(237 days)
Not Found
No
The summary mentions "image processing" and "data analysis" but does not explicitly mention AI, ML, or related terms. The performance studies described are standard clinical comparison and precision tests, not typical evaluations of AI/ML model performance.
No.
The device is used for imaging and diagnosis of retinal diseases, not for treatment.
Yes
The "Intended Use / Indications for Use" section explicitly states that the system is an "aid in the diagnosis and management of retinal diseases."
No
The device description explicitly states it is an "AC-powered device containing illumination and viewing optics" and includes a "main unit, which houses the optical system used for imaging the retina, the light source, and a power supply." This indicates significant hardware components beyond just software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use states "in vivo imaging and measurement of the retina". "In vivo" means within a living organism. IVD devices are designed to be used in vitro, meaning outside of the body, typically on biological samples like blood, urine, or tissue.
- Device Description: The device description details an optical system for examining the retina of the eye, a light source, and a patient interface (chin rest/forehead rest). This is consistent with an imaging device used directly on a patient, not a device for analyzing samples in a lab.
- Anatomical Site: The anatomical site is the "Retina of the eye," which is a part of the living body.
Therefore, this device is an in vivo imaging device, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The SOCT Copernicus is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
Product codes
OBO
Device Description
The SOCT Copernicus is an AC-powered device containing illumination and viewing optics intended to examine the retina of the eye.
Like the predicate device, the SOCT Copernicus consists of a main unit, which houses the optical system used for imaging the retina, the light source, and a power supply. In addition, both systems utilize a chin rest/forehead rest assembly for patient interface and a computer with software for data analysis / image processing.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Optical Coherence Tomography
Anatomical Site
retina of the eye, fundus of the eye
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
SOCT Copernicus Precision Test (Repeatability and Reproducibility):
Study type: Structured study to determine repeatability and reproducibility precision values.
Sample size: 10 normal eyes and 10 pathological eyes.
Key results: Statistical analysis showed good agreement across devices and operators.
SOCT Copernicus Clinical Comparison:
Study type: Prospective case series comparing the Carl Zeiss Meditec Stratus™ Time Domain Optical Coherence Tomography (TDOCT) to the Reichert Copernicus™ Spectral Domain OCT (SDOCT).
Sample size: A total of 27 eyes from 21 patients with a variety of vitreo-retinal disorders, including five normal eyes.
Key results: Based on the results of the statistical analysis, no medically significant difference seems to exist between the measurements made by the two devices.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 886.1570 Ophthalmoscope.
(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
0
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K071839 .
1. Submitter's Identification:
Reichert, Inc. c/o Sandra Brown Quality Regulatory Affairs Manager 3362 Walden Avenue Depew, NY 14043 Tel: (716) 686-4599
Date Summary Prepared: January 24, 2008
2. Name of the Device: SOCT Copernicus
3. Common or Usual Name: Optical Coherence Tomographer (OCT)
4. Predicate Device Information:
- Trade Name: STRATUS OCT .
- 510(k) Number: K030433 .
- Common Name: Optical Coherence Tomographer (OCT) •
- Classification Name: Ophthalmoscope, AC-powered .
- Classification: Class II .
- Classification Panel: 86 .
- Product Code: HLI .
- Regulation Number: 886.1570 .
: 上一 7
1
న్. SOCT Copernicus Device Description:
The SOCT Copernicus is an AC-powered device containing illumination and viewing optics intended to examine the retina of the eye.
SOCT Copernicus Intended Use:
The SOCT Copernicus is an optical coherence tomography system indicated for the invivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
6. Technological Characteristics:
The Reichert SOCT Copernicus and the predicate device have similar technological characteristics. Both devices utilize Optical Coherence Tomography, which relies upon interferometry of superluminescent diode light, reflected from the fundus of the eye, to produce a cross-sectional image of the retina. Both devices have similar hardware components. Like the predicate device, the SOCT Copernicus consists of a main unit, which houses the optical system used for imaging the retina, the light source, and a power supply. In addition, both systems utilize a chin rest/forehead rest assembly for patient interface and a computer with software for data analysis / image processing.
A brief summary of major technological characteristics for both devices is presented below:
| | Factor | Subject Device
Reichert “SOCT Copernicus” | Predicate Device
Carl Zeiss “STRATUS OCT” |
|---|---------------------------------|------------------------------------------------------------|----------------------------------------------|
| 1 | Method of
Operation | Optical Coherence Tomography:
Spectral / Fourier Domain | Optical Coherence Tomography:
Time Domain |
| 2 | Exposure
Parameters | 750 μW at the cornea | 750 μW at the cornea |
| 3 | Light Source | SLED. scan beam wavelength is
840 nm. | SLED. scan beam wavelength is
820 nm. |
| 4 | OCT Image | 25,000 A-scan/second | 400 A-scan/second |
| 5 | Depth Resolution
(in tissue) | 6 μm | 10 μm |
| 6 | Transverse
Resolution | 12 μm
18 μm typical | 20 μm |
2
7. Discussion of Clinical Tests Performed:
SOCT Copernicus Precision Test (Repeatability and Reproducibility)
A structured study was conducted with 3 operators (unaware of patient pathology) and 2 devices to determine repeatability and reproducibility precision values for the SOCT Copernicus. The test population consisted of 10 normal eves (with a wide range of media opacity and other parameters affecting image quality) and 10 pathological eyes including eyes with parameters affecting image quality such as diabetic macular edema, wet AMD, dry AMD, cystoid macular edema, vitreous hemorrhage, epiretinal membrane and/or vitromacular traction, and macular holes.
Statistical analysis of the study data showed good agreement across devices and operators.
SOCT Copernicus Clinical Comparison
A prospective case series comparing the Carl Zeiss Meditec Stratus™ Time Domain Optical Coherence Tomography (TDOCT) to the Reichert Copernicus™ Spectral Domain OCT (SDOCT) was performed at the Cleveland Clinic Cole Eye Institute.
A total of 27 eves from 21 patients with a variety of vitreo-retinal disorders, including five normal eves, were scanned with both machines and analyzed. All scans were analyzed to obtain values from nine fields: central subfield, temporal inner retina, superior inner retina, nasal inner retina, inferior inner retina, temporal outer retina, superior outer retina, nasal outer retina, and inferior outer retina.
Based on the results of the statistical analysis, no medically significant difference seems to exist between the measurements made by the two devices.
9. Conclusions:
· The SOCT Copernicus and Stratus OCT are both intended for in-vivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
• The SOCT Copernicus and Stratus OCT use similar technology with the exception that SOCT Copernicus utilizes Spectral Domain technology whereas the Stratus OCT uses Time Domain technology. Spectral Domain technology results in higher resolution retinal images. In addition, the SOCT Copernicus PC software is different from the Stratus OCT, but has been completely tested and validated.
· The devices are substantially equivalent based on the precision, repeatability and clinical testing described herein.
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 27 2002
Reichert, Inc. c/o Ms. Sandra Brown Quality/Regulatory Affairs Manager 3362 Walden Avenue Depew, NY 14043
Re: K071839
Trade/Device Name: SOCT Copernicus Spectral Optical Coherence Tomographer Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: Undated Received: February 15, 2008
Dear Ms. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
4
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. B. Egelston, m.D.
Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number (if known): _ K071839
Device Name: Reichert SOCT Copernicus Spectral Optical Coherence Tomographer
Indications For Use:
An optical coherence tomography system indicated for the in vivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
|
------------ |
Ana March |
---|
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CALLERS AND LEASE OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF |
| The name
A A A A A
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
(Division Sign-Off) |
---|
Division of Ophthalmic Ear, |
Nose and Throat Devises |
510(k) Number | K071839 |
---|---|
--------------- | --------- |
Prescription Use (Per 21 CFR 801.109) | |
---|---|
Page 1 of / |