K Number
K072115

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-11-30

(121 days)

Product Code
Regulation Number
862.1225
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Creatinine and Total Protein reagents, with associated calibrators and controls, are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer to measure a variety of analytes.

ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

ABX PENTRA Total Protein 100 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in-vitro determination of Total Proteins in serum and plasma by colorimetry.

Measurements obtained by this device are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders.

The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba ABX methods on Horiba ABX clinical chemistry analyzers.

The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision.

The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.

The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

Device Description

All the reagents, controls and calibrators included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). It is composed of a 27 ml monoreagent cassette. Reagent is a chemical solution with additives.

The ABX PENTRA Total Protein 100 CP is an in vitro diagnostic assay for the quantitative determination of total proteins in human serum and plasma based on a colorimetric test (Biuret reaction). It is composed of a 28 ml mono-reagent cassette. Reagent is a chemical solution with additives.

The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin.

The ABX PENTRA N Control and ABX PENTRA P Control are quality control products consisting of lyophilized human serum with chemical additives and materials of biological origin added as required to obtain given component levels.

The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ABX PENTRA Creatinine 120 CP and ABX PENTRA Total Protein 100 CP devices, based on the provided text:

Acceptance Criteria and Device Performance

The devices are in vitro diagnostic assays, and their performance is described in terms of analytical characteristics. The stated performance data implicitly serve as the acceptance criteria for the devices to be considered substantially equivalent to their predicate devices.

ABX PENTRA Creatinine 120 CP

Acceptance Criteria CategoryAcceptance Criteria (Implied)Reported Device Performance
Sample typeSerum, Plasma and Urine compatibilitySerum, Plasma and Urine
Detection limitSpecified limits for serum/plasma and urineSerum/Plasma: 0.18 mg/dl; Urine: 1.39 mg/dl
Accuracy and PrecisionCV Total below specified percentagesSerum/Plasma CV Total < 5.83%; Urine CV Total < 6.00%
Measuring rangeSpecified ranges for serum/plasma and urineSerum/Plasma: 0.18 mg/dl - 22.60 mg/dl; Urine: 1.39 mg/dl - 282.5 mg/dl
Upper linearity limitSpecified limits with and without automatic post-dilutionSerum/Plasma: 22.60 mg/dl (67.8 mg/dl with post-dilution); Urine: 282.5 mg/dl (857.5 mg/dl with post-dilution)
CorrelationHigh correlation coefficient (r²) and slope/intercept close to ideal line (Y=X)Serum/Plasma (n=122): Y = 0.98 x - 0.04 with r² = 0.9991; Urine (n=119): Y = 0.96 x - 0.73 with r² = 0.9975
Calibration stabilitySpecified stability periodSerum/Plasma: 24 hours; Urine: 24 hours
Reagent stabilitySpecified closed and on-board stabilityclosed stability: 24 months at 2-8°C; on-board stability: 10 days

ABX PENTRA Total Protein 100 CP

Acceptance Criteria CategoryAcceptance Criteria (Implied)Reported Device Performance
Sample typeSerum/Plasma compatibilitySerum/Plasma
Detection limitSpecified limit0.01 g/dl
Accuracy and PrecisionCV Total below specified percentageCV Total < 1.62%
Measuring rangeSpecified range0.10 g/dl – 10.0 g/dl
Upper linearity limitSpecified limit with and without automatic post-dilution10.0 g/dl (20.0 g/dl with automatic post-dilution)
CorrelationHigh correlation coefficient (r²) and slope/intercept close to ideal line (Y=X)Y = 1.03 x - 0.20 with a correlation coefficient r² = 0.9921
Calibration stabilitySpecified stability period1 day
Reagent stabilitySpecified closed and on-board stabilityclosed stability: 26 months at 2-25°C; on-board stability: 14 days

Study Information

The document describes the performance data for in vitro diagnostic reagents. The studies performed are analytical performance studies, not clinical studies involving human patients or ground truth established by medical experts for diagnostic interpretation.

  1. Sample size used for the test set and the data provenance:

    • ABX PENTRA Creatinine 120 CP:
      • Serum/Plasma correlation study: n=122 samples.
      • Urine correlation study: n=119 samples.
    • ABX PENTRA Total Protein 100 CP:
      • Correlation study: n=178 samples.
    • Data Provenance: The document states "HORIBA ABX, FRANCE" as the company location, implying the studies were conducted or samples sourced from France. The data is retrospective in the sense that it's laboratory performance data generated to characterize the assay, not new prospective patient data collected for a clinical trial.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. For in vitro diagnostic reagents, "ground truth" is typically established by reference methods or validated comparative assays. The correlation studies are comparing the performance of the new device against a comparative method. The text does not specify the comparative method used, nor does it mention any human experts establishing ground truth for these analytical measurements.
  3. Adjudication method for the test set:

    • Not applicable. Analytical studies for these types of reagents do not typically involve human adjudication of results. The "truth" or reference values for the samples would be determined by the comparative method.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. These are in vitro diagnostic reagents that measure chemical analytes. They are not AI-powered image analysis devices or clinical decision support tools that would involve human readers or AI assistance.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, these are standalone analytical performance studies of reagents and their performance on an automated analyzer (ABX PENTRA 400). The "algorithm" here refers to the chemical reaction and measurement process of the diagnostic assay itself. Human intervention is limited to sample loading, instrument operation, and quality control, but the measurement itself is automated.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for these analytical performance studies is implicitly the results obtained from a comparative method or reference method against which the new device's measurements are correlated. The document states "Correlation" with r² values, indicating a comparison against another measurement method. The nature of this comparative method (e.g., another FDA-cleared creatinine assay) is not explicitly detailed but is standard practice for IVD submissions.
  7. The sample size for the training set:

    • Not applicable. These are chemical reagents following established principles (Jaffé method, Biuret reaction). They are not machine learning or AI-based algorithms that require "training sets" in the conventional sense. The development of such reagents involves extensive research and development, but not "training data" in the AI context.
  8. How the ground truth for the training set was established:

    • Not applicable (as explained above).

{0}------------------------------------------------

Premarket Notification [510(k)] Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is:K072115
-----------------------------------------
Company:Horiba ABX
Parc Euromédecine
Rue du Caducée - BP 7290
34184 Montpellier cedex 4
FRANCE
Telephone:+ (33) 4 67 14 73 92
Fax:+ (33) 4 67 14 18 75
Contact Persons:Olivier Ducamp (oducamp@fr.abx.fr)
Caroline Ferrer (cferrer@fr.abx.fr)

NOV 3 0 2007Date Prepared: 26th July 2007

Device Names:

The following reagents, controls & calibrators are for use in conjunction with the ABX PENTRA 400, cleared to market under K052007.

REAGENTS :

Trade/Proprietary Name:ABX PENTRA Creatinine 120 CP
Common or Usual Name:Creatinine
Device ClassClass II
Classification Name:§862.1225 : Creatinine Test System
Product Code:CGX ; alkaline picrate, colorimetry, creatinine
Trade/Proprietary Name:ABX PENTRA Total Protein 100 CP
Common or Usual Name:Total Protein
Device ClassClass II
Classification Name:§862.1635 : Total Protein Test System
Product Code:CEK ; Biuret (colorimetric) total protein
CALIBRATORS:Trade/Proprietary Name:Common or Usual Name:Device ClassClassification Name:Product Code:ABX PENTRA Multical (K052007)MulticalClass II§862.1150 : CalibratorJIX ; Calibrator, Multi-Analyte Mixture

HORIBA ABX, FRANCE

{1}------------------------------------------------

CONTROLS :
Trade/Proprietary Name:ABX PENTRA N Control (K052007)
Common or Usual Name:N Control
Device ClassClass I
Classification Name:§862.1660 : Quality control material (assayed)
Product Code:JJY ; Multi-Analyte Controls, All Kinds (Assayed)
Trade/Proprietary Name:ABX PENTRA P Control (K052007)
Common or Usual Name:P Control
Device ClassClass I
Classification Name:§862.1660 : Quality control material (assayed)
Product Code:JJY ; Multi-Analyte Controls, All Kinds (assayed)
Trade/Proprietary Name:ABX PENTRA Urine Control L/H (K070249)
Common or Usual Name:Urine control
Device ClassClass I
Classification Name:§862.1660 : Quality control material (assayed)
Product Code:JJY ; Multi-Analyte Controls, All Kinds (Assayed)

Substantial Equivalence:

The data and information supplied in this submission demonstrates substantial equivalence to their respective predicate devices :

Submission deviceSubstantially equivalentPredicate device
ABX PENTRA Creatinine 120 CPK973869 (serum/plasma samples)K941837 (urine samples)
ABX PENTRA Total Protein 100 CPK973869
ABX PENTRA MulticalK052007
ABX PENTRA N ControlK052007
ABX PENTRA P ControlK052007
ABX PENTRA Urine Control L/HK070249

Description :

All the reagents, controls and calibrators included in this submission are for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). It is composed of a 27 ml monoreagent cassette. Reagent is a chemical solution with additives.

{2}------------------------------------------------

The ABX PENTRA Total Protein 100 CP is an in vitro diagnostic assay for the quantitative determination of total proteins in human serum and plasma based on a colorimetric test (Biuret reaction). It is composed of a 28 ml mono-reagent cassette. Reagent is a chemical solution with additives.

The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. This calibrator is provided in ten vials of 3 ml.

The ABX PENTRA N Control and ABX PENTRA P Control are quality control products consisting of lyophilized human serum with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexes, ensuring control of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. Each control is provided in ten vials of 5 ml.

The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the enclosed annexe, ensuring control of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. Each control level is provided in one vial of 10 ml.

Intended Use :

All reagents in this submission are intended for use on the ABX PENTRA 400 for the quantitative in-vitro determination of their respective analytes (Creatinine, Total Protein) using human serum, plasma and/or urine.

The controls and calibrator are intended for use in association with the above reagents.

{3}------------------------------------------------

Discussion of Performance Data:

ABX PENTRA Creatinine 120 CP :
Sample typeSerum, Plasma and Urine
Detection limitSerum/Plasma : 0.18 mg/dlUrine : 1.39 mg/dl
Accuracy and PrecisionSerum/Plasma CV Total < 5.83%Urine CV Total < 6.00%
Measuring rangeSerum/Plasma : 0.18 mg/dl - 22.60 mg/dlUrine : 1.39 mg/dl - 282.5 mg/dl
Upper linearity limitSerum/Plasma : 22.60 mg/dl, and with automatic post-dilution :67.8 mg/dlUrine : 282.5 mg/dl, and with automatic post-dilution : 857.5mg/dl
CorrelationSerum/Plasma (n=122) : Y = 0.98 x - 0.04 with r² = 0.9991Urine (n=119) : Y = 0.96 x - 0.73 with r² = 0.9975
Calibration stabilitySerum/Plasma : 24 hoursUrine : 24 hours
Reagent stabilityclosed stability: 24 months at 2-8°Con-board stability : 10 days

{4}------------------------------------------------

ABX PENTRA Total Protein 100 CP :
Sample typeSerum/Plasma
Detection limit0.01 g/dl
Accuracy and PrecisionCV Total < 1.62%
Measuring range0.10 g/dl – 10.0 g/dl
Upper linearity limit10.0 g/dl, and with automatic post-dilution : 20.0 g/dl
Correlation (n=178)$Y = 1.03 x - 0.20$ with a correlation coefficient r2 = 0.9921.
Calibration stability1 day
Reagent stabilityclosed stability: 26 months at 2-25°Con-board stability (refrigerated area): 14 days

{5}------------------------------------------------

CALIBRATORS

ABX PENTRA Multical:
AnalytesAlready cleared(K052007, K060205, K060318,K060325, K060854, K062180,K062737, K060434)Included in this submission
Alkaline phosphatase
Alanine aminotransferase
Amylase
Aspartate aminotransferase
Creatine kinase
GGT
Lipase
Albumin
Direct Bilirubin
Total Bilirubin
Calcium
Cholesterol
Creatinine
Creatinine 120
Glucose HK
Glucose PAP
Iron
Lactic acid
Magnesium
Phosphorus
Total Protein
Total Protein 100
Triglycerides
Urea / Blood Urea Nitrogen
Uric acid
FormatLyophilized human serum with chemical additives and materialsof biological origin
StabilityClosed stability: 24 months at 2-8°COpen stability:Once opened, the calibrator components** are stable for :8 hours at 15°C to 25°C2 days at 2°C to 8°C2 weeks at -25°C to -15°C**ExceptionsDirect Bilirubin3 hours at 15°C to 25°C8 hours at 2°C to 8°C2 weeks at -25°C to -15°CTotal Bilirubin6 hours at 15°C to 25°C

HORIBA ABX, FRANCE

.

{6}------------------------------------------------

ABX PENTRA Multical:

1 day at 2°C to 8°C 2 weeks at –25°C to –15°C

CONTROLS

ABX PENTRA N Control:
AnalytesAlready cleared(K052007, K060205, K060318,K060325, K060854, K062180,K062737, K060434)Included in this submission
Alkaline phosphatase
Alanine aminotransferase
Amylase
Aspartate aminotransferase
Creatine kinase
GGT
Lipase
Albumin
Direct Bilirubin
Total Bilirubin
Calcium
Chloride
Cholesterol
HDL
LDL
Creatinine
Creatinine 120
Glucose HK
Glucose PAP
Iron
Lactic acid
Magnesium
Phosphorus
Potassium
Sodium
Total Protein
Total Protein 100
Triglycerides
Urea / Blood Urea Nitrogen
Uric acid
FormatLyophilized human serum with chemical additives and materialsof biological origin

HORIBA ABX, FRANCE

{7}------------------------------------------------

ABX PENTRA N Control:
Open stability:
Once opened, the control components** are stable for :
12 hours at 15°C to 25°C
5 days at 2°C to 8°C
1 month at -25°C to -15°C
**Exceptions
Direct Bilirubin
4 hours at 15°C to 25°C
8 hours at 2°C to 8°C
2 weeks at -25°C to -15°C
Total Bilirubin
8 hours at 15°C to 25°C
1 day at 2°C to 8°C
2 weeks at -25°C to -15°C

:

:

.

{8}------------------------------------------------

ABX PENTRA P Control:
AnalytesAlready cleared(K052007, K060205, K060318,K060325, K060854, K062180,K062737, K060434)Included in this submission
Alkaline phosphatase
Alanine aminotransferase
Amylase
Aspartate aminotransferase
Creatine kinase
GGT
Lipase
Albumin
Direct Bilirubin
Total Bilirubin
Calcium
Chloride
Cholesterol
HDL
LDL
Creatinine
Creatinine 120
Glucose HK
Glucose PAP
Iron
Lactic acid
Magnesium
Phosphorus
Potassium
Sodium
Total Protein
Total Protein 100
Triglycerides
Urea / Blood Urea Nitrogen
Uric acid
FormatLyophilized human serum with chemical additives and materialsof biological origin
StabilityClosed stability: 30 months at 2-8°COpen stability:Once opened, the control components** are stable for :
12 hours at 15°C to 25°C
5 days at 2°C to 8°C
1 month at -25°C to -15°C**Exceptions
Direct Bilirubin
4 hours at 15°C to 25°C
ABX PENTRA P Control:
8 hours at 2°C to 8°C2 weeks at -25°C to -15°C
Total Bilirubin8 hours at 15°C to 25°C1 day at 2°C to 8°C2 weeks at -25°C to -15°C

HORIBA ABX, FRANCE

{9}------------------------------------------------

ABX PENTRA Urine Control L/H:
AnalytesAlready cleared (K070249)Included in this submission
Amylase
Calcium
Creatinine
Creatinine 120
Phosphorus
Glucose*
Urea / Blood Urea Nitrogen*
Uric acid*
Urinary proteins*
FormatLiquid solution prepared from human urine with chemicaladditives and materials of biological origin
StabilityClosed stability: 2 years at 2-8°COpen stability: 30 days at 2-8°C
  • Not cleared as of date of submission

Conclusions for Performance Testing :

The performance testing data conclude that the safety and effectiveness of the devices are not compromised, and that they met all acceptance criteria, demonstrating that the devices are substantially equivalent to their respective predicate devices.

{10}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. The text is arranged around the circumference of the circle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

NOV 3 0 2007

Horiba ABX c/o Olivier DUCAMP Regulatory Affairs Manager Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4, France

Re: K072115

Trade Name: ABX PENTRA Creatinine 120 CP Reagent, ABX PENTRA Total Protein 100 CP Reagent, ABX PENTRA Multical, ABX PENTRA N Control, ABX PENTRA P Control, ABX PENTRA Urine Control L/H Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system. Regulatory Class: Class II Product Codes: CGX, CEK, JIX, JJY Dated: November 07, 2007 Received: November 13, 2007

Dear Olivier DUCAMP:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{11}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Jean M. Cooper, M.S., D.V.M.

Sean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

510(k) Number (if known): K072115

Device Name: New Creatinine and Total Protein on ABX PENTRA 400 Clinical Chemistry Analyzer

Indication For Use:

Creatinine and Total Protein reagents, with associated calibrators and controls, are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer to measure a variety of analytes.

ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

ABX PENTRA Total Protein 100 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in-vitro determination of Total Proteins in serum and plasma by colorimetry.

Measurements obtained by this device are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Carol C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K072115

{13}------------------------------------------------

510(k) Number (if known): K072115

Device Name: ABX PENTRA Multical

Indication For Use:

The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba ABX methods on Horiba ABX clinical chemistry analyzers.

Prescription Use _ X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Carol C. Bennm

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K072115

{14}------------------------------------------------

510(k) Number (if known): K072115

Device Name: ABX PENTRA N Control

Indication For Use:

The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision.

Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carol C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K072115

{15}------------------------------------------------

510(k) Number (if known): K072115

Device Name: ABX PENTRA P Control

Indication For Use:

The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use __ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carl Benan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K072115

Page 4 of 5

{16}------------------------------------------------

510(k) Number (if known): K072115

Device Name: ABX PENTRA Urine Control L/H

Indication For Use:

The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carol Benam

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K072115

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.