K Number
K062883
Manufacturer
Date Cleared
2007-06-13

(260 days)

Product Code
Regulation Number
868.2375
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For adult use. The Gereonics Ultra-Piezo Limb Movement Sensor (LMS) is used to detect periodic limb movements for recording or monitoring on a physiological amplifier. The LMS is intended for use in sleep disorders studies.

Device Description

The Pro-Tech predicate device and the Gereonics Ultra-Piezo Limb Movement Sensor (LMS) utilize a piezo disc (sometimes referred to as a piezo element, diaphragm or bender). There is no external voltage supply -- the sensor directly converts motion to voltage. The piezo disc and output resistor convert minute taps to voltage. A lengthy lead wire extends from this assembly and terminates with two electrode style safety connectors. The LMS is applied to the patients limb using 3M Coban LF Latex Free Self-Adherent Wrap. The sensor is reusable and the self-adherent wrap is disposable. The predicate device is applied using a reusable Velcro strap with a pouch that holds the LMS.

AI/ML Overview

The provided document describes the Gereonics Ultra-Piezo Limb Movement Sensor (LMS) and its comparison to a predicate device (Pro-Tech PLM SENSOR). The study conducted is primarily a comparative performance and temperature test to demonstrate substantial equivalence, rather than establishing specific acceptance criteria for a new clinical performance claim.

Here's a breakdown of the information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" in a quantitative sense with specific thresholds that the device had to meet. Instead, the study aimed to show similarity to the predicate device.

Feature/TestAcceptance Criteria (Implicit - Similarity to Predicate)Reported Device Performance (Gereonics LMS)
Comparative Performance Test:Similar sensitivity and identical output pulse waveform to predicate device.Peak to Peak Voltage Output: 600 millivolts (Predicate: 300 millivolts). Within normal variability of other test devices.
Output Pulse Waveform: Identical to predicate device on the test system, except for amplitude.
Identical signal waveform and sensitivity when attached to ankle and tapped on floor.Signal Waveform and Sensitivity (Ankle Test): Identical to predicate device.
Comparative Temperature Test:Similar performance across temperature range of 8°C to 43°C. Similar temperature coefficient.Temperature Coefficient: -6% per °C in the 20-30°C range (Predicate: -5% per °C). Performed similarly to predicate.
Electrical Design:Essentially identical to predicate device.Essentially identical to predicate device, with modification of output resistor value (50K vs 75K in predicate).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify a numerical sample size for the "test set" in terms of number of devices or number of human subjects. The comparative tests refer to "the Pro-Tech test unit" and "the Gereonics LMS." This implies a very small sample, likely one unit of each for the bench tests.
  • Data Provenance: The study was conducted by Gereonics, Inc., the manufacturer of the device. The data is retrospective in the sense that it's a comparison to an existing predicate device. No country of origin for diverse patient data is relevant as human subjects data is not presented.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. This study did not involve human expert review to establish ground truth for clinical cases. It was a bench-testing comparison of device performance.

4. Adjudication Method for the Test Set

Not applicable. There was no need for adjudication as the study focused on objective measurements from bench testing, not subjective interpretations of clinical data by multiple experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This study's purpose was to demonstrate substantial equivalence through technical comparison, not to evaluate human reader performance with or without AI assistance.

6. Standalone Performance Study

Yes, a standalone performance assessment akin to an algorithm-only evaluation was implicitly done through the bench tests. The "Comparative Performance Test" and "Comparative Temperature Test" assessed the Gereonics LMS's direct output (voltage, waveform, temperature coefficient) and compared it to the predicate without human intervention in the signal interpretation.

7. Type of Ground Truth Used

The ground truth for this study was the performance of the predicate device (Pro-Tech PLM SENSOR) as measured through controlled bench tests and a simple "tapping on the floor" test. The goal was to show that the Gereonics LMS performed "similarly" or "identically" under these conditions.

8. Sample Size for the Training Set

Not applicable. This device is a passive, non-software-driven sensor. There is no "training set" in the context of machine learning or AI algorithms. Its function is based on piezoelectric principles, not learned from data.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for this device, no ground truth needed to be established in this context.

{0}------------------------------------------------

K062883

GEREONICS, INC.

244 La Barranca Drive Solana Beach, California 92075 Telephone: 1-800-654-6266 Fax: (858) 481-3654

510(k) Summary:

Date: Submitter: May 7, 2007 OUN 1 3 2007 Gereonics, Inc. 244 La Barranca Drive Solana Beach, CA 92075 Contact Person: Gerald A. Rost Tel: 760 728-4640 Fax: 760 728-4640

Device Identification:

Proprietary Name: Gereonics Ultra-Piezo Limb Movement Sensor Classification Name: Monitor, Breathing Frequency Device Classification: II Regulation Number: 868.2375 Product Code: BZQ

Predicate device is:

PLM SENSOR - 510(k) Number K940014 Pro-Tech Services, Inc. 4338 Harbour Pointe Blvd., Mukilteo, WA 98275

Device Description:

The Pro-Tech predicate device and the Gereonics Ultra-Piezo Limb Movement Sensor (LMS) utilize a piezo disc (sometimes referred to as a piezo element, diaphragm or bender). There is no external voltage supply -- the sensor directly converts motion to voltage. The piezo disc and output resistor convert minute taps to voltage. A lengthy lead wire extends from this assembly and terminates with two electrode style safety connectors.

The LMS is applied to the patients limb using 3M Coban LF Latex Free Self-Adherent Wrap. The sensor is reusable and the self-adherent wrap is disposable. The predicate device is applied using a reusable Velcro strap with a pouch that holds the LMS.

Intended Use:

For adult use. The Gereonics Ultra-Piezo Limb Movement Sensor (LMS) is used to detect periodic limb movements for recording or monitoring on a physiological amplifier. The LMS is intended for use in sleep disorders studies.

{1}------------------------------------------------

Information Supportive of Substantially Equivalent Claim:

FeaturesGereonicsPro-Tech
Piezo disc sensor - 12mm diameterYesYes
Built-in passive output chip resistorYes - 50K(at sensor)Yes - 75K(at sensor)
Termination of about 11 foot cable withDIN 45-802 Safety ConnectorsYesYes
Connects to physiological recorderYesYes
Comparable voltage output on test fixtureYesYes
Completely passive systemNo batteries or power supplyYesYes
Body attachment methodSelf-AdherentWrap (Tape)Velcro Strapand Pouch
For Sleep Disorders testingYesYes

Electrically, the devices are nearly identical and very simple. There is a piezo disc, a chip resistor, a 11 foot lead wire terminated with safety connectors and three solder points. The chip resistor value was modified to reduce noise. The sensors are encapsulated differently. The essential difference is the method of attachment to the patient's body -- Pro-Tech uses a Velcro strap and pouch arrangement and the Gereonics device uses 3M Coban LF latex free self-adherent wrap.

Data Supportive of Substantially Equivalent Claim:

Bench tests were conducted on the Pro-Tech predicate device and the Gereonics LMS for comparative performance test and comparative temperature test.

Comparative Performance Test:

The performance tests as performed on the Gereonics Test System show that the sensitivity to a controlled tapping device is very similar. The Pro-Tech test unit yielded a peak to peak voltage output of 300 millivolts compared with the Gereonics 600 millivolts. This is within the normal variability of other test devices at Gereonics. The output pulse waveform is identical on the test system, except for amplitude.

The sensors were also attached to the ankle using 3M Coban LF latex free self-adherent wrap and tested for sensitivity by lightly tapping the floor. The signal waveform and sensitivity were identical.

{2}------------------------------------------------

Comparative Temperature Test:

Both devices were tested across the temperature range of 8℃ to 43℃. Both performed similarly. The Pro-Tech temperature coefficient was about -5% per C and Gereonics -6% per C in the typically used range of 20 to 30°C.

Conclusions:

The electrical design of the Gereonics LMS is essentially identical to the predicate Pro-Tech device with the exception of output resistor value.

The performance as tested on the Gereonics tapping test system is very similar for both devices with identical pulse waveforms. Waveform height (sensitivity) varies somewhat between devices but can readily be accommodated by bioamplifier gain controls.

The performance of both devices was identical when attached to the ankle and tested by small repetitive taps on the floor.

The devices are encapsulated differently but perform similarly.

Gerald Rost Gerald Rost

Date: May 7, 2007

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three human figures intertwined beneath it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850

Mr. Gerald A. Rost Manager Gereonics, Incorporated 244 La Barranca Drive Solana Beach, California 92075

JUN 1 3 2007

Re: K062883

Trade/Device Name: Gereonics Ultra-Piezo Limb Movement Sensor Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: BZO Dated: May 19, 2007 Received: May 22, 2007

Dear Mr. Rost:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Rost

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sujata Y. Michaud

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number: K062883

Device Name: Gereonics Ultra-Piezo Limb Movement Sensor

INDICATIONS FOR USE:

For adult use.

The Gereonics Ultra-Piezo Limb Movement Sensor (LMS) is used to detect periodic limb movements for recording or monitoring on a physiological amplifier. The LMS is intended for use in sleep disorders studies.

CAUTION:

This product is for diagnostic purposes only and is not to be used in a life-sustaining circumstance. Federal law restricts this device to sale by or on the order of a Physician.

WARNING:

The sensor should not be immersed in fluids.

Prescription Use X (21 CFR 801 Subpart D) AND/OR

Over-The-Counter-Use (21 CFR 807 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

Consion Sign-Off) Civision of Anesthesiology, General Hospital, Intection Control, Dental Devices

510(k) Number: K062883

Page 1 of 1

June 2007

§ 868.2375 Breathing frequency monitor.

(a)
Identification. A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.(b)
Classification. Class II (performance standards).