(116 days)
EMS-XL is indicated to acquire, filter, digitize, amplify, display and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals and pressure recordings. Physiological parameters such as the diastolic, systolic and mean blood pressure, heart rate and cycle length are derived from the signal data, displayed and recorded.
The system allows the user to monitor, display and record the signal data.
The system allows the user to monitor the acquisition data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. The system may display and record data received from other medical devices typically used during these procedures, such as Ablation RF generators. The system incorporates a stimulator intended to be used for diagnostic cardiac stimulation during electrophysiological testing of the human heart.
The EMS-XL Cardiac Electrophysiology System consists of a software driven, multichannel amplifier and stimulator that are connected to a computer.
Input signals include 12 lead surfaces ECG, two Blood pressure channels and 18 or 50 channel intra-cardiac ECG.
The signals are digitized and sent to the computer for analysis, display and storage.
Signals are displayed with sweep speeds of 25 to 300mm/sec in either predefined or customized configurations.
The system consists of a computer, Front end amplifiers with integrated stimulator.
Two display screens, one used for Real Time signal display and the other for Non real Time display for analysis and review and playback of stored signals.
A Laser printer is used to print surface and intra-cardiac ECG and a CD or DVD is used for archive of signals.
The two display screens are used to show the signals. The real time display, RT, shows the real time signal waveforms. Each waveform channel is color-coded for easy identification. The non real time display, NRT, serves as a Review monitor allowing the manipulation and processing of data, including caliper measurements, event marking, snapshot storage and final report processing.
The stimulator is computer controlled and provides basic stimulation rate plus up to 4 stimulations delayed after the basic stimuli. Stimulation can be either synchronized to the cardiac electrical activity or non synchronized. Pacing protocols with automatic increment and decrement functions provides refractory period measurements.
A variety of stimulation sequences can be created and stored for future use.
The EMS-XL can be used for all types of electrophysiological procedures, including His bundle recording, Sinus Node Recovery Time - SNRT, Overdrive, Wenckebach, and tachy-arrhythmias. An interface to ablation generators provide Ablation Start and End information with data on the ablation parameters.
Each waveform is color-coded for easy identification.
The review monitor allows the manipulation and processing of data, including caliper measurements, event marking, snapshot storage and final report processing.
Special display options support the signal analysis
This includes Trigger mode and template comparison. Event marking. Holter mode: quick. minute-by-minute scrolls and jumps to event, On screen calipers: both real-time and review screen with auto measurement , Auto Tachycardia detection. Free text labeling, and Customized reports in Word™.
The input to the amplifiers is via a patient connection box, to which the intra-cardiac catheter electrodes (not manufactured and not supplied by Mennen Medical) are connected.
This 510(k) summary for the Mennen Medical EMS-XL Cardiac Electrophysiology System focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific performance acceptance criteria through clinical trials. The provided document is a predicate device comparison, which is a common approach for 510(k) submissions.
Therefore, the acceptance criteria and study detailed below are interpreted from the provided predicate comparison tables and the accompanying "Rational for claims of equivalence".
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the specifications of the predicate devices. The reported device performance (EMS-XL) is presented in direct comparison to these predicate specifications. The table below synthesizes information from the various comparison tables provided in the document (Amplifier Spec comparison, Stimulator Spec Comparison, Comparison of Systems).
| Feature | Acceptance Criteria (Predicate) | Reported Device Performance (EMS-XL) | Deviations & Rationale (Effect on Safety/Intended Use/Performance) |
|---|---|---|---|
| Amplifier Specifications | Cardiolab EP System (GE Prucka) | EMS-XL Amplifier | |
| Physical (HxWxD, 32/64 ch) | 9.5x 14.0x 14.0 in (64/32 ch) | 4.5x 11.5x 8 in (32 ch); 9.0x 11.5x 8 in (64 ch) | No (Smaller physical size, no effect) |
| Temperature Operating | 0°C to +35°C | Same | Same |
| Temperature Transport/Storage | -15°C to +50°C | -30°C to +65°C | No (Wider range, no effect on safety/intended use) |
| Humidity Operating | <95% RH at 35°C non-condensing | 30% - 75% non-condensing | No (Within predicate range, no effect) |
| Humidity Transport/Storage | <95% RH at 35°C non-condensing | 5% - 95% non-condensing | No (Wider range, no effect) |
| Power Requirements | 100-240 V AC; 50-60 Hz | Same | Same |
| Sampling Rate | 1K, 2K and 4K | 1KHz | No (Meets clinical requirements, Note 2) |
| CMMR | 100 dB min | Same | Same |
| Input Impedance | >1 Billion Ohms | Typical 20 MΩ (Above 2.5 MOhm req by ANSI/AAMI EC11 par 3.2.9) | No (Meets standard, no effect) |
| Leakage Current (Patient Source) | <10 uA | Same - Per ISO 60601-1 | Same |
| Leakage Current (Patient Sink) | <10 uA | Same - Per ISO 60601-1 | Same |
| Leakage Current (Chassis) | <100 uA | Same - Per ISO 60601-1 | Same |
| ECG Outputs | 12 lead ECG produced | Same | Same |
| ECG High Pass Filter | 0.05 Hz, 0.5 Hz, 5 Hz | 0.05, 0.2, 40, 80 Hz | No (Provides necessary filtration, Note 3) |
| ECG Low Pass Filter | 100 Hz | Same | Same |
| Gain | 50-10,000 in 8 settings | Between 0 and 255 mm/mVolt – Continuous | No (Equivalent gain, different terms/control, Note 4) |
| Saturation Recovery | Less than 1 sec | Same | Same |
| Notch Filter | Power line (50/60 Hz) | Same | Same |
| Intracardiac Inputs (32ch) | 32 intracardiac, 4 pressure, 10 ECG | 36 intracardiac (18 ch), 2 pressure, 10 ECG (12 ch) | No (Covers wide range of clinical applications, Note 5) |
| Channel Switching | Any input can be switched to any output within a Block | Each channel can be Bipolar or Unipolar with Manual switching | No (Difference overcome by manual switching, Note 6) |
| IECG High Pass Filter | DC, 0.05 Hz, 0.5 Hz, 5.0 Hz, 30 Hz, 100 Hz | 0.05, 0.2, 40, 80 Hz | No (Provides necessary filtration, Note 7) |
| IECG Low Pass Filter | 500 Hz, 2,000 Hz | 500 Hz | No (500Hz same, 2000Hz not relevant for 1KHz sampling, Note 8) |
| Pressure Inputs | Compatible with all standard external pressure transducers. 5uV/V/mmHg | Same | Same |
| Excitation Voltage | 5V DC | 10 V DC (+5 to -5V) | No (Sensitivity is important parameter, which is same, Note 9) |
| Pressure Channels Outputs | Up to 4 pressure channels | 2 pressure channels | No (Covers wide range of clinical applications, Note 5) |
| Stimulator Specifications | Micropace model EPS 320 | EMS-XL (integrated) | |
| Isolated Stimulus channels | 2 | 2 | Same |
| Pulse Amplitude Range | 0.1 to 25 mA into 800 Ω load | 0.1 to 25 mA into 1500 Ω load | No (Same current range, higher load impedance source, Note 10) |
| Pulse Amplitude Increment | 0.1 mA up to 1 mA Amplitude | 0.1 mA within whole range | No |
| Pulse Amplitude Accuracy | +/- 2% or +/- 0.2 mA (whichever is larger) | Same | Same |
| Pulse Duration Range | 0.5 to 10 msec | 0.1 to 9.9 msec | No |
| Pulse Duration Increment | 0.5 - 10 msec, increment of 1-10 msec | 0.1 msec | No (Smaller increments, no effect on performance) |
| ISI Range (Pace) | 180 msec to 9990 msec | Same | Same |
| ISI Range (Burst) | 30 - 9900 msec | Same | Same |
| ISI Increment | 1 msec | 10 msec | No (Easier control, Note 11) |
| Sequential Delay (AV) Range | 10-1000+/- 1 msec (maximum ISI - 50 msec) | 11 - 250 msec | No (No clinical significance for longer intervals, Note 12) |
| Number of Extra-stimuli | 6 (S2-S7) | 4 (S2-S5) | No (4 extra sufficient for most clinical apps, Note 13) |
| Sensing (ECG sync) | External: 50 – 2000 mV; Internal: Pacing catheter tip | Internal from any surface or intra-cardiac channel | No |
| Sync Output | Multiple sync outputs for external devices | No | No (Integrated system doesn't need external sync, Note 14) |
| Power Source (Stimulator) | Mains 220/110 to 14.5 VDC, 750 mA low voltage power supply. Backup battery | Integrated with Amplifier power (+5V, -+12V, -12V) | Yes (Integrated, but safety covered by other precautions, Note 15) |
| Programmed Protocols (Auto TP/Ectopic/Stop Tachycardia) | Yes | No | Yes (Physician decision used instead due to safety burden, Note 16) |
| Backup Stimulator | Integrated | No backup stimulator (Requires external backup) | Yes (User manual requires external backup, Note 17) |
| System Specifications | Carto® XP System (GE Prucka) | EMS-XL | |
| Processor | Intel® 2.66GHz Pentium IV Xeon or greater | Intel® 2.4GHz Pentium IV or greater | No |
| RAM | 512 MB | Same | Same |
| Hard drives | 2 x 40 GB | 80 GB | No |
| Monitors | 1 or 2 | 2 | No |
| Holter Mode | Yes | Integrated Holter mode with full disclosure | No |
| Ablation Connectivity | Connectivity to RF Ablation generator(s) | Same | Same |
| Statistics and Inventory | Xi2 Data Management System | No | No (Administrative package, no direct use on EP procedure, Note 18) |
| Mapping | Isochronal and Isopotential Maps | No | No (External devices can be used for mapping, Note 19) |
| Vital Signs | Integrated Vitals Monitoring with Audible Indicators | Not integrated (Integrated tachycardia message; recommends external monitor, Note 20) | No |
2. Sample Size Used for the Test Set and Data Provenance
This document describes a predicate comparison for a 510(k) submission, not a direct clinical study with a specified "test set" as one might find for an AI/ML device.
- Sample Size: Not applicable in the context of a prospective or retrospective clinical study of user data. The "sample" here refers implicitly to the specifications and functionalities of the two predicate devices being compared.
- Data Provenance: The data provenance is the technical specifications and reported features of the two predicate devices:
- K993414 Cardiolab EP system (Prucka Engineering, Inc)
- K011826 Micropace model EPS 320 (Micropace Pty Ltd.)
The submission itself is from Mennen Medical Ltd., Israel. The clinical experience mentioned for EMS-XL notes use by clinics "out of the USA" since December 2003, implying international (likely European) usage.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable in this type of submission. The "ground truth" for this submission is the established and legally marketed performance and specifications of the predicate devices. The review and interpretation of these specifications for equivalence determination would be performed by the manufacturer's regulatory affairs team and then by FDA reviewers.
4. Adjudication Method for the Test Set
Not applicable. This is a technical specification comparison, not a clinical trial requiring adjudication of patient outcomes or expert reads. The "adjudication" is essentially the manufacturer's justification (Rational for claims of equivalence) and the FDA's acceptance of those justifications for "substantial equivalence."
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
No, an MRMC study was not done. This device is a cardiac electrophysiology system, which includes signal acquisition, display, recording, and stimulation capabilities. It is not an AI/ML diagnostic interpretation tool that would typically undergo an MRMC study to assess reader performance with and without AI assistance. The submission confirms that the EMS-XL system is a software-driven, multichannel amplifier and stimulator connected to a computer, focused on acquiring and analyzing physiological signals.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
Yes, implicitly. The comparison tables are a standalone comparison of the technical specifications and functions of the EMS-XL device against the technical specifications and functions of its predicate devices. The "performance" being evaluated is the device's inherent functional capabilities as designed and manufactured, not an AI algorithm's interpretive ability.
7. The Type of Ground Truth Used
The "ground truth" used is the published and legally established technical specifications and performance characteristics of the predicate devices (Cardiolab EP system and Micropace model EPS 320) as detailed in their 510(k) clearances and product documentation. This is a regulatory "truth" rather than a clinical "truth" (like pathology or outcomes data).
8. The Sample Size for the Training Set
Not applicable. This document does not describe the development of an AI/ML algorithm that would require a "training set." The EMS-XL is a hardware and software system for electrophysiological studies.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as no training set for an AI/ML algorithm is described.
{0}------------------------------------------------
K07/348
Image /page/0/Picture/2 description: The image shows the logo for "mennen medical". The logo consists of a black circle with a white "m" inside it on the left. To the right of the circle, the word "mennen" is written in black, and below it, the word "medical" is written in gray.
SEP - 7 2007
Date: July 30, 2007
Topic: 510K Summery EMS-XL Cardiac Electrophysiology system
Establishment Name, Registration Number and Address:
| Name: | Mennen Medical Ltd. |
|---|---|
| Registration Number: | 9611022 |
| Operator Number: | 9069173 |
| Address: | 4 Hayarden Street, Yavne, 81228, Israel |
| Postal Address: | PO Box 102,Rehovot, 76100, Israel |
| Tel: | +972-8-9323333 |
| Fax: | +972-8-9328510 |
| Contact person: | Micha Oestereich, Regulatory Affairs |
The following information is being submitted in conformance with [807.92(a)(2)].
| 1. Device Name: | EMS-XL - Cardiac Electrophysiology System |
|---|---|
| 2. Classification Name: | Programmable Diagnostic Computer |
| 3. Trade/Proprietary Name: | EMS-XL Cardiac Electrophysiology System |
| 4. Common/Usual Name | Cardiac Electrophysiology system |
| 5. Classification Number: | 870.1425 computer, diagnostic, programmable |
| 870.1750 External programmable pacemaker pulse generator | |
| 6. FDA Classification | Class II |
| 7. Product Code | DQK |
8. Predicated Devices:
Substantial Equivalence is claimed to the following devices:
- K993414 (4 July 2000) Cardiolab EP system. Prucka Engineering, Inc -
- K011826 (24 Jan 2002) -- Micropace model EPS 320. Micropace Pty Ltd. -
{1}------------------------------------------------
Device description
The EMS-XL Cardiac Electrophysiology System consists of a software driven, multichannel amplifier and stimulator that are connected to a computer.
Input signals include 12 lead surfaces ECG, two Blood pressure channels and 18 or 50 channel intra-cardiac ECG.
The signals are digitized and sent to the computer for analysis, display and storage.
Signals are displayed with sweep speeds of 25 to 300mm/sec in either predefined or customized configurations.
The system consists of a computer, Front end amplifiers with integrated stimulator.
Two display screens, one used for Real Time signal display and the other for Non real Time display for analysis and review and playback of stored signals.
A Laser printer is used to print surface and intra-cardiac ECG and a CD or DVD is used for archive of signals.
The two display screens are used to show the signals. The real time display, RT, shows the real time signal waveforms. Each waveform channel is color-coded for easy identification. The non real time display, NRT, serves as a Review monitor allowing the manipulation and processing of data, including caliper measurements, event marking, snapshot storage and final report processing.
The stimulator is computer controlled and provides basic stimulation rate plus up to 4 stimulations delayed after the basic stimuli. Stimulation can be either synchronized to the cardiac electrical activity or non synchronized. Pacing protocols with automatic increment and decrement functions provides refractory period measurements.
A variety of stimulation sequences can be created and stored for future use.
The EMS-XL can be used for all types of electrophysiological procedures, including His bundle recording, Sinus Node Recovery Time - SNRT, Overdrive, Wenckebach, and tachy-arrhythmias. An interface to ablation generators provide Ablation Start and End information with data on the ablation parameters.
Each waveform is color-coded for easy identification.
The review monitor allows the manipulation and processing of data, including caliper measurements, event marking, snapshot storage and final report processing.
Special display options support the signal analysis
This includes Trigger mode and template comparison. Event marking. Holter mode: quick. minute-by-minute scrolls and jumps to event, On screen calipers: both real-time and review screen with auto measurement , Auto Tachycardia detection. Free text labeling, and Customized reports in Word™.
The input to the amplifiers is via a patient connection box, to which the intra-cardiac catheter electrodes (not manufactured and not supplied by Mennen Medical) are connected.
{2}------------------------------------------------
Indication for use:
EMS-XL is indicated to acquire, filter, digitize, amplify, display and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals and pressure recordings. Physiological parameters such as the diastolic, systolic and mean blood pressure, heart rate and cycle length are derived from the signal data, displayed and recorded.
The system allows the user to monitor, display and record the signal data.
The system allows the user to monitor the acquisition data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. The system may display and record data received from other medical devices typically used during these procedures, such as Ablation RF generators. The system incorporates a stimulator intended to be used for diagnostic cardiac stimulation during electrophysiological testing of the human heart.
The intended use
The EMS-XL is a computerized Electrophysiological Measurement System, designed for conducting regular or experimental electrophysiology (EP) studies.
The EMS-XL System is intended to be used for performing computerized Electrophysiological and Ablation procedures.
The EMS-XL is intended for sale as a system for performing EP clinical studies Prescription Use under the supervision of a physician
Control and Integration
The EMS-XL integrated stimulator allows complete control of all stimulator functions at all times, featuring programmable pacing protocols with automatic increment and decrement functions, comprising all basic modes. Furthermore, the system includes online Holter and Event windows that allow for simple and immediate tracking of ectopic events. Fully editable final reports are generated in a Microsoft Word™ format, easily integrating measured intervals and procedure notes, to improve efficiency and productivity while saving valuable staff time.
Additional Features
- · Trigger mode and template comparison
- · Customized reports in MS WordTM
- · Auto Tachycardia detection: Jump to event
- · SNRT Protocol
- · Automatic snapshot after pacing
- · Event marking
- · Holter mode: Quick, minute-by-minute scrolls and jumps to event
- · On screen calipers: Both real-time and review screens with auto measurement
{3}------------------------------------------------
- · Export snapshots as image file for a PowerPoint presentation
- · Multiple interfaces to Ablation RF generators for display of ablation parameters and ablation stop/start events
- · Total case full disclosure archiving on hard drive
- · Easily upgradeable software
Product Highlights
- · Built-in integrated stimulator:
- · Total full disclosure: Guaranties data integrity and long term archiving
- · Off-line review station: Review, analysis and printing of entire case from remote location (optional)
Product Comparison to predicated devices
The EMX-XL system combined signal amplification, display and recording features with stimulation capabilities.
We thus have two predicated devices:
- Cardiolab EP system. Prucka Engineering, Inc K993414 (4 July 2000) | for the signal amplification, display and recording
- Micropace model EPS 320. Micropace Pty Ltd. K011826 (24 Jan 2002) --for the stimulation specification and features.
The following are the comparison tables:
- Amplifier Spec comparison .
- Comparison of Systems .
- Stimulator Spec Comparison .
{4}------------------------------------------------
| Manufacturer | Predicated - GE Pruka | Mennen Medical | Effectonsafety | Effect onIntended use |
|---|---|---|---|---|
| System | CardioLab II PlusAmplifier | EMS-XL amplifier | ||
| Item | Description | Description | ||
| Physical Specifications | ||||
| 64/32 channel HxWxD | 9.5x 14.0x 14.0 in | 32 = 4.5x 11.5x 8 in x64 = 9.0x 11.5x 8 in x | No | No |
| 128/96 channel HxWxD | 14.0 x 14.0 x 14.0 in | Not used | N/A | N/ANote 1 |
| Environmental Specifications | ||||
| Temperature Operating | 0°C to +35°C | Same | ||
| Transport/Storage: | -15°C to + 50°C | -30°C to + 65°C | No | No |
| Humidity Operating: | < 95% relative at 35°C non-condensing | 30% - 75% noncondensing | No | No |
| HumidityTransport/Storage: | < 95% relative at 35°C non-condensing | 5% - 95% noncondensing | No | No |
| rower Specifications: | ||||
| Power Requirements | 100-240 V AC; 50-60 Hz | Same | Same | |
| Power Input (64/32channel) | 0.0-0.5 Amps Class I, TypeCF Continuous operation | + 5 V , 0.0 - 0.2 A+ 12 V, 0.0 - 0.3 A- 12 V , 0.0 - 0.3 A | No | No |
| Power Input (128/96channel) | 0.5-0.75 Amps Class I,Type CF Continuousoperation | Not used | N/A | N/ANote 1 |
| Design | ||||
| Sampling and Hold | Each channel sampled priorto acquisition. | Same | Same | |
| Sampling Rate | 1K, 2K and 4K | 1KHz | No | NoNote 2 |
| CMMR | 100 dB min | Same | Same | |
| Typical 20 MΩ | No | No | ||
| Input Impedance | > 1 Billion Ohms | Above the 2.5 MOhmrequired by theANSI/AAMI EC11 par3.2.9 | Per standard | |
| akage Current | ||||
| Patient Source: | < 10 uA | Same - Per ISO 60601- | Same | |
| ...anufacturer | Predicated – GE Pruka | Mennen Medical | Effect on safety | Effect on Intended use |
| System | CardioLab II PlusAmplifier | EMS-XL amplifier | ||
| Item | Description | Description | ||
| 1 | ||||
| Patient Sink: | < 10 uA | Same – Per ISO 60601-1 | Same | |
| Patient Sink (measured atpatient leads under singlefault conditions): | < 50 uA | Same – Per ISO 60601-1 | Same | |
| Chassis Leakage: | < 100 uA | Same – Per ISO 60601-1 | Same | |
| ECG Input | ||||
| Outputs | 12 lead ECG produced | Same | Same | |
| High Pass Filter | 0.05 Hz, 0.5 Hz, 5 Hz | 0.05, 0.2, 40, 80 Hz | No | NoNote 3 |
| Low Pass Filter | 100 Hz | Same | Same | |
| RF Filtering | All inputs | Same | Same | |
| ain | 50-10,000 in 8 settings | Between 0 and 255 mm/mVolt – Continuous | No | Note 4 |
| Saturation Recovery | Less than 1 sec. | Same | Same | |
| Notch Filter | Power line (50/60 Hz) | Same | Same | |
| No data | +/- 5 mVoltAAMI EC11 par 3.2.3 | Per standard | ||
| Dynamic range | ||||
| No data | +/- 300 mVoltAAMI EC11 par 3.2.3 | Per standard | ||
| Baseline correction | ||||
| Input / Output | ||||
| Inputs 32 channels | 32 intracardiac inputs, 4pressure inputs, 10 ECGinputs | 36 intracardiac inputs(18 channels), 2pressure inputs, 10 ECGinputs (12 channels) | No | NoNote 5 |
| Inputs 64 channels | 96 intracardiac inputs, 4pressure inputs, 10 ECGinputs | 100 intracardiac inputs,(50 channels), 2pressure inputs, 10 ECGinputs (12 channels) | No | NoNote 5 |
| Inputs 96 channels | 160 intracardiac inputs, 4pressure inputs 10 ECGinputs | Not used | No | NoNote 1 |
| Inputs 128 channels | 224 intracardiac inputs, 4 | Not used | No | No |
| ...anufacturer | Predicated - GE Pruka | Mennen Medical | Effectonsafety | Effect onIntended use |
| System | CardioLab II PlusAmplifier | EMS-XL amplifier | ||
| Item | Descriptionpressure inputs, 10 ECGinputs | Description | Note 1 | |
| Outputs | 16 channels/Block. Up to112 channels. 32 channel:Block A 64 channel: BlocksA-C 96 channel: Blocks A-E128 channel: Blocks A-G | No outputs | No | NoNote 1 |
| Switching | Any input can be switchedto any output within a Block. | Each channel can beeither Bipolar or Unipolarwith Manual switching | No | NoNote 6 |
| High Pass Filter | DC, 0.05 Hz, 0.5 Hz, 5.0 Hz,30 Hz, 100 Hz | 0.05, 0.2, 40, 80 Hz | No | NoNote 7 |
| Low Pass Filter | 500 Hz, 2,000 Hz | 500 Hz | No | NoNote 8 |
| RF Filtering | All inputs | Same | Same | |
| Gain | 50-10,000 in 8 settings | Between 0 and 255mm/mVolt - Continuous | No | NoNote 4 |
| Saturation Recovery | Less than 1 sec. | Same | Same | |
| Notch Filter | Power line (50/60 Hz) | Same | Same | |
| Dynamic range | No data | +/- 5 mVoltAAMI EC11 par 3.2.3 | Per standard | |
| Baseline correction | No data | +/- 300 m VoltAAMI EC11 par 3.2.3 | Per standard | |
| Pressure Inputs | ||||
| Inputs | Compatible with all standardexternal pressuretransducers. 5uV/V/mmHg | Same | Same | |
| Input Impedance | > 1 Billion Ohms | Typical 20 MΩAbove the 2.5 MOhmrequired by the ANSI/AAMIEC11 par 3.2.9 | No | NoPer standard |
| Outputs | Up to 4 pressure channels | 2 pressure channels | No | NoNote 4 |
| Low Pass Filter | 6 Hz, 25 Hz, 100 Hz, 400Hz | DC to 12 Hz(DC to 40 Hz optional).IEC 60601-2-34 Standardpar. 51.103 requests atleast DC to 10 Hz. | No | NoPer standard |
| Excitation Voltage | 5V DC | 10 V DC (+5 to -5V). | No | Note 9 |
| ...anufacturer | Predicated – GE Pruka | Mennen Medical | Effectonsafety | Effect onIntended use |
| System | CardioLab II PlusAmplifier | EMS-XL amplifier | ||
| Item | Description | Description | ||
| Output Impedance | < 0.050 Ohms | Same | Same | |
| Range | -400mm Hg to +400 mmHg | Same | Same |
Amplifier Technical Specifications comparison
Page 5 of 21
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Stimulator Specification comparison
| Manufacturer | Micropace Pty. Ltd.Austria | Mennen Medical Ltd.Israel | Effectonsafety | Effect onPerformance |
|---|---|---|---|---|
| Predicated device | ||||
| Subject/Parameter | EPS320 | (integrated in) EMS-XL | ||
| JA Concurrence | Yes K011826 | Pending | ||
| Isolated Stimuluschannels | 2 | 2 | ||
| Pulse Amplitude | ||||
| Range | 0.1 to 25 mA into 800 Ωload | 0.1 to 25 mA into 1500Ω load | No | Note 10 |
| Increment | 0.1 mA up to 1 mAAmplitude | 0.1 mA within wholerange | No | No |
| Accuracy | +/- 2% or +/- 0.2 mA(whichever is larger) | Same | Same | |
| Pulse duration | ||||
| Range | 0.5 to 10 msec | 0.1 to 9.9 msec | No | No |
| Increment | 0.5 - 10 msec, incrementof 1-10 msec | 0.1 msec | No | No, smallerincrements |
| Accuracy | +/- 0.15 msec | Same | same | |
| Inter-stimulus Interval (ISI) | ||||
| Range | 180 msec to 9990 msec +/-1msec or 0.1% (whicheveris greater) | Same | Same | |
| Manufacturer | Micropace Pty. Ltd.AustriaPredicated device | Mennen Medical Ltd.Israel | Effectonsafety | Effect onPerformance |
| Subject/Parameter | EPS320 | (integrated in) EMS-XL | ||
| Range (Burst) | 30 - 9900 msec +/- 1msecor 0.1% (whichever isgreater) | Same | Same | |
| Increment | 1 msec | 10 msec | No | Note 11 |
| Sequential Delay (AV) | ||||
| Range | 10-1000+/- 1 msec(maximum ISI - 50 msec) | 11 - 250 msec | No | Note 12 |
| Increment | 1 msec | same | ||
| ProgrammedProtocols | Threshold | Threshold panel key | No | No |
| SNRT | SNRT panel key | |||
| Burst | Vent. Burst key | |||
| Overdrive | Atrial Burst key | |||
| Vent. Overdrive key | ||||
| Atrial Overdrive key | ||||
| Multi- Sx | Same | |||
| ProgrammedProtocols | Pace | Same | No | No |
| Wenkebach | User defined protocol | |||
| Nodal ERP / RSync_S2 | User defined protocol | |||
| ProgrammableProtocol Key | 5 | Same | ||
| Number of Extra-stimuli | 6(S2-S7) | 4(S2-S5) | No | Note 13 |
| Sensing (ECG synchronization) | ||||
| Automatic or Manualtrigger setting -sensitivity | External: 50 – 2000 mVInternal: Pacing cathetertip | Internal from anysurface or intra-cardiacchannel | No | No |
| Automatic or Manualtrigger setting -Trigger lockup(refractory time) | 5 - 5000 msec | Same | Same | |
| Automatic or Manualgger setting - ECGdelay | 5 - 5000 msec | Same | Same | |
| Manufacturer | Micropace Pty. Ltd.AustriaPredicated device | Mennen Medical Ltd.Israel | Effectonsafety | Effect onPerformance |
| Subject/Parameter | EPS320 | (integrated in) EMS-XL | ||
| Additional Outputs | Stimulus channel A:Sync 3Input markerStimulus channel A:Sync 3Input markerProgrammable Auxiliary /Paper advance: Sync 1Input markerECG trigger marker: Sync1, Sync 2 . | No | No | Note 14 |
| Power Source | Mains 220/110 to 14.5VDC, 750 mA low voltagepower supply.Backup Power: battery | Integrated withAmplifier+ 5 V , 0.0-0.2 A+ 12 V, 0.0-0.3 A- 12 V , 0.0 - 0.3 A | Yes | Note 15 |
| Simulation module | N/A | |||
| Dimension | 3.0x12.9x14.0 inch | N/A | ||
| Weight | 18.7 lbs. | N/A | ||
| Control terminal | ||||
| IsolationTransformer | Yes | Note 15 | ||
| Environment | Stimulator integratedwith amplifier | |||
| OperatingTemperature | +10 to +40 Celsius | 0°C to +35°C | No | No, Note 15 |
| Storage Temperature | -20 to +60 Celsius | -30°C to + 65°C | No | No, Note 15 |
| Operating Relativehumidity | 20% to 90% non-condensing | 30% - 75% noncondensing | No | No, Note 15 |
| Storage Relativehumidity | 20% to 90% non-condensing | 5% - 95% non condensing | No | No, Note 15 |
| Operating altitude | 0 to 4572 meter | - 400 to 3050 meter | No | No |
| Storage Altitude | 0 to 7620 meter | - 400 to 5000 meter | No | No |
| Controller | Embedded micro-controller | Integrated into theAmplifier and computer | No | No, Note 15 |
| *Manufacturer | Micropace Pty. Ltd.AustriaPredicated device | Mennen Medical Ltd.Israel | Effectonsafety | Effect onPerformance |
| Subject/Parameter | EPS320 | (integrated in) EMS-XLcontrol | ||
| User Interface | PC windows-stylegraphical display | Same on the EMS-XLdisplay | No | Note 15 |
| Power supply | Class II | See EMS-XL computer | No | Note 15 |
| Isolation transformer | Toroidal 1 : 1 medicalgrade, 110-240 VAC, 100VA max | See EMS-XL computer | No | Note 15 |
| 0/S Software: | Custom RTOS / DatalightROM-DOS | See EMS-XL computer | No | Note 15 |
| Backup power | 12V 2.1 Ah sealed leadacid | See Note A. | No | Note A |
| Emergency power: | 9V LiMn02i PP3, 10y life | See Note A. | No | Note A |
| Operating time | Two (2) hours on backupbattery | See Note A. | No | Note A |
| Pacing Channels | ||||
| Isolated Channels(3) | (i) Atrial and(ii) Ventricular via greenRedel 4 pin socket(iii) Emergency FixedPace Output to Ventricle,via red Redel 4 pin socket | Same function withdifferent connectors | No | Same |
| Power Source: | Internal DC-DC converters | See Note A. | No | No, Note A |
| Circuit Isolation: | Compliant with IEC601-1,Class CF, 5kV, common &differential mode | Compliant withIEC60601-1, Class CF,5kV, common &differential mode | No | No |
| Computer Controlled Stimulus Pulses | ||||
| Current: | 0.1 to 25 mA into 800 Ωload | 0.1 - 25.5 mA, into1500 Ω load 40V | No | Note 10 |
| Current Steps: | 0.1 mA | 0.1 mA | Same | |
| "Manufacturer | Micropace Pty. Ltd.AustriaPredicated device | Mennen Medical Ltd.Israel(integrated in) EMS-XL | Effect onsafety | Effect onPerformance |
| Subject/Parameter | EPS320 | |||
| Accuracy: | +/- 2% or +/- 0.2 mA, which ever is greater | +/- 0.1 mA | ||
| Pulse Duration: | 0.5 ms, 1-10 ms in steps of 0.1ms | Pulse width 0.1 – 9.9 mSec steps of 0.1 mS | No | No |
| Accuracy: | +/- 0.15 ms | +/- 0.1 mA | Same | |
| Load Impedance: | 200-1000 Ω, $<$ 700 Ω for max current | 1500 Ω | No | Note 10 |
| Max Output Voltage | 27V | 40 V | No | Note 10 |
| Inter-stimulus Intervals | ||||
| S1 Range: | 180 - 9990 ms (Pace)30 - 9990 ms (Burst Pace) | Same | Same | |
| ability: | Quartz computer clock,+/- 30 parts per million @ 25° C | Same | ||
| Extra-Stimuli: | 6 max, S2-S7, independent | 4 max, S2-S5, independent | No | Note 13 |
| Coupling interval: | 30 - 9990 ms | Same | Same | |
| Accuracy | +/- 1 ms or 0.1 % whichever is greater | +/- 10 ms | No | Note 11 |
| Protocol Automation | ||||
| Auto decrement / increment: | Yes | Same | Same | |
| SNRT S1 intervals and RT calculation | Same | Same | ||
| Auto pace and sense - | Same | Same | ||
| His-coincident extra-stimulus timing calculation | Caliper with labels | No | Same function | |
| ATP S1 calculation from % of TCL | No | Yes | Note 16 | |
| Trigger output on sensed ectopic beats | No | Yes | Note 16 | |
| Manufacturer | Micropace Pty. Ltd.AustriaPredicated device | Mennen Medical Ltd.Israel | Effect onsafety | Effect onPerformance |
| Subject/Parameter | EPS320 | (integrated in) EMS-XL | ||
| Stop On Tachycardia | No | Yes | Note 16 | |
| All automation subject toinstant operatoradjustment | Same | No | Same | |
| Backup ManuallyControlledStimulation | No backup stimulatorUse external backupstimulator | Yes | Note 17 | |
| Emergency BackupPacing | No backup stimulator.Use external backupstimulator | Yes | Note 17 |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
:
{11}------------------------------------------------
{12}------------------------------------------------
Comparison of Systems
.
| Manufacturer | GE - PruckaPredicateddevice | Mennen MedicalLtd. | Effectonsafety | Effect onperformance |
|---|---|---|---|---|
| Subject/Parameter | Carto® XPSystem Bi-directionalInterface | EMS-XL | ||
| Processor/Data Storage | ||||
| Processor | Intel® 2.66GHzPentium IV Xeon orgreater | • Intel® 2.4GHzPentium IV or greater | No | No |
| RAM | 512 MB | Same | Same | |
| Hard drives | 2 x 40 GB | 80 GB | No | No |
| Diskette Drive | No | 1.44 MB | No | No |
| DVD-RAM/CD-RV drive | • 9.4 GB DVD-RAM/CD-RV drive | DVD-RW / CD R | No | No |
| Magnetic Optical Drive | • 2.6 GB MagneticOptical Drive | Optional | No | No |
| Mouse | • Optical ScrollMouse | same | Same | |
| Operating System | • OS: Windows XP | Same | Same | |
| Manufacturer | GE - PruckaPredicateddevice | Mennen MedicalLtd. | Effectonsafety | Effect onperformance |
| Subject/Parameter | Carto® XPSystem Bi-directionalInterfaceProfessional | EMS-XL | ||
| Office | • Microsoft Office XP Professional | Same | Same | |
| Modem | Modem | |||
| • 56K B. 90 BaudData | Same | Same | ||
| Networking | ||||
| Ethernet | • 100 Base-TEthernet, TCP/IP | Same | Same | |
| Flat panel display | • 20" flat panel ultrahigh-resolutioncolor | Same | Same | |
| Monitors | ||||
| solution | • 1600 x 1200 | Same | Same | |
| No of monitors | 1 or 2 | 2 | No | No |
| Printer Options | ||||
| Black/White | • HP Black/WhiteLaserJet 2300 | HP LaserJet 4100 orgreater | No | No |
| • Color LaserJet | Yes | No | No | No |
| Compliance with Standards | ||||
| standards | • UL 2601-1, IEC60601-1-2,European UnionMedical | Same | Same | |
| Device Directive (CEMarked) | Device Directive(CE Marked) | Same | Same | |
| Environmental/Electrical Specifications | ||||
| • Operating Temperature: | +10°C to +30°C | 0°C to 35°C | No | No |
| • Storage Temperature: - | -10°C to +45°C | -20°C to 65°C | ||
| • Humidity: operating | 30%-75% (non-condensing) | Same | Same | |
| Humidity: storage | 10%-95%, (non-condensing) | 5%-95%, (non-condensing) | No | No |
| Manufacturer | GE - PruckaPredicated device | Mennen MedicalLtd. | Effect onsafety | Effect onperformance |
| Subject/Parameter | Carto® XPSystem Bi-directionalInterface | EMS-XL | ||
| • Maximum Current Draw: | 15A/115V,7A/230V | Same | Same | |
| • Chassis Leakage Current: | <100μA | Same | Same | |
| Advance Features | ||||
| IT Interface | Yes | Yes | Same | |
| Holter | Yes | Integrated Holtermode with fulldisclosure,. | No | No |
| Ablation | Connectivity to RFAblationgenerator(s) | Same | Same | |
| Statistics and Inventory | Xi2 DataManagementSystem. | No | No | Note 18 |
| Documentation | Configurableelectronicdocumentation | Configurable Worddocumentation | No | No - Same |
| Networking | Remote waveformreview and Nurses'Workstation, forsimultaneous dataentry withwaveform archive | Remote Reviewstation | No | No - Same |
| Security | Standard securityfeatures | Same | ||
| Access to raw data | Easy access to rawwaveform data(data extraction),binary &ASCII formats | Yes , in FullDisclosure | No | No, Same |
| Manufacturer | GE - PruckaPredicateddevice | Mennen MedicalLtd. | Effectonsafety | Effect onperformance |
| Subject/Parameter | Carto® XPSystem Bi-directionalInterface | EMS-XL | ||
| Mapping | Isochronal andIsopotential Maps | No | No | Note 19 |
| Vital Signs | Integrated VitalsMonitoring withAudible Indicators | Not integrated | No | Note 20 |
{13}------------------------------------------------
:
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
| Certification | ||
|---|---|---|
| MDD Device Class: | Class IIb | Class IIb |
| IEC60601-1 | certified | certified |
| EMC compliance | certified | certified |
| CE Marking | certified | Certified CE 0473 |
| US Regulations | 510(k) Accepted | Applied |
Note 1: The stimulator is integrated with the amplifier and controlled by the EMS-XL software – Thus it does not have a separated power supply and battery backup.
| Dimensions | ||||||
|---|---|---|---|---|---|---|
| System Dimension | Widthcm/inch | Depthcm/inch | Heightcm/inch | Widthcm/ inch | Depthcm/inch | Heightcm/inch |
| Workstation (65") | Cart with display | |||||
| 165/65 | 76/30 | 172/77 | 61/24 | 61/24 | 162/64 | |
| Workstation (45") | Consol with display | |||||
| 114/47 | 76/30 | 75/30 | 90/35 | 85/33 | 125/50 | |
| Workstation (Box) | ||||||
| 61/24 | 71/28 | 75/30 | ||||
| Amplifier Dimension | 32/64 Channel | 32 Channel | ||||
| 36/14 | 36/14 | 24/10 | 29/11 | 22/9 | 11/5 | |
| 64 channel | ||||||
| 29/11 | 22/9 | 22/9 |
{17}------------------------------------------------
| Predicated - GE Pruka | Mennen Medical -EMS-XL | |
|---|---|---|
| Certification | ||
| MDD Device Class: | Class IIb | Class IIb |
| IEC60601-1 | certified | certified |
| EMC compliance | certified | certified |
| CE Marking | certified | Certified CE 0473 |
| US Regulations | 510(k) Accepted | Applied |
Note A : The stimulator is integrated with the amplifier and controlled by the EMS-XL software - Thus it does not have a separated power supply and battery backup.
{18}------------------------------------------------
Rational for claims of equivalence of Mennen Medical EMS-XL amplifier and GE Prucka Plus Amplifier
| Subject | Reducedsafety | Is EMS-XL as safe and effective as the Plus Amplifier(Rational) |
|---|---|---|
| Note 1-No. ofchannels | No | Assumed equivalence to the 32/64 channel Plus Amplifier and not to the128 channel amplifier |
| Note 2-Samplingrate | No | Assumed equivalence to the 1 KHz sampling rate. This sampling ratemeets the requirements for clinical practice and sufficiently showsconduction pathway potentials clearly. |
| Note 3 -ECG HighPass filter | No | Both systems provide 0.05 Hz diagnostic filter that show high fidelityECG. 0.5 Hz (Plus Amplifier) and 0.2 Hz (EMS-XL) provides a fasterresponse to electrode movement, with same QRS detection capability.The EMS-XL 40 and 80 Hz filters yields the best deflection of HRA, HISbundle and ventricle electrocardiogram- |
| Note 4 -ECG gain | No | The gain of the two systems is equivalent and only uses different terms.The term mm/mVolt used by EMS-XL shows the relation between thesignal amplitude in mVolt and the display amplitude in mm.EMS-XL has continuous gain control that covers the whole gain rangeinstead of the 8 steps of the Plus Amplifier |
| Note 5 -Number ofchannels | No | Assumed equivalence to the 32/64 channel Plus Amplifier and to 2 of the 4pressure channels. 32/64 channel system is clamed to cover a wide range ofclinical application. |
| Note 6 -Channelswitching | No | The difference in this switching capability is practically overcome bymanual switching of electrodes between inputs that using the Bipolar input |
| Note 7 -IECG highpass filter | No | The set of filters provided by EMS-XL provide the necessary filtration fordetection of fast changing intra-cardiac signals and cover the same range offilters of the Plus Amplifier and the changes between 30 and 40, andbetween 80 and 100 are not significant. |
| Note 8IECG lowpass filter | No | EMS-XL has same 500 Hz filter as the Plus Amplifier.As stated in Note 2, Mennen claims equivalence to the 1 KHz samplingrate. This sampling rate meets the requirements for clinical practice andshows intra-cardiac potential clearly. The low pass filter of 2000Hz is notrelevant for sampling rate of 1 KHz |
| Note 9Excitationvoltage | No | The important parameter for transducers is the sensitivity, that is5μV/V/mmHg (industry standard), and this is same for both amplifiers |
| Subject | Reducedsafety | Is EMS-XL as safe and effective as the EPS320(Rational) |
| Note 10Stimulation current | No | The EMS-XL stimulator and the predicated EPS320 stimulator are both constant current stimulators and cover the same current range with the EMS-XL source is up to 1500 Ω, and EPS320 only 800 ΩThis effects the maximal stimulation voltage. |
| Note 11-Stimulationincrement | No | The EMS-XL stimulator has increments of 10 mSec that provide easier control than 1 mSec increment of the EPS320. |
| Note 12-AV delay | No | EPS320 =10-1000+/- 1 msec (maximum ISI - 50 msec)EMS-XL =11 - 250 msec - No clinical significance in usin longer intervals |
| Note 13No. ofstimuli | No | EMS-XL has S1 and 4 extra stimuli, EPS320 has 6 extra.In almost all clinical applications 4 extra are sufficient to achieve the intended use and diagnostic requirement (none of the EPS320 protocols in Par 10.5 use more than 4) |
| Note 14SyncOutput | No | The EPS320 uses external sync, to synchronize between the amplifier and the stimulator.The EMS-XL has amplifier and stimulator integrated within the same system and thus external sync is not necessary. |
| Note 15Stimulatorpower | No | The EMS-XL has integrated amplifiers, stimulator and control software, thus the requirement for separate power supply is irrelevant.The stimulator safety issues are covered by the following precautions:• The system includes an Isolation transformer (see User manual page IX)• Warning in User manual page X requires a defibrillator and external pacer in the EP procedure room• The EMS-XL amplifier and stimulator unit provides fixed rate stimulation in case of a fault of the cable connection of the unit to the computer. (page 2-1 of the user manual) |
| Note 16 | No | Mennen Medical finds that the EPS320 use of the automation for ATP -- Anti-tachycardia protocol, Trigger on ectopic beat and Stop on tachycardia may create a safety burden and thus decided to use the physician decision for these function as provided also on the EPS320 |
| Note 17 | No | Warning in EMS-XL User Manual page X requires a backup stimulator to be available in the EP procedure room |
| Note 18 | No | The EMS-XL does not claim to contain an integrated Statistic and Inventory software package as does the GE.Prucka system. Such |
| administrative package has no direct use on the EP procedure. | ||
| Note 19 | The EMS-XL does not claim to have mapping capabilities. DuringEP procedures external devices can be used for mapping. | |
| Note 20 | The Prucka provides Vital signs audible indicator. The EMS-XLprovides integrated tachycardia message.For other vital signs the User manual recommends to have anexternal patient monitor. |
. .
{19}------------------------------------------------
Rational for claims of equivalence of Mennen Medical EMS-XL stimulator and Micropace Pty EPS320
·
. .
{20}------------------------------------------------
Clinical experience
· ।
The EMS-XL system has CE mark since December 2003 and is used by clinics out of the USA
Conclusion:
On the basis of the specification comparison we reach the conclusion that the EMS-XL is as safe, as effective, and performs as well as the predicate device
{21}------------------------------------------------
Image /page/21/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around a symbol. The symbol consists of a stylized caduceus-like design, with three abstract human profiles facing right, overlaid with three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 7 2007
Mennen Medical Ltd. c/o Mr. Micha Oestereich QA & RA Manager 4 Ha-Yarden St. POB 102 Rehovot Yavne 76100 ISRAEL
Re: K071348
Trade Name: EMS-XL Cardiac Electrophysiology System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: August 29, 2007 Received: September 5, 2007
Dear Mr. Oestereich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{22}------------------------------------------------
Page 2 - Mr. Micha Oestereich
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Bhimmarfor
Bram B. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{23}------------------------------------------------
Indications for Use
510(k) Number (if known): K071348
Device Name:_EMS-XL Cardiac Electrophysiology System
Indications For Use:
EMS-XL is indicated to acquire, filter, digitize, amplify, display and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals and pressure recordings. Physiological parameters such as the diastolic, systolic and mean blood pressure, heart rate and cycle length are derived from the signal data, displayed and recorded.
The system allows the user to monitor, display and record the signal data.
The system allows the user to monitor the acquisition data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. The system may display and record data received from other medical devices typically used during these procedures, such as Ablation RF generators. The system incorporates a stimulator intended to be used for diagnostic cardiac stimulation during electrophysiological testing of the human heart.
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B.Bumman
16n Sian-Off. scular Devices
Page 1 of 1
§ 870.1425 Programmable diagnostic computer.
(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).