K Number
K061162

Validate with FDA (Live)

Date Cleared
2006-07-26

(91 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BAR Pedicle Screw Spinal Fixation System is intended for posterior non cervical pedicle screw fixation in the lumbo-sacral spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar or sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition the BAR Pedicle Screw Spinal Fixation System, used as a pedicle screw fixation system is intended for:

a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint.

Device Description

The BAR Pedicle Screw Fixation System consists of pedicle screws, rods, and connectors. It can be used for single level fixation. All components are manufactured from titanium (ASTM F-67), titanium alloy (ASTM F-136), Cobalt Chromium Molybdenum alloy (ASTM F-1547.

The system consists of pedicle screw anchors which are inserted into the pedicle, a connector that has a ball and socket feature that connects to the pedicle screws. The ball and socket feature which allows for adjustment during implantation. The motion allowed by the ball and socket serves to offload the pedicle screws and distribute the stress across the vertebral body. The connector also contains a clamping mechanism which clamps the rigid rod into place.

AI/ML Overview

The provided text describes the "BAR Pedicle Screw Spinal Fixation System" and its 510(k) summary. However, it does not include information about a study proving the device meets acceptance criteria via clinical performance data, nor does it detail acceptance criteria themselves in a quantifiable manner for such a study.

The document primarily focuses on the device description, its intended use, a summary of non-clinical (mechanical) tests, and the basis for substantial equivalence to predicate devices for regulatory clearance. It does not mention any clinical study, reader study, or AI-based performance evaluation.

Therefore, I cannot fulfill your request for the specific details outlined (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC study, standalone performance, ground truth types, training set details) because this information is not present in the provided text.

The only "study" mentioned is "Testing was performed per ASTM F1717 and the results are comparable to the predicate devices." This refers to non-clinical mechanical testing, not a study involving human readers or clinical data for performance assessment as implied by your questions.

{0}------------------------------------------------

Bar Pedicle Screw Spinal Fixation System

510(k) Summary

JUL 26 2006

July 25, 2006

Submitter:Applied Spine Technologies, Inc.300 George St.New Haven, CT 06511
---------------------------------------------------------------------------------------

Contact Person: Michele Lucey

Trade Name: BAR Pedicle Screw Spinal Fixation System

Common Name: Posterior Pedicle Screw System

Classification Name: Class II per 21 CFR Section 888.3070

Product Code: MNI/MNH

Equivalent Devices: Dynesys® Dynamic Stabilization System (K031511) Manufactured and distributed by: Zimmer, Inc.

ISOBAR TTL Semi-Rigid Pedicle Screw Fusion System (K990118) SCIENT'X USA

CD Horizon® Spinal System (K043488) Medtronic Sofamor Danek

Device Description:

The BAR Pedicle Screw Fixation System consists of pedicle screws, rods, and connectors. It can be used for single level fixation. All components are manufactured from titanium (ASTM F-67), titanium alloy (ASTM F-136), Cobalt Chromium Molybdenum alloy (ASTM F-1547.

The system consists of pedicle screw anchors which are inserted into the pedicle, a connector that has a ball and socket feature that connects to the pedicle screws. The ball and socket feature which allows for adjustment during implantation. The motion allowed by the ball and socket serves to offload the pedicle screws and distribute the stress across the vertebral body. The connector also contains a clamping mechanism which clamps the rigid rod into place.

{1}------------------------------------------------

Intended Use:

The BAR Pedicle Screw Spinal Fixation System is intended for posterior non cervical pedicle screw fixation in the lumbo-sacral spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar or sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition the BAR Pedicle Screw Spinal Fixation System, used as a pedicle screw fixation system is intended for:

a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint.

Summary Nonclinical Tests:

Testing was performed per ASTM F1717 and the results are comparable to the predicate devices.

Basis for Substantial Equivalence:

The BAR Pedicle Screw Spinal Fixation System has similar design characteristics, i.e. material, screw size, rod based, clamping connection mechanisms, and indications as the ISOBAR TTL Spinal System (K990119) and the Dynesys Dynamic Stabilization System (K031511). All three systems stabilize the spine to support fusion while allowing some motion or flexibility which serve to distribute loads across the segment being fused.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health and Human Services. The seal features a stylized eagle with three lines representing its body and wings. The eagle is facing right and appears to be in flight. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 26 2006

Applied Spine Technologies, Inc. % Ms. Michele Lucey Vice President, Regulatory Affairs and Quality Assurance 300 George Street, Suite 511 New Haven, Connecticut 06511

Re: K061162

Trade/Device Name: BAR Pedicle Screw Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI Dated: June 13, 2006 Received: June 19, 2006

Dear Ms. Lucey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Michele Lucey

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Hucher Lermer us

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K061162

Device Name: BAR Pedicle Screw Spinal Fixation System

Indications for Use:

Intended Use/Indications for Use

The BAR Pedicle Screw Spinal Fixation System is intended for posterior non cervical pedicle screw fixation in the lumbo-sacral spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar or sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition the BAR Pedicle Screw Spinal Fixation System, used as a pedicle screw fixation system is intended for:

a. Patients receiving only autogenous bone graft. b. Patients having the device fixed or attached to the lumbar and sacral spine and having severe spondylolisthesis grade 3 or 4 at the fifth lumbar-first sacral (L5-S1) vertebral joint.

Warning

The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the lumbar and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L551 vertebra, degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.

Precautions

The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.

Prescription Use (Part 21 CFR 801 Subpar

lukin Lenu (w)
(DIRECTOR Sign Of)

(Division Sign-Off (21 CFR 801 Subpart C))

(PLEASE DO NOT WR

510(k) Number K061162

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.