K Number
K062033

Validate with FDA (Live)

Date Cleared
2007-03-23

(248 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BP101 Digital Blood Pressure Monitor is intended to be used to measure blood pressure (systolic and diastolic) and heart rate from the upper arm using the oscillometric method. The device is intended for use in only adult population, not applied to the other populations such as neonatal baby. It can not be used while the arm has bleeding or wound to avoid the blood flowing from the wound in pressurizing. It can be used as medical assistant instrument at home or in medical center for adult population for measuring systolic and diastolic blood pressure and heart rate.

Device Description

BP101 Digital Blood Pressure Monitor is a fully automatic non-invasive blood pressure monitor which measures systolic and diastolic blood pressure and heart rate of adult population using the oscillometric method by inflating an inflatable cuff on the upper arm. BP101 Digital Blood Pressure Monitor uses an inflated cuff which is The systolic and diastolic blood pressures are determined by Oscillometric method. The deflation rate is controlled by an automatic air-release valve at a constant rate 2~7mmHg/second. The user can deflate the cuff to stop measuring by pressing the "START/STOP" button at any time while measuring. The measuring result is displayed in LCD.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the BP101 Digital Blood Pressure Monitor:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Performance Standards: Conformity to applicable standardsConformity to ANSI/AAMI SP-10:2002, EN 1060-1, EN-1060-3, and IEC60601-1-1-2 requirements.
Clinical Repeatability: Statistically and clinically acceptableComparison study demonstrated that the clinical repeatability of BP101 is statistically and clinically acceptable.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "a comparison study," but does not provide the number of participants or measurements.
  • Data Provenance: Not explicitly stated regarding country of origin or whether it was retrospective or prospective. It is implied to be a prospective clinical study given the nature of a "comparison study" to validate performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not explicitly stated. The document refers to "trained observers" who used the auscultatory method.
  • Qualifications of Experts: "Trained observers." No further specific qualifications (e.g., years of experience, medical specialty) are provided.

4. Adjudication Method for the Test Set

  • Not explicitly stated. The document mentions a "comparison study with a device that uses auscultatory method used by trained observers." This implies the auscultatory readings by trained observers served as the reference standard, but it doesn't detail any adjudication process (e.g., if multiple observers' readings were compared or averaged).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • No. This was not an MRMC comparative effectiveness study involving AI assistance for human readers. This device is a standalone blood pressure monitor, not an AI-powered diagnostic aid for clinicians.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes. The primary study described is a standalone performance validation of the BP101 device. The device itself (algorithm and hardware) performed the measurements, which were then compared to the auscultatory method.

7. The Type of Ground Truth Used

  • Expert Consensus/Reference Standard: The ground truth for the comparison study was established using the auscultatory method performed by trained observers. This is widely considered a clinical reference standard for blood pressure measurement.

8. The Sample Size for the Training Set

  • Not applicable/Not stated. The document does not describe the use of a training set as would be relevant for machine learning or AI algorithms that "learn" from data. This device likely relies on established oscillometric principles and hardware calibration, rather than a data-driven training approach.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable/Not stated. As mentioned above, a training set based on data with established ground truth is not described for this type of device and validation.

{0}------------------------------------------------

K062033/s2
p1/3

510(k) Summary

  1. Submitter's Name: Mr. Dacheng Gong, Manager

MAR 2 3 2007

KINGYIELD HONGKONG LIMITED

Address: 12 Floor, No. 180 Electric Road, North Point, Hong Kong.

Phone: +86-755-27371997

Fax: +86-755-27331856

E-mail:kyield@szonline.net

Contact: Mr. Dacheng Gong, Manager

Date of Summary: Jan. 18 2006

  1. Trade Name: BP101 Digital Blood Pressure Monitor

  2. Classification Name

Non-invasive Blood Pressure Measurement System.

Regulation Number: 21 CFR 870.1130

Class: II

4.Product Code: DXN

5.Classification Panel: Cardiovascular

  1. Predicate Device: HEM-757 Automatic Digital Blood Pressure Monitor with Intelligence marketed by Omron Healthcare INC.

K 001670 (Predicate)

7.Device Description:

BP101 Digital Blood Pressure Monitor is a fully automatic non-invasive blood pressure monitor which measures systolic and diastolic blood

{1}------------------------------------------------

K06.20.33/52
p 2/3

pressure and heart rate of adult population using the oscillometric method by inflating an inflatable cuff on the upper arm.

The Intended use:

BP101 Digital Blood Pressure Monitor is intended to be used to measure blood pressure (systolic and diastolic) and heart rate from the upper arm using the oscillometric method.

The device is intended for use in only adult population, not applied to the other populations such as neonatal baby.

It can not be used while the arm has bleeding or wound to avoid the blood flowing from the wound in pressurizing.

8. Performance and Technological Characteristics :

8.1 Performance Summary

In terms of operating specification, Safety & EMC requirements, the device conform to applicable standards including ANSI/AAMI SP-10:2002, EN 1060-1, EN-1060-3, and IEC60601-1-1-2 requirements.

A comparison study with a device that uses auscultatory method used by trained observers was performed to validate the performance of BP101. The comparison study demonstrated that the clinical repeatability of BP101 is statistically and clinically acceptable.

8.2 Technological Characteristics

BP101 Digital Blood Pressure Monitor uses an inflated cuff which is

{2}------------------------------------------------

The systolic and diastolic blood pressures are determined by Oscillometric method. The deflation rate is controlled by an automatic air-release valve at a constant rate 2~7mmHg/second. The user can deflate the cuff to stop measuring by pressing the "START/STOP" button at any time while measuring. The measuring result is displayed in LCD.

Clark -

9.Conclusion

BP101 Digital Blood Pressure Monitor has the same intended use, the same technological characteristics as the predicate device HEM-757, Moreover, non-clinical testing & clinical testing contained in this submission demonstrated that any difference in their technological characteristics doesnot raise any new issues of safety and effectiveness. In a word, BP101 is substantial equivalent to the predicate device HEM-757.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wings and body. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 3 2007

Welch Allyn Protocol, Inc. c/o Mr. Tamas Borsai Program Manager, Third Party Review Program 12 Commerce Rd. Newton, CT 06470

Re: K062033

Trade/Device Name: BP101 Digital Blood Pressure Measurement System Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: March 8, 2007 Received: March 12, 2007

Dear Mr. Borsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Borsai

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Blymnemon for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K062033

Device Name: BP101 Digital Blood Pressure Monitor

Indications For Use:

It can be used as medical assistant instrument at home or in medical center for adult population for measuring systolic and diastolic blood pressure and heart rate.

The device is intended for use in only adult population, not applied to the other populations such as neonatal baby.

It can not be used while the arm has bleeding or wound to avoid the blood flowing from the wound in pressurizing.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

B. Almon

Page 1 of 1

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).