K Number
K072096
Date Cleared
2007-08-20

(20 days)

Product Code
Regulation Number
864.8625
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is an established laboratory procedure to use stable controls to monitor the performance of diagnostic tests. The CBC-5D Plus Retics Hematology Control is designed to document and monitor values obtained from Coulter® hematology instruments. For in vitro Diagnostic Use Only

Device Description

CBC-5D Plus Retics is an in vitro diagnostic control composed of human erythrocytes, simulated leukocytes, and mammalian platelets in a plasma-like fluid with preservatives. It is composed of stable materials that provide a means of monitoring the performance of Coulter® hematology analyzers. It is sampled in the same manner as a patient specimen.

AI/ML Overview

The CBC-5D Plus Retics Hematology Control is an in vitro diagnostic control device. The provided text outlines the acceptance criteria and the study conducted to demonstrate the device's performance.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
PerformanceSubstantially equivalent to the predicate device.
PrecisionDemonstrated with small standard deviation and % CV's.
Stability (Closed Vial)Remained within the assay range for 105 days.
Stability (Open Vial)Remained within the assay range for 14 days.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The study utilized "3 validation lots" of the CBC-5D Plus Retics Hematology Control.
  • Data Provenance: Not explicitly stated, but the testing was conducted in a "laboratory setting." The K number (K072096) indicates a premarket notification for a new device, suggesting the data is prospective, generated specifically for this submission. The manufacturer is R&D Systems, Inc., located in Minneapolis, MN, USA, implying the data likely originated from the USA.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This device is a hematology control, not a diagnostic device that requires expert interpretation of images or clinical data. Therefore, the concept of "experts used to establish ground truth" as typically applied to imaging or clinical decision support AI is not relevant here. The 'ground truth' for a control device is its expected analytical range and stability.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation or complex disease diagnosis to resolve discrepancies. This is not applicable to an analytical control device where performance is measured against established analytical methods and specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study investigates the impact of a device (often AI) on human reader performance for diagnostic tasks, which is not applicable to a hematology control device.

6. If a Standalone (i.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done

While the device's performance was evaluated independently, the term "standalone" in the context of an algorithm's performance without human intervention is not fully applicable. This device is a control material used by humans to monitor instruments. The performance described (precision, stability) is inherent to the control material itself, rather than an algorithm's output.

7. The Type of Ground Truth Used

The ground truth for this hematology control device is its established analytical range and stability specifications. The device's performance was measured against these predefined criteria, which are derived from the chemical and biological properties of the control and its interaction with hematology analyzers. The goal was to demonstrate that it remained within the assay range over the life of the product.

8. The Sample Size for the Training Set

The concept of a "training set" is not relevant for this medical device. The CBC-5D Plus Retics Hematology Control is a physical control material, not an AI/ML algorithm that requires training data. Its performance is validated through laboratory testing of manufactured lots.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for this device, this question is not applicable.

{0}------------------------------------------------

K072096

AUG 2 0 2007

510(k) Summary

Submitter Information

R&D Systems, Inc. 614 McKinley Place N.E. Minneapolis, MN 55413

Contact: Nancy Ring Phone: 612-656-4533 Fax: 612-379-6580

Date Prepared: 7/30/07

Device Information

Proprietary Name: CBC-5D Plus Retics Hematology Control Common Name: Hematology Controls Classification 21 CFR 864.8625 Classification Name: Hematology Quality Control Mixture Product Code: JPK Device Class: Panel: Hematology (81)

Predicate Device

R&D Systems CBC-5D Hematology Control, (K983032) manufactured by R&D Systems, Inc. 614 McKinley Place N.E., Minneapolis, MN 55413.

Description of Device

CBC-5D Plus Retics is an in vitro diagnostic control composed of human erythrocytes, simulated leukocytes, and mammalian platelets in a plasma-like fluid with preservatives. It is composed of stable materials that provide a means of monitoring the performance of Coulter® hematology analyzers. It is sampled in the same manner as a patient specimen.

Intended Use:

The CBC-5D Plus Retics is a control designed to monitor values on Coulter hematology analyzers. Refer to assay sheet for specific instrument models.

Technological Comparison to Predicate

The new device has the same technological characteristics as the legally marketed predicate device. Both products are used to monitor Coulter® hematology instruments. Both are used to perform quality controls assays. The CBC-5D Plus Retics Hematology Control has an additional parameter.

{1}------------------------------------------------

Summary of Performance Data

Laboratory testing of 3 validation lots has shown the CBC-5D Plus Retics Hematology Control to have substantial equivalence in performance, precision and stability to the predicate device. The CBC-5D Plus Retics Hematology Control passed the acceptance criteria of remaining within the assay range over the life of the product. The CBC-5D Plus Retics Hematology Control has demonstrated precision as indicated by the small standard deviation and % CV's obtained during laboratory testing. Expiration dating will be established at 105 days (closed vial) and 14 days (open vial) when stored at 2 - 8° C and handled according to instructions for use.

Substantial Equivalence Conclusion

The data demonstrates that the CBC-5D Plus Retics Hematology Control is substantially equivalent to the legally marketed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. The eagle is depicted in black, and the text is also in black. The overall design is simple and recognizable.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG 2 0 2007

Nancy Ring R & D Systems, Inc. 614 McKinley Place, NE Minneapolis, Minnesota 55413

Re: K072096

Trade/Device Name: CBC-5D Plus Retics Hematology Control Regulation Number: 21 CFR 864.8625 Regulation Name: Hematology Quality Control Mixture Regulatory Class: Class II Product Code: JPK Dated: July 30, 2007 Received: July 31, 2007

Dear Ms. Ring:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{3}------------------------------------------------

Page 2 -

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

· Sincerely yours,

Robert A. Becker

Robert L. Becker, Jr., MD, Ph Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

July 30, 2007

Indications for Use

K072096 510(k) Number (if known): __

Device Name: CBC-5D Plus Retics Hematology Control

Indications for Use:

It is an established laboratory procedure to use stable controls to monitor the performance of diagnostic tests. The CBC-5D Plus Retics Hematology Control is designed to document and monitor values obtained from Coulter® hematology instruments.

For in vitro Diagnostic Use Only

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Dorochine Bautista
Division Sign-Off

Division

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K072096

Page 5 of 26

Page 1 of 1

§ 864.8625 Hematology quality control mixture.

(a)
Identification. A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).(b)
Classification. Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.