K Number
K071631

Validate with FDA (Live)

Device Name
NICOM ELECTRODES
Date Cleared
2008-01-18

(218 days)

Product Code
Regulation Number
870.1435
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cheetah NICOM system is a portable, non-invasive Cardiac Output monitoring device based on bio-impedance Cardiography. The Cheetah NICOM system is intended to monitor and display a patient's Cardiac Output in units of Ltr/Min, and was cleared under K042144.

NICOM Electrodes are disposable electrodes used in conjunction with the NICOM signal processing product line.

Device Description

The Cheetah NICOM is a portable, non-invasive, Cardiac Output detector system. The NICOM system measures the Cardiac Output by employing the electrical bio-impedance measurement technique. Electrical bio-impedance is the characteristic impedance of a volume of tissue and fluid. In the case of Cardiac Output measurements, the relevant tissue includes the heart and the immediate surrounding volume of the thorax. The relevant fluid is blood.

Cheetah Medical's NICOM electrode is a double electrode sticker. Within each sticker, one electrode is used to inject a high frequency sine wave into the body, while the resulting voltage is measured at the adjacent electrode. An array of such electrodes is placed at specific areas of the thorax, the impedance to the current flow calculated, and the electrical bio-reactance waveform constructed.

AI/ML Overview

The provided text describes a 510(k) summary for the Cheetah NICOM System and its electrodes. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with specific acceptance criteria and detailed performance metrics for the device itself.

However, based on the information provided, we can infer and construct some of the requested elements, particularly regarding the equivalence of the NICOM electrodes.

Here's an analysis based on the supplied text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria in terms of sensitivity, specificity, accuracy, or other performance metrics for the Cheetah NICOM System's cardiac output measurement functionality, nor for the NICOM electrodes in isolation of their primary function.

Instead, the primary "acceptance criterion" demonstrated in the clinical testing of the NICOM electrodes is substantial equivalence to the 3M Health Care Red Dot electrodes when used with the NICOM System.

CriterionReported Performance
NICOM Electrodes Equivalence (Clinical)"The results demonstrate that the NICOM electrodes are substantially equivalent to the 3M Health Care Red Dot electrodes when used with the NICOM System."
No Adverse Events Reported (Clinical)"119 consecutive patients requiring PAC in the immediate postoperative period following cardiac surgery - No adverse events were reported.""36 patients undergoing cardiac output evaluation during stress testing in Lonestar Heart Center, Amarillo, Texas, US. No adverse events were reported." (This implies safety, a key aspect of acceptance for medical devices)
ANSI/AAMI EC12:2000/(R)2005 Compliance (Bench)"Electrodes were tested per ANSI/AAMI EC12:2000/(R)2005; Disposable ECG electrodes, 3ed." (Implies electrical performance and safety acceptance)
ISO-10993 Compliance (Biocompatibility)"Biocompatibility was confirmed per ISO-10993 standard; Biological Evaluation of Medical Devices." (Implies biocompatibility and safety acceptance)

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Study for NICOM Electrodes Equivalence:
    • Sample Size: 10 subjects.
    • Data Provenance: Not explicitly stated, but the context of the other clinical studies (Lonestar Heart Center, Amarillo, Texas, US) suggests a prospective study conducted in the United States. The study involved measuring cardiac output in these subjects using both NICOM electrodes and the predicate 3M Red Dot electrodes.
  • Clinical Safety Studies (though not performance test sets per se):
    • Sample Size:
      • 119 consecutive patients (postoperative cardiac surgery).
      • 36 patients (stress testing).
    • Data Provenance: Lonestar Heart Center, Amarillo, Texas, US for the 36 patients. The location for the 119 patients is not specified but is likely also in the US given the context of a US-based submission. These appear to be prospective studies focused on safety/adverse events.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish a ground truth for the clinical studies. For the electrode equivalence study, "cardiac output was measured in 10 subjects." The ground truth for cardiac output itself would likely be derived from a reference method (e.g., PAC - Pulmonary Artery Catheter, mentioned in another study) or the comparative device (3M Red Dot electrodes). There is no mention of expert consensus for ground truth on the electrodes' performance.

4. Adjudication Method for the Test Set

Not specified in the provided text. The nature of the clinical "testing" for electrode equivalence ("Cardiac output was measured in 10 subjects with the NICOM System using the NICOM electrodes and using the 3M Health Care Red Dot electrodes") suggests a direct comparison, rather than an adjudication process requiring multiple reviewers for a diagnostic finding.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This document describes a medical device (cardiac output monitor and its electrodes), not an AI algorithm intended to assist human readers in image interpretation or diagnosis. Therefore, an MRMC comparative effectiveness study with human readers and AI assistance is outside the scope of this submission.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The Cheetah NICOM System is a medical device that measures physiological parameters. Its "performance" is inherent to the device's ability to measure cardiac output, not an algorithm's standalone diagnostic ability in the way an AI might operate. The system itself operates "standalone" in its measurement function, but it's a device measurement, not an algorithm interpreting data that a human would then review.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the NICOM electrodes' equivalence study, the ground truth appears to be based on the measurements obtained using the predicate device (3M Health Care Red Dot electrodes) when used with the NICOM System. The direct comparison between the new electrodes and the predicate electrodes provides the basis for demonstrating equivalence in cardiac output measurements. Another clinical study mentions "patients requiring PAC," suggesting that a Pulmonary Artery Catheter (PAC) might serve as a reference standard for cardiac output measurements in some contexts, but this isn't explicitly tied to the ground truth for electrode equivalence.

8. The Sample Size for the Training Set

  • Not Applicable. This document does not mention any machine learning or AI components that would require a distinct training set. The device's operation is based on bio-impedance measurements and signal processing, not a trained algorithm in the modern AI sense.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As there is no mention of a training set or machine learning, this question does not apply.

{0}------------------------------------------------

K071631/52
p1/3

Image /page/0/Picture/1 description: The image shows the logo for Cheetah Medical. The logo features a stylized drawing of a cheetah leaping forward. Below the cheetah drawing, the words "Cheetah Medical" are printed in a simple, sans-serif font.

10291 N M Suito 10 Indianapolis, IN . I ISA

Image /page/0/Picture/3 description: The image is a gradient that transitions from white to black. The left side of the image is predominantly white, with a gradual increase in black speckles and shading as it moves towards the right. The right side of the image is almost entirely black, with only a few scattered white pixels. The transition between the two colors is smooth and gradual.

SECTION 5: 510(k) Summary

Submitter

Cheetah Medical, Inc. 2138 East 62nd Street, Suite 132 Indianapolis, IN 46250 USA Toll Free: (+1) 866 751 9097 Fax: (+1) 916 367 6717

Contact Person

Rhona Shanker Z & B Enterprises, Inc. 12154 Darnestown Road, #236 Gaithersburg, MD 20878 Ph: 301-251-9570 Fax: 301-251-9571 rhonashanker@comcast.net

Date Prepared

January 10, 2008

Device Information

Trade name: Cheetah NICOM System

Common name: Cheetah NICOM

Classification Name: Pre-Programmable Diagnostic Computer (21 CFR 870.1435, Product Code: DXG; 21 CFR 870.2360 Product Code DRX)

Tel. (+1) 317-660-6649 ● Fax.(+1) 317-660-6650 新闻 发布时间: 上一篇:

JAN 18 2008

{1}------------------------------------------------

K071631/52
p2/3

Image /page/1/Picture/1 description: The image shows the logo for Cheetah Medical. The logo includes a stylized image of a cheetah leaping forward. The text "Cheetah Medical" is written in a simple, sans-serif font below the cheetah image.

Indianapolis, IN 46290

Image /page/1/Picture/3 description: The image is mostly black with some white noise on the left side. The white noise appears to be concentrated on the left side of the image, gradually fading into the black background. The black background is uniform and covers most of the image. The image has a grainy texture.

Devices to which substantial equivalence is claimed:

The Cheetah NICOM system was found substantially equivalent to the VLIGILANCE CCO/CEDV 7 VIGILANCE CCO/SVO2/CEDV Monitor (K000664) under K042144.

The NICOM Electrodes are substantially equivalent to the electrodes cleared for marketing with the original NICOM Monitoring System (K042144), when used with the NICOM system. The electrodes cleared in the original submission were the Red Dot™ electrodes by 3M Health Care.

Device Description:

The Cheetah NICOM is a portable, non-invasive, Cardiac Output detector system. The NICOM system measures the Cardiac Output by employing the electrical bio-impedance measurement technique. Electrical bio-impedance is the characteristic impedance of a volume of tissue and fluid. In the case of Cardiac Output measurements, the relevant tissue includes the heart and the immediate surrounding volume of the thorax. The relevant fluid is blood.

Cheetah Medical's NICOM electrode is a double electrode sticker. Within each sticker, one electrode is used to inject a high frequency sine wave into the body, while the resulting voltage is measured at the adjacent electrode. An array of such electrodes is placed at specific areas of the thorax, the impedance to the current flow calculated, and the electrical bio-reactance waveform constructed.

Indications for Use:

The Cheetah NICOM system is a portable, non-invasive Cardiac Output monitoring device based on bio-impedance Cardiography. The Cheetah NICOM system is intended to monitor and display a patient's Cardiac Output in units of Ltr/Min, and was cleared under K042144.

NICOM Electrodes are disposable electrodes used in conjuction with he NICOM signal processing product line.

Brief Description of Non-Clinical Testing

    1. Electrodes were tested per ANSI/AAMI EC12:2000/(R)2005; Disposable ECG electrodes, 3ed.
    1. Biocompatibility was confirmed per ISO-10993 standard; Biological Evaluation of Medical Devices.

Tel. (+1) 317-660-6649 · Fax.(+1) 317-660-6650 · www.cheetah.medical.com

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows handwritten text that appears to be a combination of alphanumeric characters. The first line reads 'K071631/S2', with a mix of letters and numbers separated by a forward slash. Below this, the text 'P3/3' is written, possibly indicating a page number or a similar identifier. The handwriting is clear and legible, with a simple, unadorned style.

Image /page/2/Picture/1 description: The image shows the logo for Cheetah Medical. The logo includes a stylized image of a cheetah leaping forward. The text "Cheetah Medical" is written in a simple, sans-serif font below the cheetah image.

10291 N Meric Suite 100 Indianapolis, IN 46290 USA

Image /page/2/Picture/3 description: The image is a black and white gradient. The left side of the image is white, and the right side is black. The gradient is smooth, and there are no sharp edges. The image is simple and abstract.

Brief Description of Clinical Testing

NICOM Electrodes were tested on:

    1. 119 consecutive patients requiring PAC in the immediate postoperative period following cardiac surgery - No adverse events were reported.
    1. 36 patients undergoing cardiac output evaluation during stress testing in Lonestar Heart Center, Amarillo, Texas, US. No adverse events were reported.
    1. Cardiac output was measured in 10 subjects with the NICOM System using the NICOM electrodes and using the 3M Health Care Red Dot electrodes. The results demonstrate that the NICOM electrodes are substantially equivalent to the 3M Health Care Red Dot electrodes when used with the NICOM System.

Conclusion

The NICOM Electrode is a dual electrode sticker, used in conjunction with its NICOM system to monitor cardiac output by measuring electrical impedance changes in the thorax region. The Cheetah NICOM system is intended for prescription use only.

NICOM electrodes were extensively tested (bench, animal, clinical) and proved to be safe and effective.

The data demonstrate that the NICOM electrodes are substantially equivalent to the 3M Red Dot Electrodes that were cleared with the original NICOM System (K042144) when used with the NICOM system.

Tel. (+1) 317-660-6649 · Fax.(+1) 317-660-6650 · www.cheetah-medical.con

· Dr. Dr. 257 255 255 255 255 255 2

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird with three wing-like strokes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 18 2008

Cheetah Medical c/o Ms. Rhona Shanker Z & B Enterprises, Inc. 12154 Darnestown Road # 236 Gaithersburg, MD 20878

Re: K071631

Cheetah NICOM System Regulation Number: 21 CFR 870.1435 Regulation Name: Single-function preprogrammed diagnostic computer Regulatory Class: Class II (two) Product Code: DXG Dated: June 14, 2007 Received: June 14, 2007

Dear Ms. Shanker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Rhona Shanker

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Dma R. Vochner

A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K071631/52
pi of l

Image /page/5/Picture/1 description: The image shows the logo for Cheetah Medical. The logo features a stylized image of a cheetah leaping, with the text "Cheetah Medical" printed below the image. A trademark symbol is located to the right of the cheetah image.

10291 N. Meri Suite 100 Indianapolis, IN 46 USA

Image /page/5/Picture/3 description: The image appears to be a gradient that transitions from white to black. The left side of the image is predominantly white, with a high density of white pixels. As one moves towards the right, the white gradually fades into shades of gray, eventually becoming completely black on the right side. The transition is smooth, creating a visual effect of fading or blending.

SECTION 4: Indications for Use

510(k) Number: K6716 3)

Device Name: Cheetah NICOM System

Indications for Use:

The Cheetah NICOM system is a portable, non-invasive Cardiac Output monitoring device based on bio-impedance Cardiography. The Cheetah NICOM system is intended to monitor and display a patient's Cardiac Output in units of Ltr/Min, and was cleared under K042144.

NICOM Electrodes are disposable electrodes used in conjunction with the NICOM signal processing product line.

AND/OR Over-The-Counter Use Prescription Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

sumar R. de Amez

ardiovascular Devices

10(k) Number. Tel. {+1) 317-660-6649 @ Fax.(+1) 317-660-6650 . www.cheetah-medical.com A STAND STATE A

§ 870.1435 Single-function, preprogrammed diagnostic computer.

(a)
Identification. A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.(b)
Classification. Class II (performance standards).