(357 days)
CEDIA® Carbamazepine II, CEDIA® Valproic acid II
Not Found
No
The summary describes a standard in vitro diagnostic assay for measuring drug concentrations using a clinical chemistry analyzer. There is no mention of AI, ML, image processing, or any other technology typically associated with AI/ML applications in medical devices. The performance studies focus on traditional analytical metrics like precision and method comparison.
No
The device is an in vitro diagnostic intended for quantitative determination of drug concentrations, which helps in diagnosis and monitoring, but does not directly provide therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states that the device is "intended for the quantitative in vitro diagnostic determination" of drug concentrations and that "Measurements are used in the diagnosis and treatment" of overdose and for monitoring therapy.
No
The device is an in vitro diagnostic (IVD) assay kit intended for use on a clinical chemistry analyzer (T60). It involves the quantitative determination of substances in human serum, which is a chemical process, not a software-only function.
Yes, based on the provided information, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use/Indications for Use: The document explicitly states that the Carbamazepine and Valproic Acid assays are "intended for the quantitative in vitro diagnostic determination" of these substances in human serum. It also mentions their use in "diagnosis and treatment" and "monitoring levels," which are typical applications for IVDs.
- TDM Calibration set B: This component is also described as being "intended for in vitro diagnostic use as a calibrator."
The core function of these devices is to analyze biological samples (human serum) outside of the body to provide information for diagnostic and treatment purposes. This aligns perfectly with the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
Carbamazepine
For in vitro diagnostic use in the quantitative determination of the carbamazepine concentration in human serum on T60 analyzer. Measurements are used in the diagnosis and treatment of carbamazepine overdose and in monitoring levels of carbamazepine to help ensure proper therapy.
Valproic Acid
For in vitro diagnostic use in the quantitative determination of the valproic acid concentration in human serum on T60 instrument. Measurements are used in the diagnosis and treatment of valproic acid overdose and in monitoring levels of valproic acid to help ensure proper therapy.
TDM Calibration set B
For in vitro diagnostic use as a calibrator in the quantitative measurement of the kit code 981645 Carbamazepine and kit code 981650 Valproic acid assays on T60 Analyzer.
Product codes (comma separated list FDA assigned to the subject device)
KLT, LEG, DKB
Device Description
Not Found
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
human serum
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Carbamazepine
- Precision:
- Within run: Level 3.0 µg/ml (SD = 0.09, CV(%) = 2.8), Level 9.5 µg/ml (SD = 0.14, CV(%) = 1.5), Level 15.0 µg/ml (SD = 0.16, CV(%) = 1.1)
- Between run: Level 3.0 µg/ml (SD = 0.07, CV(%) = 2.4), Level 9.5 µg/ml (SD = - , CV(%) = -), Level 15.0 µg/ml (SD = 0.14, CV(%) = 0.9)
- Total: Level 3.0 µg/ml (SD = 0.19, CV(%) = 6.3), Level 9.5 µg/ml (SD = 0.32, CV(%) = 3.3), Level 15.0 µg/ml (SD = 0.42, CV(%) = 2.8)
- Method Comparison:
- Deming: y = 0.98 x + 0.02
- r = 0.993
- Range 1.9 - 19.6 µg/ml
- N = 134
- Limitations: No interference found with Hemoglobin up to 1000 mg/dl (10 g/l), Bilirubin up to 58 mg/dl (1000 µmol/l), Lipemia up to 1000 mg/dl (10 g/l) of Intralipid®.
Valproic Acid
- Precision:
- Within run: Level 35.0 µg/ml (SD = 0.43, CV(%) = 1.2), Level 81.1 µg/ml (SD = 0.81, CV(%) = 1.0), Level 113.6 µg/ml (SD = 1.01, CV(%) = 0.9)
- Between run: Level 35.0 µg/ml (SD = 0.61, CV(%) = 1.8), Level 81.1 µg/ml (SD = 1.08, CV(%) = 1.3), Level 113.6 µg/ml (SD = 1.07, CV(%) = 0.9)
- Total: Level 35.0 µg/ml (SD = 1.90, CV(%) = 5.4), Level 81.1 µg/ml (SD = 3.15, CV(%) = 3.9), Level 113.6 µg/ml (SD = 3.11, CV(%) = 2.7)
- Method Comparison:
- Deming: y = 0.996 x + 1.4
- r = 0.993
- Range 3.2 - 143.4 µg/ml
- N = 136
- Limitations: No interference found with Hemoglobin up to 1000 mg/dl (10 g/l), Bilirubin up to 58 mg/dl (1000 µmol/l), Lipemia up to 1000 mg/dl (10 g/l) of Intralipid®.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
CEDIA® Carbamazepine II, CEDIA® Valproic acid II
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.3645 Neuroleptic drugs radioreceptor assay test system.
(a)
Identification. A neuroleptic drugs radioceptor assay test system is a device intended to measure in serum or plasma the dopamine receptor blocking activity of neuroleptic drugs and their active metabolites. A neuroleptic drug has anti-psychotic action affecting principally psychomotor activity, is generally without hypnotic effects, and is a tranquilizer. Measurements obtained by this device are used to aid in determining whether a patient is taking the prescribed dosage level of such drugs.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image contains the logo for Thermo Fisher Scientific. The logo is composed of two lines of text. The first line is the company name, "ThermoFisher", in a bold, sans-serif font. The second line is the word "SCIENTIFIC" in a smaller, sans-serif font.
0CT 5 " 2007
510(k) SUMMARY
This Summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is: K063131
Introduction: A.
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
B. Submitter
Thermo Fisher Scientific Oy Ratastie 2 P.O. Box 100 FIN-01621 Vantaa Finland +358 (9) 329 100 tel +358 (9) 3291 0500 fax Contact: Päivi Sormunen, Vice President of QRC Date of Preparation: September 25, 2007
C. Device name
Proprietary name: Carbamazepine Common name: Carbamazepine test system Classification: II Class: Toxicology Product Code: KLT
Proprietary name: Valproic Acid Common name: Valproic Acid test system Classification: II Class: Toxicology Product Code: LEG
Proprietary name: TDM Calibration set B Common name: Calibrator Classification: II Class: Toxicology Product Code: DKB
Y-tunnus 0921547-0 VAT No FI09215470 Domicile Helsinki
1
Image /page/1/Picture/0 description: The image shows the logo for Thermo Fisher Scientific. The words "Thermo Fisher" are in bold, and the word "SCIENTIFIC" is in a smaller font size below. The logo is black and white.
D. Intended Use
Carbamazepine
For in vitro diagnostic use in the quantitative determination of the carbamazepine concentration in human serum on T60 analyzer. Measurements are used in the diagnosis and treatment of carbamazepine overdose and in monitoring levels of carbamazepine to help ensure proper therapy.
Valproic Acid
For in vitro diagnostic use in the quantitative determination of the valproic acid concentration in human serum on T60 instrument. Measurements are used in the diagnosis and treatment of valproic acid overdose and in monitoring levels of valproic acid to help ensure proper therapy.
TDM Calibration set B
For in vitro diagnostic use as a calibrator in the quantitative measurement of the kit code 981645 Carbamazepine and kit code 981650 Valproic acid assays on T60 Analyzer.
2
Image /page/2/Picture/0 description: The image shows the logo for Thermo Fisher Scientific. The words "Thermo Fisher" are on the top line in a bold, sans-serif font. The word "SCIENTIFIC" is on the second line, in a smaller, sans-serif font.
E. Indications for use
The Carbamazepine is intended for the quantitative in vitro diagnostic determination of the carbamazepine concentration in human serum using T60 Clinical Chemistry Analyzers. Measurements are used in the diagnosis and treatment of carbamazepine overdose and in monitoring levels of carbamazepine to help ensure proper therapy.
The Valproic Acid is intended for the quantitative in vitro diagnostic determination of the valproic acid concentration in human serum using T60 Clinical Chemistry Analyzers. Measurements are used in the diagnosis and treatment of valproic acid overdose and in monitoring levels of valproic acid to help ensure proper therapy.
TDM Calibration set B is intended for in vitro diagnostic use as a calibrator in the quantitative measurement of the kit code 981645 Carbamazepine and kit code 981650 Valproic acid assays on T60 Analyzer.
F. Substantial Equivalence
CEDIA® Carbamazepine II Microgenics Corporation
CEDIA® Valproic acid II Microgenics Corporation
Substantial equivalence -similarities G.
T60 Carbamazepine and Valproic Acid are substantially equivalent to other devices legally marketed in United Staes. We claim equivalence to the Microgenics Corporation CEDIA® Carbamazepine II and CEDIA® Valproic acid II
Y -tunnus 0921547-0 VAT No FI09215470 Domicile Helsinki
3
Attribute | New device #1 | Predicate device #1 |
---|---|---|
Intended Use | For in vitro diagnostic use in the | |
quantitative determination of the | ||
carbamazepine concentration in human | ||
serum on T60 analyzer. Measurements | ||
are used in the diagnosis and treatment of | ||
carbamazepine overdose and in | ||
monitoring levels of carbamazepine to | ||
help ensure proper therapy. | The CEDIA® Carbamazepine II | |
homogeneous enzyme immunoassay | ||
is for the quantitation of | ||
carbamazepine in human serum or | ||
plasma using automated clinical | ||
chemistry analyzers. Measurements | ||
are used in the diagnosis and | ||
treatment of carbamazepine | ||
overdose and in monitoring levels of | ||
carbamazepine to ensure proper | ||
therapy. | ||
Indication for Use | The Carbamazepine is intended for | |
the quantitative in vitro diagnostic | ||
determination of the carbamazepine | ||
concentration in human serum using | ||
T60 Clinical Chemistry Analyzers. | ||
Measurements are used in the | ||
diagnosis and treatment of | ||
carbamazepine overdose and in | ||
monitoring levels of carbamazepine | ||
to help ensure proper therapy. | See Intended Use | |
Assay Protocol | Assay uses recombinant DNA | |
technology (US Patent no. 4708929) | ||
to produce a unique homogenous | ||
enzyme immunoassay system. | Assay uses recombinant DNA | |
technology (US Patent no. 4708929) | ||
to produce a unique homogenous | ||
enzyme immunoassay system. | ||
Traceability/Standardiza | ||
tion | The calibration values are traceable to | |
USP reference materials prepared | ||
gravimetrically to drug-free human | ||
serum. | The calibration values are traceable | |
to USP reference materials prepared | ||
gravimetrically to drug-free human | ||
serum. | ||
Sample Type | Human Serum | Serum or plasma (Na or Li heparin, |
Na EDTA) | ||
Reagent Storage | The unopened reagents are stable at | |
2...8 °C until the expiration date | ||
stated on the label. Refer to the | ||
Application Notes of your T60 | ||
analyzer for the on board stability of | ||
reagents. | ||
DO NOT FREEZE the unopened | ||
reagents or the reconstituted reagents. | Store CEDIA Carbamazepine II | |
reagents at 2-8 °C. Do not freeze. | ||
For stability of the unopened | ||
components refer to the box or | ||
bottle labels for the expiration date |
The following table compares the Carbamazepine with the predicate test system | ||||
---|---|---|---|---|
4
t
Attribute | New device #1 | Predicate device #1 |
---|---|---|
Expected Values | Therapeutic range for adults: | |
According to different sources the | ||
suggested ranges are: | ||
4 - 12 µg/ml or 17 - 51 µmol/l (1) | ||
4 - 10 µg/ml or 17 - 42 µmol/l (2) | Therapeutic Range (µg/ml) | |
Penry and Newmark 5-12 | ||
Scheuer and Pedley 8-12 | ||
Troupin et al 8-12 | ||
Strandjord and Johannessen 3-12 | ||
Simonsen et al 6-10 | ||
Larkin et al 4-10 | ||
Shorvon et al 4-8 | ||
MacKichan and Kutt 4-12 | ||
Instrument | T60, DPC T60i, and DPC T60i Kusti | Roche Hitachi 911/912 |
Measuring Range | From 1.0 µg/ml to 19.0 µg/ml. | Between 0.5 µg/ml and the value of |
the Core TDM Multi-Cal High | ||
Calibrator (approximately 20 µg/ml | ||
or 84.6 µmol/l). | ||
Precision | Within run | |
Level 3.0 µg/ml | ||
SD = 0.09 | ||
CV(%) = 2.8 | ||
Level 9.5 µg/ml | ||
SD = 0.14 | ||
CV(%) = 1.5 | ||
Level 15.0 µg/ml | ||
SD = 0.16 | ||
CV(%) = 1.1 | ||
Between run | ||
Level 3.0 µg/ml | ||
SD = 0.07 | ||
CV(%) = 2.4 | ||
Level 9.5 µg/ml | ||
SD = - | ||
CV(%) = - | ||
Level 15.0 µg/ml | ||
SD = 0.14 | ||
CV(%) = 0.9 | ||
Total | ||
Level 3.0 µg/ml | ||
SD = 0.19 | ||
CV(%) = 6.3 | ||
Level 9.5 µg/ml | ||
SD = 0.32 | ||
CV(%) = 3.3 | ||
Level 15.0 µg/ml | ||
SD = 0.42 | ||
CV(%) = 2.8 | Within run | |
Level 4.2 µg/ml | ||
SD = 0.06 | ||
CV(%) = 1.5 | ||
Level 10.6 µg/ml | ||
SD = 0.08 | ||
CV(%) = 0.8 | ||
Level 16.8 µg/ml | ||
SD = 0.12 | ||
CV(%) = 0.7 | ||
Total | ||
Level 4.2 µg/ml | ||
SD = 0.15 | ||
CV(%) = 3.5 | ||
Level 10.6 µg/ml | ||
SD = 0.21 | ||
CV(%) = 2.0 | ||
Level 16.8 µg/ml | ||
SD = 0.29 | ||
CV(%) = 1.7 | ||
Attribute | New device #1 | Predicate device #1 |
Method Comparison | (Unit µg/ml) | |
Deming: | ||
$y = 0.98 x + 0.02$ | ||
$r = 0.993$ | ||
Range 1.9 - 19.6 µg/ml | ||
N = 134 | Previous CEDIA Carbamazepine | |
assay (x). Correlation (µg/ml) | ||
(Deming's): | ||
$Y =1.04x -0.04$ | ||
$r = 0.999$ | ||
Sy.x = 0.26 | ||
Range 1.3 - 19.8 µg/ml | ||
N = 103 | ||
Limitations | No interference found | |
Hemoglobin: | ||
up to 1000 mg/dl (10 g/l). | ||
Bilirubin: | ||
up to 58 mg/dl (1000 µmol/l) | ||
Lipemia: | ||
up to 1000 mg/dl (10 g/l) of | ||
Intralipid® | Samples containing carbamazepine | |
and the following concentrations of | ||
potential interference substances | ||
were quantitated accurately by the | ||
CEDIA® Carbamazepine II assay: | ||
Hemoglobin up to 1000 mg/dl, | ||
Bilirubin up to 66 mg/dl, | ||
Triglyceride up to 1000 mg/dl, | ||
Total protein up to 12 g/dl, | ||
Rheumatoid factor up to 180 IU/ml | ||
Attribute | New device #1 | Predicate device #1 |
Intended Use | For in vitro diagnostic use in the | |
quantitative determination of the valproic | ||
acid concentration in human serum on | ||
T60 instrument. Measurements are used | ||
in the diagnosis and treatment of valproic | ||
acid overdose and in monitoring levels of | ||
valproic acid to help ensure proper | ||
therapy. | The CEDIA® Valproic Acid II | |
homogeneous enzyme immunoassay | ||
is for the quantitation of valproic | ||
acid in human serum or plasma | ||
using automated clinical chemistry | ||
analyzers. Measurements are used as | ||
an aid in the diagnosis and treatment | ||
of valproic acid overdose and in | ||
monitoring levels of valproic acid to | ||
ensure proper therapy. | ||
Indication for Use | The Valproic Acid is intended for the | |
quantitative in vitro diagnostic | ||
determination of the valproic acid | ||
concentration in human serum using | ||
T60 Clinical Chemistry Analyzers. | ||
Measurements are used in the | ||
diagnosis and treatment of valproic | ||
acid overdose and in monitoring | ||
levels of valproic acid to help ensure | ||
proper therapy. | See Intended Use | |
Assay Protocol | Assay uses recombinant DNA | |
technology (US Patent no. 4708929) | ||
to produce a unique homogenous | ||
enzyme immunoassay system. | Assay uses recombinant DNA | |
technology (US Patent no. 4708929) | ||
to produce a unique homogenous | ||
enzyme immunoassay system. | ||
Traceability/Standardiza | ||
tion | The calibration values are traceable to | |
USP reference materials prepared | ||
gravimetrically to drug-free human | ||
serum. | The calibration values are traceable | |
to USP reference materials prepared | ||
gravimetrically to drug-free human | ||
serum. | ||
Sample Type | Human Serum | Serum or plasma (Na or Li heparin, |
Na EDTA) | ||
Reagent Storage | The unopened reagents are stable at | |
2...8 °C until the expiration date | ||
stated on the label. | ||
DO NOT FREEZE the unopened | ||
reagents or the reconstituted reagents. | Store CEDIA® Valproic acid II | |
reagents at 2-8 °C. Do not freeze. | ||
For stability of the unopened | ||
components refer to the box or | ||
bottle labels for the expiration date |
Ratastie 2
P.O. Box 100
FIN-01621 Vantaa
Finland
+358 (9) 329 100 tel
· 358 (9) 3291 0500 fax
Y-tunnus 0921547-0
VAT No FI09215470
Domicile Helsinki
5
SCIENTIFIC
Thermo Fisher Scientific Oy
Y-tunnus 0921547-0
VAT No FI09215470
Domicile Helsinki
www.thermo com
6
Predicate Test System | Valproic Acid |
---|---|
$\frac{d[Drug]}{dt} = k_{in} - k_{out}*[Drug]$ | $\frac{d[Drug]}{dt} = k_{in} - k_{out}[Drug] - k_{met}[Drug]$ |
7
ਪ
Attribute | New device #1 | Predicate device #1 |
---|---|---|
Expected Values | Therapeutic range for adults: | |
50 - 100 µg/ml or | ||
347 - 693 µmol/l (1,2) | Therapeutic | |
Schobben et al 50-100 µg/ml | ||
Cloyd and Leppik 50-100 µg/ml | ||
Klotz and Schweizer 40-90 µg/ml | ||
Turnbull et al 50-100 µg/ml | ||
Toxic | ||
Schobben et al - µg/ml | ||
Cloyd and Leppik >100 µg/ml | ||
Klotz and Schweizer - | ||
Turnbull et al >100 µg/ml | ||
Instrument | T60, DPC T60i, and DPC T60i Kusti | Roche Hitachi 911/912 |
Measuring Range | From 3.0 µg/ml to 142.5 µg/ml. | Between 3.0 µg/ml and the value of |
the Core TDM High Calibrator | ||
(approximately 150 µg/ml or 1039.5 µmol/l) | ||
Precision | Within run | |
Level 35.0 µg/ml | ||
SD = 0.43 | ||
CV(%) = 1.2 | ||
Level 81.1 µg/ml | ||
SD = 0.81 | ||
CV(%) = 1.0 | ||
Level 113.6 µg/ml | ||
SD = 1.01 | ||
CV(%) = 0.9 | ||
Between run | ||
Level 35.0 µg/ml | ||
SD = 0.61 | ||
CV(%) = 1.8 | ||
Level 81.1 µg/ml | ||
SD = 1.08 | ||
CV(%) = 1.3 | ||
Level 113.6 µg/ml | ||
SD = 1.07 | ||
CV(%) = 0.9 | ||
Total | ||
Level 35.0 µg/ml | ||
SD = 1.90 | ||
CV(%) = 5.4 | ||
Level 81.1 µg/ml | ||
SD = 3.15 | ||
CV(%) = 3.9 | ||
Level 113.6 µg/ml | ||
SD = 3.11 | ||
CV(%) = 2.7 | Within run | |
Level 24.4 µg/ml | ||
SD = 0.59 | ||
CV(%) = 2.4 | ||
Level 95.0 µg/ml | ||
SD = 1.43 | ||
CV(%) = 1.5 | ||
Level 136.8 µg/ml | ||
SD = 1.81 | ||
CV(%) = 1.3 | ||
Total | ||
Level 24.4 µg/ml | ||
SD(µg/ml) = 0.83 | ||
CV(%) = 3.4 | ||
Level 95.0 µg/ml | ||
SD = 1.93 | ||
CV(%) = 2.0 | ||
Level 136.8 µg/ml | ||
SD = 2.48 | ||
CV(%) = 1.8 | ||
Attribute | New device #1 | Predicate device #1 |
Method Comparison | (Unit µg/ml) | |
Deming: | ||
$y = 0.996 x + 1.4$ | ||
$r = 0.993$ | ||
Range 3.2 - 143.4 µg/ml | ||
N = 136 | Commercially available | |
fluorescence polarization | ||
immunoassay (x) | ||
Correlation (µg/ml) | ||
(Linear regression): | ||
$Y = 1.08x -0.61$ | ||
$r = 0.972$ | ||
$Sy.x = 7.042$ | ||
Range 2.6 - 119.8 µg/ml | ||
N = 77 | ||
Limitations | No interference found | |
Hemoglobin: | ||
up to 1000 mg/dl (10 g/l). | ||
Bilirubin: | ||
up to 58 mg/dl (1000 µmol/l) | ||
Lipemia: | ||
up to 1000 mg/dl (10 g/l) of | ||
Intralipid® | Samples containing valproic acid | |
and the following concentrations of | ||
potential interference substances | ||
were quantitated accurately by the | ||
CEDIA Valproic Acid II assay: | ||
Hemoglobin up to 1000 mg/dl, | ||
Bilirubin up to 60 mg/dl, | ||
Triglyceride up to 1000 mg/dl, | ||
Total protein up to 10 g/dl, | ||
IgA up to 790 mg/dl | ||
IgG up to 4300 mg/dl, | ||
IgM up to 840 mg/dl and | ||
Rheumatoid factor up to 200 IU/ml |
Thermo Fisher Scientific Oy
Ratastie 2
P.O. Box 100
FIN-01621 Vantaa
Finland
+358 (9) 329 100 tcl
+358 (9) 3291 0500 fax
Y-tunnus 0921547-0
VAT No FI09215470
Domicile Helsinki
8
S C
+358 (9) 329 100 tel
+358 (9) 3291 0500 fax
u www.thermo com
9
Image /page/9/Picture/1 description: The image shows a circular logo with an eagle in the center. The eagle is stylized with three curved lines forming its body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. The logo appears to be a seal or emblem, possibly representing a government agency.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Thermo Fisher Scientific Oy c/o Mr. Päivi Sormunen Vice President of ORC Ratastie 2, P.O. Box 100 FIN-01621 Vantaa, Finland
OCT 5 2007
Re: K063131 Trade Name: Carbamazepine, Valproic Acid and TDM Calibration set B Regulation Number: 21 CFR 862.3645 Regulation Name: Neuroleptic drugs radioreceptor assay test system. Regulatory Class: Class II Product Code: KLT, LEG, DKB Dated: September 25, 2007 Received: September 27, 2007
Dear Mr. Sormunen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
10
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Coopes, M.S., D.v.M.
Jéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
11
Indications for Use
510(k) Number (if known): K063131
Device Names: Carbamazepine Valproic Acid TDM Calibration set B
Indications for Use:
The Carbamazepine is intended for the quantitative in vitro diagnostic determination of the carbamazepine concentration in human serum using T60 Clinical Chemistry Analyzers. Measurements are used in the diagnosis and treatment of carbamazepine overdose and in monitoring levels of carbamazepine to help ensure proper therapy
The Valproic Acid is intended for the quantitative in vitro diagnostic determination of the valproic acid concentration in human serum using T60 Clinical Chemistry Analyzers. Measurements are used in the diagnosis and treatment of valproic acid overdose and in monitoring levels of valproic acid to help ensure proper therapy.
TDM Calibration set B is intended for in vitro diagnostic use as a calibrator in the quantitative measurement of the kit code 981645 Carbamazepine and kit code 981650 Valproic acid assays on T60 Analyzer.
Prescription Use × (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)
Sign-Off
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
S10(k) |K06313|