K Number
K060949
Date Cleared
2006-05-04

(28 days)

Product Code
Regulation Number
892.1560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The system is a general-purpose, fully digital ultrasound system for abdominal, gynecologic and obstetric, small parts, and cardiac applications.

The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, cardiac, transvaginal, peripheral vascular, and musculo-skeletal (both conventional and superficial). This device is intended to adult, pregnant woman, pediatric and neonate. The system is a prescription device intended to be used by or on the order of a physician or similarly qualified health care professional. This Device is not intended for home use.

Device Description

The DP-6600 Digital Ultrasonic Diagnostic Imaging System with added transducers is a general purpose, portable, software controlled, ultrasound diagnostic system. This ultrasonic device is designed to project ultrasound waves into body tissue and to present the returned echo information on the monitor. The resulting information is displayed in B-Mode, M-Mode, or in the combined mode (i.e. B/M-Mode). This system is a Track 1 device that employs an array of probes that include linear array and convex linear array with a frequency range of approximately 2 MHz to 10 MHz. Changing acoustic transducers of probes and addition of some transducers enriches the diversity of the product's configuration and expands the freedom of choices for customers. At the same time, new software functions have been added.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the DP-6600 Digital Ultrasonic Diagnostic Imaging System with added transducers. However, it explicitly states that "No clinical testing was required". This means there is no study detailed in this document that proves the device meets specific acceptance criteria based on human or AI performance.

Therefore, many of the requested sections about acceptance criteria, device performance, sample sizes, ground truth, experts, and comparative effectiveness studies cannot be fulfilled from the provided text.

Here's what can be extracted based on the document:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Acceptance Criteria (if mentioned)Reported Device Performance (if mentioned)Met/Not Met/Not Applicable
Safety and EffectivenessSubstantially equivalent to predicate devices (K052113, K992663, K042540)The device is as safe and effective as the legally marketed predicate devices.Met (based on manufacturer's conclusion)
Design SpecificationsMet all design specificationsVerified in laboratory testingMet (based on manufacturer's statement)
Acoustic OutputMeasured and calculated per NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment: 2004Measured and calculated per NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment: 2004.Met (based on manufacturer's statement)
BiocompatibilityPatient contact materials are biocompatibleAll patient contact materials are biocompatibleMet (based on manufacturer's statement)
Standards ConformanceConforms to NEMA UD 2, IEC 60601-1, IEC 60601-1-2Conforms to NEMA UD 2, IEC 60601-1, IEC 60601-1-2Met (based on manufacturer's statement)
Clinical Performance(No specific clinical performance criteria mentioned as no clinical testing was required)(No clinical performance data reported as no clinical testing was required)Not Applicable

2. Sample size used for the test set and the data provenance:

  • Not applicable. The document explicitly states "No clinical testing was required," meaning there was no test set of patient data used for performance evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. As no clinical testing was performed, there was no test set requiring ground truth established by experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. No clinical test set was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This device is a diagnostic ultrasound system, not an AI-assisted diagnostic tool as described in the context of MRMC studies. The document states "No clinical testing was required."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a hardware/software system, not a standalone algorithm. The document states "No clinical testing was required."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not applicable. No clinical data requiring ground truth was used for performance validation. The device's "ground truth" for demonstrating substantial equivalence relied on its technical specifications meeting existing standards and being comparable to predicate devices.

8. The sample size for the training set:

  • Not applicable. As "No clinical testing was required," there's no mention of a training set of data for an algorithm or AI model. The system's "training" would be its development and engineering based on established ultrasound physics and medical imaging principles.

9. How the ground truth for the training set was established:

  • Not applicable. No training set requiring ground truth was used as per the document.

{0}------------------------------------------------

MAY 4 2006 Exhibit A 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is:

Submitter:

Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

Tel: +86 755 2658 2888

Fax: +86 755 2658 2680

● Contact Person:

Li Dongling Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

● Date Prepared:

Mar 28, 2006

Name of the device:

  • Trade/Proprietary Name:
    DP-6600 Digital Ultrasonic Diagnostic Imaging System with added transducers

Common Name: Diagnostic Ultrasound System and Transducers ●

● Classification

Regulatory Class: II Review Category: Tier II

21CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-1YO) 21CFR 892.1570 Diagnostic Ultrasound Transducer (90-ITX)

{1}------------------------------------------------

Legally Marketed Predicate Device:

K052113DP-6600 Digital Ultrasonic Diagnostic Imaging System
K992663SSD-5000 Diagnostic Ultrasound System, Aloka Co. Ltd.
K042540IU22 Diagnostic Ultrasound System, Phillips Ultrasound, Inc

Description:

The DP-6600 Digital Ultrasonic Diagnostic Imaging System with added transducers is a general purpose, portable, software controlled, ultrasound diagnostic system. This ultrasonic device is designed to project ultrasound waves into body tissue and to present the returned echo information on the monitor. The resulting information is displayed in B-Mode, M-Mode, or in the combined mode (i.e. B/M-Mode). This system is a Track 1 device that employs an array of probes that include linear array and convex linear array with a frequency range of approximately 2 MHz to 10 MHz. Changing acoustic transducers of probes and addition of some transducers enriches the diversity of the product's configuration and expands the freedom of choices for customers. At the same time, new software functions have been added.

Statement of intended Use:

The system is a general-purpose, fully digital ultrasound system for abdominal, gynecologic and obstetric, small parts, and cardiac applications.

The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, cardiac, transvaginal, peripheral vascular, and musculo-skeletal (both conventional and superficial). This device is intended to adult, pregnant woman, pediatric and neonate. The system is a prescription device intended to be used by or on the order of a physician or similarly qualified health care professional. This Device is not intended for home use.

Technological Characteristics:

The DP-6600 digital ultrasonic diagnostic imaging system with added transducers incorporates the same fundamental technology as the predicate devices. The device has been tested as Track 1 Device per the FDA Guidance document "Information for Manufacturers Seeking Marketing Clearance of

{2}------------------------------------------------

Diagnostic Ultrasound Systems and Transducers" issued September 1997. The acoustic output is measured and calculated per NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment: 2004. All transducers used with the DP-6600 digital ultrasonic diagnostic imaging system are track 1. All patient contact materials are biocompatible. The technology characteristics of the DP-6600 digital ultrasonic diagnostic imaging system with these modifications do not affect the safety or efficacy of the device.

Testing:

Laboratory testing was conducted to verify that the DP-6600 digital ultrasonic diagnostic imaging system with added transducers met all design specification and was substantially equivalent to the currently marketed Predicate Device as above. The device has been found to conform to applicable medical device safety standards in regards to thermal, mechanical and electrical safety as well as biocompatibility. Acoustic output is measured and calculated according to "Acoustic Output Measuring Standard for Diagnostic Ultrasound Equipment"

Applicable Standards

The DP-6600 digital ultrasonic diagnostic imaging system with added transducers conforms to the following Standards:

NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic ultrasound Equipment IEC 60601-1 IEC 60601-1-2

Clinical Test:

No clinical testing was required

Conclusion:

The conclusions drawn from testing of the DP-6600 Digital Ultrasonic Diagnostic Imaging System with added transducers demonstrates that the device is as safe, as effective as well as the legally marketed predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

MAY 4 2006 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Shenzhen Mindray Bio-Medical Electronics Co., Ltd % Ms. Susan D. Goldstein-Falk Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 GREAT NECK NY 11021

Re: K060949

.

Trade Name: DP-6600 Digital Ultrasonic Imaging System with added Transducers Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYO and ITX Dated: March 20, 2006 Received: April 6, 2006

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the DP-6600 Digital Ultrasonic Imaging System with added Transducers, as described in your premarket notification:

Transducer Model Number

$\frac{65C15EA}{75L60EA}$

Image /page/3/Picture/13 description: The image is a black and white circular logo. The logo is for the FDA Centennial, celebrating 1906-2006. The letters FDA are in bold in the center of the logo. There are three stars at the bottom of the logo.

{4}------------------------------------------------

Page 2 - Ms. Goldstein-Falk

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

{5}------------------------------------------------

Page 3 - Ms. Goldstein-Falk

If you have any questions regarding the content of this letter, please contact Andrew Kang, M.D. at (301) 594-1212.

and

Sincerely yours,

Daniel A. Leggett

~ Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

SystemxTransducer
Model:DP-6600
510(k) Number(s)K060949
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationABMPWDCWDColor DopplerAmplitude DopplerColor Velocity ImagingCombined (specify)Other (specify)
Ophthalmic
FetalPPPP
AbdominalPPPP
Intraoperative (specify)
Intraoperative Neurological
PediatricPPPP
Small organ(specify)PPPP
Neonatal CephalicPPPP
Adult Cephalic
CardiacPPPP
Transesophageal
TransrectalPPPP
TransvaginalPPPP
Transurethral
Intravascular
Peripheral VascularPPPP
Laparoscopic
Musculo-skeletalPPPP
Conventional
Musculo-skeletal SuperficialPPPP
Other (specify)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
-------------------------------------------------------------------------

Additional comments: Combined mode: B+M

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDIED)

Concurrence of CDRH, Office of Device Evaluation(ODE)

Prescription USE (Per 21 CFR 801.109)

Demid A. Sypm
(Division Sign-Off)
Division of Remediation

Division of Reproductive, and Radiological Device 510k) Number

{7}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

SystemTransducerx
Model:65C15EA
510(k) Number(s)K060949
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
ABMPWDCWDColor DopplerAmplitude DopplerColor Velocity ImagingCombined (specify)Other (specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (specify)
Intraoperative Neurological
PediatricNNN
Small organ(specify)
Neonatal CephalicNNN
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral Vascular
Laparoscopic
Musculo-skeletal
Conventional
Musculo-skeletal Superficial
Other (specify)

N=new indication; P=previously cleared by FDA; E=added under Appendix E

Additional comments: Combined mode: B+M

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation(ODE)

Prescription USE (Per 21 CFR 80 J-109)

くなって

David A. Ingram

... . . . . . .

(Division Sign-Off) Division of Reproductive, Abd and Radiological Devices 51 R., Number _

{8}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

SystemTransducer ×
Model:75L60EA
510(k) Number(s)K060949
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
ABMPWDCWDColor DopplerAmplitude DopplerColor Velocity ImagingCombined (specify)Other (specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (specify)
Intraoperative Neurological
Pediatric
Small organ(specify)NNN
Neonatal CephalicNNN
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral VascularNNN
Laparoscopic
Musculo-skeletal ConventionalNNN
Musculo-skeletal SuperficialNNN
Other (specify)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined mode: B+M

combined model Q-TL

.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE 1E NEEDED)

Concurrence of CDRH, Office of Device Evaluation(ODE)

Prescription USE (Per 21 CFR 801. 109)

R 801.109)

(D vision Sign-Off)

David h. Syam

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

0031

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.