(43 days)
The subject devices are sterile, single use devices. They are intended for use during cemented or cementless primary or revision hip arthroplasty and trochanteric osteotomy. The Stainless Steel alloy Grip and Grip Plates are intended for use with 2.0mm Stainless Steel beaded and nonbeaded Cables only. Wrought CoCr alloy Grip and Grip Plates are intended for use with 2.0mm CoCr beaded and non-beaded Cables only. Optional Stainless Steel and Titanium screws are also available for use with the Stainless Steel and Wrought CoCr Grip Plates, respectively.
- The Dall-Miles® System is indicated for reattachment of the trochanter in any hip . procedure using the trochanteric osteotomy (total or partial) approach.
- The Dall-Miles® Trochanteric Grips and Grip Plates are indicated for use in the . fixation of the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
- The Dall-Miles® Trochanteric Grip Plate is additionally indicated for use in the . fixation of the greater trochanter due to extended trochanteric osteotomies.
The Dall-Miles® product line will now be available with Trochanteric Grip Plates and Trochanteric Grips that facilitate reattachment and fixation of the Greater Trochanter. The Trochanteric Grips will be offered in small, medium, and large sizes. The Trochanteric Grip Plates will be offered in medium sizes ranging from 100-200mm and large sizes ranging from 110-210mm in length. Both the Trochanteric Grips and Grip Plates are compatible with the existing components of the Dall-Miles® System and will be available in Stainless Steel or CoCr alloy materials and packaging options with or without Cables.
The provided text is a 510(k) summary for the Dall-Miles® System, which is a bone fixation accessory. This type of document is a premarket notification to the FDA for medical devices and focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical study data for acceptance criteria.
Therefore, the information requested in your prompt regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment is not available in the provided document. The document primarily focuses on:
- Device Description: What the Dall-Miles® System is and its components (Trochanteric Grip Plates, Trochanteric Grips).
- Intended Use and Indications for Use: How the device is meant to be used and for what specific conditions.
- Substantial Equivalence: A claim that the device is similar enough to previously cleared devices (predicate devices) that it does not require a new Preamarket Approval (PMA) application. The predicate devices mentioned are the Modified BMP™ Trochanteric Plates (K993510) and the Dall-Miles® Trochanter Cable Super Grip (K962162).
- FDA Clearance Letter: Confirmation from the FDA that the device has been cleared for marketing based on the substantial equivalence claim.
In summary, this document does not contain the information required to populate the tables and answer the questions about acceptance criteria and study data because it pertains to a 510(k) submission for a traditional medical device (bone fixation accessories), not an AI/ML-driven diagnostic device that would typically involve the detailed performance metrics and study designs you've inquired about.
{0}------------------------------------------------
MAR 0 2 2007
510(k) Summary of Safety and Effectiveness for the Dall-Miles® System
| Proprietary Name: | Dall-Miles® System |
|---|---|
| Common Name: | Bone Fixation Accessories |
| Classification Name and Reference | Single/multiple component metallic bone fixationappliances and accessoriesTitle 21 CFR §888.3030 |
| Bone fixation cerclageTitle 21 CFR § 888.3010 | |
| Regulatory Class: | Class II |
| Device Product Code(s): | HRS - Plate, fixation, bone |
| LYT -- Fixation accessory | |
| JDQ - Cerclage, fixation | |
| For Information Contact: | Patricia Setti-LaPerchRegulatory Affairs AssociateStryker Orthopaedics325 Corporate DriveMahwah, New Jersey 07430Phone: (201) 831-5938Fax: (201) 831-4938E-Mail: Patricia.LaPerch@stryker.com |
| Date Summary Prepared: | January 16, 2007 |
Device Description
The Dall-Miles® product line will now be available with Trochanteric Grip Plates and Trochanteric Grips that facilitate reattachment and fixation of the Greater Trochanter. The Trochanteric Grips will be offered in small, medium, and large sizes. The Trochanteric Grip Plates will be offered in medium sizes ranging from 100-200mm and large sizes ranging from 110-210mm in length. Both the Trochanteric Grips and Grip Plates are compatible with the existing components of the Dall-Miles® System and will be available in Stainless Steel or CoCr alloy materials and packaging options with or without Cables.
{1}------------------------------------------------
Intended Use:
The subject devices are sterile, single use devices. They are intended for use during cemented or cementless primary or revision hip arthroplasty and trochanteric osteotomy. The Stainless Steel alloy Grip and Grip Plates are intended for use with 2.0mm Stainless Steel beaded and nonbeaded Cables only. Wrought CoCr alloy Grip and Grip Plates are intended for use with 2.0mm CoCr beaded and non-beaded Cables only. Optional Stainless Steel and Titanium screws are also available for use with the Stainless Steel and Wrought CoCr Grip Plates, respectively.
Indications for Use
- The Dall-Miles® System is indicated for reattachment of the trochanter in any hip . procedure using the trochanteric osteotomy (total or partial) approach.
- The Dall-Miles® Trochanteric Grips and Grip Plates are indicated for use in the . fixation of the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
- The Dall-Miles® Trochanteric Grip Plate is additionally indicated for use in the . fixation of the greater trochanter due to extended trochanteric osteotomies.
Substantial Equivalence:
The Dall-Miles® Trochanteric Grips and Grip Plates are substantially equivalent to the Modified BMP™ Trochanteric Plates, cleared by Biomet (K993510) and the Dall-Miles® Trochanter Cable Super Grip, cleared by Howmedica Osteonics (K962162).
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized emblem resembling an eagle or bird-like figure with flowing lines, positioned to the right. Encircling the emblem are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 0 2 2007
Stryker Orthopaedics % Ms. Patricia Setti-LaPerch Regulatory Affairs Associate 325 Corporate Drive Mahwah, New Jersey 07430
Re: K070170
Trade/Device Name: Dall-Miles System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Codes: HRS, LYT, JDQ Dated: January 16, 2007 Received: January 18, 2007
Dear Ms. Setti-LaPerch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Patricia Setti-LaPerch
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark McMullen
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): _K070170
Device Name: Dall-Miles® System
Indications for Use
- The Dall-Miles® System is indicated for reattachment of the trochanter in any hip . procedure using the trochanteric osteotomy (total or partial) approach.
- The Dall-Miles® Trochanteric Grips and Grip Plates are indicated for use in the . fixation of the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
- The Dall-Miles® Trochanteric Grip Plate is additionally indicated for use in the . fixation of the greater trochanter due to extended trochanteric osteotomies.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
Over-the-Counter Use OR (Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Milliken
Division of Gener 1, Restorative, and Neurological Devices
510(k) Number K07 0170
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.