K Number
K071542
Date Cleared
2007-07-05

(30 days)

Product Code
Regulation Number
892.1710
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Hologic Digital StereoLoc II is an optional accessory for the Selenia full field digital mammography system. It is designed to allow the accurate location of lesions in the breast in three dimensions, using information extracted from stereotactic pairs of twodimensional images. It is intended to provide guidance for interventional purposes (such as biopsy, pre-surgical localization or treatment devices).

Device Description

The Digital StereoLoc II is an optional accessory to the Selenia Full Field Digital Mammography System. It is attached to the Selenia to perform stereotactic biopsies and is composed of the following elements:

  • The Biopsy Unit or Stage used for patient compression and vertical mounting of the . biopsy needle.
  • The SmartWindow control module for X and Y-axis motion control and Z-axis . positioning guidance.
  • DSM computer workstation and software for stereotactic targeting. .
    The breast is compressed between the compression paddle and the carbon fiber breast platform of the biopsy unit. The biopsy unit has its own manual compression mechanism that is independent of the Selenia FFDM compression device. There are two compression knobs; one at each side of the biopsy unit. The Smart Window control module gets the Cartesian coordinates (X, Y, Z) of the target from the DSM workstation.
  • X and Y motion is motorized. Controlled by the operator via the SmartWindow. .
  • . Z-axis motion is manual and controlled by the operator.
AI/ML Overview

Although the provided text doesn't contain a detailed study with acceptance criteria and reported device performance in a tabular format, nor does it include information on sample sizes, ground truth establishment, or specific study designs (like MRMC or standalone), I can extract the relevant information regarding the device's assessment and general conclusions.

Here's a summary based on the provided 510(k) summary:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria or a table of performance metrics. Instead, it relies on a qualitative assessment of substantial equivalence to predicate devices and compliance with relevant standards.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not provided in the document. The filing focuses on the device's function as an accessory for guidance during interventional procedures, not on AI-assisted diagnostic improvement.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not provided in the document. The device is described as an accessory that provides guidance for human operators during interventional procedures.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not provided in the document.

8. The sample size for the training set

This information is not provided in the document.

9. How the ground truth for the training set was established

This information is not provided in the document.


Summary of Device Assessment from the Provided K071542 Document:

The 510(k) summary for the Hologic Digital StereoLoc II primarily establishes substantial equivalence to predicate devices rather than proving performance against specific quantitative acceptance criteria through a dedicated study.

Study that Proves the Device Meets the Acceptance Criteria:

The "Summary of Testing" section indicates:

  • The Digital StereoLoc II System was tested by UL to IEC 60601-1 Medical Electrical Equipment Standards. This implies an assessment of electrical safety and essential performance, not clinical accuracy or diagnostic performance.
  • The DSM Workstation complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0. This refers to adherence to DICOM standards for image communication.
  • Hologic performs design control verification and validation tests under 21 CFR Part 820 (Quality System Regulation).

The core of the submission for demonstrating safety and effectiveness relies on:

  • Comparison With Predicate Devices: The Digital StereoLoc II is stated to be "substantially equivalent" to:
    • K950220 Hologic Lorad StereoLoc II
    • K040884 Hologic Lorad DSM
    • K040125 Senographe Stereo General Electric Medical Systems
      These predicate devices are also "Mammographic X-Ray Systems which aid in the performance of breast biopsies and pre-surgical localization."
  • Conclusion: "The Digital StereoLoc II principles of operation, clinical effectiveness, design, construction and materials are similar to existing marketed devices. This system poses no additional risk."

Therefore, the "proof" is based on demonstrating that the device is functionally and operationally similar to existing legally marketed devices, rather than a new clinical study with specific performance metrics against a defined ground truth. The FDA's letter confirms a "substantial equivalence" determination.

{0}------------------------------------------------

HOLOGIC

K07/542

Digital StereoLoc II

JUL - 5 2007

510(k) Summary of Safety Effectiveness

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87 (b)

Product Name: Digital StereoLoc II

Product Classification Name: Mammographic X-Ray System

Product Classification Code: 90 IZH CFR Section: 892.1710

Classification Panel: Radiology Class II

Manufacturer: Hologic, Inc. 36 Apple Ridge Road Danbury, CT 06810 USA

Contact Person:Gail Yaeker-Daunis
Telephone Number: (203) 731-8337
Fax Number: (203) 731-8440

Date Prepared: May 4, 2007

Predicate Devices: K952210 Hologic Lorad StereoLoc II K040884 Hologic Lorad DSM K040125 Senographe Stereo General Electric Medical Systems

Device Description:

The Digital StereoLoc II is an optional accessory to the Selenia Full Field Digital Mammography System. It is attached to the Selenia to perform stereotactic biopsies and is composed of the following elements:

  • The Biopsy Unit or Stage used for patient compression and vertical mounting of the . biopsy needle.
  • The SmartWindow control module for X and Y-axis motion control and Z-axis . positioning guidance.
  • DSM computer workstation and software for stereotactic targeting. .

The breast is compressed between the compression paddle and the carbon fiber breast platform of the biopsy unit. The biopsy unit has its own manual compression mechanism

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Hologic. The logo consists of the word "HOLOGIC" in large, bold, black letters. Below the word "HOLOGIC" is the phrase "CLARITY OF VISION" in smaller, black letters.

that is independent of the Selenia FFDM compression device. There are two compression knobs; one at each side of the biopsy unit. The Smart Window control module gets the Cartesian coordinates (X, Y, Z) of the target from the DSM workstation.

  • X and Y motion is motorized. Controlled by the operator via the SmartWindow. .
  • . Z-axis motion is manual and controlled by the operator.

Safety Features include:

  • Automatic lockout of Biopsy Unit movement with C-Arm when patient is under . compression.
  • . Automatic detection of biopsy unit and cables attached.
  • . Audible alert when switched from Stereo to C-Arm mode.
  • Audible alert when Selenia compression device may interfere with stereo imaging . movement.

How the Device Works:

Digital StereoLoc II for Selenia uses the Selenia detector for stereo image acquisition similar to the CCD camera used on the DSM system for the M-IV version of StereoLoc II (SLI). Stereotactic images are acquired at +/-15°. These stereo images are then automatically forwarded to the DSM Workstation for automatic calculation of the stereo coordinates and targeting of the lesion by the user (same as DSM operation with M-IV). The Compression Paddle and biopsy needle stage are the same as the StereoLoc II used with the Hologic Lorad M-IV, as are the X and Y-axis motion controllers. Both Compression and Z-axis movements are manually applied as with the M-IV. The stereo coordinates are then sent from the DSM Workstation to the Smart Window Control Module on the Digital StereoLoc LCD display (same as with StereoLoc II operation on M-IV). As with previous DSM Workstation versions, stereo images may be sent to PACS for archiving via DICOM protocol.

Indications for Use:

The Hologic Digital StereoLoc II is an optional accessory for the Selenia full field digital mammography system. It is designed to allow the accurate location of lesions in the breast in three dimensions, using information extracted from stereo tactic pairs of twodimensional images. It is intended to provide guidance for interventional purposes (such as biopsy, pre-surgical localization or treatment devices).

Comparison With Predicate Devices:

The Hologic Digital StereoLoc II is substantially equivalent to the following devices:

K950220 Hologic Lorad StereoLoc II K040884 Hologic Lorad DSM K040125 Senographe Stereo General Electric Medical Systems

These devices are Mammographic X-Ray Systems which aid in the performance of breast biopsies and pre-surgical localization.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Hologic. The logo is in black and the words "Clarity of Vision" are written below the logo. There is a registered trademark symbol to the right of the "C" in Hologic.

Summary of Testing

The Digital StereoLoc II System was tested by UL to IEC 60601-1 Medical Electrical Equipment Standards. The DSM Workstation complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0.

Hologic performs design control verification and validation tests under 21 CFR Part 820.

Conclusion

The Digital StereoLoc II principles of operation, clinical effectiveness, design, construction and materials are similar to existing marketed devices. This system poses no additional risk.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a circular seal with text around the perimeter and a symbol in the center. The text reads "U.S. Department of Health & Human Services - USA" in a circular arrangement. The symbol in the center consists of three stylized, curved lines that resemble a bird or a flowing design. The image is in black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

2007

Ms. Gail Yaeker-Daunis Senior Regulatory Specialist Hologic, Inc. 36 Apple Ridge Road DANBURY CT 06810

Re: K071542

Trade/Device Name: Digital StereoLoc II Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray system Regulatory Class: II Product Code: IZH Dated: June. 1, 2007 Received: June 5, 2007

Dear Ms. Yeaker-Daunis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for. use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/3/Picture/10 description: The image is a black and white circular logo. The logo has the text "1906-2006" at the top, "PA" in the center, and "Centennial" below that. The letters "USA" are at the bottom of the logo.

roloting and Promoting Publio Hoalth

{4}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) No. _

Indications For Use

Device Name: Digital StereoLoc II

The Hologic Digital StereoLoc II is an optional accessory for the Selenia full field digital mammography system. It is designed to allow the accurate location of lesions in the breast in three dimensions, using information extracted from stereotactic pairs of twodimensional images. It is intended to provide guidance for interventional purposes (such as biopsy, pre-surgical localization or treatment devices).

Concurrence of CDRH, Office of Device Evaluation (ODE)
--------------------------------------------------------
Prescription Use21 CFR 801.109OROver-the-Counter Use __________
-------------------------------------------------------------------------
(Division Sign-Off)
Division of Reproductive, Abdominal andRadiological Devices
510(k) NumberK071542

Section A
12

§ 892.1710 Mammographic x-ray system.

(a)
Identification. A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.