(44 days)
Not Found
No
The document describes a microprocessor-based system using standard oscillometric, temperature, and pulse oximetry methods. There is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies.
No.
The device is used for measuring and displaying vital signs (blood pressure, heart rate, temperature, SpO2) and is described as a "measurement system," indicating it is a diagnostic or monitoring device, not a therapeutic one.
Yes
The device is indicated for "measuring and displaying Systolic and Diastolic blood pressures, heart rate, temperature, and functional saturation of arterial hemoglobin (SpO2)", which are all measurements used to aid in diagnosis.
No
The device description explicitly mentions hardware components such as a microprocessor, electric pump, valves, and optical finger sensor, indicating it is not a software-only device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The SunTech Medical 247 device measures vital signs (blood pressure, temperature, and SpO2) directly from the patient's body using non-invasive methods (cuff, probes, finger sensor). It does not analyze samples taken from the body.
- Intended Use: The intended use describes measuring and displaying physiological parameters of patients, not analyzing biological samples.
Therefore, the device falls under the category of a non-invasive patient monitoring device, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The SunTech Medical 247 NIBP , Temperature, and Pulse Oximeter device is indicated for use in measuring and displaying Systolic and Diastolic blood pressures, heart rate, temperature, and functional saturation of arterial hemoglobin (SpO2) of adult and pediatric patients in hospitals, medical facilities, clinics, physicians offices, and other subacute environments.
Product codes
DXN, MUD, FLL
Device Description
The 247 BP, Temperature and Pulse Oximeter device, a microprocessor based noninvasive blood pressure, predictive temperature and oxygen saturation measurement system, uses oscillometric signals to determine blood pressure, oral, rectal or auxiliary probes to determine temperature and an optical finger sensor for oxygen saturation. An internal electric pump is used to inflate the cuff. Deflation of the cuff is controlled by two valves. Implementation of Pulse Oximetry and temperature is achieved through the incorporation of OEM devices from companies with currently marketed products and accessories.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult and pediatric patients
Intended User / Care Setting
hospitals, medical facilities, clinics, physicians offices, and other subacute environments.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The 247 has been tested to the applicable requirements of the following standards and requirements documents. These tests have indicated passing results.
- AAMI SP10: 2002 .
- IEC 60601-1:1996 .
- IEC 60601-1-2:2001 ●
- IEC 60601-1-4:2000 .
- ISO 9919:1992 .
- E1112-00:2006 .
- Functional Specification, (SunTech document # 99-0049-XX-FS) .
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo for SunTech Medical. The logo consists of a stylized sun-like graphic to the left of the text "SunTech Medical". A registered trademark symbol is located to the right of the word "Medical".
K07075D
SunTech Medical, Inc. Abbreviated 510(k) Submission 247 510(k) Summary March 16, 2007
Page :
MAY - 2 2007
(1) Submitter information
Name: | SunTech Medical, Inc |
---|---|
Address: | 507 Airport Boulevard |
Suite 117 | |
Morrisville, North Carolina 27560-8200 | |
Telephone: | 919.654.2332 |
FAX: | 919.654.2301 |
Contact person: | David Gallick (Official Correspondent). |
Date prepared: | March 16, 2007 |
(2) Name of Device
Trade Name: | 247 |
---|---|
Common Name: | NIBP Monitor |
Classification name: | Noninvasive blood pressure measurement system, 870.1130 |
(3) Legally-marketed predicate devices
Sun Tech Medical has identified Welch Allyn's Spot Vital Signs, K002530, as the predicate device for the 247.
The 247 is substantially equivilent to this device.
1
(4) Description
The 247 BP, Temperature and Pulse Oximeter device, a microprocessor based noninvasive blood pressure, predictive temperature and oxygen saturation measurement system, uses oscillometric signals to determine blood pressure, oral, rectal or auxiliary probes to determine temperature and an optical finger sensor for oxygen saturation. An internal electric pump is used to inflate the cuff. Deflation of the cuff is controlled by two valves. Implementation of Pulse Oximetry and temperature is achieved through the incorporation of OEM devices from companies with currently marketed products and accessories.
(5) Intended Use
The SunTech Medical 247 NIBP , Temperature, and Pulse Oximeter device is indicated for use in measuring and displaying Systolic and Diastolic blood pressures, heart rate, temperature, and functional saturation of arterial hemoglobin (SpO2) of adult and pediatric patients in hospitals, medical facilities, clinics, physicians offices, and other subacute environments.
(6) Comparison to Predicate Devices
The 247 has the same basic construction as the predicate device. Both devices are microprocessor controlled and the devices utilize similar circuitry. Both devices measure NIBP, SpO2, and body temperature. Neither device is intended to monitor the patient's vital signs. Both devices utilize the same measurement ranges for the various vital signs parameters being measured. Both devices are intended for adult and pediatric populations.
(7) Testing and Validations
The 247 has been tested to the applicable requirements of the following standards and requirements documents. These tests have indicated passing results.
- AAMI SP10: 2002 .
- IEC 60601-1:1996 .
- IEC 60601-1-2:2001 ●
- IEC 60601-1-4:2000 .
- ISO 9919:1992 .
- E1112-00:2006 .
- Functional Specification, (SunTech document # 99-0049-XX-FS) .
(8) Conclusion
The 247 is equivalent in safety and efficacy to the legally-marketed predicate device.
2
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 2 2007
SunTech Medical, Inc. c/o Mr. Chuck Setzer Regulatory Manager 507 Airport Blvd., Suite 117 Morrisville, NC 27560-8200
Re: K070750
Trade Name: 247 BP Device Regulation Number: 21 CFR 870.1130, 21 CFR 870.2700, and 21 CFR 880.2910 Regulation Name: Noninvasive Blood Pressure Measurement System; Oximeter; and, Clinical Electronic Thermometer Regulatory Class: Class II (two) Product Code: DXN, MUD, and FLL Dated: March 16, 2007 Received: March 19, 2007
Dear Mr. Setzer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 – Mr. Chuck Setzer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmumor for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Indications for use
K070750 510(k) Number (if known):
Device Name: 247
Indications for Use:
The SunTech Medical 247 NIBP , Temperature, and Pulse Oximeter device is indicated for use in measuring and displaying Systolic and Diastolic blood pressures, heart rate, temperature, and functional saturation of arterial hemoglobin (SpO2) of adult and pediatric patients in hospitals, medical facilities, clinics, physicians offices, and other subacute environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummenta
(Division Sign Off)
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number
Page_1_of_1