K Number
K060709
Manufacturer
Date Cleared
2006-06-15

(91 days)

Product Code
Regulation Number
862.3450
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Multigent® Gentamicin assay is intended for the quantitative determination of Gentamicin in human serum or plasma on the Architect C8000 System. The results obtained are used in the diagnosis and treatment of Gentamicin overdose and in monitoring levels of Gentamicin to ensure appropriate therapy.

Device Description

The Multigente Gentamicin assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle. The assay consists of reagents R1: anti-gentamicin monoclonal antibody and R2: gentamicin-coated microparticles. A six-level set of Multigent" Gentamicin Calibrators (A through F) is used to calibrate the assay.

AI/ML Overview

This document describes the acceptance criteria and the studies performed to demonstrate the performance of the Multigent® Gentamicin assay.

1. Table of Acceptance Criteria and Reported Device Performance:

Performance MetricAcceptance CriteriaReported Device Performance
Accuracy (Recovery)100 ± 10% or 0.1 µg/mLConcentration (µg/mL)
Linearity (Recovery)100 ± 10%Concentration (µg/mL)
Sensitivity (LDD)Not explicitly stated in the "Acceptance Criteria" column, but the goal is to claim 0.1 µg/mL.The average LDD is 0.09 µg/mL. This supports a claim of 0.1 µg/mL.
Interference (Bilirubin)100 ± 10%Mean Recovery: 3.42 µg/mL for a target of 3.44 µg/mL. (% Recovery for 20mg/dL Bilirubin: 99.42% (calculated from data). The table has an error in displaying this value.)
Interference (Hemoglobin)100 ± 10%Mean Recovery: 3.38 µg/mL for a target of 3.44 µg/mL. (% Recovery for 2g/dL Hemoglobin: 98.26%)
Interference (Triglyceride)100 ± 10%Mean Recovery: 3.30 µg/mL for a target of 3.44 µg/mL. (% Recovery for 1691 mg/dL Triglyceride: 95.83%)
Interference (Total Protein)100 ± 10%Mean Recovery: 3.21 µg/mL for a target of 3.44 µg/mL. (% Recovery for 12 g/dL Total Protein: 93.41%)
Interference (Rheumatoid Factor)100 ± 10%Mean Recovery: 3.26 µg/mL for a target of 2.46 µg/mL. (% Recovery for 582 IU Rheumatoid Factor: 132.34%) - This value exceeds the acceptance criteria of 100 ± 10%.
Interference (HAMA Type-1)100 ± 10%Mean Recovery: 3.30 µg/mL. (% Recovery: 99.10%)
Interference (HAMA Type-2)100 ± 10%Mean Recovery: 3.08 µg/mL. (% Recovery: 93.34%)
Method Comparison (Correlation with Predicate)High correlation (e.g., R-squared close to 1)N = 55, Slope = 1.165, y-intercept = -0.719, R = 0.996, R² = 0.992. "Results show excellent correlation between the two assays."
PrecisionCV (%) ranges from 1.07% to 5.69% for various control levels (details in document for within-run, between-day, between-run, and total precision).Low Control (2.68 µg/mL): Total CV 5.69%, SD 0.15Mid Control (6.47 µg/mL): Total CV 2.44%, SD 0.16High Control (9.41 µg/mL): Total CV 2.15%, SD 0.20(Full details for within-run, between-day, and between-run are available in the provided text, and all appear to be within acceptable limits for an immunoassay.)
On-Board Stability (Calibration Curve)Data supports the stability period28 days
On-Board Stability (Reagent)Data supports the stability period40 days
AnticoagulantsNo significant difference in recovery with various anticoagulants"The results indicate that there is no significant difference between the recovery of Gentamicin in serum or plasma. The collection tubes evaluated show no adverse effects on the recovery of Gentamicin, within the experimental error for the spiking study."

Note regarding Rheumatoid Factor interference: The reported % Recovery (132.34%) for Rheumatoid Factor at 582 IU exceeds the stated acceptance criteria of 100 ± 10%. This indicates potential interference from high levels of Rheumatoid Factor, which should be noted in the device labeling.

2. Sample Size and Data Provenance:

  • Accuracy: 5 samples (0.25, 1.00, 2.25, 4.50, 8.00 µg/mL) each run in triplicate (total of 15 measurements in the table shown). Data provenance not specified (retrospective/prospective, country of origin).
  • Linearity: 4 serially diluted samples (6.88, 5.16, 3.44, 1.72 MG/ML) each run in triplicate (total of 12 measurements in the table shown). Data provenance not specified.
  • Precision: 3 control levels (low, mid, high) tested, with N=80 for each (number of replicates over time). Data provenance not specified.
  • Sensitivity: Calibrator A (0 µg/mL) run for a total of 20 replicates. Data provenance not specified.
  • Interferences (Endogenous Substances):
    • Bilirubin: 3 replicates for "N=3" (interpreted as the number of independent samples or measurement replicates, as the "N" column is inconsistent).
    • Hemoglobin: 2 replicates for "N=2".
    • Triglyceride, Total Protein: "N" is listed as a non-numeric character (presumably an error in transcription, but the text mentions "run in triplicate" for triglyceride and total protein).
    • Rheumatoid Factor: "N" is listed as a non-numeric character (text mentions "run in triplicate").
    • Data provenance not specified.
  • Interferences (HAMA): Duplicate HAMA samples (Type-1 and Type-2) and duplicate control samples. Data provenance not specified.
  • Interferences (Common Co-Administered Drugs): Test samples and control samples assayed in duplicate for each drug. Data provenance not specified.
  • Anticoagulants: Blood drawn from at least ten healthy donors for each of the 9 tube types. Samples were spiked with Gentamicin and run in duplicate. Data provenance not specified.
  • Method Comparison: 55 serum and plasma samples, ranging from 0.78 to 9.02 µg/mL Gentamicin. Data provenance not specified.

3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts:

This device is an in vitro diagnostic (IVD) assay for quantitative determination of a drug concentration. The "ground truth" for chemical concentration assays is typically established by reference methods, primary standards, or gravimetric/volumetric preparation of known concentrations. There is no mention of human "experts" establishing ground truth in the traditional sense of clinical opinion (e.g., radiologists, pathologists). The ground truth for performance studies like accuracy and linearity is based on the theoretical concentrations of prepared samples or reference materials. For method comparison, the predicate device (Abbott TDx®/TDxFLx® Gentamicin assay) serves as the reference method.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

Not applicable. Adjudication methods are typically used in studies involving subjective assessment (e.g., image interpretation) where multiple readers provide independent evaluations that might conflict. For quantitative chemical assays, the result is a numerical value, and "adjudication" is not a standard practice. Statistical methods are used to compare results to expected values or reference methods.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study or assessment of human reader improvement with AI assistance is irrelevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The performance studies described (Precision, Accuracy, Linearity, Sensitivity, etc.) represent the standalone performance of the Multigent® Gentamicin assay system. The device itself generates a quantitative result without direct human interpretation of a visual output. The human-in-the-loop would be the laboratory personnel operating the Architect C8000 System and interpreting the numerical result in a clinical context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth for this device's performance relies on:

  • Theoretical Concentrations: For accuracy, linearity, and sensitivity studies, known concentrations of Gentamicin in control matrices serve as the ground truth. These are typically prepared using gravimetric or volumetric methods with highly pure reference standards.
  • Reference Method: For the method comparison study, the Abbott TDx®/TDxFLx® Gentamicin assay served as the reference (or comparative) method, with its results considered the "ground truth" for evaluating the new device's correlation.

8. The sample size for the training set:

Not applicable. This device is a chemical immunoassay, not a machine learning or artificial intelligence system that requires a "training set" in the computational sense. The assay works based on established biochemical principles and reagents rather than being "trained" on data.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of in vitro diagnostic device.

{0}------------------------------------------------

510K SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is;

COMPANY/CONTACT PERSON

Seradyn. Inc 7998 Georgetown Road, Suite 1000 Indianapolis, IN 46268

JUN 15 2006

Establishment registration No: 1836010

Earl E. Knight III, MPA Regulatory Affairs Associate Telephone: (317) 544-0639 Fax: (317) 610-0018

DATE PREPARED

March 14, 2006

DEVICE NAME

Trade Name:

Multigent® Gentamicin

Common Name: Enzyme Immunoassay, Gentamicin

Device Classification:

21 CFR 862.3450; Gentamicin Test System; ●■■■■

INTENDED USE

The Multigent" Gentamicin assay is intended for the quantitative determination of Gentamicin in human serum or plasma on the Architect C8000 System.

LEGALLY MARKETED DEVICE TO WHICH EQUIVALENCY IS CLAIMED

Abbott TDx®/TDxFLx® Gentamicin (K904226)

DESCRIPTION OF DEVICE

The Multigente Gentamicin assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle. The assay consists of reagents R1: anti-gentamicin monoclonal antibody and R2: gentamicin-coated microparticles. A six-level set of Multigent" Gentamicin Calibrators (A through F) is used to calibrate the assay.

{1}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
DeviceMultigent® GentamicinPredicateAbbott TDx®/TDxFLx® Gentamicin
Intended UseThe Multigent® Gentamicin assay isintended for the quantitativedetermination of Gentamicin in humanserum or plasma on the ArchitectC8000 System.The TDx®/TDxFLx® Gentamicin assay isa reagent system for the quantitativemeasurement of Gentamicin, anantibiotic drug in human serum orplasma.
Indicationsfor UseThe results obtained are used in thediagnosis and treatment ofGentamicin overdose and inmonitoring levels of Gentamicin toensure appropriate therapy.The measurements obtained are usedin the diagnosis and treatment ofGentamicin overdose to ensureappropriate therapy.
MethodologyHomogeneous particle-enhancedturbidimetric immunoassay (particleagglutination)Fluorescence PolarizationImmunoassay (FPIA) technology.
ReagentComponentsTwo (2) reagent system:Anti-Gentamicin Antibody Reagent(R1) in buffers containingstabilizers with sodium azide Gentamicin-coated MicroparticleReagent (R2) in buffer containingstabilizers with sodium azideThree (3) reagent system:Pretreatment Solution (P) Surfactantin buffer containing protein stabilizerand sodium azide. S Gentamicin Antiserum (Sheep) inbuffer with protein stabilizer andSodium azide. T Gentamicin Fluorescein Tracer inbuffer with protein stabilizer,surfactant and Sodium azide
CalibrationMultigent® GentamicinCalibrators - six levelsGentamicin Calibrators - six levels

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

SUMMARY OF CLINICAL TESTING

Precision

A precision study was performed using the Clinical and Laboratory Standards Institute (CLSI) Guideline NCCLS EP5: Evaluation of Precision Performance of Clinical Chery Canadial Include

.

NMeanµg/mLWithin RunBetween DayBetween RunTotal
SDCV (%)SDCV (%)SDCV (%)SDCV (%)
Low Control802.680.082.930.124.370.062.170.155.69
Mid Control806.470.071.090.121.910.071.070.162.44
High Control809.410.141.530.131.340.070.700.202.15

{2}------------------------------------------------

Accuracy

Accuracy was determined using a procedure described in CLSI Guideline NCCLS EP6-A. The samples were Accuracy was dolomined abing a problement in assay. A mean of the replicates for each sample was determined and percent recovery calculated.

% Recovery
THEORETICALCONC.(µg/mL)Rep 1Rep 2Rep 3MeanSDCVAcceptanceCriteria:100±10% or0.1 µg/mL
0.250.230.280.260.260.0259.82102.67%
1.001.020.990.981.000.0212.0999.67%
2.252.242.222.212.220.0150.6998.81%
4.504.444.554.424.470.0701.5799.33%
8.008.127.998.158.090.0851.05101.08%
Mean Percent Recovery100.31%

Image /page/2/Figure/3 description: This image shows a graph titled "Accuracy". The x-axis is labeled "Theoretical Conc. ug/mL" and ranges from 0 to 8. The y-axis is labeled "Mean Recovered Conc. ug/mL" and ranges from 0 to 10. A line is plotted on the graph, and the equation for the line is y = 1.01x + 0.024, with an R-squared value of 0.9999.

Linearity

A Gentamicin linearity standard was serially diluted and run in triplicate using the Multigent® Gentamicin assay. A mean of the replicates for each sample was determined and percent recovery was calculated.

THEORETICALCONC.(MG/ML)Rep 1Rep 2Rep 3MeanRecoveredConc.SD%CV% RecoveryAcceptanceCriteria:$100\pm10%$
6.887.147.317.147.200.0981.36104.60
5.165.205.235.395.270.1021.94102.20
3.443.503.653.703.620.1042.88105.14
1.721.591.671.661.640.0442.6695.35
Mean Percent Recovery101.82

{3}------------------------------------------------

Image /page/3/Figure/0 description: The image shows a linearity graph with the title "Linearity". The graph plots "Recovered conc. µg/mL" on the y-axis against "Theoretical conc. µg/mL" on the x-axis. The plot includes data points for "Theoretical vs. Recovered" and a linear regression line for "Linear (Theoretical vs. Recovered)", with the equation y = 1.0657x - 0.15 and an R-squared value of 0.9989.

Sensitivity

The Analytical Sensitivity or Least Detectable Dose (LDD) of the assay is defined as the concentration at which the lowest concentration is distinguishable from zero with 95% confidence.

A calibration curve was run. Calibrator A (0 µg/mL.) was run for a total of 20 replicates. The LDD was calculated using the following formula:

$$\text{'LDD} \qquad = \frac{2 \times \text{(SD rate of Zero Cal)}}{\text{(rate of Zero Cal - rate of 4} \text{'' non-zero cal)}} \qquad \text{' } \qquad \text{'Cuce of 4} \text{'' non-zero cal)}$$

Where:

  • . Zero Cal = Cal A (0 uq/mL)
  • . SD Zero Cal = standard deviation of the multiple determinations
  • 1st Non-Zero Cal = Cal B (0.5 µg/mL) .

The average LDD is 0.09 µg/mL, supporting a claim of 0.1µg/mL.

Specificity

The Multigent® Gentamicin assay utilizes a mouse derived (ascites) Gentamicin monoclonal antibody directed against Gentamicin. There are no metabolites of Gentamicin.

Interferences

Interference were assessed using the Dose Response Method.

A. Endogenous Substances

Bilirubin

A Bilirubin Stock was prepared by adding bilirubin to normal human serum at 400 mg/dL. Dilutions were then made to produce bilirubin levels of 20, 40, and 60mg/dL. An Analyte Stock was prepared by adding Gentamicin to 10 mL of normal human serum. A 1:100 dilution of the Analyte Stock was prepared using each level of bilirubin. The control was prepared by diluting the Analyte Stock 1:100 with normal hurnan serum. Each level was assayed. Only the 20 mg/dl. data is shown.

{4}------------------------------------------------

Hemoglobin

A Hemoglobin stock solution was prepared from a human blood hemolysate. The hemolysate was diled to nomal human serum at 1.0 and 2.0g/dL hemoglobin. The Analyte Stock from above was diluted 1:100 with each level of hemoglobin. A control was prepared using normal human senum. Each level was assayed. Only the 2g/dL data is shown.

Triglyceride

A 1691 mg/dL patient pool was spiked with Gentamicin stock. The spiked sample was run in
tripliedto triplicate.

Total Protein

A 12 g/dL stock of Human Serum Albumin (HSA) in saline was prepared. The sample was spiked with Gentamicin stock. The spiked sample was run in triplicate.

Rheumatoid Factor

A 582 IU patient pool was spiked with Gentamicin stock and run in triplicate.

Results are shown in the table Below.

{5}------------------------------------------------

ENDOGENOUS INTERFERING SUBSTANCE

InterferingSubstanceInterferentConcentrationNTarget(No Interferent)ug/mLMeanRecoveryug/mL% RecoveryAcceptanceCriteria:100±10%
Bilirubin20mg/dL33.443.42ರಿಗಿ ನಿರ್ವಿಸಿದ
Hemoglobin2g/dL23.443.3898.26
Triglyceride1691 mg/dL3.443.3095.83
Total Protein12 g/dL3.443.2193.41
Rheumatoid Factor582 IUತಿ2.463.26132.34

B. HAMA

As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (HAMA) in the sample, which could cause falsely elevated results.

For this study, HAMA Type-1 and Type-2 samples were spiked with Gentamicin. The Mean Recovery for each (Type-1 and Type-2) of the duplicate HAMA samples was compared to the Mean Recovery of each respective Control (normal human serum).

Results are shown in the table below.

HAMA

Rep 1µg/mLRep 2µg/mLMeanRecoveryµg/mLSDCV% RecoveryAcceptance Criteria:$100\pm10%$
HAMAType-13.313.293.300.0140.4399.10
Control3.403.263.330.0992.97100.00
HAMAType-23.083.073.080.0070.2393.34
Control3.403.263.330.0992.97100.00

{6}------------------------------------------------

C. Common Co-Administered Drugs

Gentamicin was spiked into normal human serum. Co-Administered drug stock concentrates were prepared at 100X of the initial concentration tested. A test aliquot was then prepared for each cross-reactant by combining 99 (or 9) volumes of the Gentamicin-spiked whole serum with 1 volume of the first and of of the United on the Gentament-spited while series while in the many

A control aliquot was prepared for each solvent system by combining 99 (or 9) volumes of the stock analyte solution with 1 volume of only of online by combining as (of sy voluntes of the stocure of the corresponding solvent used for the cross-reactant concentrate (no cross-reactant).

The test samples and control samples were then assayed in duplicate. Percent cross- reactivity was calculated using the following formula:

Percent Cross-Reactivity = $((D_A - D_T) / C) X 100$

Where:

$D_T$ = average observed concentration of the control solution

$D_A$ = average of the observed concentration of the cross-reactant test solution

C = concentration at which the cross-reactant is tested

Tabulated data is shown in the table below.

{7}------------------------------------------------

Cross-reactant DrugConc. Testedµg/mLPercent Cross-Reactivity
5-Fluorocytosine300.5333
Acetaminophen200ND
Acetyl Cysteine1000ND
Acetylsalicylic Acid300ND
Amikacin300ND
Amphotericin B100ND
Ampicillin50ND
Ascorbic Acid30-0.2333
Carbenicillin2500ND
Cefamandole Naftate250ND
Cefoxitin1000ND
Cephalexin320ND
Cephalosporin C1000ND
Cephalothin1000ND
Chloramphenicol250ND
Clindamycin2000ND
Cyclosporin6000ND
Erythromycin500ND
Ethacrynic Acid400ND
Furosemide100ND
Fusidic Acid1000ND
Ibuprofen7000ND
Kanamycin A4000.1017
Kanamycin B400ND
Levodopa1000ND
Lincomycin2000ND
Methicillin200ND
Methotrexate50ND
Methylprednisolone200ND
Metronidazole1000ND
Neomycin1000ND
Netilmicin1250.2533
Oxytetracycline2000ND
Penicillin V10ND
Phenylbutazone1000ND
Prednisolone12-0.2920
Rifampin50ND
Sisomicin1050.35
Spectinomycin100ND
Streptomycin400ND
Sulfadiazine1000ND
Sulfamethoxazole400ND
Tetracycline2000ND
Theophylline200ND
Ticarcillin100-0.4400
Tobramycin1000.1633
Trimethoprim200.3000
Vancomycin400ND

*ND = Not Detected

and the second and the state of the states

.

.

:

{8}------------------------------------------------

D. Anticoagulants

Studies were conducted to determine the performance characteristics of the assay for both serum and plasma samples containing Gentamicin.

Blood was drawn from at least ten healthy donors (no Gentamicin therapy) for each tube type listed below:

  • plastic K2 EDTA tube
  • · giass K3 EDTA tube
  • · glass Plasma separator lithium heparin tube
  • glass sodium heparin tube
  • · glass lithium heparin tube
  • · glass serum separator tube
  • plastic tube with clot activator
  • · glass tube; no additive (served as the control)
  • · plastic tube; no additive

All tubes were processed per the manufacturers instructions. The serum or plasma was removed from the collection tubes and aliquoted into new tubes for testing. Serum or plasma from the removed type was spiked with Gentamicin. The samples were analyzed on the Architect C8000 analyzer in duplicate. Baseline results were obtained on day zero for each type of tube.

The results indicate that there is no significant difference between the recovery of Gentamicin in serum or plasma. The collection tubes evaluated show no adverse effects on the recovery of Gentamicin, within the experimental error for the spiking study.

A claim for assay application to both serum and plasma samples is thus supported.

Method Comparison

A study was conducted according to CLSI Guideline NCCLS EP9-A2: Method Comparison and Bias Estimation Using Patient Samples to compare accuracy of Gentamicin in serum assayed by the Multigent® Gentamicin assay to the Abbott TDx® TDxFLx® Gentamicin assay.

Serum and plasma samples, ranging from 0.78 to 9.02µg/mL Gentamicin, were first tested using Abbott's TDx Gentamicin assay. The same samples were then tested by the Multigent Gentamicin assay on the Architect C8000 analyzer.

Mean values for the Abbott TDx TDxFLx® Gentamicin assay reference method were plotted against those for the Multigent Gentamicin assay on the Architect C8000 (Figure 3). The results, using Passing – Bablok parameters, are:

$N$ = 55 Slope = 1.165 y-intercept = -0.719 $R$ = 0.996 $R^2$ = 0.992

Results show excellent correlation between the two assays.

{9}------------------------------------------------

On-Board Stability

1) Calibration Curve stability

Calibration curve stability of a period of 28 days is supported by the data.

2) Reagent On-Board Stability

A 40 day on-board reagent stability claim is supported by the data.

CONCLUSION

The Multigent" Gentamicin assay has been shown to be substantially equivalent to the Abbot
TDxYTDxFLx" Gentamicin assay. The performance testing verifies that the device func and that design specifications have been satisfied.

{10}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing wings or feathers. The symbol is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circumference of the circle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 15 2006

Earl E. Knight III, MPA Regulatory Affairs Associate Seradyn, Inc. 7998 Georgetown Road Suite 1000 Indianapolis, IN 46268-5260

K060709 Re:

Trade/Device Name: Multigent® Gentamicin Regulation Number: 21 CFR§862.3450 Regulation Name: Gentamicin test system Regulatory Class: Class II Product Code: LCD Dated: May 23, 2006 Received: May 24, 2006

Dear Mr. Knight:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application(PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registred - not liising (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{11}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free no (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Alberto Guti

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

Indications for Use

510(k) Number (if known):

Multigent® Gentamicin Device Name:

K060709 Indications for Use:

The Multigent Gentamicin assay is intended for the quantitative determination of · Gentamicin in human serum or plasma on the Architect C8000 System.

The results obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.

Prescription Use(Part 21 CFR 801 Subpart D)XAND/OROver-The-Counter Use(21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Office of In Vitro Diagnostic Device

K060709

xii

§ 862.3450 Gentamicin test system.

(a)
Identification. A gentamicin test system is a device intended to measure gentamicin, an antibiotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.(b)
Classification. Class II.