K Number
K071188
Date Cleared
2007-05-21

(21 days)

Product Code
Regulation Number
866.3045
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The JBAIDS Anthrax Detection System is a real-time polymerase chain reaction (PCR) test system intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences on the pXO1 plasmid (Target 1) and the pXO2 plasmid (Target 2) from Bacillus anthracis. The system can be used to test human whole blood collected in sodium citrate from individuals suspected of having anthrax, positive blood cultures, and cultured organisms grown on blood agar plates. The JBAIDS Anthrax Target 2 assay is used as a supplementary test only after a positive result with the Target 1 Assay.

The JBAIDS Anthrax Target 1 and Target 2 Assays are run on the JBAIDS instrument using the Diagnostic Wizard.

Results are for the presumptive identification of B. anthracis, in conjunction with culture and other laboratory tests. The following considerations also apply:

  • The diagnosis of anthrax infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of pXO1 and pXO2 targets either from cultures or from direct blood specimens.
  • The assays have not been evaluated with blood from individuals without clinical signs or symptoms who were presumed exposed and who subsequently developed anthrax (inhalation or other forms of the disease), or from individuals with any form of anthrax (inhalational, cutaneous, or gastrointestinal).
  • The level of plasmid targets that would be present in blood from individuals with early systemic infection is unknown.
  • The definitive identification of B. anthracis from colony growth, liquid blood culture growth, or from blood specimens requires additional testing and confirmation procedures in consultation with public health or other authorities for whom reports are required.

The safety and effectiveness of other types of tests or sample types (not identified as "For in vitro diagnostic use") have not been established.

Device Description

The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Anthrax Detection System is a fully integrated in vitro diagnostic (IVD) system composed of the following:

  • JBAIDS instrument with laptop computer
  • Software
  • Two different freeze-dried reagent assays (in one kit) for the qualitative detection of pathogenic Bacillus anthracis
  • Four different sample preparation protocols, two for isolating target DNA from whole blood, one for processing blood culture, and another for processing colonies.

The JBAIDS instrument, using Polymerase Chain Reaction (PCR) technology, is a portable thermocycler and real-time fluorimeter. The JBAIDS Anthrax Detection Kit is specially designed for PCR in glass capillaries using the JBAIDS instrument and hydrolysis probes for sequence-specific detection of B. anthracis DNA found on the pX01 plasmid (Target 1) and the pX02 plasmid (Target 2).

The reagent kit contains four different types of freeze-dried reagent vials: Positive Controls, Negative Controls, Inhibition Controls, and Unknowns (used for testing the patient sample). Each JBAIDS assay requires a Positive and Negative Control, and each sample is tested using both an Inhibition Control vial and an Unknown reagent vial.

Before testing, whole-blood samples are purified using the Idaho Technology IT 1-2-3 FLOW or QFLOWdir Sample Purification Kit (or validated equivalent), while blood culture and direct culture specimens are prepared using the IT 1-2-3 SWIPE Sample Purification Kit (or validated equivalent). The resulting purified sample is added to an Unknown reagent vial and an Inhibition Control reagent vial, along with reconstitution buffer. A Positive Control and a Negative Control vial are prepared using reconstitution buffer and reagent grade water. Aliquots from each reagent vial are transferred to two reaction capillaries that are tested together in the JBAIDS instrument. The instrument is programmed to perform heating and cooling cycles that drive the PCR process. The heating and cooling cycles are generated using a heating coil and varying fan speeds. Fluorescence emission is monitored over one of three wavelengths, and the instrument software interprets the change in fluorescence to determine whether the target DNA is present.

When the organism is present, a fragment of B. anthracis DNA is amplified using specific primers. The amplicon is detected by fluorescence using a specific hydrolysis probe. The hydrolysis probe contains a short oligonucleotide that hybridizes to an internal sequence of the amplified fragment during the annealing phase of the PCR cycle. This probe has the 5' and 3' ends labeled with a reporter dye and a quenching dye, respectively. When the probe hybridizes to the specific DNA target, the Taq polymerase enzyme, replicating the target-specific DNA, hydrolyzes the probe, which separates the two fluorophores, thus allowing the reporter dye to fluoresce.

The level of fluorescence from each unknown sample and control is measured by the JBAIDS instrument. JBAIDS Software analyzes fluorescence amplification curves and reports results as "Positive," "Negative," "Inhibited," or "Uncertain." A failure of the Positive or Negative Control will result in the entire run being called "Invalid." Failure of the Inhibition Control yields an Inhibited result for the associated sample and requires retesting of that sample.

AI/ML Overview

The JBAIDS Anthrax Detection System is a real-time PCR test system intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences on the pXO1 plasmid (Target 1) and the pXO2 plasmid (Target 2) from Bacillus anthracis.

1. Acceptance Criteria and Reported Device Performance:

The provided document describes a method comparison and carry-over study rather than specific acceptance criteria thresholds with a pass/fail outcome. The studies aim to demonstrate equivalence between a new sample purification kit (IT 1-2-3 QFLOWdna) and an existing one (IT 1-2-3 FLOW).

Acceptance Criteria (Implied)Reported Device Performance
Method Comparison (Equivalence in detecting B. anthracis at LOD)All six samples spiked at LOD were positive with both Target 1 and Target 2 assays using both IT 1-2-3 QFLOWdna and IT 1-2-3 FLOW. Samples purified with QFLOWdna showed equivalent or better recovery and purity of DNA.
Method Comparison (Specificity/Negative Sample Performance)All 16 normal healthy donor samples processed with IT 1-2-3 QFLOWdna protocol gave negative results for both assays. For IT 1-2-3 FLOW, 15 samples were negative for Target 1 (one inhibited), and all 16 were negative for Target 2. The QFLOWdna showed better or equivalent performance in negative samples.
Carry-over (Rate of False Positives from Strong Positive Samples)For QFLOW™ protocol, carry-over observed in 2.4% (1/42) of negative capillaries for Target 1 assay and 0% (0/42) for Target 2. The carry-over rate for IT 1-2-3 FLOW protocol was 0% (0/12) for both Target 1 and Target 2. The rates were determined to be equivalent.

2. Sample Sizes and Data Provenance:

  • Method Comparison - Spiked Samples: 6 citrated whole blood samples spiked with B. anthracis at the Limit of Detection (LOD).
  • Method Comparison - Negative Samples: 16 normal healthy donors (citrated whole blood samples).
  • Carry-over Study:
    • Strongly positive samples (spiked at 5 x 10" CFU/mL) processed next to negative (unspiked) samples.
    • QFLOW™ protocol: 42 negative capillaries were assessed for carry-over.
    • IT 1-2-3 FLOW protocol: 12 negative capillaries were assessed for carry-over.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be prospective as they were specifically performed to evaluate the new sample purification kit.

3. Number of Experts and Qualifications for Ground Truth: No information provided. This is a molecular diagnostic test, and ground truth would typically be established based on the known spiked concentration of B. anthracis or confirmed negative status of healthy donor samples, rather than expert interpretation of images or clinical data.

4. Adjudication Method: Not applicable. For this type of molecular test, objective results (positive/negative/inhibited) are generated by the instrument based on fluorescence curves and software analysis, not human interpretation requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No. This is not an imaging device or one that relies on human interpretation for diagnosis; therefore, an MRMC study comparing human readers with and without AI assistance is not applicable.

6. Standalone Performance: Yes, the device's performance results (detection of B. anthracis DNA, specificity in negative samples, carry-over rate) are reported for the algorithm (JBAIDS Anthrax Detection System) based on its analysis of samples. The system reports "Positive," "Negative," "Inhibited," or "Uncertain" results without human intervention in the interpretation of the raw data.

7. Type of Ground Truth Used:

  • For the method comparison using spiked samples: The ground truth was based on the known presence of B. anthracis DNA at a defined concentration (LOD).
  • For the method comparison using healthy donor samples: The ground truth was the known absence of B. anthracis DNA in these healthy individuals.
  • For the carry-over study: The ground truth for positive samples was their known high concentration of B. anthracis, and for negative samples, it was the known absence of B. anthracis.

8. Sample Size for the Training Set: Not specified. This document describes performance validation for a specific modification (new sample purification kit) rather than the initial development and training of the PCR assay's core algorithms. PCR assays are typically developed based on laboratory experiments to define primer/probe specificity and amplification efficiency, rather than machine learning on a "training set" in the traditional sense.

9. How the Ground Truth for the Training Set Was Established: Not specified. As mentioned above, for a PCR assay, "training" involves optimizing molecular biology parameters, and ground truth is based on known biological characteristics of the target organism and non-target organisms.

{0}------------------------------------------------

K07//88

MAY 2 1 2007

510(k) Summary 7 .. 1. remissioner portugues por esta programman por portugalites por de la colo lo concertain

510(k) Summary Idaho Technology Inc. JBAIDS Anthrax Detection System

Introduction:According to the requirements of 21 CFR 807.92, the following information providessufficient detail to understand the basis for a determination of substantial equivalence.
Submitted by:Idaho Technology Inc.390 Wakara WaySalt Lake City, UT 84108
Telephone: 801-736-6354Facsimile: 801-588-0507
Contact Person: Beth Lingenfelter, ext. 407
Date Prepared: April 27, 2007
Device Name:Trade Name:JBAIDS Anthrax Detection System
Common Name:Real-time PCR amplification and detection system for targeted Bacillus anthracisDNA sequences
Classification Name:System: Microorganism differentiation and identification device; 21 CFR 866.2660
Instrument: Micro Chemistry Analyzer for Clinical Use; 21 CFR 862.2170, productcode JJF
Reagent Kit: (B. anthracis) NHT

{1}------------------------------------------------

Device The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Anthrax Detection System is a fully integrated in vitro chagnostic (IVD) system composed of the Description: following:

  • D JBAIDS instrument with laptop computer
  • . Software
  • Two different freeze-dried reagent assays (in one kit) for the qualitative � detection of pathogenic Bacillus anthracis
  • . Four different sample preparation protocols, two for isolating target DNA from whole blood, one for processing blood culture, and another for processing colonies.

The JBAIDS instrument, using Polymerase Chain Reaction (PCR) technology, is a portable thermocycler and real-time fluorimeter. The JBAIDS Anthrax Detection Kit is specially designed for PCR in glass capillaries using the JBAIDS instrument and hydrolysis probes for sequence-specific detection of B. anthracis DNA found on the pX01 plasmid (Target 1) and the pX02 plasmid (Target 2).

The reagent kit contains four different types of freeze-dried reagent vials: Positive Controls, Negative Controls, Inhibition Controls, and Unknowns (used for testing the patient sample). Each JBAIDS assay requires a Positive and Negative Control, and each sample is tested using both an Inhibition Control vial and an Unknown reagent vial.

Before testing, whole-blood samples are purified using the Idaho Technology IT 1-2-3TM FLOW or QFLOWdir® Sample Purification Kit (or validated equivalent), while blood culture and direct culture specimens are prepared using the IT 1-2-3 SWIPE Sample Purification Kit (or validated equivalent). The resulting purified sample is added to an Unknown reagent vial and an Inhibition Control reagent vial, along with reconstitution buffer. A Positive Control and a Negative Control vial are prepared using reconstitution buffer and reagent grade water. Aliquots from each reagent vial are transferred to two reaction capillaries that are tested together in the JBAIDS instrument. The instrument is programmed to perform heating and cooling cycles that drive the PCR process. The heating and cooling cycles are generated using a heating coil and varying fan speeds. Fluorescence emission is monitored over one of three wavelengths, and the instrument software interprets the change in fluorescence to determine whether the target DNA is present.

When the organism is present, a fragment of B. anthracis DNA is amplified using specific primers. The amplicon is detected by fluorescence using a specific hydrolysis probe. The hydrolysis probe contains a short oligonucleotide that hybridizes to an internal sequence of the amplified fragment during the annealing phase of the PCR cycle. This probe has the 5' and 3' ends labeled with a reporter dye and a quenching dye, respectively. When the probe hybridizes to the specific DNA target, the Taq polymerase enzyme, replicating the target-specific DNA, hydrolyzes the probe, which separates the two fluorophores, thus allowing the reporter dye to fluoresce.

The level of fluorescence from each unknown sample and control is measured by the JBAIDS instrument. JBAIDS Software analyzes fluorescence amplification curves and reports results as "Positive," "Negative," "Inhibited," or "Uncertain." A failure of the Positive or Negative Control will result in the entire run being called "Invalid." Failure of the Inhibition Control vields an Inhibited result for the associated sample and requires retesting of that sample.

Idaho Technology Inc. Special 510(k) JBAIDS Anthrax Detection Kit

Confidential

{2}------------------------------------------------

Intended Use: The JBAIDS Anthrax Detection System is intended for the qualitative IVD detection of targeted DNA sequences on the pX01 plasmid (Target 1) and the pX02 plasmid (Target 2) from the Bacillus anthracis pathogen. The system can be used to test human whole blood collected in sodium citrate, positive blood cultures, and cultured organisms grown on blood agar plates.

The JBAIDS Anthrax Target 1 Assay, when run on the JBAIDS instrument, is a qualitative IVD test for the detection of one of two DNA sequence targets, both of which are essential for the organism's pathogenicity. The JBAIDS Anthrax Target 2 Assay, when run on the JBAIDS instrument, is a qualitative IVD test for the detection of the second DNA sequence target and is run as a confirmatory test after obtaining a Positive result from the Target 1 Assay. The results from these tests are used in conjunction with culture and other laboratory tests and clinical information as an aid in the diagnosis of systemic anthrax infection in individuals suspected of having the disease.

These tests must be run on the JBAIDS instrument using the Diagnostic option for valid clinical results. Reports from the instrument are identified with the phrase "For In Vitro Diagnostic Use" when the Diagnostic option is used. Results that are not so identified should not be reported or used in patient diagnosis decisions.

Predicate The JBAIDS Anthrax Detection System is substantially equivalent to unmodified Device: system 510(k) number K0561713.

Table 1 Provides a comparison between the modified and unmodified device.

ELEMENTPREDICATE:Unmodified JBAIDS AnthraxDetection System (K051713)New Device:Modified JBAIDS AnthraxDetection System
Intended UseIdentification of anthrax infectionthrough the detection of 2 DNAsequence targets, which are bothessential for Bacillus anthracispathogenicity. Results are used inconjunction with clinical information,culture, and other laboratory tests asan aid in the diagnosis of systemicanthrax infection in individualssuspected of having the disease.Same
SpecimenWhole blood (collected in 3.2%sodium citrate), blood culture (grownin soybean-casein digest broth), orbacterial culture (grown on bloodagar)Same
Table 1. Comparison of the modified and unmodified JBAIDS Anthrax Detection System.
-----------------------------------------------------------------------------------------------

{3}------------------------------------------------

ELEMENTPREDICATE:Unmodified JBAIDS AnthraxDetection System (K051713)New Device:Modified JBAIDS AnthraxDetection System
SpecimenPreparationPurified with IT 1-2-3 FLOW SamplePurification Kit orIT 1-2-3 SWIPE Sample PurificationKit (or validated equivalent)Same, plus the addition of the IT 1-2-3 QFLOWdna Sample PurificationKit as an alternate method for thepurification of whole bloodsamples.
Testing PlatformThe JBAIDS instrument which is areal-time PCR thermocycler.Same
Time Requiredfor Analysis ofSpecimenLess than 3 hoursSame
PhysicalPropertiesFreeze dried reagents withreconstitution water and bufferprovided in kitSame
Test ResultIdentification of two plasmidsrequired for pathogenicity of theorganism; both plasmids are foundtogether in virulent strains of B.anthracis.Same

Performance Summary:

A method comparison and carry-over study were performed and demonstrated that whole blood samples purified using the IT 1-2-3 QFLOWare Sample Purification Kit provide equivalent results as whole blood samples purified using the IT 1-2-3 FLOW Sample Purification Kit.

Method Comparison:

Six (6) citrated whole blood samples spiked at the assay's limit of detection (LOD) with B. anthrocis were processed using the IT 1-2-3 QFLOWdia Sample Purification Kit and the II 1-2-3 FLOW Sample Purification Kit. All six samples processed using both methods were positive with the both the B. anthracis Target 1 and Target 2 assays. Samples purified using the IT 1-2-3 QFLOWdana protocol demonstrated equivalent, or better, recovery and purity of DNA.

Citrated whole blood samples from 16 normal healthy donors were processed using the IT 1-2-3 QFLOW400 and IT 1-2-3 FLOW Sample Purification Kits and then tested using both assays for Q LOW - and II - 2 - 2 - 2 - 1 - and 11 - 1 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 protocol gave negative test results for both assays. One sample processed using the IT 1-2-3 FLOW protocol gave an inhibited result with the Target 1 assay while the other 15 were negative. All 16 samples gave negative results with the Target 2 assay.

{4}------------------------------------------------

Carrv-over:

To determine the rate of carry-over for the two sample purification kits, strongly positive samples (spiked at 5 x 10" CFUmL) were purified next to negative (unspiked) samples. The negative (spired at 9 x 10 °C L Crimi) white p Target 1 and Target 2 assays. For samples purified using the QFLOW™ protocol, carry-over was observed in 2.4% (1/42) of negative capillaries for the Target Q1 LOW - precodes, carry he Target 2 assay. The carry-over rate for the same samples processed r assay and 0% (0/12) for both the Target 1 and Target 1 and Target 2 assays. The rate of asing the 1 2011 protoce. the wo sample purification methods was determined to be equivalent and is also equivalent to the reported carry-over rate reported for blood culture samples purified using the IT 1-2-3 SWIPE Sample purification kit.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Beth Lingenfelter Regulatory Affairs Manager Idaho Technology Inc. 390 Wakara Way Salt Lake City, UT 84108

MAY 2 1 2007

Re: K071188

Trade/Device Name: JBAIDS Anthrax Detection System Regulation Number: 21 CFR 866.2660 Regulation Name: Microorganism differentiation and identification device Regulatory Class: Class II Product Code: NHT Dated: April 27, 2007 Received: April 30, 2007

Dear Ms. Lingenfelter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 10gally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its tollsfree number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Sally, attorn

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): Device Name: JBAIDS Anthrax Detection System Indications for Use:

The JBAIDS Anthrax Detection System is a real-time polymerase chain reaction (PCR) test system intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences on the pXO1 plasmid (Target 1) and the pXO2 plasmid (Target 2) from Bacillus anthracis. The system can be used to test human whole blood collected in sodium citrate from individuals suspected of having anthrax, positive blood cultures, and cultured organisms grown on blood agar plates. The JBAIDS Anthrax Target 2 assay is used as a supplementary test only after a positive result with the Target 1 Assay.

The JBAIDS Anthrax Target 1 and Target 2 Assays are run on the JBAIDS instrument using the Diagnostic Wizard.

Results are for the presumptive identification of B. anthracis, in conjunction with culture and other laboratory tests. The following considerations also apply:

  • K The diagnosis of anthrax infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of pXO1 and pXO2 targets either from cultures or from direct blood specimens.
  • 트 The assays have not been evaluated with blood from individuals without clinical signs or symptoms who were presumed exposed and who subsequently developed anthrax (inhalation or other forms of the disease), or from individuals with any form of anthrax (inhalational, cutaneous, or gastrointestinal).
  • The level of plasmid targets that would be present in blood from individuals with early systemic infection is unknown.
  • . The definitive identification of B. anthracis from colony growth, liquid blood culture growth, or from blood specimens requires additional testing and confirmation procedures in consultation with public health or other authorities for whom reports are required.

The safety and effectiveness of other types of tests or sample types (not identified as "For in vitro diagnostic use") have not been established.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use
(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Freddie Lee Poole

ision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

ldaho Technology Inc. Special 510(k) JBAIDS Anthrax Detection System

510(k) Ko71188

Page vi

§ 866.3045 In vitro diagnostic device for

Bacillus spp. detection.(a)
Identification. An in vitro diagnostic device forBacillus species (spp.) detection is a prescription device used to detect and differentiate amongBacillus spp. and presumptively identifyB. anthracis and otherBacillus spp. from cultured isolates or clinical specimens as an aid in the diagnosis of anthrax and other diseases caused byBacillus spp. This device may consist ofBacillus spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to presumptively identify bacillus-like organisms in clinical specimens; bacteriophage used for differentiatingB. anthracis from otherBacillus spp. based on susceptibility to lysis by the phage; or antigens used to identify antibodies toB. anthracis (anti-toxin and anti-capsular) in serum. Bacillus infections include anthrax (cutaneous, inhalational, or gastrointestinal) caused byB. anthracis, and gastrointestinal disease and non-gastrointestinal infections caused byB. cereus. (b)
Classification. Class II (special controls). The special controls are set forth in FDA's special controls guideline document entitled “In Vitro Diagnostic Devices forBacillus spp. Detection; Class II Special Controls Guideline for Industry and Food and Drug Administration Staff.” For availability of the guideline document, see § 866.1(e).(c)
Restriction on Distribution. The distribution of these devices is limited to laboratories that follow public health guidelines that address appropriate biosafety conditions, interpretation of test results, and coordination of findings with public health authorities.(d)
Restriction on Use. The use of this device is restricted to prescription use and must comply with the following:(1) The device must be in the possession of:
(i)(A) A person, or his agents or employees, regularly and lawfully engaged in the manufacture, transportation, storage, or wholesale or retail distribution of such device; or
(B) A practitioner, such as a physician, licensed by law to use or order the use of such device; and
(ii) The device must be sold only to or on the prescription or other order of such practitioner for use in the course of his professional practice.
(2) The label of the device shall bear the statement “Caution: Federal law restricts this device to sale by or on the order of a ____”, the blank to be filled with the word “physician” or with the descriptive designation of any other practitioner licensed by the law of the State in which he practices to use or order the use of the device.
(3) Any labeling, as defined in section 201(m) of the Federal Food, Drug, and Cosmetic Act, whether or not it is on or within a package from which the device is to be dispensed, distributed by, or on behalf of the manufacturer, packer, or distributor of the device, that furnishes or purports to furnish information for use of the device contains adequate information for such use, including indications, effects, routes, methods, and frequency and duration of administration and any relevant hazards, contraindications, side effects, and precautions, under which practitioners licensed by law to employ the device can use the device safely and for the purposes for which it is intended, including all purposes for which it is advertised or represented. This information will not be required on so-called reminder-piece labeling which calls attention to the name of the device but does not include indications or other use information.
(4) All labeling, except labels and cartons, bearing information for use of the device also bears the date of the issuance or the date of the latest revision of such labeling.