(28 days)
The MEDRAD Continuum MR Compatible Infusion System (3T Proximity) is designed for patients who require maintenance medications and fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during, and immediately after the MR procedure, functioning while either stationary or mobile. It is not intended to provide long-term patient care outside the MR environment. The system is to be used by trained medical staff, primarily critical care, emergency room and radiology nursing staff.
The Medrad Continuum Infusion System maintains the same intended use, similar operational parameters, similar labeling and is essentially used in a manner similar to the predicate device. A label will be affixed to the system stating that when using the Continuum Infusion System in the MR environment it's casters brakes should be on and the system should not be placed in a static magnetic field exceeding 2000 gauss. The user's manual has also been updated to include this warning. The Continuum MR Compatible Infusion System (3.0T) is comprised of the same components as the predicate device.
The provided text describes a 510(k) premarket notification for an infusion pump, the MEDRAD Continuum MR Compatible Infusion System (3T Proximity). This submission aims to demonstrate substantial equivalence to a predicate device, the MEDRAD Continuum MR Compatible Infusion System - 1.5T. The information primarily focuses on comparing the new device to its predicate, rather than providing detailed acceptance criteria and study results for a novel device.
The study presented is not a typical clinical trial to assess diagnostic or treatment effectiveness, but rather a benchmarking comparison and safety assessment to demonstrate that the new device (3T Proximity) performs equivalently to an existing, legally marketed predicate device (1.5T). The goal is to show that the minor modifications, particularly its compatibility with a 3.0T MR environment, do not introduce new safety or effectiveness concerns while maintaining the same performance characteristics.
Here's an analysis based on the provided text, addressing your points:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criterion for the new device is substantial equivalence to the predicate device. This means the new device must have the "same intended use, similar operational parameters, similar labeling and is essentially used in a manner similar to the predicate device."
The "reported device performance" is a direct comparison of parameters between the new device (MEDRAD Continuum MR Compatible Infusion System - 3.0T) and the predicate device (MEDRAD Continuum MR Compatible Infusion System - 1.5T). The key finding is that all listed performance parameters are "Same" between the two devices, except for the explicit mention of MRI compatibility up to "2000 gauss" for the new device, whereas the predicate was simply "Yes".
| Parameter (Acceptance Criteria: Must be "Same" as Predicate unless otherwise specified) | Predicate Device (1.5T) Performance | New Device (3.0T Proximity) Performance | Meets Acceptance Criteria? |
|---|---|---|---|
| Classification | Type CF | Same | Yes (Implicitly) |
| Compatible IV Administration Set | Standard PVC Single Use Y-Site Set Free Flow Protection | Same | Yes (Implicitly) |
| Flow Rate | Continuous: 1 to 99.9 ml/h in 0.1 ml/h increments, 100 to 1200 ml/h in 1 ml/h increments | Same | Yes (Implicitly) |
| Total Infused Volume | Continuous: 1 to 9,999 ml | Same | Yes (Implicitly) |
| Housing | ABS | Same | Yes (Implicitly) |
| Microprocessor Controlled | Yes | Same | Yes (Implicitly) |
| KVO Infusion Rate | 1 to 5 ml/h | Same | Yes (Implicitly) |
| System Accuracy | +/- 10% | Same | Yes (Implicitly) |
| Occlusion Pressure | User Selectable: Low-0.3 bar (5 psi) Medium-0.5 bar (7.5 psi) High-0.7 bar (10 psi) | Same | Yes (Implicitly) |
| Battery | Li-Ion Rechargeable | Same | Yes (Implicitly) |
| Battery Operation | 4 hr @ 1200 ml/h | Same | Yes (Implicitly) |
| Battery Charging | ~6 hr to 95% maximum battery capacity | Same | Yes (Implicitly) |
| Air-In-Line Detector | User Selectable | Same | Yes (Implicitly) |
| Total Time Setting | 100 hours | Same | Yes (Implicitly) |
| Air Sensor | Ultrasonic | Same | Yes (Implicitly) |
| Prime Rate | 1600 to 1800 ml/h | Same | Yes (Implicitly) |
| Bolus Rate | 0.1 to 1200 ml/h | Same | Yes (Implicitly) |
| Bolus Volume | 0.1 to 100 ml | Same | Yes (Implicitly) |
| Alarm Conditions (All sub-points) | Yes (for each: Air-In-Line, Down Occlusion, Pump Unattended, Low Battery, End Battery, Fatal Error, End Program, Missing Key, Lock Mode) | Yes (for each) | Yes (Implicitly) |
| Program Modes (All sub-points) | Yes (for each: Continuous, Dose, Non-Weight Dose) | Yes (for each) | Yes (Implicitly) |
| Integrated Mounting Clamp for IV Pole | Yes | Same | Yes (Implicitly) |
| Built in Free Flow Protection | Yes | Same | Yes (Implicitly) |
| Automatic Bolus Function (Safety Feature) | Yes | Same | Yes (Implicitly) |
| User Selectable Alarm Volume | Yes | Same | Yes (Implicitly) |
| Comprehensive Safety Features | Yes | Same | Yes (Implicitly) |
| Low Battery | Yes | Same | Yes (Implicitly) |
| MRI Compatible | Yes | Yes (up to 2000 gauss) | Yes (Explicitly extended compatibility) |
Note: The primary "study" proving acceptance is the detailed comparison table that asserts the physical and functional identity of the new device with the predicate, with the added MR compatibility for 3.0T environments (up to 2000 gauss static magnetic field) being the key difference and point of assessment.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable in the context of human clinical data. This is a technical comparison for substantial equivalence. The "test set" would refer to the new device itself (MEDRAD Continuum MR Compatible Infusion System - 3.0T), which was designed and tested against the specifications of the predicate.
- Data Provenance: The data provenance is from the manufacturer's (MEDRAD, Inc.) engineering and testing documentation, comparing the design and performance specifications of their existing 1.5T compatible device with the proposed new 3.0T compatible device. This is inherently prospective in the sense of designing and testing the new device to meet predicate performance, but the comparison is based on retrospective knowledge of the predicate's performance. The country of origin of the data is implicitly the USA, as MEDRAD, Inc. is located in Pennsylvania.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable for this type of submission. "Ground truth" in this context is established by the engineering specifications and performance characteristics of the predicate device, which have already been deemed safe and effective by the FDA. The submission relies on the manufacturer's internal engineering and quality assurance processes to confirm the new device meets these established specifications and the new MRI compatibility requirements. There are no "experts" establishing a clinical ground truth for a test set in the way one would for diagnostic imaging.
4. Adjudication Method for the Test Set
Not applicable. There is no multi-reader adjudication process described, as it's not a clinical diagnostic performance study. The adjudication is essentially the FDA's review determining if the provided technical comparison demonstrates substantial equivalence to the predicate device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
No MRMC study was done. This is not an AI-enabled device or a diagnostic imaging device requiring human reader interpretation. It is an infusion pump.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Study Was Done
Yes, in a conceptual sense. The "study" here is the technical comparison and engineering validation of the device itself (the "algorithm only" being the device's operational design and components). The device functions somewhat autonomously according to its programming (flow rate, alarms, etc.), but it requires human users (trained medical staff) for setup and monitoring. The "performance" is based on its ability to meet the established operational parameters, not on human-in-the-loop diagnostic accuracy.
7. The Type of Ground Truth Used
The ground truth used is the engineering specifications and established performance characteristics of the legally marketed predicate device (MEDRAD Continuum MR Compatible Infusion System - 1.5T), combined with new safety testing to confirm compatibility with the higher magnetic field environment (3.0T). The implicit ground truth is that if the new device performs identically to a predicate already cleared for market, and addresses the new environmental concerns, it is also safe and effective.
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning or AI device that requires a "training set." The device is hardware-based with embedded software, and its design principles are based on established medical device engineering.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the context of this device.
{0}------------------------------------------------
510(k) Summary
OFFICIAL CONTACT:
John M Kiste Regulatory Affairs Analyst MEDRAD, Inc. One Medrad Drive Indianola, PA 15051 (412) 767-2400 ext. 3444
Infusion Pump & Accessories [21 CFR 880.5725] CLASSIFICATION NAME:
MEDRAD
Peristaltic Pump System
COMMON NAME(S):
PROPRIETARY NAME:
PREDICATE DEVICE:
INTENDED USE:
(K032771) MEDRAD Continuum MR Compatible The Infusion System is designed for patients who require maintenance medications and fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during, and immediately after the MR procedure, functioning while either stationary or mobile. It is not intended to provide long-term patient care outside the MR environment. The system is to be used by trained medical staff, primarily critical care, emergency room and radiology nursing staff.
MR Continuum Infusion System
MEDRAD Continuum MR Infusion System
DEVICE DESCRIPTION AND COMPARISON TO UNMODIFIED PREDICATE:
The Medrad Continuum Infusion System maintains the same intended use, similar operational parameters, similar labeling and is essentially used in a manner similar to the predicate device.
A label will be affixed to the system stating that when using the Continuum Infusion System in the MR environment it's casters brakes should be on and the system should not be placed in a static magnetic field exceeding 2000 gauss. The user's manual has also been updated to include this warning.
The Continuum MR Compatible Infusion System (3.0T) is comprised of the same components as the predicate device. Differences between the predicate device and the new Continuum MR Compatible Infusion System (3.0T) are detailed in the table below. The following Comparison Matrix identifies the similarities and differences between the new device and the predicate device.
- 10 -
{1}------------------------------------------------
| Parameter | MEDRAD ContinuumMR Compatible InfusionSystem - 1.5T | MEDRAD ContinuumMR Compatible InfusionSystem - 3.0T |
|---|---|---|
| Classification | Type CF | Same |
| Compatible IVAdministration Set | Standard PVCSingle UseY-Site SetFree Flow Protection | Same |
| Flow Rate | Continuous: 1 to 99.9 ml/hin 0.1 ml/h increments,100 to 1200 ml/h in 1 ml/hincrements | Same |
| Total Infused Volume | Continuous: 1 to 9,999 ml | Same |
| Housing | ABS | Same |
| MicroprocessorControlled | Yes | Same |
| KVO Infusion Rate | 1 to 5 ml/h | Same |
| System Accuracy | +/- 10% | Same |
| Occlusion Pressure | User Selectable:Low-0.3 bar (5 psi)Medium-0.5 bar (7.5 psi)High-0.7 bar (10 psi) | Same |
| Battery | Li-Ion Rechargeable | Same |
| Battery Operation | 4 hr @ 1200 ml/h | Same |
| Battery Charging | ~6 hr to 95% maximumbattery capacity | Same |
| Air-In-Line Detector | User Selectable | Same |
| Total Time Setting | 100 hours | Same |
| Air Sensor | Ultrasonic | Same |
| Prime Rate | 1600 to 1800 ml/h | Same |
| Bolus Rate | 0.1 to 1200 ml/h | Same |
| Bolus Volume | 0.1 to 100 ml | Same |
| Alarm Conditions | ||
| > Air-In-Line | Yes | Yes |
| > DownOcclusion | Yes | Yes |
| > PumpUnattended | Yes | Yes |
| > Low Battery | Yes | Yes |
| > End Battery | Yes | Yes |
| > Fatal Error | Yes | Yes |
| > End Program | Yes | Yes |
| > Missing Key | Yes | Yes |
| > Lock Mode | Yes | Yes |
| Program Modes | ||
| > Continuous | Yes | Yes |
| > Dose | Yes | Yes |
| > Non-WeightDose | Yes | Yes |
| Integrated MountingClamp for IV Pole | Yes | Yes |
| Built in Free FlowProtection | Yes | Yes |
| Automatic Bolus Function(Safety Feature) | Yes | Yes |
| User Selectable AlarmVolume | Yes | Yes |
| Comprehensive SafetyFeatures | Yes | Yes |
| Low Battery | Yes | Yes |
| MRI Compatible | Yes | Yes (up to 2000 gauss) |
.
:
{2}------------------------------------------------
and the comments of the comments of
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 22 2006
Mr. John M. Kiste Regulatory Affairs Analyst MEDRAD, Incorporated One Medrad Drive Indianola, Pennsylvania 15051
Re: K061128
Trade/Device Name: MEDRAD Continuum MR Compatible Infusion System (3T Proximity) Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: April 20, 2006 Received: April 25, 2006
Dear Mr. Kiste:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Kiste
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Syette y. Michael Smis.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: __ MEDRAD Continuum MR Compatible Infusion System (3T Proximity)
Indications For Use:
The MEDRAD Continuum MR Compatible Infusion System (3T Proximity) is designed for patients who require maintenance medications and fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during, and immediately after the MR procedure, functioning while either stationary or mobile. It is not intended to provide long-term patient care outside the MR environment. The system is to be used by trained medical staff, primarily critical care, emergency room and radiology nursing staff.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Antan ank
Colody, General Hospits). .. Lartal Devic
Page 1 of
§ 880.5725 Infusion pump.
(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).