K Number
K062959

Validate with FDA (Live)

Date Cleared
2006-10-16

(17 days)

Product Code
Regulation Number
870.2800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The "Indications for Use" of the modified DMS 300-2, 300-3 and 300–4 recorders are indicated for use in a clinical setting, by qualified medical professionals only, for recording multi-lead ECG data of patients during a minimum ambulatory time period of 24-hours. It is not a life supporting system, and is not connected to an AC power source. The "Intended Uses" of the modified 300-2, 300-3, and 300-4 Holter ECG Recorders are exactly the same as the predicate devices (DMS 300-7 and 300 ECG Recorder).Ambulatory multi-day electrocardiography is used for the below indications:

. Evaluation of patients with symptoms suggesting arrhythmias.

  • . Evaluation of patients with pacemakers.
    . Evaluation of patient heart rate changes and QRS interval changes.

. Evaluation of patient response to drug therapy treatment.

Device Description

The modified DMS 300-2 Holter ECG Recorder is intended for use as a part of a Holter ECG analysis system for continuous and cardiac event monitoring, and is designed to be used with the DMS Premier Holter system. The DMS 300-2 provides three (3) leads of continuous ECG recording. With its memory built into its circuit board, the 300-2 can store up to thirty (30) days of Long-Term Ambulatory ECG, whether it be continuous ECG or patient symptomatic event ECG data. The 300-2 acquires, digitizes, and stores ECG data that can be later analyzed by the Premier Holter system. Short-term, two-minute ECG's can also be phone transmitted from the Holter recorder to a host PC for symptomatic ECG's. The Premier Holter system processes pre-recorded patient ECG data that has been stored in the DMS 300-2. The cardiac data provided by the 300-2 and the Premier Holter system is used by trained medical personnel to assist in the diagnosis of patients with various ECG rhvthm patterns.

The modified DMS 300-3 Holter ECG Recorder is intended for use as a part of a Holter analysis system, and is designed to be used with the DMS Premier Holter system. The DMS 300-3 provides three (3) channels of continuous ECG recording. With its memory built into its circuit board, the 300-3 can store up to two (2) days of continuous Holter ECG data. The 300-3 acquires, digitizes, and stores data to be later analyzed by the Premier Holter system. The Premier Holter system processes pre-recorded patient ECG data that has been stored in the DMS 300-3. The cardiac data provided by the 300-3 and the Premier Holter system is used by trained medical personnel to assist in the diagnosis of patients with various ECG rhythm patterns.

The modified DMS 300-4 Holter ECG Recorder is intended for use as a part of a Holter analysis system, and is designed to be used with the DMS Premier Holter system. The DMS 300-4 provides twelve (12) leads of continuous ECG recording. With its memory built into its circuit board, the 300-4 can store up to two (2) days of continuous Holter ECG data. The 300-4 acquires, digitizes, and stores data to be later analyzed by the Premier Holter system. The Premier Holter system processes nre-recorded patient ECG data that has been stored in the DMS 300-4. The cardiac data provided by the 300-4 and the Premier Holter system is used by trained medical personnel to assist in the diagnosis of patients with various ECG rhythm patterns.

AI/ML Overview

The provided document is a 510(k) notification for bundling Holter ECG recorders. It focuses on demonstrating substantial equivalence to predicate devices rather than presenting a performance study with detailed acceptance criteria and testing results against those criteria in the way a clinical trial would.

Therefore, many of the requested details about acceptance criteria, specific performance metrics, sample sizes, expert involvement, and ground truth establishment are not explicitly present in this type of regulatory submission. The document relies on comparing specifications and stating that "the same input signals produce the same results in the processing of the ECG data with the same Holter playback device." This implies a functional equivalence test rather than a clinical performance study with predefined accuracy or clinical endpoint targets.

Here's an attempt to answer the questions based on the information available:

1. A table of acceptance criteria and the reported device performance

The document doesn't define explicit acceptance criteria in terms of numerical performance thresholds (e.g., sensitivity, specificity for arrhythmia detection) as a clinical study would. Instead, the "acceptance criteria" are implied by demonstrating substantial equivalence to predicate devices. The performance is reported by showing that the new devices have identical or improved specifications compared to the predicates and produce the "same results in the processing of the ECG data."

Feature/SpecificationAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (DMS 300-2, 300-3, 300-4)
ECG LeadsMust be 3-Lead ECG (for 300-2, 300-3) or justified if different300-2: 3-Lead ECG300-3: 3-Lead ECG300-4: 12-Lead ECG (justified as a modification with no significant safety/effectiveness impact)
Resolution8-Bit8-Bit
Recording DurationEquivalent to or greater than predicate (24-hours)300-2: 30-days300-3: 48-hours300-4: 48-hours
Bandwidth0.05 to 100 Hz0.05 to 100 Hz
Common Mode Rejection> 60 dB> 60 dB
Power SourceFunctionally adequate (AA Alkaline or 9V)AAA Alkaline (change justified by reduced size)
Average Current DrainFunctionally adequate (5 mA or 9 mA)3 mA (improvement)
Event ECG ButtonYesYes
Operating Temperature0 to 60 C (or 0 to 50 C for one predicate)0 to 60 C
1-MV Input1-CM square wave1-CM square wave
DimensionsSize reduction compared to predicate (4.94 x 2.75 x 0.94 in or 5.20 x 2.90 x 1.16 in)2.90 x 2.16 x 0.75 in. (smaller)
WeightWeight reduction compared to predicate (4 oz or 8 oz w/o battery)2 oz. w/o battery (lighter)
Processing SystemPremier Holter SystemPremier Holter System
ECG Data TransferPC's hard diskPC's hard disk
Functional EquivalenceProduce the same results in processing ECG data with the same Holter playback device as the predicate."Testing of the devices with the same input signals produce the same results in the processing of the ECG data with the same Holter playback device."
Intended Use/IndicationsSame as predicate devicesSame as predicate devices

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in terms of patient data. The testing described is functional testing on the devices themselves using "the same input signals." Therefore, there is no mention of a sample size of patients or data provenance (country of origin, retrospective/prospective). This is a technical performance comparison between devices, not a clinical study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for this type of submission is the functional output of the predicate device when given standard input signals. Clinical ground truth (e.g., verified diagnoses by experts) is not discussed.

4. Adjudication method for the test set

Not applicable. There's no clinical test set requiring adjudication in the context of this submission. The comparison is against known technical specifications and functional output of the predicate.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This submission is for Holter ECG recorders, which are hardware devices for data acquisition, not AI-powered analysis tools. No MRMC study or AI assistance is mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is for a hardware device, not an algorithm. The device itself is "standalone" in recording ECG data, but the data is then processed by a "Premier Holter system" which is mentioned as part of the overall system. The focus is on the recorder's ability to acquire and store data equivalently.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the comparison is the known and established functional output and specifications of the legally marketed predicate device (DMS 300-7 Holter ECG Recorder K062007) when subjected to the "same input signals." It's a technical equivalency rather than a clinical ground truth.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As above, no training set is relevant for this hardware submission.

{0}------------------------------------------------

K062959
-pi/6

Special 510(k) Notification: Device Summary

Bundling of 300-2, 300-3, and 300-4 Holter ECG Recorders

Date: September 19, 2006

OCT 16 2006

Submitter:

William Parsons, Official Correspondent Diagnostic Monitoring Software 292 Kingsbury, 32B P.O. Box 3109 Stateline, NV 89449 Fax: 714 953 0260 714 953 2653 Tel: e-mail: holterdms@hotmail.com Contact: William Parsons

Trade Name: DMS 300-2 DMS 300-3 DMS 300-4

Common Name: Holter ECG Recorder Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: II (two) Product Code: MWJ

Establishment Registration Number: 2028190 Owner/Operator Number: 9003252 Payment Identification Number: MD6027393-956733

Legally marketed device to which S.E. is claimed:

This bundling of three (3) Holter ECG recorders (DMS 300-2, DMS 300-3, and DMS 300-4) are an addition to its predecessor device, Model 300-7, which has its 510(k) number issued by the FDA. The modifications are a reduction in size, a change in memory from removable flash cards to memory mounted on the circuit board, and a change in battery from AA alkaline to AAA alkaline. For these three bundled Holter ECG recorders, the intended use, indications for use, functions, and recording output ECG data are the same as the predicate device, the Model 300-7 Holter ECG recorder. The final test results of the predicate DMS 300-7 Recorder (K062007) and

{1}------------------------------------------------

K062959
P2/6

the modified submission devices 300-2, 300-3, and 300-4, show that the testing of the devices with the same input signals produce the same results in the processing of the ECG data with the same Holter playback device. These modified DMS 300-2, 300-3, and 300-4 Holter recorders are substantially equivalent (SE) to the following legally marketed predicate devices that were cleared 510(k)s under 21 CFR 870.2800, Class II (two):

  • Diagnostic Monitoring Software DMS 300-7 Holter ECG Recorder (K062007). .
  • Diagnostic Monitoring Software (Scole) 300 ECG Recorder (K923664). .

The decision to submit this Special 510(k) as Bundling for these three (3) modified Holter ECG recorders was a result of a phone conversation on July 21, 2006 with Elias Mallis (301 443 8517), who recommended that this Bundling would be appropriate.

Description

DMS 300-2 Holter ECG Recorder: (Bundled device 1 of 3)

The modified DMS 300-2 Holter ECG Recorder is intended for use as a part of a Holter ECG analysis system for continuous and cardiac event monitoring, and is designed to be used with the DMS Premier Holter system. The DMS 300-2 provides three (3) leads of continuous ECG recording. With its memory built into its circuit board, the 300-2 can store up to thirty (30) days of Long-Term Ambulatory ECG, whether it be continuous ECG or patient symptomatic event ECG data. The 300-2 acquires, digitizes, and stores ECG data that can be later analyzed by the Premier Holter system. Short-term, two-minute ECG's can also be phone transmitted from the Holter recorder to a host PC for symptomatic ECG's. The Premier Holter system processes pre-recorded patient ECG data that has been stored in the DMS 300-2. The cardiac data provided by the 300-2 and the Premier Holter system is used by trained medical personnel to assist in the diagnosis of patients with various ECG rhvthm patterns.

The modification differences between the predicate DMS 300-7 Holter ECG Recorder (K062007) and the 300-2 are as follows: (a) the 300-2 is a reduction in size, (b) the 300-2 has its digital memory built into the pcb; whereas, the 300-7 uses removable compact flash cards. (c) the 300-2 uses a single AAA alkaline battery, whereas the 300-7 uses a single AA alkaline battery, and (d) the 300-2 outputs a sound signal that converts to an ECG signal for testing the quality of the ECG signal and for sending symptomatic event ECG's through a phone transmission; whereas, the 300-7 uses a cable to output an ECG signal to a patient isolation source to test the quality of the ECG signal. Since digital memory has become low cost, it is a natural evolution for Holter ECG recorders to be multi-day, ambulatory, ECG recording devices. A potential problem in multi-day monitoring is that the electrode signal can deteriorate to the point that the ECG data is of no value. There is no need to waste the patient's

{2}------------------------------------------------

time in wearing a Holter recorder if the ECG signal is not of acceptable quality. The reception of the sound signal the standard telephone at once or more per day. provides a standard ECG modality to assure that useful ECG data is being captured and being sent by the 300-2 Holter recording device.

The indications for use are the same for the predicate 300-7 device and the 300-2, the same ECG signal quality is generated by both the 300-7 and 300-2, the intended use is the same for both the 300-7 and 300-2. and the product modifications do not alter the fundamental scientific technology of the device. Data generated by the design control procedures are maintained by the manufacturer, and are available for FDA inspection. The product modifications do not significantly affect the safety and effectiveness of the 300-2.

The 300-2 is a 3-Lead ECG recorder which includes a patient activated ECG Event button that can be pressed by the patient to flag a symptomatic cardiac event ECG, as is the same with both of the listed predicate devices; namely, the DMS 300-7 (K062007) and Model 300 ECG Recorder (K923664).

DMS 300-3 Holter ECG Recorder: (Bundled device 2 of 3)

The modified DMS 300-3 Holter ECG Recorder is intended for use as a part of a Holter analysis system, and is designed to be used with the DMS Premier Holter system. The DMS 300-3 provides three (3) channels of continuous ECG recording. With its memory built into its circuit board, the 300-3 can store up to two (2) days of continuous Holter ECG data. The 300-3 acquires, digitizes, and stores data to be later analyzed by the Premier Holter system. The Premier Holter system processes pre-recorded patient ECG data that has been stored in the DMS 300-3. The cardiac data provided by the 300-3 and the Premier Holter system is used by trained medical personnel to assist in the diagnosis of patients with various ECG rhythm patterns.

The modification differences between the predicate DMS 300-7 Holter ECG Recorder (K062007) and the 300-3 are as follows: (a) the 300-3 is a reduction in size. (b) the 300-3 has its digital memory built into the circuit board, whereas the 300-7 uses removable compact flash cards, (c) the 300-3 uses a single AAA alkaline battery. whereas the 300-7 uses a single AA alkaline battery, and (d) the 300-3 records for 48 hours, whereas the 300-7 records for 24 hours.

The 300-3 is a 3-Lead ECG recorder which includes a patient activated ECG Event button that can be pressed by the patient to flag a symptomatic cardiac event ECG, as is the same with both of the listed predicate devices; namely, the DMS 300-7 (K062007) and Model 300 ECG Recorder (K923664).

The indications for use are the same for the predicate 300-7 device and the 300-3. the same ECG signal quality is generated by both the 300-7 and 300-3, the intended use is the same for both the 300-7 and 300-3, and the product modifications do not alter

{3}------------------------------------------------

the fundamental scientific technology of the device. Data generated by the design the fanaamonuments are maintained by the manufacturer, and are available for FDA inspection. The product modifications do not significantly affect the safety and effectiveness of the 300-3.

Model 300-4 Holter ECG Recorder: (Bundled device 3 of 3)

The modified DMS 300-4 Holter ECG Recorder is intended for use as a part of a Holter analysis system, and is designed to be used with the DMS Premier Holter system. The DMS 300-4 provides twelve (12) leads of continuous ECG recording. With its memory built into its circuit board, the 300-4 can store up to two (2) days of continuous Holter ECG data. The 300-4 acquires, digitizes, and stores data to be later analyzed by the Premier Holter system. The Premier Holter system processes nre-recorded patient ECG data that has been stored in the DMS 300-4. The cardiac data provided by the 300-4 and the Premier Holter system is used by trained medical personnel to assist in the diagnosis of patients with various ECG rhythm patterns.

The modification differences between the predicate DMS 300-7 Holter ECG Recorder (K062007) and the 300-4 are as follows: (a) the 300-4 is a reduction in size, (b) the 300-4 has its digital memory built into the circuit board, whereas the 300-7 uses removable compact flash cards, (c) the 300-4 uses a single AAA alkaline battery, whereas the 300-7 uses a single AA alkaline battery, (d) the 300-4 records 12-Leads of ECG data, whereas the 300-7 records 3-Lead of ECG data, and (e) the 300-4 records for 48 hours, whereas the 300-7 records for 24 hours.

The 300-4 is a 12-Lead ECG recorder which includes a patient activated ECG Event button that can be pressed by the patient to flag a symptomatic cardiac event ECG, as is the same with both of the listed predicate devices; namely, the DMS 300-7 (K062007) and Model 300 ECG Recorder (K923664).

The indications for use are the same for the predicate 300-7 device and the 300-4, the same ECG signal quality is generated by both the 300-7 and 300-4, the intended use is the same for both the 300-7 and 300-4, and the product modifications do not alter the fundamental scientific technology of the device. Data generated by the design control procedures are maintained by the manufacturer, and are available for FDA inspection. The product modifications do not significantly affect the safety and effectiveness of the 300-4.

Diagnostic Monitoring Software maintains its file per the FDA's guidance document entitled "Deciding When to Submit a 510(k) for a change to an Existing Device, 510(k) Memorandum # K97-1, January 10, 1997, from the Director, Office of Device Evaluation." The process since 2000 has documented that the DMS Holter recorders have met the requirements of "Documentation" rather than "New 510(k)." Forms 2891 and 2892 have been submitted annually to the FDA. In June of 2006, DMS voluntarily asked the FDA to review our "Documentation" versus "New 510(k)" files,

{4}------------------------------------------------

K062959
P5/6

and the FDA requested that we submit a New 510(k). On July 10, 2006, we spoke on the phone with the FDA's Rod Perez (800 636 2041), and he confirmed that we should submit a New Special 510(k) for the modified Holter ECG recorder. On August 16, 2006 the FDA granted to us final 510(k) marketing approval for the DMS 300-7 Holter Recorder with K062007 . In a phone conversation with Elias Mallis on July 21, 2006, he recommended that we should submit the 300-2, 300-3, and 300-4 Holter recorders as a Bundled submission.

Comparison to the Sponsor's Predicate Devices:

SpecificationsDMS 300-2DMS 300-7300 ECG Recorder
Predicate DeviceNoYesYes
OwnerDMSDMSDMS
510(k) numberK062007K923664
ECG Leads3-Lead ECG3-Lead ECG3-Lead ECG
Resolution8-Bit8-Bit8-Bit
Recording Duration30-days24-hours24-hours
Bandwidth0.05 to 100 Hz0.05 to 100 Hz0.05 to 100 Hz
Common Mode Rej> 60 db> 60 db> 60 db
Power SourceAAA AlkalineAA Alkaline9V or AA Alkaline
Average Current Drain3 mA5 mA9 mA
Event ECG ButtonYesYesYes
Operating Temp0 to 60 C0 to 60 C0 to 50 C
1-MV Input =1-CM square wave1-CM square wave1-CM square wave
Dimensions2.90 x 2.16 x 0.75 in.4.94 x 2.75 x 0.94 in.5.20 x 2.90 x 1.16 in.
Weight2 oz. w/o battery4 oz. w/o battery8 oz. w/o battery
Processing SystemPremier Holter SystemPremier Holter SystemPremier Holter System
ECG Data TransferPC's hard diskPC's hard diskPC's hard disk

The DMS 300-2 is substantially equivalent (SE) to the predicate below devices.

The DMS 300-3 is substantially equivalent (SE) to the predicate below devices.

SpecificationsDMS 300-3DMS 300-7300 ECG Recorder
Predicate DeviceNoYesYes
OwnerDMSDMSDMS
510(k) numberK062007K923664

{5}------------------------------------------------

ECG Leads3-Lead ECG3-Lead ECG3-Lead ECG
Resolution8-Bit8-Bit8-Bit
Recording Duration48-hours24-hours24-hours
Bandwidth0.05 to 100 Hz0.05 to 100 Hz0.05 to 100 Hz
Common Mode Rej> 60 db> 60 db> 60 db
Power SourceAAA AlkalineAA Alkaline9V or AA Alkaline
Average Current Drain3 mA5 mA9 mA
Event ECG ButtonYesYesYes
Operating Temp0 to 60 C0 to 60 C0 to 50 C
1-MV Input =1-CM square wave1-CM square wave1-CM square wave
Dimensions2.90 x 2.16 x 0.75 in.4.94 x 2.75 x 0.94 in.5.20 x 2.90 x 1.16 in.
Weight2 oz. w/o battery4 oz. w/o battery8 oz. w/o battery
Processing SystemPremier Holter SystemPremier Holter SystemPremier Holter System
ECG Data TransferPC's hard diskPC's hard diskPC's hard disk

The DMS 300-4 is substantially equivalent (SE) to the predicate below devices.

SpecificationsDMS 300-4DMS 300-7300 ECG Recorder
Predicate DeviceNoYesYes
OwnerDMSDMSDMS
510(k) numberK062007K923664
ECG Leads12-Lead ECG3-Lead ECG3-Lead ECG
Resolution8-Bit8-Bit8-Bit
Recording Duration48-hours24-hours24-hours
Bandwidth0.05 to 100 Hz0.05 to 100 Hz0.05 to 100 Hz
Common Mode Rej> 60 db> 60 db> 60 db
Power SourceAAA AlkalineAA Alkaline9V or AA Alkaline
Average Current Drain3 mA5 mA9 mA
Event ECG ButtonYesYesYes
Operating Temp0 to 60 C0 to 60 C0 to 50 C
1-MV Input =1-CM square wave1-CM square wave1-CM square wave
Dimensions2.90 x 2.16 x 0.75 in.4.94 x 2.75 x 0.94 in.5.20 x 2.90 x 1.16 in.
Weight2 oz. w/o battery4 oz. w/o battery8 oz. w/o battery
Processing SystemPremier Holter SystemPremier Holter SystemPremier Holter System
ECG Data TransferPC's hard diskPC's hard diskPC's hard disk

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads, representing the department's focus on health, human services, and well-being. The eagle is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 6 2006

Diagnostic Monitoring Software c/o Mr. William Parsons Official Correspondent P.O. Box 3109 Stateline. NV 89449

Re: K062959

Trade Name: DMS Holter Recorders: Models 300-2, 300-3, and 300-4 Regulation Number: 21 CFR 870.2800 Regulation Name: Ambulatory Electrocardiograph without analysis Regulatory Class: Class II Product Code: 74 MWJ Dated: September 19, 2006 Received: September 29, 2006

Dear Mr. Parsons:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Mr. William Parsons

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276- 2041. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

B/flimman for

Bran D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

510(k) Number (if known):

Device Name: DMS HOLTER RECORDERS: MODELS 300-2, 300-3 and 300-4

Indications For Use: The "Indications for Use" of the modified DMS 300-2, 300-3 and 300–4 recorders are indicated for use in a clinical setting, by qualified medical professionals only, for recording multi-lead ECG data of patients during a minimum ambulatory time period of 24-hours. It is not a life supporting system, and is not connected to an AC power source. The "Intended Uses" of the modified 300-2, 300-3, and 300-4 Holter ECG Recorders are exactly the same as the predicate devices (DMS 300-7 and 300 ECG Recorder).Ambulatory multi-day electrocardiography is used for the below indications:

. Evaluation of patients with symptoms suggesting arrhythmias.

  • . Evaluation of patients with pacemakers.
    . Evaluation of patient heart rate changes and QRS interval changes.

. Evaluation of patient response to drug therapy treatment.

AND/OR

X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

B.Bammicamor

Division Sign-Off
Division of Cardiovascular Devices
510(k) Number K060959

Page 1 of

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).