(170 days)
No
The document describes image analysis capabilities but does not explicitly mention the use of AI, ML, or related terms like deep learning or neural networks. The analysis is based on "color, intensity, size, pattern and shape," which can be achieved with traditional image processing algorithms.
No
The device is described as an aid to the pathologist for diagnosis (display, detection, counting, and classification of tissues and cells; semi-quantitative measurement of HER2/neu) and assessment for treatment consideration, not for treating a condition itself. Therefore, it is a diagnostic device, not a therapeutic one.
Yes
The device "is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest" and includes an "IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2)". These statements clearly indicate its use in providing information for diagnosis.
No
The device description explicitly states that the system components include an automated digital microscope slide scanner, computer, color monitor, keyboard, and digital pathology information management software. This indicates the device is a system with both hardware and software components, not software-only.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use Statement: The "Intended Use / Indications for Use" section explicitly states: "It is intended for in vitro diagnostic use as an aid to the pathologist..." This is a clear declaration of its intended purpose as an IVD.
- Accessory to an IVD: The IHC HER2 Image Analysis application is described as an "accessory to the Dako HercepTest™," which is a known IVD used for HER2 testing. This further reinforces its role within the in vitro diagnostic process.
- Clinical Decision Making: The device is intended to aid pathologists in the "display, detection, counting and classification of tissues and cells of clinical interest" and specifically in the "semi-quantitative measurement of HER2/neu" to aid in the assessment of breast cancer patients for whom HERCEPTIN® treatment is being considered. This directly relates to providing information for clinical decision-making based on in vitro analysis of biological specimens.
- Analysis of Biological Specimens: The system analyzes digital images derived from "formalin-fixed, paraffin-embedded normal and neoplastic tissue," which are biological specimens.
All these points align with the definition of an In Vitro Diagnostic device, which is used to examine specimens taken from the human body to provide information for diagnosis, monitoring, or treatment.
N/A
Intended Use / Indications for Use
The ScanScope System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.
The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue.
The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established.
Product codes
NOT
Device Description
System: The ScanScope® XT System is an automated digital slide creation, management, viewing and analysis system. The ScanScope® XT System components consist of an automated digital microscope slide scanner, computer, color monitor, keyboard and digital pathology information management software. The system capabilities include digitizing microscope slides at high resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing facilities for annotating digital slides and entering and editing metadata associated with digital slides, and facilities for image analysis of digital slides. Image analysis capabilities include the ability to quantify characteristics useful to Pathologists, such as measuring and scoring immunohistochemical stains applied to histology specimens, such as the Dako HerceptTest"M, which reveal the presence of proteins such as Human Epidermal growth factor Receptor 2 (HER2), which may be used to determine patient treatment for breast cancer.
Hardware Operation: The ScanScope XT digital slide scanner creates high resolution, color digital slide images of entire glass slides in a matter of minutes. High numeric aperture 20x or 40x objectives, as found on conventional microscopes, are used to produce high-quality images. The ScanScope XT employs a linear-array scanning technique that generates digital slide images that have no tiling artifacts and that are essentially free from optical aberrations along the scanning axis.
Software Operation: The Spectrum™ software is a full-featured digital pathology information management system. The software runs on a server computer called a Digital Slide Repository (DSR), which stores digital slide images on disk storage such as a RAID array, and which hosts an SQL database that contains digital slide metadata. Spectrum includes a web application and services which encapsulate database and digital slide image access for other computers. The Spectrum server supports the capability of running a variety of digital slide image analysis algorithms on digital slides, and storing the results of analysis into the database. Spectrum also includes support for locally or remotely connected image workstation computers, which run digital slide viewing and analysis software provided as part of Spectrum.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Microscopic Digital Slide Scans
Anatomical Site
Formalin-fixed, paraffin-embedded normal and neoplastic tissue, specifically breast cancer specimens.
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Pathologist / In vitro diagnostic use
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
A multi-site study was conducted at two clinical sites to compare the performance of Aperio's IHC HER2 Image Analysis to manual microscopy. 180 formalin-fixed, paraffinembedded breast tissue specimens immunohistochemically stained using Dako's HerceptTest™ were used for this study; 80 specimens with approximately equal HER2 score distribution from site 1, and 100 routine specimens from site 2. At each site, three pathologists performed a blinded read of the glass slides using a microscope and reported the HER2 score for each of the slides. The glass slides were scanned at Aperio using a different ScanScope for each site, and after a wash-out period of over one week and randomization of the slides, the same three pathologists remotely viewed and outlined a representative set of tumor regions to be analyzed by the IHC HER2 image analysis. The pathologists received feedback on the way they outlined tumor regions for their first 3 to 7 slides before the slides were analyzed. The algorithm itself was run in batch mode blinded from the pathologists to avoid any influence of the pathologists in their choice of the tumor regions. The algorithm was used "out of the box". The algorithm reported the HER2 score for each of the three pathologists for each of the slides.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Comparison studies:
a. Method comparison with predicate device:
The substantial equivalence study was based on comparison of image analysis to conventional manual microscopy.
A multi-site study was conducted at two clinical sites to compare the performance of Aperio's IHC HER2 Image Analysis to manual microscopy. 180 formalin-fixed, paraffinembedded breast tissue specimens immunohistochemically stained using Dako's HerceptTest™ were used for this study; 80 specimens with approximately equal HER2 score distribution from site 1, and 100 routine specimens from site 2. At each site, three pathologists performed a blinded read of the glass slides using a microscope and reported the HER2 score for each of the slides. The glass slides were scanned at Aperio using a different ScanScope for each site, and after a wash-out period of over one week and randomization of the slides, the same three pathologists remotely viewed and outlined a representative set of tumor regions to be analyzed by the IHC HER2 image analysis. The pathologists received feedback on the way they outlined tumor regions for their first 3 to 7 slides before the slides were analyzed. The algorithm itself was run in batch mode blinded from the pathologists to avoid any influence of the pathologists in their choice of the tumor regions. The algorithm was used "out of the box". The algorithm reported the HER2 score for each of the three pathologists for each of the slides.
The statistical analyses are presented across all slides for each of the methods; manual microscopy and image analysis, and comparatively between methods for manual microscopy against image analysis.
Statistical analyses are provided for a trichotomous categorization of the HER2 scores combining 0 and 1+ and leaving 2+ and 3+ uncombined. Percentage Agreement (PA) along with an exact 95% Confidence Interval (CI) are presented overall for all trichotomous HER2 score categories combined and for each of the trichotomous HER2 score categories separately using a dichotomous outcome of that category vs. the two other categories.
Manual Microscopy - Inter-Pathologists - Agreements:
Clinical Site 1: Pathologist 1 v 2 PA: 91.3% (82.8, 96.4); Pathologist 1 v 3 PA: 77.5% (66.8, 86.1); Pathologist 2 v 3 PA: 76.3% (65.4, 85.1)
Clinical Site 2: Pathologist 1 v 2 PA: 84.0% (75.3, 90.6); Pathologist 1 v 3 PA: 82.0% (73.1, 89.0); Pathologist 2 v 3 PA: 90.0% (82.4, 95.1)
Image Analysis - Inter-Pathologists - Agreements:
Clinical Site 1: Pathologist 1 v 2 PA: 88.8% (79.7, 94.7); Pathologist 1 v 3 PA: 93.8% (86.0, 97.9); Pathologist 2 v 3 PA: 86.3% (76.7, 92.9)
Clinical Site 2: Pathologist 1 v 2 PA: 87.0% (78.8, 92.9); Pathologist 1 v 3 PA: 92.0% (84.8, 96.5); Pathologist 2 v 3 PA: 89.0% (81.2, 94.4)
Manual Microscopy vs Image Analysis - same Pathologist - Agreements:
Clinical Site 1: Pathologist 1 PA: 92.5% (84.4, 97.2); Pathologist 2 PA: 90.0% (81.2, 95.6); Pathologist 3 PA: 77.5% (66.8, 86.1)
Clinical Site 2: Pathologist 1 PA: 90.0% (82.4, 95.1); Pathologist 2 PA: 79.0% (69.7, 86.5); Pathologist 3 PA: 90.0% (82.4, 95.1)
Key Results: The inter-pathologists agreements for the performed (blinded) image analysis (PA: 86.3-93.8%) were comparable to the inter-pathologists agreements for manual microscopy (PA: 76.3-91.3%). The agreements between the pathologists' manual microscopy and performed (blinded) image analysis (PA: 77.5-92.5%) were comparable to the interpathologists agreements for manual microscopy (PA: 76.3-91.3%).
Analytical Performance:
a. Precision:
The precision of the ScanScope XT System was determined in a suite of intra-run/intrasystem, inter-run/intra-system and inter-systems studies. Eight HER2 slides from the comparison study were selected to provide two slides in each of the HER2 score classes 0, 1+, 2+ and 3+.
Intra-run/intra-system: The slide scores provided by Image Analysis over ten consecutive scans were analyzed for all eight HER2 slides. The data show perfect agreement (100%) for the calculated HER2 scores across all runs.
Inter-run/intra-system: The slide scores provided by Image Analysis over twenty scans on different days were analyzed for all eight HER2 slides. The data show perfect agreement (100%) for the calculated HER2 scores across all runs.
Inter-systems: The slide scores provided by Image Analysis over ten consecutive scans on three different ScanScope scanner instruments were analyzed for all eight HER2 slides. The data show perfect agreement (100%) for the calculated HER2 scores across all systems and across all runs.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Percentage Agreement (PA)
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 864.1860 Immunohistochemistry reagents and kits.
(a)
Identification. Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.(b)
Classification of immunohistochemistry devices. (1) Class I (general controls). Except as described in paragraphs (b)(2) and (b)(3) of this section, these devices are exempt from the premarket notification requirements in part 807, subpart E of this chapter. This exemption applies to IHC's that provide the pathologist with adjunctive diagnostic information that may be incorporated into the pathologist's report, but that is not ordinarily reported to the clinician as an independent finding. These IHC's are used after the primary diagnosis of tumor (neoplasm) has been made by conventional histopathology using nonimmunologic histochemical stains, such as hematoxylin and eosin. Examples of class I IHC's are differentiation markers that are used as adjunctive tests to subclassify tumors, such as keratin.(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
(3) Class III (premarket approval). IHC's intended for any use not described in paragraphs (b)(1) or (b)(2) of this section.
(c)
Date of PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the Federal Food, Drug, and Cosmetic Act is required for any device described in paragraph (b)(3) of this section before this device may be commercially distributed. See § 864.3.
0
Ko7iiz8
510(k) Summary of Substantial Equivalence Aperio Technologies, Inc. (ScanScope® XT System)
OC7 1 0 2007
21 CFR 807.92(a):
21 CFR 807.92(a) (1):
Submitter's name and address:
Aperio Technologies, Inc. 1430 Vantage Court, Suite 106 Vista, CA 92081
Submitter's telephone and fax numbers:
Phone: | (760) 304-6211 |
---|---|
Fax: | (760) 304-6211 |
Contact person:
Kim L. Bloom Manager of Regulatory Affairs Aperio Technologies, Inc. 1430 Vantage Court, Suite 106 Vista, CA 92081 kbloom@aperio.com
Date this 510(k) summary was prepared:
September 28, 2007
21 CFR 807.92(a)(2):
Trade Name of Device: | ScanScope® XT System |
---|---|
Regulatory Section: | 21 CFR 864.1860 Immunohistochemistry reagents and kits |
Classification: | Class II |
Product Code: | NOT (microscope, automated, image analysis, operator |
intervention) |
1
21 CFR 807.92(a)(3): Leally marketed predicate device to which substantial equivalence is claimed:
| Predicate Device: | Automated Cellular Imaging System (“ACIS”) and ACIS HER2
software application |
|----------------------|----------------------------------------------------------------------------------|
| Manufacturer: | ChromaVision Medical Systems, Inc. |
| Predicate Device k#: | K032113 |
21 CFR 807.92(a)(4): Description of the device that is the subject of this premarket notification:
System: The ScanScope® XT System is an automated digital slide creation, management, viewing and analysis system. The ScanScope® XT System components consist of an automated digital microscope slide scanner, computer, color monitor, keyboard and digital pathology information management software. The system capabilities include digitizing microscope slides at high resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing facilities for annotating digital slides and entering and editing metadata associated with digital slides, and facilities for image analysis of digital slides. Image analysis capabilities include the ability to quantify characteristics useful to Pathologists, such as measuring and scoring immunohistochemical stains applied to histology specimens, such as the Dako HerceptTest"M, which reveal the presence of proteins such as Human Epidermal growth factor Receptor 2 (HER2), which may be used to determine patient treatment for breast cancer.
Hardware Operation: The ScanScope XT digital slide scanner creates high resolution, color digital slide images of entire glass slides in a matter of minutes. High numeric aperture 20x or 40x objectives, as found on conventional microscopes, are used to produce high-quality images. The ScanScope XT employs a linear-array scanning technique that generates digital slide images that have no tiling artifacts and that are essentially free from optical aberrations along the scanning axis.
Software Operation: The Spectrum™ software is a full-featured digital pathology information management system. The software runs on a server computer called a Digital Slide Repository (DSR), which stores digital slide images on disk storage such as a RAID array, and which hosts an SQL database that contains digital slide metadata. Spectrum includes a web application and services which encapsulate database and digital slide image access for other computers. The Spectrum server supports the capability of running a variety of digital slide image analysis algorithms on digital slides, and storing the results of analysis into the database. Spectrum also includes support for locally or remotely connected image workstation computers, which run digital slide viewing and analysis software provided as part of Spectrum.
2
Overview of System Operation: The laboratory technician or operator loads glass microscope slides into a specially designed slide carrier with a capacity of up to 120 slides. The scanning process begins when the operator starts the ScanScope scanner and finishes when the scanner has completed scanning of all loaded slides. As each glass slide is processed, the system automatically stores individual "striped" images of the tissue contained on the glass slide and integrates the striped images into a single digital slide image, which represents a histological reconstruction of the entire tissue section. After scanning is completed, the operator is able to view and perform certain analytical tests on the digital slides.
21 CFR 807.92(a)(5): Intended use and labeled indications for use:
The ScanScope System is an automated digital slide creation, management, viewing and analysissystem. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.
The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue.
The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established.
21 CFR 807.92(a)(6): Technological characteristics:
The design, construction, energy source and other characteristics of the ScanScope System candidate device are considered to be substantially equivalent to the relevant features of the predicate device. A summary of the technological characteristics of the ScanScope System candidate device in comparison to the predicate device follows:
3
Method of cell detection. The method of cell detection is by colorimetric pattern recognition by microscopic examination of prepared cells by size, shape, hue and intensity as observed by a computer-automated, microscopic digital slide scanner system and/or by visual observation by a health care professional.
System Components. The system components comprising the ScanScope System candidate device are substantially equivalent to those in the predicate device; i.e., a computer-automated digital microscope slide scanner, computer, color monitor, and keyboard.
Energy Source. The electrical service is 100vAC - 240vAC, 50Hz/60 Hz, 2 amp, which is similar to the predicate device electrical service requirements.
21 CFR 807.92(b): 510(k) summaries for those premarket submissions in which determination of substantial equivalence is also based on an assessment of performance data shall contain the following information:
21 CFR 807.92(b)(1): Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification:
There are no nonclinical tests submitted, referenced or relied on in this submission.
21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:
Comparison studies:
a. Method comparison with predicate device:
The substantial equivalence study was based on comparison of image analysis to conventional manual microscopy.
A multi-site study was conducted at two clinical sites to compare the performance of Aperio's IHC HER2 Image Analysis to manual microscopy. 180 formalin-fixed, paraffinembedded breast tissue specimens immunohistochemically stained using Dako's HerceptTest™ were used for this study; 80 specimens with approximately equal HER2 score distribution from site 1, and 100 routine specimens from site 2. At each site, three pathologists performed a blinded read of the glass slides using a microscope and reported the HER2 score for each of the slides. The glass slides were scanned at Aperio using a different ScanScope for each site, and after a wash-out period of over one week and randomization of the slides, the same three pathologists remotely viewed and outlined a representative set of tumor regions to be analyzed by the IHC HER2 image analysis. The pathologists received feedback on the way they outlined tumor regions for their first 3 to 7 slides before the slides were analyzed. The algorithm itself was run in batch mode
4
blinded from the pathologists to avoid any influence of the pathologists in their choice of the tumor regions. The algorithm was used "out of the box". The algorithm reported the HER2 score for each of the three pathologists for each of the slides.
The statistical analyses are presented across all slides for each of the methods; manual microscopy and image analysis, and comparatively between methods for manual microscopy against image analysis.
Statistical analyses are provided for a trichotomous categorization of the HER2 scores combining 0 and 1+ and leaving 2+ and 3+ uncombined. Percentage Agreement (PA) along with an exact 95% Confidence Interval (CI) are presented overall for all trichotomous HER2 score categories combined and for each of the trichotomous HER2 score categories separately using a dichotomous outcome of that category vs. the two other categories.
Pathologist 1 v 2 | Pathologist 1 v 3 | Pathologist 2 v 3 | ||||
---|---|---|---|---|---|---|
PA | PA 95% CI | PA | PA 95% CI | PA | PA 95% CI | |
Clinical Site 1 | 91.3% | (82.8, 96.4) | 77.5% | (66.8, 86.1) | 76.3% | (65.4, 85.1) |
Clinical Site 2 | 84.0% | (75.3, 90.6) | 82.0% | (73.1, 89.0) | 90.0% | (82.4, 95.1) |
Manual Microscopy - Inter-Pathologists - Agreements.
Pathologist 1 v 2 | Pathologist 1 v 3 | Pathologist 2 v 3 | ||||
---|---|---|---|---|---|---|
PA | PA 95% CI | PA | PA 95% CI | PA | PA 95% CI | |
Clinical Site 1 | 88.8% | (79.7, 94.7) | 93.8% | (86.0, 97.9) | 86.3% | (76.7, 92.9) |
Clinical Site 2 | 87.0% | (78.8, 92.9) | 92.0% | (84.8, 96.5) | 89.0% | (81.2, 94.4) |
Pathologist 1 | Pathologist 2 | Pathologist 3 | ||||
---|---|---|---|---|---|---|
PA | PA 95% CI | PA | PA 95% CI | PA | PA 95% CI | |
Clinical Site 1 | 92.5% | (84.4, 97.2) | 90.0% | (81.2, 95.6) | 77.5% | (66.8, 86.1) |
Clinical Site 2 | 90.0% | (82.4, 95.1) | 79.0% | (69.7, 86.5) | 90.0% | (82.4, 95.1) |
Image Analysis - Inter-Pathologists - Agreements.
Manual Microscopy vs Image Analysis - same Pathologist - Agreements.
The inter-pathologists agreements for the performed (blinded) image analysis (PA: 86.3-93.8%) were comparable to the inter-pathologists agreements for manual microscopy (PA: 76.3-91.3%). The agreements between the pathologists' manual microscopy and performed (blinded) image analysis (PA: 77.5-92.5%) were comparable to the interpathologists agreements for manual microscopy (PA: 76.3-91.3%).
Analytical Performance:
.
a. Precision:
The precision of the ScanScope XT System was determined in a suite of intra-run/intrasystem, inter-run/intra-system and inter-systems studies. Eight HER2 slides from the
5
comparison study were selected to provide two slides in each of the HER2 score classes 0, 1+, 2+ and 3+.
Intra-run/intra-system: The slide scores provided by Image Analysis over ten consecutive scans were analyzed for all eight HER2 slides. The data show perfect agreement (100%) for the calculated HER2 scores across all runs.
Inter-run/intra-system: The slide scores provided by Image Analysis over twenty scans on different days were analyzed for all eightHER2 slides. The data show perfect agreement (100%) for the calculated HER2 scores across all runs.
Inter-systems: The slide scores provided by Image Analysis over ten consecutive scans on three different ScanScope scanner instruments were analyzed for all eight HER2 slides. The data show perfect agreement (100%) for the calculated HER2 scores across all systems and across all runs.
21 CFR 807.92(b)(3): Conclusions drawn from the nonclinical and clinical tests:
Based on the results of the clinical studies described in this 510(k) submission, it is concluded that the ScanScope System device is as safe and effective (therefore substantially equivalent) as the predicate device as an aid in the assessment of specimens from breast cancer patients for whom Herceptin® (Trastuzumab) treatment is being considered.
... End of 510(k) Summary ....
6
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, along with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black against a white background.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
OCT 1 0 2007
Perry Johnston Aperio Technologies 1430 Vantage Court Suite 106 Vista, California 92081
Re: K071128
Trade/Device Name: Scanscope XT System Regulation Number: 21 CFR 864.1860 Regulation Name: Immunohistochemistry reagents and kits Regulatory Class: Class II Product Code: NOT Dated: April 20, 2007 Received: April 23, 2007
Dear Mr. Johnston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter
7
Page 2 – Perry Johnston
will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. 'Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Robert H. Beckerf
Robert L. Becker, Jr., M.D., Ph.D., Ph.B. Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
8
Indications for Use
510(k) Number (if known): _ K07 | | 28
Device Name: ScanScope® XT System
Indications for Use:
The ScanScope System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.
The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue.
The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established.
Prescription Use X (Part 21CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Josephine Bautista
Division Office of In Vitro Diagnostic Device Evaluation and Safety510(k)___________________________________________________________________________________________________________________________________________________