K Number
K061549
Device Name
AMNIOSTAT-FLM-PG
Date Cleared
2006-10-18

(135 days)

Product Code
Regulation Number
862.1455
Panel
CH
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AmnioStat-FLM-PG Kit is intended to be used for the immunological detection of phosphatidylglycerol (PG) in amniotic fluid to determine fetal lung maturity of the fetus collected from transabdominal or vaginal pool samples.

Device Description

The AmnioStat-FLM-PG Kit is an immunological semi-quantitative agglutination test for determining the presence of phosphatidylglycerol in human amniotic fluid at concentrations indicative of fetal lung maturity. The AmnioStat-FLM-PG Kit is comprised of five (5) components and a disposable agglutination cards (cardboard with laminate). The five (5) components that comprise the kit include three (3) kit controls, Negative Control, Catalog #91013, Low Positive Control, Catalog #91027, High Positive Control, Catalog #91028, two (2) reagents, Reagent A, Catalog #91012 and Reagent B, Catalog #91029, and one (1) Buffer, Catalog #91011. The three (2) controls (negative, low positive, and high positive) in the AmnioStat-FLM-PG Kit are run with each test. The AmnioStat-FLM-PG Kit can be used to run six (6) patient test samples. Each patient sample is run on an individual agglutination card with a set of controls. The AmnioStat-FLM-PG Kit is a next generation kit based upon the same immunological agglutination test as AmnioStat-FLM (K822150) previously cleared for marketing. The immunological agglutination reaction is the same between the AmnioStat-FLM (K822150) and the AmnioStat-FLM-PG kit. The AmnioStat-FLM-PG Kit incorporates the use of Sudan Black B into one of the kit reagents which makes the differentiation of the three (3) kit controls more distinguishable. The AmnioStat-FLM (K822150) is performed on a glass slide that requires the use of a mirror to distinguish and differentiate the kit controls based upon clarity of the background and the size of the agglutinated particles. The AmnioStat-FLM-PG is performed on a disposable agglutination card that is white and does not require the use of a mirror to distinguish and differentiate the kit controls.

AI/ML Overview

The provided text describes a 510(k) submission for the AmnioStat-FLM-PG Kit, an immunological semi-quantitative agglutination test for determining the presence of phosphatidylglycerol (PG) in human amniotic fluid to assess fetal lung maturity.

Here's an analysis of the acceptance criteria and study information provided in the document:


1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly define quantitative acceptance criteria for the AmnioStat-FLM-PG Kit. Instead, the performance is described qualitatively and through a comparative study.

Acceptance Criteria (Implicit)Reported Device Performance
Functional Equivalence to Predicate Device (AmnioStat-FLM K822150)- "The immunological agglutination reaction is the same between the AmnioStat-FLM (K822150) and the AmnioStat-FLM-PG kit."
  • Clinical comparison data for 73 patient samples yielded "no false positive test results" and "demonstrated that the AmnioStat-FLM-PG kit performs in accordance with the intended use of the product." This implies comparable accuracy to the predicate for determining the presence of PG. |
    | Ease of Use / User-Friendliness (Improvement over predicate) | - "The AmnioStat-FLM-PG Kit incorporates the use of Sudan Black B into one of the kit reagents which makes the differentiation of the three (3) kit controls more distinguishable."
  • "The AmnioStat-FLM-PG is performed on a disposable agglutination card that is white and does not require the use of a mirror to distinguish and differentiate the kit controls."
  • "The response and feedback from the laboratories that evaluated the kit indicates it was well received in terms of its application, ease of use..." |
    | Effectiveness in Determining Fetal Lung Maturity (as indicated by PG presence in amniotic fluid) | - "...effectiveness in the determining fetal lung maturity in human amniotic fluid."
  • The comparative study with 73 patient samples reported "no false positive test results" and confirmed performance "in accordance with the intended use," which is to identify PG for fetal lung maturity assessment. |
    | Reliable Control Functionality | - "Functionality of the three (3) kit controls (negative, low positive, and high positive) in the AmnioStat-FLM-PG Kit with regards to differentiation in appearance is performed as a condition of release." (This is a manufacturing quality control, not a study outcome). |
    | Compliance with Regulation (21 CFR Section 862.1455) | - "meets the criteria as outlined in 21 CFR Section 862.1455, Lecithin/Sphingomyelin ratio in amniotic fluid test system." (This is a self-assessment statement). |

Summary of Acceptance: The acceptance is primarily based on demonstrating a substantially equivalent performance to the predicate device, with an improvement in ease of use and clarity of results. Specific quantitative performance metrics like sensitivity, specificity, or PPV/NPV are not provided in the summary.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 73 patient samples during the clinical comparison study.
  • Data Provenance: The study involved two (2) clinical laboratories. The country of origin is not specified but implied to be the US given the submission to the FDA. The data is retrospective in nature, as indicated by "clinical comparison data was also generated for seventy-three (73) patient samples that were tested between the two (2) clinical laboratories." This suggests pre-existing samples or samples collected and analyzed for comparison, rather than a prospective study designed to evaluate the new device exclusively from scratch.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not specify the number of experts used to establish ground truth or their qualifications. The "clinical comparison data" likely refers to results from the predicate device, AmnioStat-FLM (K822150), being used as the comparative "truth" or gold standard.


4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set. It describes a comparison between the new device and the predicate device's results.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not explicitly performed or described. The study was a comparison of results between two devices on patient samples, not an evaluation of human readers' performance with and without AI assistance (as this is a diagnostic test kit, not an AI-based imaging interpretation tool). Therefore, no effect size for human reader improvement with AI assistance is applicable or reported.


6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, this was a standalone performance evaluation of the diagnostic test kit itself. The AmnioStat-FLM-PG Kit is an in-vitro diagnostic device designed to provide a result directly, not an algorithm that assists a human in interpretation. The "performance data" section describes the kit's performance rather than human performance with the kit.


7. The Type of Ground Truth Used

The ground truth used was implied to be the results obtained from the predicate device, AmnioStat-FLM (K822150). The document states, "The results of the comparison study yielded no false positive test results and demonstrated that the AmnioStat-FLM-PG kit performs in accordance with the intended use of the product." This indicates the predicate device's findings were used as the reference to assess the new device's accuracy (specifically, in terms of false positives). The ultimate ground truth relates to the presence of phosphatidylglycerol, which the predicate device is also designed to detect.


8. The Sample Size for the Training Set

The document does not mention a separate training set. This is not an AI/machine learning device that typically requires distinct training, validation, and test sets. It's a chemical/immunological diagnostic assay. The samples mentioned (73 patient samples) are for the clinical comparison study, analogous to a test or validation set.


9. How the Ground Truth for the Training Set Was Established

Since no training set is described for this type of device, this question is not applicable.

§ 862.1455 Lecithin/sphingomyelin ratio in amniotic fluid test system.

(a)
Identification. A lecithin/sphingomyelin ratio in amniotic fluid test system is a device intended to measure the lecithin/sphingomyelin ratio in amniotic fluid. Lecithin and sphingomyelin are phospholipids (fats or fat-like substances containing phosphorus). Measurements of the lecithin/sphingomyelin ratio in amniotic fluid are used in evaluating fetal maturity.(b)
Classification. Class II.