Oral Radiology System was most recently cleared on June 7, 2004 under K041385. Other prior clearances include K022953 and K933455. The device and its predicates are small digital imaging receptors that may be used in place of dental x-ray film.

B. Modification Description

A new control mechanism differs from the predicate in that image acquisition is initiated via a new hardware trigger residing on the CMOS ASIC. Direct triggering via an x-ray tube continues to be supported as was cleared in K041385. The modification offers an improvement in system-level support.

The modification in no way alters the fundamental technology nor intended use,

C. Intended Use

The Computed Oral Radiology System is intended for intra-oral x-rav examinations and indicated for dental patients. It produces instant, digital, intraoral x-ray images of a patient's mouth while reducing the necessary x-rav dosage. This device modification in no way alters the indications for use of this machine beyond what was most recently cleared in K041385.

D. Substantial Equivalence Summary

A risk analysis established the areas of concem. The principal risk is unintended x-ray exposure. These areas have been evaluated following imaging, software validation, and third-party safety testing. All validation activities have demonstrated that the predetermined acceptance criteria were met. Where appropriate, warnings have been incorporated within the user manual.

{1}------------------------------------------------

KC72134
Page 2 of 2

E. Conclusions

. . *

Schick Technologies has demonstrated through a risk analysis and validation studies that the device modification is substantially equivalent to the already cleared and marketed device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo consists of two parts: a symbol on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The symbol is a stylized representation of a human figure, with three blue lines forming the head, body, and legs. The text is in a blue sans-serif font.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Daniel Michaeli Director of Scientific Development Schick Technologies, Inc. 30-30 47" Avenue LONG ISLAND CITY NY 11101

AUG 2 3 2013

Re: K072134

Trade/Device Name: Computed Oral Radiology System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH, EHD, and MOB Dated: October 15, 2007 Received: October 17, 2007

Dear Mr. Michaeli:

This letter corrects our substantially equivalent letter of November 2. 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Computed Oral Radiology System

Indications For Use:

. . ..

1

The Computed Oral Radiology System is indicated for patients undergoing an intra-oral dental x-ray examination. It produces instant, digital, intra-oral x-ray images of a palient's mouth while reducing the necessary x-ray dosage.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tori Whan
(Division Sign Off)

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Davices 510(k) Number

Page 1 of _1

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.