(141 days)
Endoscopic surgical procedures. It is a family of instruments which includes graspers, Linsectors, and scissors, which are intended to be used to grasp, manipulate, cut, and cauterize soft tissue.
The GENICON disposable electrosurgical instruments are single use sterile instruments made from biocompatible plastic and stainless steel with a working length of 33cm. Current may be supplied by an approved electrosurgical generator which provides the ability for the coagulation of tissue when used with an appropriate ground electrode.
The provided text describes the Genicon Disposable Electrosurgical Instrumentation device, its intended use, and its substantial equivalence to a predicate device. However, it does not include specific acceptance criteria for performance metrics in the way one would expect for an AI/ML device, nor does it detail a study proving the device meets such criteria through quantitative performance measures.
Instead, the document states that Genicon electrosurgical instruments were "designed to meet the following voluntary standards." These standards are related to electrical safety, surgical instrument materials, and sterilization, rather than performance metrics like sensitivity, specificity, or reader improvement.
Therefore, many of the requested information points (e.g., acceptance criteria for specific performance, sample sizes for test/training sets, ground truth establishment, MRMC study details) are not applicable or cannot be extracted from this document, as it pertains to a traditional medical device (electrosurgical instruments) and not an AI/ML device.
Here's a breakdown of what can be extracted based on the provided text, and where information is missing or not applicable:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Voluntary Standards Met) | Reported Device Performance (Implied by meeting standards) |
|---|---|
| AAMI/ANSI HF18:2001, Electrosurgical Devices | Implies electrical safety and performance as per standard. |
| IEC 60601-2-2:2000, Medical electrical equipment - Particular requirements for the safety of high frequency surgical equipment | Implies electrical safety for high-frequency surgical equipment. |
| EN 12011:1998, Instrumentation to be used in association with non-active surgical implants - General requirements | Implies suitability for use with non-active surgical implants. |
| ISO 7153-1:1991, Surgical Instruments -- Metallic Materials -- Part 1: Stainless Steel | Implies use of appropriate stainless steel for surgical instruments. |
| ISO 4957:1999, Tool Steels | Implies use of appropriate tool steels. |
| EN 868-1:1997, Packaging materials and systems for medical devices which are to be sterilized - Part 1: General requirements and test methods | Implies appropriate packaging for sterilization. |
| EN 552:1994, Sterilization of medical devices - Validation and routine control of sterilization by irradiation | Implies validated sterilization process for irradiation. |
Device Performance: The document states, "The technological characteristics of the new device are the same as the predicate device." The conclusion further states, "Based on the indications for use and technological characteristics, we conclude that the new device is substantially equivalent to the predicate device under the Federal Food, Drug, and Cosmetic Act." This implies that the device performs equivalently to the predicate device (Ethicon Endopath K984240) in its intended use, based on the adherence to the listed voluntary standards. No specific quantitative performance metrics (e.g., cutting force, coagulation time, tissue dissection efficiency) are provided in this summary.
2. Sample size used for the test set and the data provenance
- Not Applicable. This document does not describe a clinical study with a "test set" in the context of an AI/ML device. The "performance" section refers to adherence to voluntary engineering and safety standards, not a clinical data-driven performance study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. No ground truth establishment by experts for a test set is described.
4. Adjudication method for the test set
- Not Applicable. No adjudication method is mentioned as there is no "test set" in the context of an AI/ML performance study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI/ML device, and no MRMC study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an AI/ML device.
7. The type of ground truth used
- Not Applicable. For this traditional device, "ground truth" as pertaining to a performance study with data is not detailed. The "ground truth" for demonstrating safety and efficacy is essentially adherence to the listed engineering and safety standards, and demonstrating substantial equivalence to a legally marketed predicate device.
8. The sample size for the training set
- Not Applicable. This is not an AI/ML device, and no training set is mentioned.
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/ML device, and no training set is mentioned.
Summary of what the document primarily focuses on:
The 510(k) summary for the Genicon Disposable Electrosurgical Instrumentation focuses on demonstrating substantial equivalence to a predicate device (Ethicon ENDOPATH K984240). This is achieved by:
- Stating that the technological characteristics of the new device are the same as the predicate device.
- Listing adherence to a series of voluntary industry standards related to electrical safety, materials, and sterilization, which are relevant to electrosurgical devices.
The "study" in this context is the comparison against technical specifications and adherence to standards, rather than a clinical performance study with defined performance metrics and a patient dataset.
{0}------------------------------------------------
510(k) Summary
OCT 1 0 2006
(As Required By 21 CFR 807.93)
PreMarket Notification Summary
Image /page/0/Picture/5 description: The image shows a stack of books next to the words "See Related Information". The books are stacked on top of each other, and the words are in bold black font. The image is likely a button or icon that links to related information on a website or in a document. The background is white.
A 510(k) summary has been selected in lieu of a 510[k] statement.
Trade Name - GENICON Electrosurgical Instrumentation
Common Name - Electrosurgical Instrumentation
Classification Name - Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400, Product Code GEI)
Legally Marketed Device - Ethicon ENDOPATH (K984240)
Device Description - The GENICON disposable electrosurgical instruments are single use sterile instruments made from biocompatible plastic and stainless steel with a working length of 33cm. Current may be supplied by an approved electrosurgical generator which provides the ability for the coagulation of tissue when used with an appropriate ground electrode.
Intended Use - GENICON Disposable Electrosurgical Instrumentation is indicated for use in endoscopic surgical procedures. It is a family of instruments which includes graspers, dissectors, and scissors, which are intended to be used to grasp, manipulate, cut, and cauterize soft tissue.
Technological Characteristics - The technological characteristics of the new device are the same as the predicate device.
Performance – There are no mandatory performance standards or special controls in existence for this device at the date of submittal. GENICON electrosurgical instruments were designed to meet the following voluntary standards:
AAMI/ANSI HF18:2001, Electrosurgical Devices
IEC 60601-2-2:2000, Medical electrical equipment - Particular requirements for the safety of high frequency surgical equipment
EN 12011:1998, Instrumentation to be used in association with non-active surgical implants - General requirements
ISO 7153-1:1991, Surgical Instruments -- Metallic Materials -- Part 1: Stainless Steel
ISO 4957:1999, Tool Steels
EN 868-1:1997, Packaging materials and systems for medical devices which are to be sterilized - Part 1: General requirements and test methods EN 552:1994, Sterilization of medical devices - Validation and routine control of sterilization by irradiation
Conclusion - Based on the indications for use and technological characteristics, we conclude that the new device is substantially equivalent to the predicate device under the Federal Food, Drug, and Cosmetic Act.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OC1 1 0 2006
Genicon % Mr. Gary Haberland 6869 Stapoint Court, Suite 114 Winter Park, Florida 32792
Re: K061417
Trade/Device Name: Genicon Disposable Electrosurgical Instrumentation Regulation Number: 21 CFR 878.4400 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: GEI Dated: September 13, 2006 Received: September 14, 2006
Dear Mr. Haberland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 – Mr. Gary Haberland
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Barbara Buettm
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known): K061417
Device Name: Genicon Disposable Electrosurgical Instrumentation
Indications for Use:
Endoscopic surgical procedures. It is a family of instruments which includes graspers, Linsectors, and scissors, which are intended to be used to grasp, manipulate, cut, and cauterize soft tissue.
Target population definition
Patients requiring laparoscopic surgical intervention where use of an electro-surgical cutting, manipulating, and/or coagulating device is indicated.
Anatomical sites
Where laparoscopic surgical techniques are indicated, but not for fallopian tube sterilization.
Prescription Use YES X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Buettner
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number K191417
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.