K Number
K062711
Manufacturer
Date Cleared
2007-03-07

(177 days)

Product Code
Regulation Number
882.4840
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Baxano, Inc. Ultra Low Profile Rongeur is designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.

Device Description

The ULP Rongeur and Access Tools are used to access the neural foramen and decompress targeted areas. The Rongeur is pulled into the foramen with the distal handle and tissue is removed by pulling up on the distal handle to decompress the impinged nerve root. The Access accessories include a probe, guide and needle wire.

AI/ML Overview

Since this is a 510(k) premarket notification for a medical device (a surgical rongeur), the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are generally framed around demonstrating substantial equivalence to a predicate device, rather than explicit numerical performance targets like those for AI/Software as a Medical Device (SaMD).

Here's an analysis based on the provided text, structured to answer your questions where applicable, and noting where information typical for AI/SaMD studies is not present for this type of device:


Acceptance Criteria and Device Performance Study for the Baxano Ultra Low Profile Rongeur (K062711)

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Demonstrated Substantial Equivalence)Reported Device Performance and Evidence
Indications for Use: Same as predicate devices (accessing, cutting, and biting soft tissue and bone involving the spinal column).The ULP Rongeur has the same indications for use as the predicates; to access, cut, and bite soft tissue and bone involving the spinal column. The Access Tools also have the same intended uses as commercially available probes and guides for discectomy procedures.
Technological Characteristics: Same materials and similar design as predicate devices (stainless steel, curved, low profile for access to compromised neural areas).The ULP Rongeur is made of stainless steel and is curved, with a low profile, to allow access to compromised neural areas. These are the same materials and technological characteristics as the predicate devices. Access Tools (probe and guide) are substantially equivalent to existing commercial products.
Safety and Effectiveness: Demonstrate comparable safety and effectiveness to predicate devices; no new questions of safety or effectiveness.Mechanical Performance Tests: Conducted to verify the device meets design specifications and intended performance characteristics for removing bone. Cadaver Studies: Used by skilled physicians. Results demonstrated "comparable decompression with no visible damage to neural structures" and "an ease of use comparable to the standard of care," when compared to commercially available rongeurs. No specific quantitative data provided in this summary.

2. Sample Size for Test Set and Data Provenance

  • Test Set Description: The "test set" in this context refers to the cadaver studies and mechanical performance tests.
  • Sample Size:
    • Mechanical Performance Tests: Not specified.
    • Cadaver Studies: "multiple cadaver studies." The exact number of cadavers or individual tests is not quantified in the provided text.
  • Data Provenance: Not explicitly stated, but cadaver studies are typically conducted in a laboratory or research environment. It is not "retrospective or prospective" in the medical record sense, but rather an experimental study. The country of origin is not mentioned.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not explicitly stated, but referred to as "physicians skilled in procedures for decompressing neural foramen and lateral recess."
  • Qualifications: "skilled in procedures for decompressing neural foramen and lateral recess." Specific specialties (e.g., neurosurgeon, orthopedic spine surgeon) or years of experience are not provided.
  • Ground Truth Establishment: The physicians' feedback and observations during the cadaver studies, regarding "comparable decompression with no visible damage to neural structures" and "ease of use," served as the qualitative assessment against the "standard of care" (commercially available rongeurs).

4. Adjudication Method for the Test Set

  • No formal adjudication method (like 2+1, 3+1 consensus) is described. The results appear to be based on the observations and feedback of the participating "skilled physicians" during the cadaver studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a "Multi-Reader Multi-Case (MRMC) comparative effectiveness study" as understood in imaging or AI contexts was not performed. This device is a manual surgical instrument, not an AI or imaging diagnostic tool. The performance assessment was based on human expert evaluation of its use in a cadaveric setting compared to standard tools.
  • Therefore, an "effect size of how much human readers improve with AI vs. without AI assistance" is not applicable.

6. Standalone (Algorithm Only) Performance Study

  • No, a standalone (algorithm only without human-in-the-loop performance) study was not done. This device is a manual surgical rongeur and does not involve any algorithms or AI components.

7. Type of Ground Truth Used

  • The ground truth in this context was expert assessment/consensus (from skilled physicians during cadaver studies) regarding the device's ability to achieve "comparable decompression with no visible damage to neural structures" and "ease of use comparable to the standard of care," alongside mechanical performance data against design specifications. It is not pathology or outcomes data from live patients.

8. Sample Size for the Training Set

  • This concept is not applicable as the device is a manual surgical instrument and does not involve machine learning or AI models with training sets.

9. How the Ground Truth for the Training Set Was Established

  • This concept is not applicable given the nature of the device.

Summary of Device-Specific Context:

It's crucial to understand that this document describes a traditional medical device (a surgical tool), not a software or AI-driven diagnostic device. Therefore, the "acceptance criteria" and "study" focus on demonstrating substantial equivalence to existing, legally marketed predicate devices through:

  • Comparison of Indications for Use.
  • Comparison of Technological Characteristics (materials, design).
  • Non-Clinical Performance Data (mechanical tests and cadaver studies) to confirm that the device performs as intended and introduces no new safety or effectiveness concerns compared to established methods/devices.

The type of rigorous, quantitative performance metrics and study designs common for AI/SaMD are not typically found in these types of 510(k) submissions for manual surgical instruments.

{0}------------------------------------------------

K062711

MAR 0 7 2007

510(K) SUMMARY (21 CFR 807.92)

ULTRA LOW PROFILE RONGEUR

  • 510(k) Owner: Baxano, Inc. 2660 Marine Way, Suite B Mountain View, CA 94043 Tel: 650-937-1400 Fax: 650-937-1410
  • Contact Person: Jefferey Bleam Tel: 650-937-1400 ext. 408 E-mail: jbleam@baxano.com

Date Prepared: September, 2006

Trade Name: Ultra Low Profile (ULP) Rongeur and Access Tools

  • Common Name: Manual rongeur for cutting or biting bone in the skull or spinal column
    Classification: Manual rongeur (per 21 CFR section 882.4840)

Codman Laminectomy Shaver, Integra Ruggles Instrument, US Predicate Devices: Surgical Kerrison Spinal Rongeur, Zeppelin Laminectomy Rongeur Ellman Disc FX System (access tools)

  • Device Description: The ULP Rongeur and Access Tools are used to access the neural foramen and decompress targeted areas. The Rongeur is pulled into the foramen with the distal handle and tissue is removed by pulling up on the distal handle to decompress the impinged nerve root. The Access accessories include a probe, guide and needle wire.
  • Intended Use: The Baxano, Inc. Ultra Low Profile Rongeur is designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.

{1}------------------------------------------------

Substantial
Equivalence:The ULP Rongeur has the same indications for use as the predicates; to access, cut, and bite soft tissue and bone involving the spinal column. In addition, the Access Tools have the same intended uses; to access, guide, and probe in neurosurgical applications. Any differences between the ULP Rongeur and the predicates are specific in design and do not raise new questions of safety or effectiveness. The device and the predicates are labeled for open decompression procedures.
Technological
Characteristics:The ULP Rongeur is made of stainless steel and is curved, with a low profile, to allow access to compromised neural areas when exposure is otherwise difficult to obtain. These are the same materials and technological characteristics as the predicate devices. The Access Tools include a probe and guide substantially equivalent to probes and guides commercially available for discectomy procedures in the lumbar spine.
Non-Clinical
Performance Data:Mechanical performance tests were conducted to verify that the device meets design specifications and intended performance characteristics, based on the application for removing bone in compromised neural areas.
The ULP Rongeur was used in multiple cadaver studies by physicians skilled in procedures for decompressing neural foramen and lateral recess. Results demonstrated that the ULP Rongeur provided comparable decompression with no visible damage to neural structures, and an ease of use comparable to the standard of care, when compared to commercially available ronguers.
Conclusions:Baxano has determined, based on the performance testing and cadaver studies, that the ULP Rongeur and Access Tools conform to the design specifications and are at least as safe and effective as the predicate devices for accessing and decompressing bone in the spinal column.

Any statement regarding "substantial equivalence" made in this 510(k) submission and summary only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not interpreted as an admission or any other type of evidence in ary patent proceeding, including patent infringement, litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting to the scope of any patent or patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Baxano, Inc. % Rockwell & Associates Ms. Sharon Rockwell Vice-President, Regulatory and Clinical Affairs 2660 Marine Way, Suite B Mountain View, California 94043

MAR 0 7 2007

Re: K062711

Trade/Device Name: Baxano Ultra Low Profile Rongeur Regulation Number: 21 CFR 882.4840 Regulation Name: Manual rongeur Regulatory Class: II Product Code: HAE Dated: February 26, 2007 Received: February 27, 2006

Dear Ms. Rockwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Sharon Rockwell

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

For Pedro Ruiz

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Baxano, Inc. Ultra Low Profile Rongeur

TAB 4

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): ________________

Device Name: Baxano Ultra Low Profile Rongeur

Indications for Use:

The Baxano, Inc. Ultra Low Profile Rongeur is designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off) Division of Ceneral. Restorative, and Neurological Devices

Page of

510(k) Number |\ b(27)1

4-1

Confidential: This document and the information contained herein may not be reproduced, used or disclosed without written permission from Baxano, Inc.

§ 882.4840 Manual rongeur.

(a)
Identification. A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.(b)
Classification. Class II (performance standards).