K Number
K071000
Device Name
MYRIAN
Manufacturer
Date Cleared
2007-05-14

(35 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Myrian is a multi modality medical diagnostic device. It is aimed at reviewing and analysing anatomy and pathology. It also includes DICOM communication capabilities and media interchange features (printing, CD burning, storing). It runs on any standard PC including laptops that might be purchased independently by the end user. It provides user a set of tools meant to create and modify volumes of interest. This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Device Description
Myrian® system is a software suite providing the following services : Import of DICOM images from any DICOM modality, workstation or PACS Visualization of DICOM images in thin MPR, thick MPR and full 3D volume rendering Creation of VOI (Volume Of Interest) with dedicated tools Calculation of volumes, surface and of average, minimum and maximum densities of VOI Follow-up of patient examination Generation of medical reports Export of DICOM images to any format, DICOM entity or media
More Information

No
The document describes standard image processing and visualization tools, and explicitly states "Not Found" for mentions of AI, DNN, or ML.

No
The device is described as a medical diagnostic device, aimed at reviewing and analyzing anatomy and pathology, and providing tools for viewing and quantifying images, rather than for treating a condition.

Yes

The "Intended Use / Indications for Use" states, "Myrian is a multi modality medical diagnostic device."

Yes

The device description explicitly states "Myrian® system is a software suite providing the following services". While it runs on standard PCs, the device itself is presented as the software, not the hardware it runs on.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device's intended use and description: The description clearly states that Myrian is a "multi modality medical diagnostic device" aimed at "reviewing and analysing anatomy and pathology" using "DICOM images from any DICOM modality". It focuses on visualizing and analyzing medical images (like CT, MR, US, etc.) which are generated in vivo (within the living body).
  • Lack of mention of biological samples: There is no mention of the device processing or analyzing biological samples.

Therefore, Myrian falls under the category of medical imaging software or a medical image analysis device, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

Myrian® with its modules is a software application that is used for reviewing medical images. Myrian® with its modules is designed to run on standard PC hardware. The hardware is all "off-the-shelf' standard computer components and may be purchased independently by the end user. Myrian with its modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF modalities, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources like PACS).
Myrian® and its Modules can be used to communicate, to process, print and display medical images. Users have access to various image processing, Volume of Interest editing and measurement tools to assist them in viewing and quantifying images.
Myrian® and its Modules can be integrated with an institution's existing HIS or RIS for a fully integrated electronic patient record.
When interpreted by a trained physician, filmed or displayed images on the Myrian® and its Modules may be used as a basis for diagnosis, except in the case of mammography images.
Myrian® with its Modules does not support the display of mammography images for diagnosis.

Myrian is a multi modality medical diagnostic device. It is aimed at reviewing and analysing anatomy and pathology. It also includes DICOM communication capabilities and media interchange features (printing, CD burning, storing). It runs on any standard PC including laptops that might be purchased independently by the end user. It provides user a set of tools meant to create and modify volumes of interest.
This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.

Product codes

LLZ

Device Description

Myrian® system is a software suite providing the following services :
Import of DICOM images from any DICOM modality, workstation or PACS
Visualization of DICOM images in thin MPR, thick MPR and full 3D volume rendering
Creation of VOI (Volume Of Interest) with dedicated tools
Calculation of volumes, surface and of average, minimum and maximum densities of VOI
Follow-up of patient examination
Generation of medical reports
Export of DICOM images to any format, DICOM entity or media

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

CT, MR, US, RF modalities, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources like PACS.

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Typical users of Myrian® with its Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance data were verified versus the requirements of the FDA "Guidance of the Content of Pre Market Submissions for Software Contained in Medical Devices"
User Site Testing, Benchmarking and clinical data analysis demonstrate that MYRIAN meet the required specifications. No adverse affects have been detected.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K052995, K061624

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

Intrasense PREMARKET NOTIFICATION 510(K) SUBMISSION

MYRIAN
060830

.

5. 510(K) SUMMARY
[As Required by 21 CFR 807.92]
Summary of Safety and EffectivenessMAY 14 2007
Preparation dateApril 6th, 2007
Submitter
NameINTRASENSE
Registration numberTBD
AddressCAP OMEGA, Rond Point Benjamin Franklin CS 39521 34960 Montpellier FR
Tel Fax(+33) 467 130 130 / (+33) 467 130 132
Contact PersonMr Stéphane CHEMOUNY
Phone number : (+33) 467 130 130
Fax number : (+33) 467 130 132
Device name
Common NameSystem, Image Processing
Trade NameMYRIAN
Model numberN/A
Device classification
Classification nameSystem, Image Processing, Radiological
Code productLLZ
Panel892
Regulation number892.=2050
Regulatory classII
Predicate devices[K052995] Cleared [November 8, 2005] [General Electric Medical Systems]
[Advantage Workstation Version 4.3], manufactured by [General Electric Medical
Systems]
[K061624] Cleared [June 27, 2006] [Vital Images, Inc.] [Vitrea2 Version 3.9],
manufactured by [Vital Images, Inc.]
DescriptionMyrian® system is a software suite providing the following services :
Import of DICOM images from any DICOM modality, workstation or PACS
Visualization of DICOM images in thin MPR, thick MPR and full 3D volume rendering
Creation of VOI (Volume Of Interest) with dedicated tools
Calculation of volumes, surface and of average, minimum and maximum densities of
VOI
Follow-up of patient examination
Generation of medical reports
Export of DICOM images to any format, DICOM entity or media
Explanation of how the device operatesMyrian® with its modules is designed to run on standard PC hardware. The hardware
is all "off-the-shelf' standard computer components and may be purchased
independently by the end user.
Intended useMyrian® with its modules is a software application that is used for reviewing medical
images. Myrian® with its modules is designed to run on standard PC hardware. The
hardware is all "off-the-shelf' standard computer components and may be purchased
independently by the end user. Myrian with its modules receives digital images and
data from various sources (including but not limited to CT, MR, US, RF modalities,
computed and direct radiographic devices, secondary capture devices, scanners,
imaging gateways or imaging sources like PACS).
Myrian® and its Modules can be used to communicate, to process, print and display
medical images. Users have access to various image processing, Volume of Interest
editing and measurement tools to assist them in viewing and quantifying images.
Myrian® and its Modules can be integrated with an institution's existing HIS or RIS for
a fully integrated electronic patient record.
Typical users of Myrian® with its Modules are trained medical professionals, including
but not limited to radiologists, technologists and clinicians.
When interpreted by a trained physician, filmed or displayed images on the Myrian®
and its Modules may be used as a basis for diagnosis, except in the case of
mammography images.
Myrian® with its Modules does not support the display of mammography images for
diagnosis.
Performance dataPerformance data were verified versus the requirements of the FDA "Guidance of the
Content of Pre Market Submissions for Software Contained in Medical Devices"
User Site Testing, Benchmarking and clinical data analysis demonstrate that MYRIAN
meet the required specifications. No adverse affects have been detected.
Substantial equivalence summaryThe technological characteristics, features, specifications, materials, mode of
operation, and intended use of MYRIAN device are equivalent to those of the
predicate devices quoted above.
MYRIAN is the same as the predicate devices in K061624 and K052995
The differences that exist between the devices do not raise new issues of safety or
effectiveness regarding MYRIAN Device.
It is substantially equivalent in terms of safety and effectiveness to the predicate
devices.

1

Intrasense PREMARKET NOTIFICATION 510(K) SUBMISSION 060830

2

Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

INTRASENSE % Mr. Nicolas Clary Project Leader CEISO "Les Portes d'Espagne" - Bât. B - 99, route d'Espagne 31100 Toulouse 31100 FRANCE

MAY 1 4 2007

Re: K071000

Trade/Device Name: Myrian Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 6, 2007 Received: April 13, 2007

Dear Mr. Clary:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a circular logo for the FDA Centennial, celebrating 1906-2006. The logo features the letters "FDA" in a stylized font, with the word "Centennial" written below. The logo is surrounded by a circular border with text and stars.

moting Public J

3

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I toase or acressed a determination that your device complies with other requirements of the Act that I Dederal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally premainer nedicated device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you use office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21 CFF at 80 ! Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

PREMARKET NOTIFICATION 510(K) SUBMISSION MYRIAN Intrasense 060830

4. INDICATIONS FOR USE
510(k) Number (if known):K07/000
Device Name:Myrian
Indications for Use:Myrian is a multi modality medical diagnostic device. It is aimed at reviewing and analysing anatomy and pathology. It also includes DICOM communication capabilities and media interchange features (printing, CD burning, storing). It runs on any standard PC including laptops that might be purchased independently by the end user. It provides user a set of tools meant to create and modify volumes of interest.
This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Prescription Use (Part 21 CFR 801 Subpart D)1.1 AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

.

.

Nancy C Hogdon

(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 071000 510(k) Number ________________________________________________________________________________________________________________________________________________________________