(67 days)
K#042961,K#060867
K#061871
No
The description focuses on the imaging hardware, light source, dye, and the ability to capture, view, record, and replay fluorescent images. There is no mention of AI, ML, or advanced image processing beyond basic viewing and recording.
No
The device is described as an "imaging system" used for "visual assessment of blood flow," and states that it provides the "capability to view, record, and replay fluorescent images." It is specifically identified as an "adjunctive method for the evaluation of tissue perfusion," highlighting its role in providing information rather than directly treating a condition.
Yes
The SPY System is an imaging system used for the visual assessment of blood flow and evaluation of tissue perfusion to aid in surgical procedures and assess flap viability. This capability to assess physiological conditions for diagnosis or evaluation qualifies it as a diagnostic device.
No
The device description explicitly details hardware components including an imaging head with a CCD camera, laser light source, motion sensor, distance sensor, articulating arm, mobile cart, flat panel display, computer, electronics enclosure, and printer. It also includes physical components like sterile drapes and an imaging agent.
Based on the provided information, the SPY System is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: An IVD device is a medical device intended for use in vitro for the examination of specimens derived from the human body in order to provide information for diagnostic purposes. This typically involves testing samples like blood, urine, or tissue outside of the body.
- SPY System Function: The SPY System is an imaging system that uses fluorescence to visualize blood flow within the body (in vivo). It captures images of tissue perfusion and circulation directly on the patient during surgical procedures.
- Lack of Specimen Examination: The device does not examine specimens derived from the human body in a laboratory setting. It directly images the patient's tissues.
Therefore, the SPY System falls under the category of an in vivo imaging device, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The SPY System is an imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.
Product codes
90 IZI
Device Description
The Novadaq® Technologies SPY® Imaging System consists of 2 components:
- · the SPY Imaging Device; and
- = the SPY Paq™
The SPY Paq is comprised of:
- 6 Novadrape™ custom sterile drapes: and
- = 1 box of IC-Green™ (Indocvanine Green-ICG) imaging agent that contains 6 x 25 mg vials of IC-Green and 6 x 10 ml ampules of sterile Aqueous Solvent.
The IC-Green is supplied intact as received and labeled from the manufacturer of the product.
The SPY Imaging Device consists of an imaging head containing a charge coupled device (CCD) camera, a laser light source, motion sensor and distance sensor attached via an articulating arm to a mobile cart. The mobile cart contains a flat panel display, computer, electronics enclosure and printer.
The SPY System provides the surgeon with the capability to view record and replay fluorescent images of blood flow in vessels and bypass grafts of the heart. A laser light source is used to illuminate the heart surface. IC-Green™ (indocyanine green) is injected intravenously through the central venous line, bypass pump, cardioplegia line and aorto coronary graft and while it is passing through the vessels, the absorption of laser light causes excitation of the dye followed by emission of infrared energy. The result is a fluorescent image of the coronary vasculature. A CCD camera captures the image. These images are used to evaluate the integrity of the coronary vasculature and blood flow in the heart and bypass grafts.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Fluorescent images
Anatomical Site
Tissue and free flaps used in plastic, micro- and reconstructive surgical procedures (including the heart and coronary vasculature previously cleared).
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Surgeon, medical staff. Used intra-operatively and post-surgically.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Animal studies, human experience and in vitro testing were conducted to support the safe and effective use of the SPY System in its original premarket notification 510(k) application (K#042961).
In Vitro Testing:
Testing of the SPY System was completed in conformance with:
-
- Electrical per IEC 60601-1 and UL2601-1
-
- Electromagnetic Compatibility per IEC 60601-1-2
-
- Light Emitting Laser Products per 21 CFR 1040
-
- Safe Use of Lasers in Health Care Facilities per ANSI Z136.3
-
- American National Standard for Safe Use of Lasers per ANSI Z136.1
The SPY System successfully met all requirements for these standards.
- American National Standard for Safe Use of Lasers per ANSI Z136.1
In Vivo Testing:
The SPY System is commercially available in the United States of America, Japan, Europe and Canada.
Used in over 4000 CABG procedures in humans with no reports of adverse acute or long-term cellular, renal or hepatic effects.
SPY® images were useful in determining relative perfusion of the flap for harvest prior to incision, and in reconstructive surgery, the blood flow in the native vessels to provide insight into sufficient perfusion to support reconstruction and free flap inset.
Used intra-operatively to visually assess blood flow in native artery and vein related to the flap and post reconstruction to assess anastomotic integrity and tissue-transfer circulation.
Images provided demonstrate use of fluorescent images in assessing blood flow, post surgical closure, in a non-body contact manner to aid medical staff in assessing flap viability, along with other standard practices.
Results from the use of the SPY System has been the subject of 12 peer reviewed journal articles (10 related to cardiac surgery, 2 related to transplantation surgeries).
Literature reports that the SPY System was able to non-invasively, quickly and safely identify 17 conduits in 311 patients that required revision during surgical procedures. In all cases, lack of patency was visualized clearly by the SPY System using doses of IC-Green™ well below that approved for human use, allowing the surgeon to revise the graft. Cardiac, renal and hepatic function were monitored and no adverse effects reported.
Six pig studies were conducted to support the original 510(k):
-
- acquired high quality images simply and reproducibly using small doses of IC-Green.
-
- performed multiple imaging sequences with no detrimental effects on heart function, coronary flow or peripheral pressure.
-
- acquired images with no increase in myocardial tissue temperature.
-
- visualized all coronary beds with high quality images even when the heart was in a vertical position.
In vivo evidence findings:
- The exposure for the SPY® System is 35 mW/cm² which is far below the maximum permissible exposure of 327 mW/cm² established by ANSI for exposure to the skin.
- Use of the SPY System does not cause any thermal damage to the area of interest, even after repeated imaging sequences.
- For the heart, there were no changes in electrocardiograms or arterial pressures during and/or following SPY use.
- There were no acute or long-term cellular effects.
- There were no acute or long-term renal or hepatic effects.
- The SPY System was able to acquire high quality images of the entire vascular bed on each area of interest.
- The SPY System is capable of imaging through the skin to provide a visual assessment of dermal and subdermal blood flow.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
K#042961, K#060867
Reference Device(s)
K#061871
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1600 Angiographic x-ray system.
(a)
Identification. An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.
0
510(k) Summary
Trade Name: | SPY® Imaging System |
---|---|
Model Number: | SP2000 JAN 1 2 2007 |
Common Name: | Fluorescent Angiographic System |
Classification: | 21 CFR 892.1600 |
Product Code: | 90 IZI |
Classification: | Class II |
Manufacturer: | Novadaq Technologies Inc. |
2585 Skymark Avenue | |
Suite 306 | |
Mississauga, Ontario | |
Canada | |
L4W 4L5 | |
905.629.3822 ext. 240 | |
Contact Name: | Allison Manners |
Vice President – Regulatory and Clinical Affairs |
Date 510(k) Summary Prepared: November 2, 2006
Legally Marketed Predicate Devices:
The Novadaq® SPY Imaging System had received FDA 510(k) clearance for market in January 2005 (K#042961) and subsequently 510(k) clearance for a labeling change in May 2006 (K#060867).
The Leica FL800 had received FDA 510(k) clearance for market in September 2006 (K#061871). The Leica FL800 is intended for use to allow neurosurgeons to view blood flow.
Device Description:
The SPY Imaging System: SP2000 is currently cleared for intraoperative visual assessment of the coronary vasculature and grafts during coronary artery bypass graft (CABG) surgery.
1
The Novadaq® Technologies SPY® Imaging System consists of 2 components:
- · the SPY Imaging Device; and
- = the SPY Paq™
The SPY Paq is comprised of:
- 6 Novadrape™ custom sterile drapes: and
- = 1 box of IC-Green™ (Indocvanine Green-ICG) imaging agent that contains 6 x 25 mg vials of IC-Green and 6 x 10 ml ampules of sterile Aqueous Solvent.
The IC-Green is supplied intact as received and labeled from the manufacturer of the product.
The SPY Imaging Device
The SPY Imaging Device consists of an imaging head containing a charge coupled device (CCD) camera, a laser light source, motion sensor and distance sensor attached via an articulating arm to a mobile cart. The mobile cart contains a flat panel display, computer, electronics enclosure and printer.
The SPY System provides the surgeon with the capability to view record and replay fluorescent images of blood flow in vessels and bypass grafts of the heart. A laser light source is used to illuminate the heart surface. IC-Green™ (indocyanine green) is injected intravenously through the central venous line, bypass pump, cardioplegia line and aorto coronary graft and while it is passing through the vessels, the absorption of laser light causes excitation of the dye followed by emission of infrared energy. The result is a fluorescent image of the coronary vasculature. A CCD camera captures the image. These images are used to evaluate the integrity of the coronary vasculature and blood flow in the heart and bypass grafts.
There have been no significant changes or modifications made to the SPY Imaging device since the original 510(k) clearance in January 2005, premarket notification 510(k) K#042961, or for the 510(k) submitted for a label change for this device. K#060867.
This 510(k) submission describes a proposed change in intended use that does not alter the devices fundamental scientific technology or characteristics in anyway.
Proposed Intended Use of the SPY System:
Please note that the currently cleared indication for use in CABG surgery is not being amended in anyway with this submission.
2
This premarket notification 510(k) application is being made to add an additional indication for use for the SPY® System.
In addition to the already cleared indication for use of the SPY System, the SPY System is an imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.
Testing:
Animal studies, human experience and in vitro testing were conducted to support the safe and effective use of the SPY System in its original premarket notification 510(k) application (K#042961).
In Vitro Testing:
Testing of the SPY System was completed in conformance with the following standards. The SPY System successfully met all of the requirements for these standards.
-
- Electrical per IEC 60601-1 and UL2601-1
-
- Electromagnetic Compatibility per IEC 60601-1-2
-
- Light Emitting Laser Products per 21 CFR 1040
-
- Safe Use of Lasers in Health Care Facilities per ANSI Z136.3
-
- American National Standard for Safe Use of Lasers per ANSI Z136.1
In Vivo Testina:
The SPY System is commercially available in the United States of America, Japan, Europe and Canada. To date, the SPY System has been used in over 4000 CABG procedures in humans and there have been no reports of adverse acute or long-term cellular, renal or hepatic effects. Along with the data from intra-operative imaging in CABG surgery, the use of the SPY System in plastic, micro- or reconstructive surgery demonstrates the clinical utility of the device in producing high quality and resolution images of the entire vascular bed of the point of interest.
In the proposed indications for use in this application, the SPY® images were useful in determining not only the relative perfusion of the flap for harvest prior to incision, but also in reconstructive surgery, the blood flow in the native vessels to provide insight as to whether there is sufficient perfusion to support the reconstruction and the inset of a free flap. The SPY System has also been used intra-operatively to visually assess blood flow in the native artery and vein related to the flap and post reconstruction to assess anastomotic integrity and tissue-transfer circulation. The images provided in Section 19 - Clinical, of this
3
application demonstrate the use of the fluorescent images in assessing blood flow, post surgical closure, in a non-body contact manner to aid medical staff in assessing flap viability post surgery, along with other standard practices, such as evaluation of flap color, temperature, time for recapillarization and bleeding after puncture.
Results from the use of the SPY System has been the subject of 12 peer reviewed journal articles, 10 related to its use in cardiac surgery and 2 related to its use in transplantation surgeries, namely kidney and liver. Please refer to the bibliography in Section 19 - Clinical for a listing of all relevant journal articles.
The literature reports that the SPY System was able to non-invasively, quickly and safely identify 17 conduits in 311 patients that required revision during the surgical procedures. In all cases the lack of patency was visualized clearly by the SPY System using doses of IC-Green™ well below that approved for human use, allowing the surgeon to revise the graft thus decreasing subsequent myocardial infarctions and the morbidity and mortality associated with poor graft patency. Cardiac, renal and hepatic function were monitored during use of the SPY System and there were no reported adverse effects.
To support the original premarket notification 510(k) application, the system was used in six pig studies. These studies demonstrated that:
-
- it was possible to acquire high quality images in a simple and reproducible manner using small doses of IC-Green well below the concentrations approved for human use;
-
- it was possible to perform multiple imaging sequences with no detrimental effects on heart function, coronary flow or peripheral pressure; and
-
- it was possible to acquire images with no increase in myocardial tissue temperature; and
-
- it was possible to visualize all of the coronary beds with high quality images even when the heart was in a vertical position for visualizing posterior arteries.
Therefore, the in vivo evidence shows that:
- The exposure for the SPY® System is 35 mW/cm² which is far 1. below the maximum permissible exposure of 327 mW/cm2 established by ANSI for exposure to the skin.
-
- Use of the SPY System does not cause any thermal damage to the area of interest, even after repeated imaging sequences.
- ന For the heart, there were no changes in electrocardiograms or arterial pressures during and/or following SPY use.
4
- There were no acute or long-term cellular effects of using the 4. SPY System.
-
- There were no acute or long-term renal or hepatic effects of using the SPY System.
-
- The SPY System was able to acquire high quality images of the entire vascular bed on each area of interest.
- The SPY System is capable of imaging through the skin to 7. provide a visual assessment of dermal and subdermal blood flow.
Conclusions:
The above testing demonstrates that the SPY Imaging System is safe and effective in imaging blood flow indicative of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, microand reconstructive surgical procedures and is equivalent to the predicate devices.
5
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect and promote the health and well-being of Americans. The eagle is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Novadag Technologies, Iinc. % Ms. Allison Manners Vice President, Regulatory and Clinical Affairs 2585 Skymark Avenue, Suite 306 Mississauga, Ontario Canada L4W 4L5
JAN 1 2 2007
Re: K063345
Trade/Device Name: SPY® Imaging System Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: Class II Product Code: IZI Dated: November 3, 2006 Received: November 6, 2006
Dear Ms. Manners:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
6
Page 2 - Ms. Allison Manners
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
Indications for Use
510(k) Number (if known):
K 063345
Device Name:
SPY® Imaging System: SP2000
Indications for Use:
The SPY System is an imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General, Restorative,
Page of of
and Neurological bevices
510(k) Number_