K Number
K070196
Date Cleared
2007-01-26

(4 days)

Product Code
Regulation Number
872.6640
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Flight Dental Systems A-Series Operative Units are intended to supply power und serve as a base for dental devices and accessories. The use intended is in the treatment of dental patients in the dental clinic/office environment. The units are for use only by trained dentists, dental hygienists, dental technicians and dental assistants.

Device Description

The Flight A-Series Dental Operative Unit is a dental operating system. The device includes a patient chair, dentist's element, utility center and a floor box. The unit may also include a cuspidor, an assistant's element and a dental operating light. The doctor's element may be mounted on the patient chair or on a mobile cart. The unit is used to position the patient in a comfortable position and to provide the power to the dentist's instruments including dental handpieces.

AI/ML Overview

This document describes the Flight Dental Systems A-Series Dental Operative Unit, cleared under K070196. This device is a dental operating system comprising a patient chair, dentist's element, utility center, and floor box, intended to supply power and serve as a base for dental devices and accessories in a dental clinic/office environment.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

The provided 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (Sirona C8+ Dental Operative unit - K983242) rather than defining explicit, quantitative acceptance criteria for de novo performance. The acceptance criteria are implicitly met by demonstrating conformance to relevant standards and direct equivalence to the predicate.

Acceptance Criteria (Implied by Substantial Equivalence and Standards Conformance)Reported Device Performance
Safety: Device electrical safety, mechanical integrity, and biocompatibility.Bench testing against various standards:
- Electrical Safety: Comply with IEC 60601-1 / UL 60601-1 and CAN/CSA-C22.2 No. 601.1 M90 for electrical medical equipment.- Independent laboratory testing by Intertek Testing confirmed conformity with UL 60601-1 and CAN/CSA-C22.2 No. 601.1 M90.
- Mechanical Safety/Functional Standards: Comply with ISO 7494-1:1996 (Dentistry - Dental units - Part 1: General requirements and test methods) and ISO 7494-2:2003 (Dentistry - Dental units - Part 2: Water and air supply).- Bench testing against local GB standards (GB9706.1-1995 and Y Y/T1043-2004) which are near equivalent to ISO 7494-2:2003 and ISO 7494-1:1996. Additional testing performed where local standards were not equivalent.
- Compatibility with Handpieces: Comply with ISO 9168:1991 (Dental handpieces - Coupling dimensions).- Tested against ISO 9168:1991 and found to be in compliance. Device uses industry-standard 4-Hole handpiece tubings.
- Biocompatibility (Patient/Operator Contact Surfaces): Comply with ISO 7405:1997 (Dentistry - Evaluation of biocompatibility of medical and dental devices).- Patient and operator contact surfaces tested and found to be in compliance with ISO 7405:1997. Device composed of materials with a long history of use.
Effectiveness/Performance: Ability to supply power and act as a base for dental devices and accessories, and to position the patient.- Device description and comparison table assert "Equivalent" performance in all functional aspects (control of air and water, installation, components, accessories, power/utility supply, water system, cleaning system, activation, place of use, intended users) to the predicate Sirona C8.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: The submission states that "Bench testing has been performed on a sample that is in all respects the same as the device to be marketed." This indicates a single device sample (or a very small number representing the manufactured article) was used for physical bench testing.
  • Data Provenance: The bench testing was performed against local GB standards in China. Additional independent laboratory testing was performed by Intertek Testing of Mississauga, Ontario, Canada. This is prospective testing of the device itself.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This submission does not involve clinical data or expert evaluations for establishing ground truth in the context of image interpretation or diagnostic performance. Instead, ground truth is established by the compliance of the device with recognized international and national standards as assessed by engineering and testing professionals at the designated testing laboratories. Therefore, the "experts" are the qualified personnel at the testing facilities (e.g., Intertek Testing) who conducted the specified tests and verified compliance. Their specific qualifications are not detailed beyond their affiliation with the testing labs.

4. Adjudication Method for the Test Set:

  • Not applicable. The testing described is objective bench testing against engineering standards, not subjective evaluation requiring adjudication among multiple human assessments.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices or AI-assisted devices that impact human clinical decision-making. This device is a physical dental operative unit.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Not applicable. This device is a physical medical device, not an algorithm or software-only device. Its performance is inherent in its physical and electrical characteristics conforming to safety and performance standards.

7. Type of Ground Truth Used:

  • The "ground truth" used for this submission is conformance to established engineering and safety standards (e.g., ISO, UL, CSA, GB standards), combined with direct comparison of technological characteristics to a legally marketed predicate device.

8. Sample Size for the Training Set:

  • Not applicable. This device does not involve machine learning or AI, and therefore does not have a "training set."

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, as there is no training set.

{0}------------------------------------------------

K070196

510(K) SUMMARY 5.0

The 510(k) Summary follows.

510(k) SUMMARY

JAN 26 2007

Manufacturer's Name, Contact Person, Address, Phone, Fax, Email, Date Prepared

Mfr Name:HR Dental Products Inc.
Contact Name:Jim Ronalds
Postal Address.21 Kenview Blvd, Unit 9Brampton, ON CanadaL6T 5G7
Phone:905-799-0517
Fax:905-793-2406
Email:jimronalds@rogers.com

Date Prepared: September 14, 2006, 2006

Device Name, Common/Generic Name and Classification Name

Flight Dental Systems A-Series Dental Operative Unit Proprietary Name:

Common/Usual Name: Dental operative unit

Classification Name / Code: Unit, Operative Dental / ElA

Predicate Device

Sirona C8+ Dental Operative unit - K983242

Description of the Device

The Flight A-Series Dental Operative Unit is a dental operating system. The device includes a patient chair, dentist's element, utility center and a floor box. The unit may also include a cuspidor, an assistant's element and a dental operating light. The doctor's element may be mounted on the patient chair or on a mobile cart.

The unit is used to position the patient in a comfortable position and to provide the power to the dentist's instruments including dental handpieces.

Intended Use of the Device

The Flight Dental Systems A-Series Operative Units are intended to supply power and serve as a base for dental devices and accessories. The use intended is in the treatment of dental patients in

Traditional 510(k) Submission

{1}------------------------------------------------

the dental clinic/office environment. The units are for use only by trained dentists, dental hygienists, dental technicians and dental assistants.

Technological Characteristics

The general components, design, characteristics and mode of operation of the Flight A-Series Dental Operative Unit is substantially equivalent to Sirona C8 Dental Operative Unit. Both are AC powered operative units with accessories that are intended to supply power to and serve as a base for dental devices and accessories. Both systems include the doctor's element, patient chair, floor box and utility center and may include optional cuspidors, operating lights and assistants elements.

Performance Data

Bench testing has been performed on a sample that is in all respects the same as the device to be marketed. Testing was performed against local GB standards in China including GB9706.1-1995 and Y Y/T1043-2004 which are near equivalent to the recognized consensus standards ISO 7494-2:2003 and ISO 7494-1:1996. In any case in which the local standard was not equivalent to the recognized Consensus standard, additional testing has been performed in accordance with the recognized consensus standards.

Further independent laboratory testing was performed by Intertek Testing of Mississauga, Ontario Canada to confirm conformity UI. 60601-1 and CAN/CSA-C22.2 No. 601.58aagage, with the electrical safety portions of ISO 7494-1:1996. In any case in which the local standard was not equivalent to the recognized Consensus standard, additional testing has been performed in accordance with the recognized consensus standard.

The equipment has also been tested against ISO 9168:1991, and found to be in compliance with the standard. The patient and operator contact surfaces have been tested and found to be in compliance with ISO 7405:1997. The device is composed of materials that have a long instory of use in the medical and dental community.

No deviations from the recognized standards were found except those noted in the table provided in section 9 of this submission. In any case of deviation from the recognized standard alternate methods of meeting safety and effectiveness requirements were implemented.

The conclusions drawn from the bench testing are that the device is as safe and as effective as the predicate device. Furthermore, the device performs its intended tasks as well as the legally marketed predicate device and complies with the recognized standards.

Substantial Equivalence Conclusions

In all important respects the Flight dental Systems A-Series Dental Operative Units are substantially equivalent to the Sirona Dental Systems C8 (K983242). This conclusion is ased upon indications for use, technical characteristics, device users and features comparison. Any

{2}------------------------------------------------

differences in the technological characteristics do not raise any new safety and effectiveness issues.

CharacteristicSirona C8 - K983242Flight Dental Systems A-Series
Intended UseThe Sirona C8 is an electronicallycontrolled dental operative unit withaccessories that are intended to supplypower to and serve as a base for dentaldevices and accessoriesEquivalent
Control of Air andWaterUses pneumatically controlled vales tocontrol the flow of air and water. On/offand intensity controlled by foot pedalEquivalent
InstallationAvailable with chair mounted and cartmounted dentist element. Installed bytrained technicians.Equivalent.
ComponentsDoctors element, assistant's element,flex arm, utility center with optionalcuspidor, patient chair, treatment light,floor box.Equivalent
Optional accessoriesCuspidor, Assistant element with vaccumvalves, treatment light, foot switches,monitor mountsEquivalent
Power and UtilitySupply110V/220V AC electrical supply,compressed air and waterEquivalent
CompatibilityCompatible with industry standard fittingsfor Air/water syringe tips, high speed andlow speed pneumatic turbines, air scaler,air polisher, High Volume Evacuator tips.Use industry standard 4-Hole handpiecetubings.Equivalent
Electrical SafetyStandardsNot SpecifiedUL 60601-1CAN/CSA-C22.2 No. 601.1 M90See Appendix H
Water SystemUser may select Self Contained watersystem or city water supplyEquivalent
Cleaning SystemWater FlushEquivalent -- See Appendix E
ActivationMaster on/off, toggle switch and footcontrolEquivalent
Place of UseDental office, hospitalEquivalent
Intended UsersDentists, Dental Hygienists, DentalAssistantsEquivalent

Traditional 510(k) Submission

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized eagle or bird symbol, with three curved lines representing the wings or feathers. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 6 2007

HR Dental Products, Incorporated C/O Mr. Neil E. Devine Responsible Third Party Official Intertek Testing Services NA. Incorporated 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087

Re: K070196

Trade/Device Name: Flight Dental Systems A-Series Dental Operative Units Regulation Number: 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: I Product Code: EIA Dated: January 19, 2007 Received: January 22, 2007

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Devine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation. entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number:

Device Name: Flight Dental Systems A-Series Dental Operative Units

Indications for Use:

Statement of Indication for Use

The Flight Dental Systems A-Series Operative Units are intended to supply power und serve as a base for dental devices and accessories. The use intended is in the treatment of dental patients in the dental clinic/office environment. The units are for use only by trained dentists, dental hygienists, dental technicians and dental assistants.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODF)

Keii Muluy for MSR

(Division Sign-Off) Division of Anesther Nogy, General Hospital, Infection Control, Dental Devices

510(k) Number: K070196

Page 1 of 1

§ 872.6640 Dental operative unit and accessories.

(a)
Identification. A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.(b)
Classification. Class I (general controls). Except for dental operative unit, accessories are exempt from premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.