K Number
K071572

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-07-23

(45 days)

Product Code
Regulation Number
870.4350
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Capiox FX05 device is intended to be used during open heart surgical procedures to transfer oxygen and remove carbon dioxide from blood and to control the blood temperature during cardiopulmonary bypass for periods up to 6 hours. The Capiox FX05 is a Neonate/Infant oxygenator intended for use in procedures up to a maximum flow of 1.5 L/min. The patient weight and BSA should be considered upon use.

The FX05 Hardshell Reservoir is also intended for use in vacuum assisted venous drainage procedures.

The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

Device Description

The Capiox FX05 device is a modification of the Capiox RX05 device that has an arterial filter integrated into the design. The design of the oxygenator device is such that it utilizes an integrated oxygenator/heat exchanger module that provides for gas transfer (blood oxygenation and carbon dioxide removal) and for blood temperature control. The RX Oxygenator/Arterial Filter device also utilizes a hardshell reservoir that is used to collect and store blood during a cardiopulmonary bypass procedure. The arterial filter contained within the oxygenator module is comprised of 32 micron PET (polyethylene terephalate) mesh material that is wrapped around the outer circumference of the oxygenator fiber bundle.

AI/ML Overview

The provided document describes the Capiox® FX05 Hollow Fiber Oxygenator and Arterial Filter, a medical device intended for use during open heart surgical procedures. The submission is a 510(k) premarket notification, which means it aims to demonstrate substantial equivalence to legally marketed predicate devices, rather than establishing completely new safety and effectiveness. As such, the "study" described is a series of in-vitro performance evaluations comparing the new device to its predicates.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" with specific numerical targets. Instead, it states that the device "exhibited performance that is deemed to be substantially equivalent to the performance of the predicate devices." This substantial equivalence is based on a series of in-vitro tests where the new device's performance was compared to the established performance of the predicate devices. The implicit acceptance criterion is that the new device's performance on these tests is comparable to or better than the predicate devices, and that any differences do not raise new issues of safety or effectiveness.

Test CategoryReported Device Performance
Gas TransferPerformance "deemed to be substantially equivalent to predicate devices."
Effects on Blood Components (Hemolysis)Performance "deemed to be substantially equivalent to predicate devices."
Pressure DropPerformance "deemed to be substantially equivalent to predicate devices."
Mechanical IntegrityPerformance "deemed to be substantially equivalent to predicate devices."
Static Priming VolumePerformance "deemed to be substantially equivalent to predicate devices."
Heat Exchanger PerformancePerformance "deemed to be substantially equivalent to predicate devices."
Filtration EfficiencyPerformance "deemed to be substantially equivalent to predicate devices."
Air HandlingPerformance "deemed to be substantially equivalent to predicate devices."
Tubing Connection StrengthPerformance "deemed to be substantially equivalent to predicate devices."

2. Sample size used for the test set and the data provenance

The document explicitly states: "Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations."

  • Sample Size for Test Set: Not specified. The document repeatedly refers to "in-vitro performance evaluations" and "testing" without providing numerical sample sizes for each test (e.g., how many units were tested for pressure drop, or how many blood samples were used for hemolysis).
  • Data Provenance: In-vitro laboratory testing (likely conducted by Terumo Corporation or its designated testing facilities). The country of origin for the manufacturing and submission is Japan/USA, but the specific location of the in-vitro testing is not detailed. The data is prospective in the sense that these tests were conducted specifically for this submission to demonstrate equivalence.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable to this type of submission. The "ground truth" for mechanical and performance characteristics of medical devices is typically established through adherence to recognized international standards and scientifically validated test methods, rather than expert consensus on a test set. The predicate devices themselves represent the "ground truth" for comparison in a 510(k) process, as they are already legally marketed and presumed safe and effective.

4. Adjudication method for the test set

Not applicable. Device performance against technical specifications and predicate device performance is typically determined through direct measurement and comparison, not adjudication by human experts in the context of this 510(k) submission.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical medical device (an oxygenator and filter), not an AI-powered diagnostic or decision support system. Therefore, MRMC studies and AI-assisted human reader performance are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical medical device, not an algorithm.

7. The type of ground truth used

For the in-vitro performance evaluations, the "ground truth" is established by:

  • Predicate Device Performance: The primary "ground truth" for comparison is the documented and accepted performance characteristics of the predicate devices (Terumo's Capiox® RX05 Oxygenator/Reservoir K022115 and Terumo's Capiox® AF02X Arterial Filter K011804). The new device must perform comparably to these established devices.
  • Accepted Industry Standards: Implied adherence to relevant industry standards for medical device testing (e.g., for blood compatibility, mechanical integrity, gas exchange efficiency).
  • Biocompatibility Standards: Biocompatibility evaluation was done according to ISO 10993 guidelines, implying this standard serves as the ground truth for safety in this aspect.

8. The sample size for the training set

Not applicable. As this is a physical medical device and not an AI/ML algorithm, there is no "training set."

9. How the ground truth for the training set was established

Not applicable. Since there is no training set, there is no ground truth for it.

{0}------------------------------------------------

T) TERUMO

KO71572

Submitter Information:
This submission was prepared in May 2007 by:JUL 23 2007
Garry A. Courtney, MBA, RAC
Sr. Regulatory Affairs Specialist
Regulatory Management
Terumo Corporation/Terumo Cardiovascular
125 Blue Ball Road
Elkton, MD 21921
Telephone: 1-800-283-7866, Ext. 7420
This submission was prepared for:
Terumo Corporation (Ashitaka Factory)
Manufacturer/Sterilizer
150 Maimaigi-cho
Fujinomiya City, Shizuoka Pref.
Japan 418-0015
Registration #9681834
Device Names/Classifications:
Proprietary NameClassification NameCommon Name
Capiox® FX05 HollowFiber Oxygenator andArterial FilterCardiopulmonary BypassOxygenator (Code: DTZ)Oxygenator
Cardiopulmonary BypassHeat Exchanger (Code: DTR)Heat Exchanger
Cardiopulmonary BypassArterial Line Blood Filter(Code: DTM)Arterial Filter
Cardiopulmonary BypassBlood Reservoir (Code DTN)Blood Reservoir

Predicate Device:

The device submitted in this 510(k) maintains characteristics that are substantially equivalent in intended use, design, technology/principles of operation, materials and specifications to the following devices:

  • Terumo's Capiox® RX05 Oxygenator/Reservoir K022115. .
  • Terumo's Capiox® AF02X Arterial Filter K011804. .

Intended Use:

The Capiox FX05 device is intended to be used during open heart surgical procedures to transfer oxygen and remove carbon dioxide from blood and to control the blood temperature during cardiopulmonary bypass for periods up to 6 hours. The Capiox FX05 is a Neonate/Infant oxygenator intended for use in procedures up to a maximum flow of 1.5 L/min. The patient weight and BSA should be considered upon use.

{1}------------------------------------------------

The FX05 Hardshell Reservoir is also intended for use in vacuum assisted venous drainage procedures.

The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

Principles of Operation and Technology:

The Capiox FX05 Oxygenator utilizes micro-porous membrane technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary bypass surgery. A polypropylene fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion. Diffusion occurs as oxygen moves from the gas-phase into the blood-phase; and carbon dioxide moves from the blood-phase into the gas-phase. The carbon dioxide is subsequently exhausted via a gas outlet port.

The Capiox FX05 device has an integrated heat exchanger that is comprised of a stainless steel bellows assembly encased in a polycarbonate housing. The stainless steel acts as a heat transfer material that permits heat that is generated from a temperature controlled external water bath to transverse across the walls of the stainless steel to effect the necessary temperature change upon circulating blood.

The Capiox FX device relies on external pump technology to allow for proper blood flow through the device. That is, after blood has been collected into a reservoir via gravity and/or vacuum assist, a pump (either roller or non-roller) will propel the blood into the oxygenator unit. The unit itself is not electro-mechanical in nature. Note: The external pump is responsible for facilitating blood flow through the extracorporeal circuit.

With respect to the filtration of arterial blood, the Capiox FX05 Oxygenator/Arterial Filter relies upon mechanical entrapment of particulates and emboli within the polyethylene terephthalate mesh as a means to remove those particulates from the blood. Air is removed from the blood stream via diffusion of the air into the gas phase of the oxygenator (i.e., inside of the polypropylene fibers) - where it is expelled through a gas outlet port located at the base of the oxygenator module.

Design and Materials:

With respect to the design of the oxygenator, the Capiox FX05 device is a modification of the Capiox RX05 device that has an arterial filter integrated into the design. The design of the oxygenator device is such that it utilizes an integrated oxygenator/heat exchanger module that provides for gas transfer (blood oxygenation and carbon dioxide removal) and for blood temperature control. The RX Oxygenator/Arterial Filter device also utilizes a hardshell reservoir that is used to collect and store blood during a cardiopulmonary bypass procedure. Note: The Hardshell Reservoir has been previously cleared by FDA in K022115.

With respect to the design of the Arterial Filter contained within the oxygenator module is comprised of 32 micron PET (polyethylene terephalate) mesh material that is wrapped around the outer circumference of the oxygenator fiber bundle. This design permits the oxygenation of blood and removal of carbon dioxide as blood passes through the fiber bundle -

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Terumo. The logo consists of a circle with a stylized "T" inside, followed by the word "TERUMO" in bold, sans-serif font. The logo is simple and recognizable, and it is likely used to represent the company's brand.

and also facilitates blood filtration after the blood has been oxygenated. This filtration is designed to remove particulate matter and emboli prior to the blood's re-entry to the patient's body via mechanical entrapment. Air removal is accomplished by entrapment followed by permeation of the air into the hollow fibers of the oxygenator bundle -- and subsequently is exhausted (along with carbon dioxide) via the gas outlet port.

The materials that are used in the construction of the Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter includes, but are not limited to, polycarbonate, stainless steel, polyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene terephthalate, polyethylene and X-Coating™.

Performance Evaluations:

Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:

  • Gas Transfer �
  • Effects on Blood Components (Hemolysis) .
  • Pressure Drop .
  • Mechanical Integrity .
  • Static Priming Volume .
  • Heat Exchanger Performance .
  • Filtration Efficiency .
  • Air Handling .
  • Tubing Connection Strength .

Substantial Equivalence Comparison:

The information presented in this section depicts a comparison between the subject of this 510(k) submission, the Capiox FX05 Oxygenator/Arterial Filter, and the predicate Capiox RX05 Oxygenator and Capiox AF02X Arterial Filter devices.

Comparison of Intended Use: .

The Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter and the predicate Capiox RX05 devices are equivalent in their intended uses:

Use during open heart surgical procedures to transfer oxygen and remove carbon dioxide from blood and to control the blood temperature during cardiopulmonary bypass procedures for periods up to 6 hours. Both the new and predicate devices are neonate/infant oxygenators intended for use in procedures requiring a maximum blood flow rate of 1.5 L/min. Both devices indicate that the patient weight and BSA should be considered upon use.

The Hardshell Reservoir that is used with both the FX05 device and the predicate RX05 device may be used in vacuum assisted venous drainage procedures.

The Capiox FX05 Oxygenator/Arterial Filter and the predicate Capiox AF02X Arterial Filter have the same intended uses. Both filters are intended to filter non-

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "TERUMO" in bold, black letters. To the left of the word is a circle with a "T" inside. To the right of the word is a superscript symbol that looks like an asterisk.

biologic particles and emboli and to facilitate air bubble removal from the blood flowing through a cardiotomy bypass circuit.

Duration of Use: .

The Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter and the predicate devices can be used in procedures lasting up to 6 hours.

Comparison of Labeling: .

Both the Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter and the predicate Capiox devices are offered with adequate Instructions for Use and other product labeling as required The Instructions for Use for the Capiox FX05 Hollow Fiber by regulation. Oxygenator/Arterial Filter are presented in the Appendices of this submission; the Instructions for Use for the predicate Capiox devices are also presented in the Appendices of this submission.

Comparison of Principles of Operation & Technology: .

Both the Capiox FX05 Hollow Fiber Oxygenator and the predicate Capiox RX05 Oxygenator utilize the exact same technologies and principles of operation:

Each device utilizes micro-porous membrane technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary bypass surgery. A polypropylene fiber bundle provides a porous membrane surface to sufficiently allow the movement of gases through the walls of the hollow fibers via diffusion. Diffusion occurs as oxygen moves from the gas-phase into the blood-phase; and carbon dioxide moves from the blood-phase into the gas-phase. The carbon dioxide is subsequently exhausted via a gas outlet port.

The Capiox FX05 device and the predicate Capiox RX05 device each have integrated heat exchangers that are comprised of a stainless steel bellows assembly. The stainless steel acts to provide a mechanism for the transfer of heat (generated from a temperature controlled external water bath) across the walls of the stainless steel to effect the necessary temperature change on the blood.

The Capiox FX05 device and the predicate Capiox RX05 device each rely on external pump technology to allow for proper operation. That is, after blood has been collected into a reservoir, a pump (either roller) will propel the blood into the oxygenator unit. The unit itself is not electro-mechanical in nature.

With respect to the filtration of arterial blood, the Capiox FX05 Oxygenator/Arterial Filter and the predicate Capiox AF02X each rely upon mechanical entrapment of particulates and emboli. However, there are differences in their respective mechanisms for air removal:

The predicate Capiox AF02X removes air from the extracorporeal circuit by forcing the bubbles to move towards an air purge that is located in the upper region of the filter. With such a design, as air migrates to the purge port, it is easily removed via the port. By contrast, the Capiox FX05 arterial filter removes air by permeation of air into the gas phase (i.e., inside of the polypropylene fibers) of the oxygenator where it is expelled

{4}------------------------------------------------

through a gas outlet port located at the base of the oxygenator module. This difference in operation does not present any additional issues of safety and/or effectiveness - but merely represent a difference in the manner in which air is removed from the patient's blood.

The Capiox FX05 Hollow Fiber Oxygenator/ and the predicate device are substantially equivalent in operation and technology. The noted differences do not raise any new issues of safety and effectiveness.

Comparison of Design: .

With respect to the design of the oxygenator, the design of the Capiox FX05 Hollow Fiber Oxygenator has the exact same design as the predicate device. In fact, the Capiox FX05 device is a modification of the Capiox RX05 device that has an arterial filter integrated into the design. The design of the oxygenator devices is exact in that they each utilize the same integrated oxygenator/heat exchanger module that provides for gas transfer (blood oxygenation and carbon dioxide removal) and for blood temperature control. Each of the devices also utilizes a hardshell reservoir that is used to collect and store blood during a cardiopulmonary bypass procedure. The reservoirs each provide for filtration of venous and cardiotomy blood as it enters the reservoir.

Noted difference in design between the Capiox FX05 device and the predicate Capiox RX05 devices:

The Capiox FX05 device has an integrated Arterial Filter that is enclosed within the oxygenator module of the device. In essence, the oxygenator and the integrated arterial filter are combined into a singular unit; whereas the Capiox RX05 device does not have an integrated Arterial Filter.

With respect to the design of the Arterial Filter, the filter contained within the oxygenator module is comprised of a 32 micron PET (polyethylene terephalate) mesh material that is wrapped around the outer circumference of the oxygenator bundle. This design permits the oxygenation of blood and removal of carbon dioxide as blood passes through the fiber bundle - and also facilitates blood filtration after the blood has been oxygenated. This filtration is designed to remove particulate matter and emboli prior to the blood's re-entry to the patient's body via mechanical entrapment. Air removal is accomplished by entrapment followed by permeation of the air into the hollow fibers of the oxygenator bundle - and subsequently is exhausted (along with carbon dioxide) via the gas outlet port.

By contrast, the predicate Capiox AF02X filter is a unit that is not integrated into an oxygenator, but rather, is a separate device that may be included in the extracorporeal circuit. When included in an extracorporeal circuit, the predicate Capiox AF02X is positioned between the oxygenator and the patient - thereby effecting the same surgical benefits as the Capiox FX device that is the subject of this premarket notification application.

Terumo considers the design of the Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter to be comparable and substantially equivalent to that of the predicate Capiox RX05 Oxygenator and Capiox Arterial Filter devices. The noted differences do not raise new issues of safety and effectiveness.

{5}------------------------------------------------

. Comparison of Materials:

The Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter and the predicate Capiox devices are manufactured from the exact same materials. The devices (both new and predicate) are comprised of polycarbonate, stainless steel (heat exchanger), polypropylene (gas transfer material), polyvinyl chloride, polyurethane (adhesives), polyethylene, polyester and other materials that are typically used in the manufacture of medical devices.

The Capiox FX05 devices and predicate devices contain polymethoxyethyl acrylate (PMEA), which is a biocompatible coating applied to the blood contacting surfaces of the devices. The PMEA coating is effective in reducing platelet adhesion to the device as blood circulates through the device during surgical procedures.

The safety and efficacy of PMEA is well recognized and is used on several other medical devices that have been cleared by FDA. The use of PMEA on medical devices has been demonstrated as safe for use on medical devices and it's use is well documented in the medical community.

Terumo considers the Capiox FX05 device and the predicate devices to be comparable and substantially equivalent in materials of construction - especially since those materials are identical.

Comparison of Performance: .

The Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter exhibited performance that is deemed to be substantially equivalent to the performance of the predicate devices. This determination is based upon the results of the following tests:

  • Gas transfer testing. .
  • Effects upon cellular blood components of the integral oxygenator/heat exchanger. .
  • Pressure Drop Testing of the integral oxygenator/heat exchanger. ●
  • Mechanical Integrity of the integral oxygenator/heat exchanger. .
  • Static Priming Volume of the integral oxygenator/heat exchanger. .
  • Heat Exchanger performance evaluation (heat transfer). ●
  • Filtration Efficiency evaluations. .
  • Air Handling testing. .
  • Tubing Connection Strength .

Each of the above-indicated tests is reported in detail in Section 18 of this submission.

Conclusion:

In summary, Terumo deems the Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter as substantially equivalent to the predicate Capiox RX05 Oxygenator and Capiox AF02X Arterial Filter device with respect to intended use, duration of use, design, materials, principles of

1 Polymethoxyethyl acrylate is also used on X-Coated SX18/25 Oxygenators (K993772), Hardshell Reservoirs (K002238), CX*AF200X Arterial Filters (K002026), X-Coated Tubing (K003371), X-Coated Adapters (K003604), RX Hardshell Reservoirs (K013526), and X-Coated SX10 Oxygenators (K010443).

{6}------------------------------------------------

operation, performance and specifications. It is further noted that any recognized differences do not raise any new issues of patient/user safety or product effectiveness.

Substantial Equivalence Statement:

The Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter and the predicate devices are I he Capios 17100 from intended use, principles of operation and technology, design and materials, and performance. Any noted differences between the subject device and the predicate devices do not raise new issues of safety and effectiveness.

Additional Safety Information:

  • Sterilization conditions have been validated in accordance with AAMI guidelines to provide . a Sterility Assurance Level (SAL) of 10%. Terumo further asserts that the ethylene oxide residues will not exceed the maximum residue limits at the time of product distribution.
  • Terumo maintains biocompatibility studies as recommended in the FDA General Program • Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing." [External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. The blood contacting materials were found to be biocompatible.
  • The polymer coating material that is applied to the blood-contacting surfaces of the device . was also evaluated in an in-vivo animal study. No adverse conditions were noted.

Conclusion:

In summary, the Capiox FX05 Hollow Fiber Oxygenator/Arterial Filter is substantially equivalent in intended use, principles of operation and technology, design and materials, and performance to the predicate oxygenator and reservoir devices identified in this application.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 3 2007

Terumo Corporation c/o Mr. Gary A. Courtney, MBA, RAC Sr. Regulatory Affairs Specialist 125 Blue Ball Road Elkton, MD 21921

Re: K071572

Device Name: CAPIOX® FX05 Hollow Fiber Oxygenator with Integrated Arterial Filter Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: June 7, 2007 Received: June 8, 2007

Dear Mr. Courtney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{8}------------------------------------------------

Page 2 - Mr. Gary A. Courtney, MBA. RAC

CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Duna R. Lochner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Image /page/9/Picture/17 description: The image shows the word "TERUMO" in bold, black letters. To the left of the word is a circle with a "T" inside. There is a superscript symbol to the right of the word.

SECTION 4 Indications for Use

Unknown at time of Submission> k 6 7 ( 5 7 2 510(k) Number (if known):

CAPIOX® FX05 Hollow Fiber Oxygenator with Integrated Arterial Filter Device Name:

Indications For Use:

The Capiox FX05 device is intended to be used during open heart surgical procedures to transfer oxygen and remove carbon dioxide from blood and to control the blood temperature during cardiopulmonary bypass for periods up to 6 hours. The Capiox FX05 is a Neonate/Infant oxygenator intended for use in procedures up to a maximum flow of 1.5 L/min. The patient weight and BSA should be considered upon use.

The FX05 Hardshell Reservoir is also intended for use in vacuum assisted venous drainage procedures.

The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

Prescription Use XX (Part 21 CFR 801 Subpart D) OROver-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duma R. Vuhnes

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number K671532

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”