K Number
K060757

Validate with FDA (Live)

Date Cleared
2006-04-19

(29 days)

Product Code
Regulation Number
870.1220
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Inquiry Optima Catheter is a steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The Optima catheters are to be used to map the atrial regions of the heart.

Device Description

The Inquiry Optima PLUS Steerable Electrophysiology Catheter is a flexible, radiopaque catheter with a variable number of electrodes with the first electrode located at the distal tip and the other band electrodes following at predetermined distances. A connecting cable is used to connect the catheter to electrogram devices. The catheter has a distal loop in a plane perpendicular to the catheter body. The circumferential shape or loop allows the electrophysiologist to record the potentials of cardiac structures without changing the position of the catheter shaft and/or loop is steerable by manipulating the handle. The placement of the electrodes around the entire circumference of the distal loop and on the shaft of the catheter also assists the electrophysiologist during fluoroscopy with visualization. The distal loop shape is easily straightened with the thumb and forefinger to facilitate insertion into sheaths and introducers. Once the catheter is extended beyond the sheath, the catheter resumes its pre-formed shape. The device is supplied sterile and is intended for single use only.

AI/ML Overview

The provided text is a 510(k) summary for a medical device and does not contain the specific information requested about acceptance criteria, detailed study design, sample sizes, expert qualifications, or ground truth establishment in the context of device performance metrics like sensitivity, specificity, accuracy, etc. It primarily focuses on demonstrating substantial equivalence to a predicate device.

However, I can extract what is stated regarding testing and comparison:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria with specific performance metrics (e.g., sensitivity, specificity, accuracy) and corresponding reported device performance values. Instead, it refers to a non-clinical test plan and its outcome.

Acceptance CriteriaReported Device Performance
(Not explicitly stated as quantifiable metrics for device performance)"The test results indicated reliable performance when the device was used in accordance with the Instructions for Use."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a sample size for a test set, nor does it mention data provenance (country of origin, retrospective/prospective). The study described is "non-clinical tests."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided. The non-clinical tests likely involved engineering or laboratory assessments rather than human expert evaluation for "ground truth" in the clinical sense.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an electrophysiology catheter, and this type of study is typically associated with AI-driven diagnostic imaging devices.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a physical catheter, not an algorithm, and is used with a human in the loop (electrophysiologist).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The concept of "ground truth" as typically applied to diagnostic algorithms is not directly relevant to the described non-clinical performance testing of a physical medical device. The "test results" likely refer to engineering specifications and functional performance.

8. The sample size for the training set

No training set is mentioned as this device is not an AI/ML algorithm.

9. How the ground truth for the training set was established

No training set is mentioned.

{0}------------------------------------------------

K060757

: | | | | 1 0 2006 SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the SMDA and 21 CFR 807.92.

  • Submitter's name, address, telephone number, contact person, and date summary l. prepared:
    Submitter: Irvine Biomedical, Inc. 2375 Morse Avenue Irvine, CA 92614 Tel. (949) 851-3053

Contact: Dennis Hong Regulatory Affairs Manager Tel. (949) 271-1134 March 18, 2006 Date:

  1. Name of device, including trade name and classification name:
a.Trade/Proprietary Name:Inquiry™ Optima™ PLUS SteerableElectrophysiology Catheter
b.Classification names:Catheter, Electrode Recording
    1. Identification of the predicate device or legally marketed device or devices to which substantial equivalence is being claimed:
      Company: Irvine Biomedical, Inc. Device: Inquiry Optima Steerable Electrophysiology Catheter 510(k): K042775 Date Cleared: November 4, 2004

page 1 of 3

{1}------------------------------------------------

K ctc"fst Page 2

  • A description of the device that is the subject of the 510(k), including explanation of 4 . how the device functions, basic scientific concepts, significant physical and performance characteristics (design, material, physical properties):
    The Inquiry Optima PLUS Steerable Electrophysiology Catheter is a flexible, radiopaque catheter with a variable number of electrodes with the first electrode located at the distal tip and the other band electrodes following at predetermined distances. A connecting cable is used to connect the catheter to electrogram devices.

The catheter has a distal loop in a plane perpendicular to the catheter body. The circumferential shape or loop allows the electrophysiologist to record the potentials of cardiac structures without changing the position of the catheter shaft and/or loop is steerable by manipulating the handle. The placement of the electrodes around the entire circumference of the distal loop and on the shaft of the catheter also assists the electrophysiologist during fluoroscopy with visualization. The distal loop shape is easily straightened with the thumb and forefinger to facilitate insertion into sheaths and introducers. Once the catheter is extended beyond the sheath, the catheter resumes its pre-formed shape.

The device is supplied sterile and is intended for single use only.

    1. Statement of intended use:
      The Inquiry Optima Catheter is a steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The Optima catheters are to be used to map the atrial regions of the heart.
  • Statement of how the technological characteristics of the device compare to those of 6. the predicate or legally marketed device.
    The Inquiry Optima and Inquiry Optima PLUS Steerable Electrophysiology Catheters are intended for electrogram recording and stimulation during electrophysiological studies. The modification in additional electrodes on the shaft of the Inquiry Optima PLUS catheter does not affect the intended use or scientific technology of the device.

    1. Brief summary of non-clinical tests and results:
      The test plan for the Inquiry Optima PLUS Steerable Electrophysiology Catheter was based on the guidance document "Electrode Recording Catheter Preliminary Guidance, Draft Version", March 1995. The test results indicated reliable performance when the device was used in accordance with the Instructions for Use. The Inquiry Optima PLUS catheter does not raise new issues of safety, effectiveness, or performance of the product.

page 2 of 3

{2}------------------------------------------------

The Inquiry Optima Steerable EP Catheter and the Inquiry Optima PLUS Steerable EP catheter are both intended to be used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. In comparison to the Inquiry Optima Steerable EP Catheter, the Inquiry Optima PLUS Steerable EP Catheter has 4 additional electrodes. The similarities and differences between these catheters are tabularized below.

Choorg

Product DescriptionInquiry Optima Steerable EPCatheter (K042775)Inquiry Optima PLUS SteerableEP Catheter
TypeVariable loop, steerable diagnosticcatheterSame
Size7Fr Shaft, 4-5Fr LoopSame
Usable Length110cmSame
Electrode Number onLoop10 - 20 electrodes20 electrodes
Size (width, diameter) ofElectrode on Loop0.6 mm - 1.0 mm width4 Fr - 5 Fr diameter1.0 mm width5 Fr diameter
Electrode Number onShaftNone4 electrodes
Size (width, diameter) ofElectrode on ShaftNone1 mm width7 Fr diameter
Distal Tip Electrode(length, diameter)1 mm - 2 mm length,4 Fr - 5 Fr diameter1mm length5 Fr diameter
Electrodes MaterialPlatinum / IridiumSame
Electrode ProfileEach band has adhesive applied onboth ends of the electrodeSame
Type of CurveThe catheter has a uni-direction 0-180° deflectable distal end, SMcurve and an adjustable loopSame
Loop SizeAdjustable, minimum is 15mmdiameter, maximum is 25mmdiameterSame
Type of Catheter OuterShaft7 Fr size, non braided on distal,braided on proximalSame
Material of CatheterOuter ShaftPebaxSame
Handle and ConnectorThe handle has a push/pullmechanism to steer the distaldeflection curve and a rotatingknob to adjust the distal loopdiameter. The handle has aconnector with 10 to 24 pinsattached at the back of the handle.The handle has a push/pullmechanism to steer the distaldeflection curve and a rotatingknob to adjust the distal loopdiameter. The handle has aconnector with 24 pins attached atthe back of the handle.
Type of Conductor WireElectrodes use a Copper / NickelAlloySame
Single Use or UsableSingle Use onlySame

Page 3 of 3

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 9 2006

Irvine Biomedical, Inc. c/o Mr. Dennis Hong Regulatory Affairs Manager 2375 Morse Avenue Irvine, CA 92614

Re: K060757

Trade Name: Inquiry™ Optima™ PLUS Steerable Electrophysiology Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Regulatory Class: Class II (two) Product Code: DRF Dated: March 20, 2006 Received: March 21, 2005

Dear Mr. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Dennis Hong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Bhimmar for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

KO60757

Indications for Use

KOBO757 510(k) Number: Inquiry Optima PLUS Steerable Electrophysiology Catheter Device Name: The Inquiry Optima Catheter is a steerable electrophysiology Indications for Use: catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The Optima catheters are to be used to map the atrial regions of the heart.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

Blumina

(Division Sig
Division of Cardio vascular Devices
510(K) Number: K060757

Page | of 1

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).