(270 days)
Flier's Quality Water Systems, Inc. systems are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001. They remove organic and inorganic substances and microbial contaminants from water used to dilute dialysis concentrate to form dialysate, in the reprocessing of hemodialyzers, and equipment rinse and disinfection.
The Acute System #1 consists of two each carbon and mixed bed DI tanks with pre & post filtration. It is intended to be placed on the back of dialysis machine or on a handcart for single patient dialysis. Acute System #2 consists of just carbon tanks and is intended to be used as pretreatment to a single patient RO system.
The Central System #1 and #2 consists of two or more carbon and mixed bed D1 tanks with pre & post filtration. They are used as either a temporary system to feed multiple dialysis machines in normal use, or for a system to feed a biomedical technician service station. For permanent, multiple station dialysis clinics Flier's recommends using central water treatment systems including reverse osmosis.
The Back-up System #1 and #2 consists of two or more mixed bed DI tanks as a component in a "back up" condition when a RO system or major component on a water purification system for hemodialysis is in need of repair. It may also be used in DI polishing mode to remove ions from an RO stream to meet AAMI RD62:2001 requirements.
Upon exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or with new resin altogether.
These systems are components of a larger water treatment system employing adequate pre-treatment and post-treatment sections. Flier's systems are not to be used alone.
The systems are primary or temporary devices used to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001.
The Flier's Quality Water Systems, Inc. systems purify potable feed water through deionization. Deionization is used to remove 99.9% of ions from water. Deionization alone does not remove particulates, organics, bacteria, viruses or endotoxins. The systems require pretreatment and post-treatment. Deionizers are designed in a hemodialysis application to treat RO water or carbon filtered potable water. Improper use can result in the formation of nitrosamines in the effluent of the deionizer.
The purpose of the pretreatment section of the system is to condition that feed water supplying the deionizers. Conditioning the feed water will include: cartridge filters to reduce particulates and sediment and carbon filtration tarks to remove chlorine/chloramine residual.
The purpose of the post-treatment section of the system is to remove bacteria and endotoxins or lower them to acceptable levels as required by AAMI standards. The post-treatment section of the system will include: submicron/ultrafilters after the deionizers, sample ports to check for microbial contamination and to test for comprehensive AAMI analysis, a 1 megohm quality control light at mid tank and a temperature compensated audible and visual resistivity alarm set at 1M ohm or greater for final water quality. A remote alarm will be installed if the deionization system is not located in the patient treatment area. A distribution section of the system is necessary to deliver product water that meets ANSI/AAMI standards to the points of use. During use, continual monitoring of the effluent through the use of the temperature compensated audible and visual resistivity alarm is required. Routine monitoring of the mid tank quality indicator is also required on intervals during the day, at least before and after each patient treatment.
Should final water quality fall at or below 1 megohm, the alarm will sound and all patient treatments must be discontinued. The deionization tanks must be replaced per tank exchange procedure prior to resuming and patient treatment.
Divert to drain option will be provided when necessary per AAMI quidelines. Divert to drain is not supplied for acute systems located in patient room. Monitor and divert to drain not supplied for back-up systems with existing monitor/alarm/divert.
This document is a 510(k) Premarket Notification for the Flier's Quality Water Systems, Inc. line of deionization systems for hemodialysis. It is a submission to the FDA seeking clearance for these medical devices. Unlike a research paper that presents a study to prove a device meets acceptance criteria, this document declares substantial equivalence to a predicate device already on the market. Therefore, the information requested in your prompt regarding acceptance criteria and a study proving performance is largely not present in this type of regulatory submission in the way you're asking.
Here's an analysis based on the provided text, explaining why certain information is not available and what can be inferred:
Acceptance Criteria and Study for Flier's Quality Water Systems, Inc. (K071104)
This 510(k) submission primarily focuses on demonstrating substantial equivalence to a predicate device (USFilter Corporation's/Siemens, K980182) rather than presenting a standalone study with defined acceptance criteria and performance metrics. Therefore, explicit tables of acceptance criteria for the new device and a study proving it meets those criteria are not provided in this document. Instead, the "acceptance criteria" are implicitly met by demonstrating that the new device functions in a manner substantially equivalent to the predicate device, thereby fulfilling the requirements for water purification in hemodialysis as specified by ANSI/AAMI RD62:2001.
1. A table of acceptance criteria and the reported device performance
- Acceptance Criteria (Inferred): The primary "acceptance criteria" for the Flier's systems are to achieve water purity levels suitable for hemodialysis applications, specifically by removing 99.9% of ions and effectively reducing particulates, chlorine/chloramine, bacteria, viruses, and endotoxins to acceptable ANSI/AAMI RD62:2001 standards. This is achieved through deionization combined with pre- and post-treatment.
- Specific criteria mentioned:
- Remove 99.9% of ions.
- Remove chlorine/chloramine residual (minimum empty bed contact time of 10 minutes, activated carbon with Iodine number of 900 or greater).
- Submicron/ultrafilters after deionizers to remove bacteria and endotoxins or lower them to acceptable AAMI levels.
- Temperature compensated audible and visual resistivity alarm set at 1M ohm or greater for final water quality.
- Mid-tank quality indicator for routine monitoring.
- Divert to drain option when necessary.
- Specific criteria mentioned:
- Reported Device Performance (Inferred): The document states that "The Flier's Quality Water Systems, Inc. systems purify potable feed water through deionization. Deionization is used to remove 99.9% of ions from water." It also highlights that "The systems require pretreatment and post-treatment" to address other contaminants. The performance is essentially claimed to be equivalent to the predicate device, which also uses carbon and deionization with pre- and post-filtration for water purification in hemodialysis. The conclusion explicitly states: "The core water purification components and technology are exactly the same. The predicate device components utilize the exact same water purification principles."
| Acceptance Criteria (Inferred from Predicate Equivalence & AAMI Standards) | Reported Device Performance (Inferred/Stated) |
|---|---|
| Removal of 99.9% of ions from water | "Deionization is used to remove 99.9% of ions from water." |
| Removal of chlorine and chloramines to AAMI standards | Carbon filtration utilized, two carbon filters in series, minimum 10 min EBCT, Iodine number ≥ 900 |
| Reduction of bacteria and endotoxins to AAMI standards | Post-treatment includes submicron/ultrafilters. |
| Final water quality monitoring and alarm at ≥ 1 M Ohm | Temperature compensated audible and visual resistivity alarm set at 1M ohm or greater. |
| Mid-tank quality indicator for routine monitoring | 1 megohm quality control light at mid-tank. |
| Compliance with ANSI/AAMI RD62:2001 | Systems "are primary or temporary devices used to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. This document is a 510(k) premarket notification for substantial equivalence, not a clinical or performance study report. There is no "test set" in the context of clinical trial data or a formal performance evaluation with a specified sample size. The claim of equivalence is based on the technical characteristics and operating principles being the same as a legally marketed predicate device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. As there is no "test set" used in the manner of a diagnostic device requiring ground truth establishment by experts, this information is not provided. The "ground truth" here is the established standards for water quality in hemodialysis (ANSI/AAMI RD62:2001) and the demonstrated performance of the predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. This information is relevant to studies involving human interpretation or subjective assessments, which is not the nature of this submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This device is a water purification system, not a diagnostic imaging device or an AI-powered system designed to assist human readers. Therefore, an MRMC study is irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This device does not involve an algorithm or AI. It is a physical water purification system. Its performance is inherent in its engineering and components, designed to meet physical and chemical purity standards.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" in this context is the physicochemical requirements for water purity in hemodialysis as defined by established standards, specifically ANSI/AAMI RD62:2001. The performance of the predicate device (K980182) in meeting these standards serves as the benchmark for substantial equivalence.
8. The sample size for the training set
- Not Applicable. This device does not utilize machine learning or AI, and therefore does not have a "training set."
9. How the ground truth for the training set was established
- Not Applicable. As there is no "training set," this question is not relevant.
{0}------------------------------------------------
K071104
PAGE 1 OF 3
Image /page/0/Picture/1 description: The image shows the logo for Flier's Quality Water Systems Inc. The logo features the company name in a stylized font, with the words "Quality Water Systems" stacked on top of each other. To the left of the text is a black water droplet, which serves as a visual representation of the company's focus on water systems.
JAN 1 4 2008
510(k) Summary
This 510(k) Summary is provided as part of this Premarket Notification to comply with the provisions of the safe Medical Devices Act of 1990 requiring that either a summary be included in a submission or a statement that a summary is available upon request.
Submitter James B. Flier Flier's Quality Water Systems. Inc. 7425 Clyde Park Ave. SW Byron Center, MI 49315 April 13, 2006
Device Names Acute System #1, Acute System #2 (AS-1, AS-2) Central System #1, Central System #2 (CS-1, CS-2) Back-up System #1, Back-up System #2 (BS-1, BS-2)
Common or usual name Deionization system with pre & post treatment and water distribution components.
Classification name Water purification systems for hemodialysis (21CFR 876.5665)
Predicate Device
The Flier's Quality Water Systems, Inc. systems are substantially equivalent to USFilter Corporation's/Siemens (K980182) predicate marketed water treatment systems for dialysis which use carbon and deionization canisters with pre and post filtration to purify water for hemodialysis.
Device Description
The systems are primary or temporary devices used to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001.
The Flier's Quality Water Systems, Inc. systems purify potable feed water through deionization. Deionization is used to remove 99.9% of ions from water. Deionization alone does not remove particulates, organics, bacteria, viruses or endotoxins. The systems require pretreatment and post-treatment. Deionizers are designed in a hemodialysis application to treat RO water or carbon filtered potable water. Improper use can result in the formation of nitrosamines in the effluent of the deionizer.
{1}------------------------------------------------
The purpose of the pretreatment section of the system is to condition that feed water supplying the deionizers. Conditioning the feed water will include: cartridge filters to reduce particulates and sediment and carbon filtration tarks to remove chlorine/chloramine residual.
The purpose of the post-treatment section of the system is to remove bacteria and endotoxins or lower them to acceptable levels as required by AAMI standards. The post-treatment section of the system will include: submicron/ultrafilters after the deionizers, sample ports to check for microbial contamination and to test for comprehensive AAMI analysis, a 1 megohm quality control light at mid tank and a temperature compensated audible and visual resistivity alarm set at 1M ohm or greater for final water quality. A remote alarm will be installed if the deionization system is not located in the patient treatment area. A distribution section of the system is necessary to deliver product water that meets ANSI/AAMI standards to the points of use. During use, continual monitoring of the effluent through the use of the temperature compensated audible and visual resistivity alarm is required. Routine monitoring of the mid tank quality indicator is also required on intervals during the day, at least before and after each patient treatment.
Should final water quality fall at or below 1 megohm, the alarm will sound and all patient treatments must be discontinued. The deionization tanks must be replaced per tank exchange procedure prior to resuming and patient treatment.
Divert to drain option will be provided when necessary per AAMI quidelines. Divert to drain is not supplied for acute systems located in patient room. Monitor and divert to drain not supplied for back-up systems with existing monitor/alarm/divert.
Indications for Use
The Acute System #1 consists of two each carbon and mixed bed DI tanks with pre & post filtration. It is intended to be placed on the back of dialysis machine or on a handcart for single patient dialysis. Acute System #2 consists of just carbon tanks and is intended to be used as pretreatment to a single patient RO system.
The Central System #1 and #2 consists of two or more carbon and mixed bed DI tanks with pre & post filtration. They are used as either a temporary system to feed multiple dialysis machines in normal use, or for a system to feed a biomedical technician service station. For permanent, multiple station dialysis clinics Flier's recommends using central water treatment systems including reverse osmosis.
The Back-up System #1 and #2 consists of two or more mixed bed DI tanks as a component in a "back up" condition when a RO system or major component on a water purification system for hemodialysis is in need of repair. It may also be used in DI polishing mode to remove ions from an RO stream to meet AAMI RD62:2001 requirements.
Conclusion
In both the case of Flier's systems and the predicate device carbon filtration is utilized to filter out chlorine and chloramines from the water. Both companies use two carbon filters in a series configuration. A minimum empty bed contact time of 10 minutes is incorporated into the designs as recommended by the FDA for chlorine and chloramine removal. Both companies utilize an
{2}------------------------------------------------
KC711C4
PAGE 3 OF 3
activated carbon with an lodine number of 900 or greater. Flier's always recommends installing dual carbon tanks in series including single patient systems (Acute System #1, Acute System #2).
In both cases deionization tanks are utilized to remove dissolved solids from the water. Both utilize mixed bed resin, consisting of anion and cation resin, to remove the charged particles from the water. Both utilize parts and materials which are NSF and/or FDA approved.
In summary, the components of Flier's systems and those of the US Filter/Siemens predicate device are very similar to one another. The core water purification components and technology are exactly the same. The predicate device components utilize the exact same water purification principles.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three lines extending from the eagle's body.
Public Health Service
JAN 1 4 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. James Flier Quality Committee Flier's Quality Water Systems, Inc. 7425 Clyde Park Ave., S.W., Suite A BYRON CENTER MI 49315
K071104 Re:
Trade/Device Name: Acute Portable Exchange Deionization (PEDI) System Regulation Number: 21 CFR §876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: December 26, 2007 Received: December 26, 2007
Dear Mr. Flier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Flier's
Quality
Water
Systems
inc
Indications for Use
510(k) Number: K071104
Device Name:
Acute System #1, Acute System #2 (AS-1, AS-2) Central System #1, Central System #2 (CS-1, CS-2) Back-up System #1, Back-up System #2 (BS-1, BS-2)
Indications for Use:
Flier's Quality Water Systems, Inc. systems are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001. They remove organic and inorganic substances and microbial contaminants from water used to dilute dialysis concentrate to form dialysate, in the reprocessing of hemodialyzers, and equipment rinse and disinfection.
The Acute System #1 consists of two each carbon and mixed bed DI tanks with pre & post filtration. It is intended to be placed on the back of dialysis machine or on a handcart for single patient dialysis. Acute System #2 consists of just carbon tanks and is intended to be used as pretreatment to a single patient RO system.
The Central System #1 and #2 consists of two or more carbon and mixed bed D1 tanks with pre & post filtration. They are used as either a temporary system to feed multiple dialysis machines in normal use, or for a system to feed a biomedical technician service station. For permanent, multiple station dialysis clinics Flier's recommends using central water treatment systems including reverse osmosis.
The Back-up System #1 and #2 consists of two or more mixed bed DI tanks as a component in a "back up" condition when a RO system or major component on a water purification system for hemodialysis is in need of repair. It may also be used in DI polishing mode to remove ions from an RO stream to meet AAMI RD62:2001 requirements.
Upon exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or with new resin altogether.
These systems are components of a larger water treatment system employing adequate pre-treatment and post-treatment sections. Flier's systems are not to be used alone.
| Prescription Use(Part 21 CFR 801 Subpart D) | X |
|---|---|
| AND/OR | |
| Over-The-Counter Use(21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Occurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K071104 |
|---|---|
| --------------- | --------- |
§ 876.5665 Water purification system for hemodialysis.
(a)
Identification. A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.(b)
Classification. Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.