(202 days)
The Signalife Fidelity 200 Cardiac Event Recorder is a battery operated, user-activated recorder and transmitter intended to record an ECG and transmit their ECG data to medical professionals via a communication device to a cardiac monitoring service. This allows users to record and transmit their ECG data to medical professionals who are concerned about their heart rhythm and have experienced the following symptoms that are suggestive of abnormal cardiac rhythms: Pounding heart (Palpitations), Heart racing or irregular pulse, Lightheadedness or Faintness, History of Arrhythmia.
The device acquires an ECG signal and produces a single lead ECG. Upon user activation, an ECG signal is acquired and digitized producing a 45-second ECG recording utilizing the company's signal processing technology. The Signalife Fidelity 200 Cardiac Event Recorder converts the signal stream into a specific data format where it is stored on a memory chip for transtelephonic transmission to and analysis by another device such as the PaceArt System K024278 manufactured by Medtronic, Inc. Recorded data is not real time but stored for further analysis. The Signalife Fidelity 200 Cardiac Event Recorder is not intended to evaluate the signal or alarm. The fundamental technology of the Signalife Fidelity 200 Cardiac Event Recorder is the same as that of the predicate devices. The Signalife Fidelity 200 Cardiac Event Recorder employs the company's signal processing technology to minimize noise in the ECG signal created by the ambulatory nature of the signal source.
The provided document K071228 is a 510(k) summary for the Signalife Fidelity 200 Cardiac Event Recorder. It describes the device, its intended use, and claims substantial equivalence to predicate devices. However, this document does not contain information about specific acceptance criteria or a study proving the device meets particular performance metrics beyond general claims of substantial equivalence.
The document primarily focuses on:
- Device Description: A battery-operated, single-lead ECG event recorder that acquires and digitizes a 45-second ECG event for transtelephonic transmission.
- Intended Use: For individuals desiring access to a cardiac monitoring service, capable of recording and transmitting ECG data to medical professionals for analysis and diagnosis.
- Predicate Devices: HeartCard® Telephone Electrocardiograph Transmitter and Receiver (K010945), Card Guard CG-2206 1-Lead ECG Event Recorder (K963725), and CG-2211 Self Check ECG Transmitter (K012223).
- Basis for Substantial Equivalence: Claims equivalence in intended use, target population, technical characteristics, performance, and compliance to consensus standards for recording and transtelephonic transmission of ECG data.
Since the provided text does not include a performance study with acceptance criteria, the following points will indicate that the information is "Not provided in the document."
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not provided in the document | Not provided in the document |
2. Sample size used for the test set and the data provenance
- Sample size: Not provided in the document.
- Data provenance: Not provided in the document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not provided in the document.
4. Adjudication method for the test set
- Not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was described. The device is a "Cardiac Event Recorder" which records and transmits ECG data for subsequent analysis by a medical professional. It does not incorporate AI assistance for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- The document states, "The ECG recording is then transferred to another device, such as the PaceArt® System K024278 manufactured by Medtronic, Inc. for ECG analysis." and "The Signalife Fidelity 200 Cardiac Event Recorder is not intended to evaluate the signal or alarm." This indicates that the device itself is a data acquisition and transmission tool, and any "standalone" analysis would be performed by a separate, downstream system (like PaceArt®) which is not part of this 510(k) submission. Therefore, a standalone performance study for signal evaluation by the Fidelity 200 was not done, as it's not its intended function.
7. The type of ground truth used
- Not provided in the document.
8. The sample size for the training set
- Not provided in the document.
9. How the ground truth for the training set was established
- Not provided in the document.
{0}------------------------------------------------
K071228
Signalife, Inc 510(k) Summary of Safety and Effectiveness Signalife, Inc. Fidelity 200 Cardiac Event Recorder Section 5
510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
1. Sponsor:
NOV 2 1 2007 Signalife, Inc. 531 S. Main St. #301 Greenville, SC 29601 Phone (864) 233-2300 Facsimile (864) 233-2100 Contact Person: William R. Matthews, Regulatory Affairs Date Prepared: April 30, 2007 2. Device Name: Signalife Fidelity 200 Cardiac Event Recorder Trade or Proprietary Name: Fidelity 200 Cardiac Event Recorder Common or Usual Name: Cardiac Event Recorder Classification Name: 21 CFR 870.2920 - Transmitters and Receivers . Electrocardiograph, Telephone 21 CFR 870.2800 -- Medical Magnetic Tape Recorder . 3. Predicate Devices: The Signalife Fidelity 200 Cardiac Event Recorder is equivalent to: HeartCard® Telephone Electrocardiograh Transmitter and . Receiver, K010945, Instromedix, Inc. Card Guard CG-2206 1-Lead ECG Event Recorder, . K963725, Instromedix, Inc. CG-2211 Self Check ECG Transmitter, K012223, Card . Guard Scientific Survival, LTD 4. Intended Use: The Signalife Fidelity 200 Cardiac Event Recorder is a battery operated, single lead ECG event recorder capable of recording one,
ર-I
45-second event of ECG data for the purpose of transtelephonic transmission to a cardiac monitoring station for analysis and diagnosis by a medical professional where cardiac monitoring is
{1}------------------------------------------------
Signalife, Inc 510(k) Summary of Safety and Effectiveness Signalife, Inc. Fidelity 200 Cardiac Event Recorder Section S
desired by a subscriber to the service. The system includes recording of a single 45-second ECG and transtelephonic transmission of the recorded data to a monitoring station.
The ECG recording is then transferred to another device, such as the PaceArt® System K024278 manufactured by Medtronic, Inc. for ECG analysis. The Signalife Fidelity 200 Cardiac Event Recorder is not intended to sound any physiological alarms.
The Signalife Fidelity 200 Cardiac Event Recorder can be obtained over-the-counter (OTC) or direct-to-consumer by qualified individuals who wish to have access to a cardiac monitoring service.
5. Device Description:
The device acquires an ECG signal and produces a single lead ECG. Upon user activation, an ECG signal is acquired and digitized producing a 45-second ECG recording utilizing the company's signal processing technology.
The Signalife Fidelity 200 Cardiac Event Recorder converts the signal stream into a specific data format where it is stored on a memory chip for transtelephonic transmission to and analysis by another device such as the PaceArt System K024278 manufactured by Medtronic, Inc.
Recorded data is not real time but stored for further analysis. The Signalife Fidelity 200 Cardiac Event Recorder is not intended to evaluate the signal or alarm.
The fundamental technology of the Signalife Fidelity 200 Cardiac Event Recorder is the same as that of the predicate devices. The Signalife Fidelity 200 Cardiac Event Recorder employs the company's signal processing technology to minimize noise in the ECG signal created by the ambulatory nature of the signal source.
6. Basis for Substantial Equivalence:
The Signalife Fidelity 200 Cardiac Event Recorder and the referenced predicates all have substantially equivalent intended use, target population, technical characteristics, performance and compliance to consensus standards for the recording and transtelephonic transmission of ECG data for analysis.
{2}------------------------------------------------
Signalife, Inc 510(k) Summary of Safety and Effectiveness Signalife, Inc. Fidelity 200 Cardiac Event Recorder Section 5
An examination of the differences between the Signalife Fidelity 200 Cardiac Event Recorder and the predicate devices does not raise new questions of safety or effectiveness.
It is our determination that the Fidelity 200 is safe, effective and performs within its design specifications and is substantially equivalent to the referenced predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 1 2007
Signalife, Inc. c/o Mr. William Matthews Regulatory Affairs 531 South Main St. Greenville, SC 29601
Re: K071228
Trade/Device Name: Fidelity 200 Cardiac Event Recorder Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II Product Code: DXH, DSH Dated: October 31, 2007 Received: November 1, 2007
Dear Mr. Matthews:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. William Matthews
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
C. R. Zobel, M.D.
D. Zuckerman M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K071228
Device Name: Signalife Fidelity 200 Cardiac Event Recorder
Indications for Use:
The Signalife Fidelity 200 Cardiac Event Recorder is a battery operated, user-THC Olynalifo Transmitter intended to recorder and transmitter intended to record an ECG by activated EOO overil 1866, as service. This allows users to record and transmit their ECG data to medical professionals via a communication device to a cardiac their LOO data to modified Fidelity 200 Cardiac Event Recorder is intended frionitings who are concerned about their heart rhythm and have experienced the following symptoms that are suggestive of abnormal cardiac rhythms: the following cymptonding heart (Palpitations), Heart racing or irregular pulse, Lightheadedness or Faintness, History of Arrhythmia
Contraindications for Use:
In order to use this service, the patient must be able to perform all of the following: Read and understand the Instructions for Use, Place the Event Recorder on their chest and hold it steadily for at least 45 seconds, Hear the indicator beeps, Speak and understand English, Operate a telephone, Operate a simple, push button device. Due to the possible seriousness of the abnormal heart rhythms that can be associated with these conditions, persons who have been diagnosed with the following conditions should consult their physician before using this service: Blockage of the arteries of the heart, Heart valve problems, Heart transplant, Congestive heart failure, Loss of consciousness. If the user has any of these conditions, Signalife will need to obtain authorization from their physician before enrollment in the service.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mesa May for DOT
(Division Sign-Off)
Division of Cardiovascular
the Vascular Devices
510(k) Number K071228
Page 1 of 2
{6}------------------------------------------------
Warning: This device is not designed to be used with pacemakers or defibrillators. If the user has either of these, they will not be allowed to enroll in this service. The Signalife Fidelity 200 Cardiac Event Recorder cannot predict or diagnose a heart attack or be used for chest pain monitoring and is not a allegriese a nount attention. This device is for monitoring purposes only and has no therapeutic value. It provides the individual access to cardiac monitoring services.
Need for Signed Subscriber Agreement: Your agreement indicates you understand that Signalife (or affiliate) will contact your physician to verify in writing that you are their patient and that they are willing to be contacted in cases where there are clinically significant events involving your care. Your agreement indicates you understand that if written verification is not received from your physician within 35 days of your enrollment, you will not be able to utilize any aspects of the Signalife (or affiliate) service until such verification is received by Signalife (or affiliate). Your agreement certifies you understand that this service is not a substitute for physician care and that this is only a screening service. Signalife, Inc Indications for Use Signalife Model 200 Cardiac Event Recorder
Page 2 of 2
§ 870.2920 Telephone electrocardiograph transmitter and receiver.
(a)
Identification. A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.(b)
Classification. Class II (performance standards).