K Number
K062484

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-06-08

(287 days)

Product Code
Regulation Number
868.5630
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are reusable single patient use nebulizers which are used where a fine aerosol mist of medication must be delivered to a patient for the treatment of respiratory disorders. The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are intended to be used for both adult and pediatric patients who have been prescribed inhalation therapy or medication in both home health care and hospital use. The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers may be used with compressed air or an air gas source providing air flow between 4 to 10 lmin.

Device Description

The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are hand-held, reusable, single patient use and pneumatically powered nebulizers. Both devices are made using similar materials and are used by patients with the same accessories set (mouthpiece, masks etc ... ). The primary difference between the two devices is that the RF6 Plus is equipped with inspiration and expiration valves while the RF6 Basic is not.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Flaem Nuova RF6 Basic and RF6 Plus Nebulizers, based on the provided text:

Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Implicit)Reported Device Performance (Flaem Nuova RF6 Basic & Plus)
MMAD (Mass Median Aerodynamic Diameter)Equivalent to predicate devicesEquivalent to predicate devices
FPF (Fine Particle Fraction)Equivalent to predicate devicesEquivalent to predicate devices
DD (Deposited Dose)Equivalent to predicate devicesEquivalent to predicate devices
DDR (Deposited Dose Rate)Equivalent to predicate devicesEquivalent to predicate devices

Study Details

  1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: Not specified. The document states "bench tests presented" but does not give specific numbers of nebulizer units or test runs.
    • Data Provenance: Not explicitly stated, but given that Flaem Nuova is based in Italy, the testing was likely conducted in Europe, possibly Italy, and the data was submitted to the FDA in the US. The term "bench tests" implies laboratory conditions.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not applicable. This study primarily involved "performance tests" and "bench tests" to evaluate physical device parameters (MMAD, FPF, DD, DDR). These are objective measurements, not subjective evaluations requiring expert consensus for ground truth.
    • Qualifications of Experts: Not applicable.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. The performance evaluation was based on objective physical measurements, not human interpretation that would require an adjudication process.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This document describes the testing of a medical device (nebulizer), not an AI algorithm for diagnostic imaging or similar applications. Therefore, an MRMC study is not relevant to this submission.
    • Effect Size: Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Study: Not applicable. This is not an AI algorithm. The performance tests ("bench tests") are a form of standalone testing for the device's technical specifications.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: The ground truth for this study was established through objective physical measurements and standardized testing procedures for nebulizer performance parameters (MMAD, FPF, DD, DDR). The comparison was made against the published or known performance of the predicate devices.
  7. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This device is not an AI/ML algorithm that requires a training set.
  8. How the ground truth for the training set was established:

    • Ground Truth Establishment for Training Set: Not applicable. This device is not an AI/ML algorithm.

{0}------------------------------------------------

FLAEM NUOVA RF6 BASIC AND RF6 PLUS NEBULIZERS CONFIDENTIAL

K062484

510(k) Summary of Safety and Effectiveness for the Flaem Nuova RF6 Basic and RF6 Plus Nebulizers

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

1. General Information

Contact Person:

Submitter:FLAEM NUOVA
Via Colli Storici 221, 223, 225
S. Martino della Battaglia
Brescia 25010 Italy

Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Telephone: 978-207-1245 Fax: 978-207-1246

Summary Preparation Date:

August 23, 2006

  1. Names

Device Name:

RF6 Basic RF6 Plus

Classification Name:

Nebulizer Regulation number: 868.5630 Product Code: CAF

3. Predicate Devices

The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are substantially equivalent to a combination of the following devices:

  • Salter Labs, 8660 Series- K962879
  • V Pari LC STAR Reusable Nebulizer 22F50 K963924
  • V Medic-Aid - Sidestream Nebulizer - K991725

4. Device Description

The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are hand-held, reusable, single patient use and pneumatically powered nebulizers. Both devices are made using similar materials and are used by patients with the same accessories set (mouthpiece, masks etc ... ). The primary difference

{1}------------------------------------------------

FLAEM NUOVA RF6 BASIC AND RF6 PLUS NEBULIZERS

between the two devices is that the RF6 Plus is equipped with inspiration and expiration valves while the RF6 Basic is not.

5. Indications for Use

The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are reusable single patient use nebulizers which are used where a fine aerosol mist of medication must be delivered to a patient for the treatment of respiratory disorders. The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are intended to be used for both adult and pediatric patients who have been prescribed inhalation therapy or medication in both home health care and hospital use. The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers may be used with compressed air or an air gas source providing air flow between 4 to 10 lmin.

6. Performance Data

The performance tests performed by Flaem Nuova, as required by the "Reviewer guidance for nebulizers, metered dose inhalers, spacers and actuators" issued on October 1993, demonstrate that the Flaem Nuova RF6 Basic and RF6 Plus Nebulizers have the same effectiveness as the identified predicate devices because they have equivalent performance parameters as the predicate devices including: MMAD, FPF, DD and DDR. MMAD, FPF, DD and DDR are the parameters used to evaluate the performance of nebulizers. Clinical data are not submitted because they are not required.

7. Comparison to Predicate Devices

The Flaem Nuova RF6 nebulizers and their predicate devices are indicated for the same intended use and have equivalent design solutions as the predicate devices. They have the same atomization principle used by all the nebulizers. The RF6 Plus has the same the valve system as that used in the Pari LC STAR (K963924). The RF6 Basic is similar to the Medic-Aid SIDESTREAM Nebulizer (K991725). In addition, the Flaem Nuova RF6 Nebulizers performance is equivalent to the predicate devices as shown in the bench tests presented.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 8 2007

FLAEM NUOVA S.p.A. C/O Maureen O'Connell Regulatory Consultant O'Connell Regulatory Consultants, Incorporated 5 Timber Lane North Reading, Massachusetts 01864

Re: K062484

Trade/Device Name: Flaem Nuova RF6 Basic and RF6 Plus Nebulizers Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: June 05, 2007 Received: June 06, 2007

Dear Mr. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. O'Connell

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sigute Y. Michaud MD

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Eyaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Flaem Nuova RF6 Basic and RF6 Plus Nebulizers

Indications for Use:

The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are reusable single patient use nebulizers which are used where a fine aerosol mist of medication must be delivered to a patient for the treatment of respiratory disorders. The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers are intended to be used for both adult and pediatric patients who have been prescribed inhalation therapy or medication in both home health care and hospital use. The Flaem Nuova RF6 Basic and RF6 Plus Nebulizers may be used with compressed air or an air gas source providing air flow between 4 and 10 lmin.

NOTE

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

010(k) Number ...

்லா Sign-Off) ാന ാന്ദ്രന-ഗ്രീ)
Asion of Anesthesiology, General Hospital Inston Control, Dental Devices

KO6248

Page 1 of 1

510(K) FILE

§ 868.5630 Nebulizer.

(a)
Identification. A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.(b)
Classification. Class II (performance standards).