(135 days)
The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, end tidal CO2, 12-Lead ECG monitoring, non-invasive blood pressure measurement, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL E Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The device is light weight and easy to carry for transport.
The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing a patient's ECG signal via the Advisory feature on the device The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in the event of a shock advised determination. The user will be prompted to re-assess the patient in the event a no shock advised determination is returned.
Defibrillator Function: The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation. The E Series products must be prescribed for use by a physician or medical advisor of an emergecy response team.
Intended Use - Manual Operation: Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions: Unconsciousness, Absence of breathing, and Absence of pulse. This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. In manual mode, the E Series unit may also be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardias (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate. The advisory function should be used to confirm ventricualr fibrillation and wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (previously listed).
Intended Use - Semiautomatic Operation (AED): The E Series AED unit is designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient. They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol. Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation.
Intended Use - CPR Monitoring: The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients. The CPR monitoring function is not intended for use on patients under 8 years of age.
External Pacemaker (Pacer Version Only) Intended Use - Pacemaker: This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include: Resuscitation from standstill or bradycardia of any etiology, As a standby when standstill or bradycardia might be expected, Suppression of tachycardia, Pediatric Pacing.
Monitor Intended-Use Multi-parameter Monitoring: This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP). ECG monitoring is performed by connecting the patient to the unit via the 3 or 5 lead patient cable. MFE Pads, or through the paddles. SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions. EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO2) and respiration rate for adult, pediatric and neonatal patients. 12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias. NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
The ZOLL E Series with Real CPR Help™ provides 'real-time' user CPR assistance when used in conjunction with ZOLL CPR-D-padz™ Multi-function Electrodes. CPR feedback is provided via ZOLL's unique sensor assembly that relays user compression data to the ZOLL E Series. The ZOLL E Series with Real CPR Help™ provides a metronome to encourage rescuers to perform CPR at the AHA/ECR recommended rate of 100 compressions per minute (CPM). The ZOLL E Series additionally provides chest compression performance feedback to the user through displayed symbols, text messages and voice prompts.
The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, end tidal CO2, 12-Lead ECG monitoring, non-invasive blood pressure measurement, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL E Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The device is light weight and easy to carry for transport.
The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing a patient's ECG signal via the Advisory feature on the device The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in the event of a shock advised determination. The user will be prompted to re-assess the patient in the event a no shock advised determination is returned.
The ZOLL E Series with Real CPR Help™ is a device that provides real-time CPR assistance and monitoring functions. The primary focus of the performance testing for this device, as described in the provided text, is on ensuring substantial equivalence to a predicate device, particularly concerning the new "Real CPR Help™" feature.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria or specific numerical performance metrics for the ZOLL E Series with Real CPR Help™. Instead, the performance testing focuses on demonstrating substantial equivalence to existing predicate devices and compliance with functional requirements and safety standards.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| CPR Compression Performance Feedback | The acquired depth and rate signals are used to provide CPR compression performance feedback to the user through displayed symbols, text messages, and voice prompts. This feedback is deemed equivalent to other ZOLL defibrillators with user CPR feedback. |
| Functional Requirements | Extensive performance testing ensures that the ZOLL E Series with Real CPR Help™ performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications. |
| Safety Testing | Safety testing assures the device complies with applicable sections of recognized industry and safety standards. |
| Software Modifications Impact | Functional testing of the device's features and functions was conducted to ensure that the modifications to the software (for Real CPR Help™) did not degrade or impact other product features, functions or performance specifications. |
| Substantial Equivalence (Overall Conclusion) | Performance and safety testing of the ZOLL E Series with Real CPR Help™ demonstrates that its features and functions are substantially equivalent to that of the indicated commercially distributed device with regard to performance, safety, and effectiveness. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a sample size for any test set or the data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be functional and safety testing rather than a clinical trial with a patient test set.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not mention using experts to establish ground truth for a test set. The validation seems to be based on functional and safety testing to demonstrate equivalence and compliance with standards rather than expert-adjudicated ground truth data.
4. Adjudication Method for the Test Set
The document does not specify an adjudication method. As no expert ground truth establishment is mentioned, an adjudication method would not be applicable in this context.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The focus is on the device's own performance and its substantial equivalence to predicate devices, not on the improvement of human readers with AI assistance.
6. Standalone (Algorithm Only) Performance Study
The document does not explicitly describe a "standalone" or "algorithm only" performance study in the context of typical AI/ML medical device evaluations. The "Real CPR Help™" function is an integrated feature of the ZOLL E Series device, and its performance is assessed as part of the overall device functionality and equivalence. The CPR feedback mechanism relies on a sensor and algorithm, but its performance is discussed in terms of its output (displayed symbols, text, voice prompts) being equivalent to other ZOLL defibrillators with CPR feedback, implying an integrated system test.
7. Type of Ground Truth Used
The "ground truth" for the performance testing appears to be:
- Predicate Device Performance: The functional requirements and performance specifications of the previously marketed ZOLL Defibrillators equipped with CPR Feedback.
- Industry and Safety Standards: Applicable sections of recognized industry and safety standards to which the device must comply.
- Internal Functional Requirements: The device's own design specifications for features and functions.
There is no mention of pathology, outcomes data, or expert consensus on clinical cases as the ground truth.
8. Sample Size for the Training Set
The document does not mention a training set or its sample size. This device is described as an upgrade to an existing product line (ZOLL E Series Defibrillator) with the addition of the "Real CPR Help™" feature, rather than a new AI/ML algorithm developed from scratch requiring a distinct training dataset. The "Real CPR Help™" acquired signals are used to provide feedback, implying a rule-based or engineered system rather than a machine learning model requiring a large training dataset as commonly understood today.
9. How the Ground Truth for the Training Set Was Established
Since no training set is mentioned, the method for establishing its ground truth is not applicable and therefore, not specified.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten number, which appears to be "1062177". The numbers are written in a cursive style with varying stroke thicknesses. The image also contains the partial text "OU Medical Corporation" at the bottom left.
ZOLL Me
Worldwide Headquarters
269 Mill Road
Chelmsford, MA 01824-4105
U.S.A.
978 421-9655
978 421-0025 Main Fax
510(k) Summary:
Submitter's Name and Address:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person:
Sean Reynolds
(978) 421-9655, Ext. 9386
Date Summary Prepared:
December 11, 2006
Device:
ZOLL E Series with Real CPR Help™
Classification:
Automatic External Defibrillators: Class III (21 CFR 870.5310)
Cardiopulmonary Resuscitation Aid: Class III (21 CFR 870.5200)
Description:
The ZOLL E Series with Real CPR Help™ provides 'real-time' user CPR assistance when used in conjunction with ZOLL CPR-D-padz™ Multi-function Electrodes. CPR feedback is provided via ZOLL's unique sensor assembly that relays user compression data to the ZOLL E Series. The ZOLL E Series with Real CPR Help™ provides a metronome to encourage rescuers to perform CPR at the AHA/ECR recommended rate of 100 compressions per minute (CPM). The ZOLL E Series additionally provides chest compression performance feedback to the user through displayed symbols, text messages and voice prompts.
The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, end tidal CO2, 12-Lead ECG monitoring, non-invasive blood pressure measurement, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL E Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability
Image /page/0/Picture/20 description: The image shows the word "ZOLL" in large, bold letters at the top. Below the word, the phrase "Advancing Resuscitation. Today." is written in a smaller font. The image appears to be a logo or advertisement for a company or product related to resuscitation.
DEC 1 3 2006
{1}------------------------------------------------
to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The device is light weight and easy to carry for transport.
The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing a patient's ECG signal via the Advisory feature on the device The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in the event of a shock advised determination. The user will be prompted to re-assess the patient in the event a no shock advised determination is returned.
Indications for Use:
Defibrillator Function
The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation.
The E Series products must be prescribed for use by a physician or medical advisor of an emergecy response team.
Intended Use - Manual Operation
Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions:
- . Unconsciousness
- . Absence of breathing, and
- . Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
In manual mode, the E Series unit may also be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardias (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate.
The advisory function should be used to confirm ventricualr fibrillation and wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (previously listed).
Intended Use - Semiautomatic Operation (AED)
The E Series AED unit is designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation.
{2}------------------------------------------------
Intended Use - CPR Monitoring
The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.
The CPR monitoring function is not intended for use on patients under 8 years of age.
External Pacemaker (Pacer Version Only)
Intended Use - Pacemaker
This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
Note: This device must not be connected to internal pacemaker electrodes.
The purposes of pacing include:
Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, b- blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, druq toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous therapy that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
Suppression of tachycardia:
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.
Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL pediatric MFE Pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.
{3}------------------------------------------------
Monitor
Intended-Use Multi-parameter Monitoring
This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP).
ECG monitoring is performed by connecting the patient to the unit via the 3 or 5 lead patient cable. MFE Pads, or through the paddles.
SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions.
EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO2) and respiration rate for adult, pediatric and neonatal patients.
12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias.
NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
Substantial Equivalence:
The features and functions of the ZOLL E Series with Real CPR Help™ are identical to those of the E Series Defibrillator, with the exception of the incorporation of Real CPR Help™. The characteristics, features and functions of the Real CPR Help™ are substantially equivalent to other ZOLL Defibrillators equipped with CPR Feedback.
Comparison of Technological Characteristics
The ZOLL E Series with Real CPR Help™ acquires compression depth and rate signals from the CPR sensor incorporated in the currently marketed CPR-D.padz™ Multi-function Electrodes. The acquired depth and rate signals are used to provide CPR compression performance feedback to the user through displayed symbols, text messages and voice prompts which are equivalent to other ZOLL defibrillators with user CPR feedback.
Performance Testing:
Extensive performance testing ensures that the ZOLL E Series with Real CPR Help™ performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications. Safety testing assures the device complies with applicable sections of recognized industry and safety standards.
Functional testing of the device's features and functions was conducted to ensure that the modifications to the software did not degrade or impact other product features, functions or performance specifications.
Conclusion
Performance and safety testing of the ZOLL E Series with Real CPR Help™ demonstrates that its features and functions are substantially equivalent to that of the indicated commercially distributed device with regard to performance, safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines forming its body and wings. The seal is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 3 2006
ZOLL Medical Corporation c/o Sean Reynolds Regulatory Affairs Engineer 269 Mill Road Chelmsford, MA 01824-4105
Re: K062177
Trade/Device Name: ZOLL E Series Defibrillator with Real CPR Help Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, LIX Dated: November 13, 2006 Received: November 29, 2006
Dear Mr. Reynolds:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Reynolds
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
B.Zimmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
SECTION 4 - INDICATIONS FOR USE
510(k) Number (if known): K062177
ZOLL E Series Device Name:
Defibrillator Function
The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation.
The E Series products must be prescribed for use by a physician or medical advisor of an emergecy response team.
Intended Use - Manual Operation
Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions:
- . Unconsciousness
- . Absence of breathing, and
- Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
In manual mode, the E Series unit may also be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardias (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate.
The advisory function should be used to confirm ventricualr fibrillation and wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (previously listed).
Intended Use - Semiautomatic Operation (AED)
The E Series AED unit is designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
(continued on next page)
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|---|
| All man concerner |
| (Division Sign-Off) |
|---|
| --------------------- |
Division of Cardiovascular Devices
| 510(k) Number | K062177 |
|---|---|
| --------------- | --------- |
{7}------------------------------------------------
Indications for Use (continued from previous page)
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation.
Intended Use - CPR Monitoring
The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.
The CPR monitoring function is not intended for use on patients under 8 years of age.
External Pacemaker (Pacer Version Only)
Intended Use — Pacemaker
This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
Note: This device must not be connected to internal pacemaker electrodes.
The purposes of pacing include:
Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, guinidine, digitalis, b- blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
Suppression of tachycardia:
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.
Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL pediatric MFE Pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.
(continued on next page)
{8}------------------------------------------------
Monitor
Intended-Use Multi-parameter Monitoring
This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP).
ECG monitoring is performed by connecting the patient to the 3 or 5 lead patient cable, MFE Pads, or through the paddles.
SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions.
EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (ECO2) and respiration rate for adult, pediatric and neonatal patients.
12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias.
NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
§ 870.5310 Automated external defibrillator system.
(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.