K Number
K060947

Validate with FDA (Live)

Date Cleared
2006-04-19

(13 days)

Product Code
Regulation Number
892.1570
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is intended for ultrasound examination of peripheral intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.

Device Description

The Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is a sterile, single-use, short rail (SR), 20 MHz imaging catheter with a 3.2F distal crossing profile. The catheter is intended to operate with the BSC ClearView Ultra™, Galaxy®, Galaxy2™, or iLab® intravascular ultrasound imaging instruments. The catheter and instrument form an imaging system that is intended for the ultrasonic examination of peripheral intravascular pathology only. The catheter consists of two (2) main assemblies: a sheath assembly and an imaging core assembly. The sheath assembly is the outer part of the catheter that comprises much of the working length of the device. The imaging core is internal to the sheath assembly and rotates independently of the sheath. The imaging core contains the piezoelectric transducer that converts electrical energy from the instrument to ultrasonic energy that is in turn transmitted. The transducer converts the returned ultrasonic energy into electrical energy for display on the instrument. The rotating imaging core drive shaft directs the path of the ultrasonic energy (beam) from the transducer. A proximal telescoping section allows the imaging core to be retracted and advanced, across an anatomical region of interest, without requiring movement of the sheath within the anatomy. The modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is identical in all aspects to the current Sonicath Ultra™ 3.2F 20MHz Imaging Catheter except for the following change: Changed catheter sheath support material from one Low Density Polyethylene (LDPE) resin formulation to another.

AI/ML Overview

The provided document K060947 is a 510(k) summary for the Boston Scientific Sonicath Ultra™ 3.2F 20MHz Imaging Catheter. This summary describes a device modification (change in catheter sheath support material) and demonstrates its substantial equivalence to a predicate device.

It's important to note that this document does not describe a study involving an AI device or a comparative effectiveness study with human readers. The device is an intravascular ultrasound imaging catheter, and the "study" referred to in the document is design verification testing and bench testing of the catheter itself, not a clinical study involving AI or human interpretation of images.

Therefore, many of the requested criteria related to AI performance, sample size for test/training sets, expert ground truth, adjudication methods, and MRMC studies are not applicable to this submission.

Here's an analysis of the available information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (What they tested for)Reported Device Performance
Physical integrity and functionality (e.g., tensile, pressure) of catheter sheath support and full, sterile device."The results demonstrated that the device satisfies all performance, physical and functional requirements." (Bench Testing section)
Biocompatibility"The results of the biocompatibility testing demonstrate that the modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is acceptable for it intended use." (Biological Safety Testing section)
Acoustic output"Acoustic Output testing was not required... has no effect on acoustic output." (Acoustic Output Testing section)
Predetermined acceptance criteria for design verification testing"Design verification testing confirms that predetermined acceptance criteria were met and no new issues of safety or efficacy have been raised with the use of this new catheter sheath support material."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified in terms of number of catheters or specific test units. The testing described is bench and biological safety testing of the device itself, not image data.
  • Data Provenance: Not applicable as it's not a data-driven study. The testing was conducted internally by Boston Scientific Corporation.
  • Retrospective/Prospective: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Number of experts: Not applicable. The ground truth for device performance (tensile strength, biocompatibility, etc.) is established through engineering standards and laboratory testing, not expert interpretation of clinical data.
  • Qualifications of experts: Not applicable in the context of clinical experts. The testing would have been conducted by engineers and laboratory personnel qualified in their respective fields.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable. This is not a study involving human interpretation of clinical data that would require an adjudication process for ground truth establishment.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This document describes a modification to a medical device (an imaging catheter), not an AI device, and therefore no MRMC study was performed or is relevant.

6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

  • Standalone Performance: No. This document does not pertain to an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: The "ground truth" here refers to established engineering specifications, material properties, and biological safety standards. For instance, tensile strength must meet a certain Newton threshold, and biocompatibility tests must show no cytotoxic or irritant effects.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. This is not an AI device, so there is no training set. The "training" here would refer to the development and testing process of the physical catheter.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable. As there is no AI training set, there is no ground truth for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Boston Scientific. The words "Boston" and "Scientific" are stacked on top of each other. The font is a serif font and the text is black.

K060947

APR 1 9 2006

510(k) Summary of Safety and Effectiveness

Boston Scientific Corporation Sonicath Ultra™ 3.2F 20MHz Imaging Catheter

Submitted By:Boston Scientific CorporationIVUS Technology Center47900 Bayside ParkwayFremont, CA 94538
Contact Person:Robert Z. PhillipsManager, Regulatory AffairsTel: (510) 624-2307Fax: (510) 624-1449phillipr1@bsci.com
Date Prepared:April 5, 2006
Proprietary Name:Sonicath Ultra™ 3.2F 20MHz Imaging Catheter
Common Name(s):Ultrasound Diagnostic Imaging CatheterDiagnostic Intravascular Catheter (74DQO)Diagnostic Ultrasonic Transducer (90ITX)
Classification Name(s):Diagnostic Intravascular Catheters, 21 CFR 870.1200(74DQO)Diagnostic Ultrasonic Transducers, 21 CFR 892.1570(90ITX)
Predicate Device(s):Sonicath Ultra™ 3.2F 20MHz Imaging Catheter issubstantially equivalent to the following device:
Predicate510(k)Clearance Date
Sonicath Ultra™ 3.2F20MHz ImagingCatheterK970049June 20, 1997

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the words "Boston Scientific" stacked on top of each other. The word "Boston" is on the top line, and the word "Scientific" is on the second line. The text is in a serif font and is black in color.

Description of the Device:

The Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is a sterile, single-use, short rail (SR), 20 MHz imaging catheter with a 3.2F distal crossing profile. The catheter is intended to operate with the BSC ClearView Ultra™, Galaxy®, Galaxy2™, or iLab® intravascular ultrasound imaging instruments. The catheter and instrument form an imaging system that is intended for the ultrasonic examination of peripheral intravascular pathology only.

The catheter consists of two (2) main assemblies: a sheath assembly and an imaging core assembly. The sheath assembly is the outer part of the catheter that comprises much of the working length of the device. The imaging core is internal to the sheath assembly and rotates independently of the sheath. The imaging core contains the piezoelectric transducer that converts electrical energy from the instrument to ultrasonic energy that is in turn transmitted. The transducer converts the returned ultrasonic energy into electrical energy for display on the instrument. The rotating imaging core drive shaft directs the path of the ultrasonic energy (beam) from the transducer. A proximal telescoping section allows the imaging core to be retracted and advanced, across an anatomical region of interest, without requiring movement of the sheath within the anatomy.

Sonicath Ultra™ 3.2F 20MHz Imaqing Catheter:

The modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is identical in all aspects to the current Sonicath Ultra™ 3.2F 20MHz Imaging Catheter except for the following change:

  • Changed catheter sheath support material from one Low Density Polyethylene . (LDPE) resin formulation to another.

{2}------------------------------------------------

Design verification testing confirms that predetermined acceptance criteria were met and no new issues of safety or efficacy have been raised with the use of this new catheter sheath support material.

Intended Use/Indications:

The Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is intended for ultrasound examination of peripheral intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.

Device Technology Characteristics and Comparison to Predicate Device:

The modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is the same as the predicate device, the Sonicath Ultra™ 3.2F 20MHz Imaging Catheter cleared June 20, 1997, in all aspects except one -- change in sheath support resin formulation.

Non-clinical Test Results:

Bench and biological safety testing demonstrate that the modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter meets performance requirements and is safe and effective for its intended use.

Bench Testing:

Bench testing was performed to evaluate the physical integrity and functionality of the modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter. This testing included dimensional and functional (e.g., tensile, pressure) testing of the catheter sheath support and the full, sterile device. The results demonstrated that the device satisfies all performance, physical and functional requirements.

{3}------------------------------------------------

K060947

Image /page/3/Picture/1 description: The image shows the words "Boston Scientific" stacked on top of each other. The word "Boston" is on the top line, and the word "Scientific" is on the bottom line. The font is a serif font, and the text is black. The background is white.

APF 1 9 2005

Biological Safety Testing:

The Sonicath Ultra™ 3.2F 20MHz Imaging Catheter was subjected to a series of biocompatibility tests per USP and EN ISO 10993 standards. The results of the biocompatibility testing demonstrate that the modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is acceptable for it intended use.

Acoustic Output Testing:

Acoustic Output testing was not required for the modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter as the formulation change to the sheath support has no effect on acoustic output.

Per the FDA Guidance, Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (September 30 1997), acoustic output measurements for the modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter remain unchanged from its predicate device, the Sonicath Ultra™ 3.2F 20MHz Imaging Catheter.

Conclusion:

The modified Sonicath Ultra™ 3.2F 20MHz Imaging Catheter utilizes the same design features and has the same intended use as the predicate device, the Sonicath Ultra™ 3.2F 20MHz Imaging Catheter. The tests conducted support a determination of substantial equivalence of the modified device to the predicate device.

Image /page/3/Picture/11 description: The image shows the word "Confidential" on the top left, followed by the word "Page" on the bottom left. To the right of the word "Page" is the number 78. The text is in a simple, sans-serif font and is likely part of a document or report cover page.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (USA). The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

APR 1 2 2006

Mr. Robert Z. Phillips Manager, Regulatory Affairs Boston Scientific Corporation IVUS Technology Center 47900 Bayside Parkway FREMONT CA 94538-6515

Re: K060947

Trade Name: Sonicath Ultra™ Imaging Catheter Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: April 5. 2006 Received: April 6, 2006

Dear Mr. Phillips:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Sonicath Ultra™ Imaging Catheter, as described in your premarket notification:

Transducer Model Number

3.2F 20MHz

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can

Image /page/4/Picture/13 description: The image shows a circular logo with the letters "FDA" in the center. The letters are in a bold, sans-serif font. Above the letters, there is the text "The New". Below the letters, there is the text "Centennial". The logo is surrounded by a dotted circle.

ling Public - Health

{5}------------------------------------------------

Page 2 - Mr. Phillips

be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reade of a , made a determination that your device complies with other requirements of the Act that + D. Has Internations and regulations administered by other Federal agencies. You must or uny 1 with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic rordly to and ation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicated wice results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rryou desire of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other generalional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Ewa Czerska at (301) 594-1212.

Sincerely yours.

David A. Ingram

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Indications for Use Statement K 060947 510(k) Number: Sonicath Ultra™ 3.2F 20MHz Imaging Catheter Device Name: The Sonicath Ultra™ 3.2F 20MHz Imaging Catheter is Indications for Use: intended for ultrasound examination of peripheral intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.

Prescription Use_ × (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

David H. Lynn

(Division Sign-Off) (Division Sign-On)
Division of Reproductive, Abde Division of neproof and Radiological Devices and Radiological Devices and Radiological Proper and (k) Number -

Confidential

Page

§ 892.1570 Diagnostic ultrasonic transducer.

(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.